Critikon Dinamap Error codes

DINAMAP™ Compact Vital Signs Monitor
6.5 Procedural & Error Alarm Code Table
Alarm Code
N99 N55 N44 N33 N00
N12 Systolic
N13 Systolic
P55 P00
E33 E11 E00
No change
Display ‘N12’
Display ‘N13’
No change
No change
LCD Display Audio Tone &
N99-NIBP FAILED
N55­TIMEOUT: PRESS
N44­TIMEOUT: TOTAL
N33­TIMEOUT: INFLATE
N00- OVER PRESSURE
N0 Display Steady tone,
No Display Steady tone,
P55-SpO2: NO SIGNAL
P00-SpO2: NO SpO2 SENSOR
E33-Temp: FAIL
E11-Temp: FAIL
E00-Temp: FAIL
Volume
High priority alarm. Volume adjustable.
maximum volume
maximum volume
High priority alarm. Volume adjustable.
High priority alarm. Volume adjustable.
Effect of Alarm Silence switch
2 minutes silence
No effect No effect System pressure whilst idle 5 or
No effect No effect Pump on/off status determined by
2 minutes silence
2 minutes silence
Effect of Clear via Rotor
Clear
Clear
Clear Clear Clear
Probable Cause
Unable to make an NIBP determination due to insufficient signal.
One cuff pressure for > 1 minute. Motion artifact.
Determination time exceeds 2 minutes. Motion artifact.
Inflation time > 40 seconds or air leak detected.
Overpressure detected.
15mmHg above ambient for 30 seconds for a neonate or an adult cuff respectively
measurement of pump current does not concur with Main board software expected status. Unit will power down within 30 seconds of detected failure.. No or very low SpO2 signal. Check or reposition sensor.
SpO2 sensor not connected. No sensor code detected. Sensor failure.
Temperature probe not connected or inoperable. Predictive temperature loss of tissue contact > 30 sec. Removal of probe before temperature determination complete.
Predictive temperature determination exceeds 60 sec.
75
DINAMAP™ Compact Vital Signs Monitor Procedural & Error Alarm Code Table continued
Alarm Code
No code
Other: N,P,E,I,S.
No change
Blank
No change
Blank
Blank Error code,
LCD Display Audio Tone &
Low Battery Flashing battery icon
Low Battery ­System Disabled
Printer - No Paper
NIBP Range error
description
Volume
3 beeps every 10 sec, adjustable volume.
Steady tone, maximum volume
High priority alarm Volume adjustable
High priority alarm Volume adjustable Steady tone, maximum volume
Effect of Alarm Silence switch
2 minutes silence
No effect
2 minute silence
2 minute silence
No effect No effect Internal system fault
Effect of Clear via Rotor
No effect
No effect
Clear
Clear
Probable Cause
Replace or recharge battery. From onset of alarm, 5 NIBP measurements available. Beep linearly increases rate as battery discharges.
Replace or recharge battery. NIBP measurement disabled.
Paper expired or printer door open.
NIBP algorithm returned value outside specified accuracy rage.
76