TXgard-Plus Operation
11
4. Operation
WARNING
Prior to carrying out any work ensure local regulations and site
procedures are followed. Never attempt to open the detector
or junction box when flammable gas is present. Ensure that the
associated control panel is inhibited so as to prevent false alarms.
4.1 Commissioning Procedure
a Open the junction box of the detector by loosening the tamper proof
grub screw and removing the lid by rotating it anti-clockwise.
b Remove the amplier and check that all electrical connections have
been made and are correct as per Diagram 3.
c Replace the amplier and close the junction box ensuring that all
screws have been re-tted correctly.
d Apply power to the detector. The status LED will show a steady green
indicating normal operation.
e Leave the detector to stabilise for 1-2 hours.
Zeroing the detector (ZERO Mode):
f Place the end of ‘CRO-Mag’ over the ‘MENU’. Green Status LED will
ash. Decimal point above ‘ZERO’ will illuminate. Display backlight
will illuminate. Decimal point above ‘INHIBIT’ may ash if the option
has been internally selected.
g With no ammable gas present at the detector, place the end of the
‘CRO-Mag’ key over the ‘UP’ or ‘DOWN’ arrow pads to make the
display read zero.
Note: Oxygen detectors may be zeroed in fresh air.
Calibrating the detector (CAL Mode):
h Place the end of the ‘CRO-Mag’ over the MENU pad. The decimal
point above the word CAL will illuminate. Other indications remain
unchanged.
i Apply calibration gas to the detector at a ow rate of 0.5 litres/min.
(contact Crowcon for the supply of calibration gas.)