INSTALLATION INSTRUCTIONS
PHYSICAL INSTALLATION
Install the autoclave on a flat, leveled and firm surface at a proper height for the operator to handle it
(approximately 80 cm / 2.6 feet from the ground) Leave enough room close to the autoclave for the proper handling ).
of the materials to be sterilized Install the autoclave in a ventilated and clean place, apart from the room where the .
patients are treated The ideal place for the installation of the autoclave should be a separate sterilization room. .
Important! .Install your autoclave where it can be easily unplugged
ELECTRICAL INSTALLATION
Check if the voltage of the autoclave is the same of the place where it will be installed. To do that, simply look at
the identification label in the back of the autoclave. See “How to Identify your Autoclave” (Fig. 24, page 18).
For the installation use a 20A grounded outlet (3 pins) connecting phase/neutral or phase/phase on the side pins
and grounding on the central pin (Fig. 2, page 8). Never connect grounding to neutral.
ATTENTION As in any other electrical device, grounding is very important for the safety of the operator and the warranty !
of your equipment. Therefore, never remove or cut the plug’s central (grounding) pin off. Non-compliance with this
requirement may damage your autoclave. Cristófoli is not responsible for damages caused by inadequate
installation, voltage and/or electricity fluctuations.
(Fig. 3, page 8).Never use extensions, voltage transformers or any kind of adapters
For the proper functioning of your autoclave, the wiring voltage must be stable, in other words, without fluctuations.
Consult a professional electrician to check the electrical wiring in your building/workplace and make sure they are in
accordance with the specifications required. It is mandatory to use an exclusive circuit breaker for the outlet
where the autoclave will be installed. If after following all the specifications, the electrical current is still fluctuating,
contact your local electricity company for repairs.
The equipment must be always carried by two people to prevent it from falling or causing an accident. The
storage/installation must be done in a place protected from the weather action (indoors) in normal temperature conditions
on a counter that can bear the weight of the autoclave.
Cristófoli Autoclave Vitale can be easily installed. Check if the wiring and voltage of your building are in accordance with
the specifications below by consulting a professional electrician or your Cristófoli local dealer. See “Warranty Terms” (Page
24) and “How to Proceed When Service is Needed” (Page 23).
In case the end user chooses an unauthorized technician to install the autoclave, such technician should follow all the
guidelines described in the instruction manual for the physical, electrical and hydraulic installation of the equipment.
Cristófoli is not responsible for any services performed in disagreement with the instruction manual that accompanies the
product.
The electrical installation must follow the data from the table below.
Note: In areas where voltage is 220 V:
PP (Phase-Phase) use a “bipolar” breaker
PN (Phase-Neutral) use a “unipolar” breaker connected to phase.
1
2
4
3
7
Table 1
MODEL
NOMINAL
CURRENT
WIRING GAUGECIRCUIT BREAKER VOLTAGE
Vitale 12 - 127 V
10 A
6 A
127 V AC » 114 V - 140 V
220 V AC » 198 V - 242 V
For distance up to 5 m (16 ft) from the
circuit breaker to the outlet used to
connect the equipment, use 2,5 mm wiring.
From 5 to 15 m (16 to 46 ft) use 4 mm,
and from 15 to 50 m (46 to 164 ft) 6 mm.
1 Breaker 15 A.
1 Breaker 15 A.
Vitale 12 - 220 V
Vitale 21 - 127 V
Vitale 21 - 220 V
12 A
8 A
1 Breaker 10 A.
1 Breaker 20 A.