Fault finding TXgard-IS+
Issue 7 28
“Cal gain error” The signal is very high or low for the
level of gas you are applying. A
scrolling message would have been
displayed on completion of the zero or
calibrate operation giving this fault.
Check you are using the right
level of gas, and that you are
inputting the correct number
through the keypad. Ensure
sensor has been properly zeroed
before calibrating.
“Gas calib. err” (Fatal) Gas calibration data stored in the
sensor module’s non-volatile memory
has been corrupted.
Sensor must be re-zeroed and
calibrated. Without calibration
data the instrument cannot
correctly measure gas.
“mA Calib error” mA calibration data stored in the
instrument’s non-volatile memory has
been corrupted.
mA calibration must be checked,
using the Ramp mA function (see
Appendix D), and re-calibrated or
instrument must be returned to
Crowcon for service.
“FRAM1 fault”
“FRAM2 fault”
FRAMs are non volatile memory. The
instrument has detected corruption in
one. FRAM1 is on the main PCB and
stores the instrument configuration.
FRAM2 is in the Personality Module
and stores sensor data (Calibration
constants, gas name etc.).
Disconnect power and reconnect
it. The FRAM with an error should
restore its data from backup in
the other FRAM.
Alternatively, use the Clear Faults
menu option (see appendix D).
Symptom / error message Cause Solution