CSA Medical truFreeze Active Venting Setup Instructions

Page 1
truFreeze
®
Active Venting Set-up Instructions
1. Position the console as desired for procedure and lock the wheels by pushing the red lever to either side.
2. Remove the power cord from the holder and plug into standard wall power. Ensure the power cord is connected to the console.
Make sure the power cord
is dry.
3. Power on the console using the green switch on the back of the device.
4. Ensure that the red
emergency stop button is off (NOT lit).
3
1
6
2
4
truFreeze
®
Active Venting Set-up Instructions
16. If using the remote, ensure the remote control
has two (2) working AA batteries installed by selecting the run tab and attempting to adjust the site counter. Reset site to “1”.
5. The computer will run a series of built-in tests (BIT). Note: the device must have adequate liquid nitrogen to pass the BIT.
P
See separate tank filling instructions.
6. While the BIT is running, position the foot pedals on the floor as desired, ensuring the foot pedal cord is plugged into the console.
Avoid touching the pins of the foot pedal
connector.
5
CSA Medical, Inc.
For service, call: 866.481.7786
www.csamedical.com
© 2014 CSA Medical, Inc. All rights reserved.
truFreeze is a registered trademark of CSA Medical, Inc.
15-00292-00
Active Venting Set-up
Liquid nitrogen expands more than 700 times when changing to a gaseous state. It is
imperative for patient safety that adequate ventilation is ensured.
note: some consoles also display a shutdown button.
15. To ensure the suction system has been set up correctly, depress the gray suction pedal and verify that the tubing draws suction. Then press the suction pedal again to turn off suction prior to the start of the procedure.
Please read the Instructions For Use and Operator’s Manual for complete
information about the truFreeze System.
Page 2
truFreeze
®
Active Venting Set-up Instructions
11. Remove the 4 inch connector piece from the suction tubing package and connect the blue end to the top port of the Bemis suction canister.
12. Place the canister in the front of the console. Attach the clear end of the 4 inch accessory tubing to the suction port on the console above the canister. Ensure a tight seal.
13. Remove the long section (72 inches) of accessory tubing and attach the blue end to the front port of the canister. The other end will be attached to the cryodecom­pression tube (CDT) AFTER it is placed in the patient by the physician.
truFreeze
®
Active Venting Set-up Instructions
10. Locate the sterile truFreeze catheters and cryodecompression kits.
Make sure the catheters and cryodecompression
kits have not passed their expiration dates.
cryodecompression kit
catheter kit
12
gray suction pedal
suction port
8. Select the “run” screen.
9
9. Confirm that the device has sufficient liquid nitrogen via the indicator on screen (in the setup tab).
If the tank indicator displays orange or red,
the device must be filled prior to use. Filling is recommended but not required if the indicator is yellow.
P
See separate tank filling instructions.
7. If the BIT fails, a red “x” and “fail – complete” will display. Press the test button at the home screen and let the system repeat the BIT.
If the system fails BIT two (2) times with the
same error, please call service at 866.481.7786
cdt tubing
(note black bands)
14. Leave the CDT (the tube with the black bands) in the packaging until it is requested by the physician.
Loading...