Dierential Pressure Test Set User Instructions
Overview
Using Crystal Engineering’s Dierential Test Set based XP2iDD and XP2i gauges, it’s possible to perform pressure dierential measurements. By simply connecting your upstream
(+) and downstream (–) pressure lines to the 1/8” FNPT ttings
you are ready to go. This test set is capable of displaying
live or average dierential pressures (average of 10 samples)
depending upon which display mode you select.
Figure 1: General piping interconnect diagram.
Operation
A typical dierential pressure measurement may be done in
the following manner.
i If averaging mode is selected, the lower line will
display the averaged dierential pressure regardless
of what display mode the left hand (+) gauge is in.
ii When only the
AVG
icon is lit, the second line display is
the averaged non-dierential reading from the gauge
and may not be appropriate for your application.
b Left hand (+) or upstream gauge: Using the
button, select live pressure readings (no HI or LO icons),
or Average pressure over 10 reading (both HI and
icons).
3 Pressure Connections: Connect upstream (+), down-
stream (–), and their corresponding vent lines to the four
1/8” FNPT connection ports located near the hinge of case.
Use pipe thread tape or pipe thread sealant as necessary to
ensure a leak free connection.
a Always use a wrench on the test set bulkhead
1
/8" FNPT
tting, as there is a limit to how much rotational force
can be applied to the case.
b Due to the robust nature of the Crystal XP2I gauges, full
scale pressure may be applied to either gauge at any
time without damage.
4 Zero Gauges: Zero both gauges in the vented condition
by pressing the
button on each gauge. Figure 2
illustrates the appropriate valve states to ensure venting
and isolation between both gauges.
1 Select Units: Select your required unit of pressure mea-
surement on each gauge by pressing the
button
on each gauge independently. You may select dierent
units for each gauge but they must be set individually. The
dierential pressure will be displayed on the lower line of
the right (–) dual display gauge and will be in the same
units as the static pressure.
2 Select Display Options: Select the appropriate display
option on each unit using the
a Right hand (–) or dierential pressure gauge: Using
the
button, select live pressure display (both
and LO icons), or Average pressure over 10 read-
ings (both
, LO, and
be display on the primary line and the dierential pres-
AVG
button:
icons). Static pressure will
sure will be displayed on the secondary or lower line;
Figure 2: Zero gauge valve states.
5 DP Measurement Mode: Change the valves as shown
C R Y S T A L
XP
2
C R Y S T A L
XP
2
HI
C R Y S T A L
XP
2
C R Y S T A L
XP
2
HI
BUREAUVERITAS
IS0900
1
C
ertification
No 191559
in gure 3 to allow line pressure and dierential pressure
measurement.
a Dierential Pressure: The lower line of the right
hand (–) gauge is the dierential measurement between
both gauges. Place the right hand (–) gauge in averaging mode if further damping is required in the event of
erratic or noisy readings.
Specifications
Pressure
Connections: 1/8" female NPT.
System Pressure Limit: Based on lowest pressure rating of
either gauge.
Valving Pressure Limit: 3000 psi maximum.
Accuracy of system using equal pressure range
gauges
20% to 100% of Full Scale: ±(0.14% of Static Line Reading).
0% to 20% of Full Scale: ±(0.028% of Full Scale).
Vacuum for 2000 kPa (300 PSI) and lower pressure gauges:
0 to -99.9 kPa (-14.5 PSIG): ±(0.35% of Full Scale),
Where F.S.: -99.9 kPa (-14.5 PSIG).
Batteries
3 each “D” cell. Rotate lid counter-clockwise 90 degrees to
remove.
Battery Life
1200 hours typical.
Figure 3: DP measurement mode (pressure gauge isolation).
6 Vent: Vent left-hand (+) gauge as shown in gure 4.
Reverse both upper valves to vent the right hand (–) gauge.
Temperature
Operating and Compensated: -10C to +50C (14F to 122F).
Storage: -40C to +75C (-40F to 167F).
Display Options
Line or Average (over 10 samples).
Display Rate
3 readings/second.
If used in Average mode, 10 sample window is data taken over
the last 3.33 seconds.
Display Height:
14.5mm (0.55”) Right hand or (–) dual display dierential gauge.
16.5mm (0.65”) If congured with XP2I gauge (left hand
location only).
Media Compatibility:
Liquids and gases compatible with 316 stainless steel.
Enclosure:
Color: Black, watertight, crushproof and dust proof.
IP Rating: IP67.
Figure 4: Vent both gauges.
7 Disconnect Connections: Disconnect connections as
appropriate.
Note: Further detailed information is available in the XP2I-DD
(Dual Display or the unit on the right side). This information
can be found at crystalengineering.net
Weight:
6.1 kg (13 lbs 8 oz)
Size:
37.1 cm x 25.8 cm x 15.2 cm
14.6 in x 10.2 in x 6.0 in
PN: 3540—Re v C