Dierential Pressure Test Set User Instructions
Overview
Using Crystal Engineering’s Dierential Test Set based XP2i­DD and XP2i gauges, it’s possible to perform pressure dier­ential measurements. By simply connecting your upstream (+) and downstream (–) pressure lines to the 1/8” FNPT ttings you are ready to go. This test set is capable of displaying live or average dierential pressures (average of 10 samples) depending upon which display mode you select.
Figure 1: General piping interconnect diagram.
Operation
A typical dierential pressure measurement may be done in
the following manner.
i If averaging mode is selected, the lower line will
display the averaged dierential pressure regardless of what display mode the left hand (+) gauge is in.
ii When only the
AVG
icon is lit, the second line display is the averaged non-dierential reading from the gauge and may not be appropriate for your application.
b Left hand (+) or upstream gauge: Using the
button, select live pressure readings (no HI or LO icons), or Average pressure over 10 reading (both HI and
icons).
3 Pressure Connections: Connect upstream (+), down-
stream (–), and their corresponding vent lines to the four 1/8” FNPT connection ports located near the hinge of case. Use pipe thread tape or pipe thread sealant as necessary to ensure a leak free connection.
a Always use a wrench on the test set bulkhead
1
/8" FNPT tting, as there is a limit to how much rotational force can be applied to the case.
b Due to the robust nature of the Crystal XP2I gauges, full
scale pressure may be applied to either gauge at any time without damage.
4 Zero Gauges: Zero both gauges in the vented condition
by pressing the
button on each gauge. Figure 2 illustrates the appropriate valve states to ensure venting and isolation between both gauges.
1 Select Units: Select your required unit of pressure mea-
surement on each gauge by pressing the
button on each gauge independently. You may select dierent units for each gauge but they must be set individually. The dierential pressure will be displayed on the lower line of the right (–) dual display gauge and will be in the same units as the static pressure.
2 Select Display Options: Select the appropriate display
option on each unit using the
a Right hand (–) or dierential pressure gauge: Using
the
button, select live pressure display (both
and LO icons), or Average pressure over 10 read-
ings (both
, LO, and
be display on the primary line and the dierential pres-
AVG
button:
icons). Static pressure will
sure will be displayed on the secondary or lower line;
Figure 2: Zero gauge valve states.
5 DP Measurement Mode: Change the valves as shown
C R Y S T A L
XP
2
C R Y S T A L
XP
2
HI
C R Y S T A L
XP
2
C R Y S T A L
XP
2
HI
BUREAUVERITAS
IS0900
1
C
ertification
No 191559
in gure 3 to allow line pressure and dierential pressure measurement.
a Dierential Pressure: The lower line of the right
hand (–) gauge is the dierential measurement between both gauges. Place the right hand (–) gauge in averag­ing mode if further damping is required in the event of erratic or noisy readings.
Specifications
Pressure
Connections: 1/8" female NPT.
System Pressure Limit: Based on lowest pressure rating of
either gauge.
Valving Pressure Limit: 3000 psi maximum.
Accuracy of system using equal pressure range gauges
20% to 100% of Full Scale: ±(0.14% of Static Line Reading).
0% to 20% of Full Scale: ±(0.028% of Full Scale).
Vacuum for 2000 kPa (300 PSI) and lower pressure gauges:
0 to -99.9 kPa (-14.5 PSIG): ±(0.35% of Full Scale),
Where F.S.: -99.9 kPa (-14.5 PSIG).
Batteries
3 each “D” cell. Rotate lid counter-clockwise 90 degrees to remove.
Battery Life
1200 hours typical.
Figure 3: DP measurement mode (pressure gauge isolation).
6 Vent: Vent left-hand (+) gauge as shown in gure 4.
Reverse both upper valves to vent the right hand (–) gauge.
Temperature
Operating and Compensated: -10C to +50C (14F to 122F).
Storage: -40C to +75C (-40F to 167F).
Display Options
Line or Average (over 10 samples).
Display Rate
3 readings/second.
If used in Average mode, 10 sample window is data taken over the last 3.33 seconds.
Display Height:
14.5mm (0.55”) Right hand or (–) dual display dierential gauge.
16.5mm (0.65”) If congured with XP2I gauge (left hand location only).
Media Compatibility:
Liquids and gases compatible with 316 stainless steel.
Enclosure:
Color: Black, watertight, crushproof and dust proof.
IP Rating: IP67.
Figure 4: Vent both gauges.
7 Disconnect Connections: Disconnect connections as
appropriate.
Note: Further detailed information is available in the XP2I-DD
(Dual Display or the unit on the right side). This information can be found at crystalengineering.net
Weight:
6.1 kg (13 lbs 8 oz)
Size:
37.1 cm x 25.8 cm x 15.2 cm
14.6 in x 10.2 in x 6.0 in
PN: 3540—Re v C