Craftsman 82174, 82140 User Manual

Owner's Manual
!
Digital MultiMeter
Model No. 82140
and
Model No. 82174
CAUTION: Read, understand and
follow Safety Rules and Operating Instructions in this manual before using this product.
© Sears, Roebuck and Co., Hoffman Estates, IL 60179 U.S.A.
www.craftsman.com 070606
Safety
! Operation ! Maintenance ! Español
1
TABLE OF CONTENTS
Page Warranty 3 Safety Instructions 4 Safety Symbols 5 Control and Jacks 6 Symbols and Annunciators 6 Specifications 7 Battery Installation 9 Operating Instructions 10 DC Voltage Measurements 10 AC Voltage Measurements 11 DC Current Measurements 12 Resistance Measurements 13 Continuity Check 13 Diode Test 14 Battery Test 14 Maintenance 15 Replacing Batteries 16 Replacing Fuses 16 Troubleshooting 17 Service and Parts 17 Model 82174 instructions 18
2
ONE YEAR FULL WARRANTY
ONE YEAR FULL WARRANTY ON CRAFTSMAN MANUAL RANGING MULTIMETER
If this CRAFTSMAN Manual Ranging MultiMeter fails to give complete satisfaction within one year from the date of purchase, RETURN IT TO THE NEAREST SEARS STORE OR OTHER CRAFTSMAN OUTLET IN THE UNITED STATES, and Sears will replace it, free of charge.
If this CRAFTSMAN Manual Ranging MultiMeter is used for commercial or rental purposes, this warranty applies for 90 days from the date of purchase.
This warranty gives you specific legal rights, and you may also have other rights which vary from state to state
Sears, Roebuck and Co., Dept. 817WA, Hoffman Estates, IL 60179
For Customer Assistance Call 9am-5 PM (EST) Monday through Friday 1-888-326-1006
WARNING: USE EXTREME CAUTION IN THE USE OF THIS DEVICE.
Improper use of this device can result in injury or death. Follow all safeguards suggested in this manual. In addition to the normal safety precautions used in working with electrical circuits. DO NOT service this device if you are not qualified to do so.
3
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the
specified maximum:
Function Maximum Input VAC 600V DC/AC V DC or V AC 600V DC/AC, 200Vrms on 200mV range mA DC 200mA 250V fast acting fuse A DC 10A 250V fast acting fuse(30 seconds max
Resistance, Continuity 250Vrms for 15sec max
2. USE EXTREME CAUTION when working with high voltages.
3. DO NOT measure voltage if the voltage on the "COM" input jack
exceeds 500V above earth ground.
4. NEVER connect the meter leads across a voltage source while the
function switch is in the current, resistance, or diode mode. Doing so can damage the meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect
the power when making resistance or diode tests.
6. ALWAYS turn off the power and disconnect the test leads before
opening the doors to replace the fuse or batteries.
7. NEVER operate the meter unless the back cover and the battery and
fuse doors are in place and fastened securely.
Input Limits
every 15 minutes)
4
SAFETY SYMBOLS
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal injury or damage to the meter.
WARNING
CAUTION
MAX 500V
This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product.
This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds (in this case) 500 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.
5
CONTROLS AND JACKS
1. LCD Display
2. Function switch
3. COM jack
4. 10A jack
5. Positive jack
Note: Tilt stand, fuse and battery compartment are on rear of unit.
SYMBOLS AND ANNUNCIATORS
•))) Continuity
Diode test
" micro (amps)
m milli ( volts, amps)
k kilo (ohms)
##$# ohms
VDC volts direct current
VAC volts alternating current
ADC amps direct current
BAT Battery test
1
2
3 4 5
6
SPECIFICATIONS
Function Range Resolution Accuracy DC Voltage (V DC)
(V AC) DC Current
(A DC)
Resistance
NOTE: Accuracy specifications consist of two elements:
! (% reading) – This is the accuracy of the measurement circuit. ! (+ digits) – This is the accuracy of the analog to digital converter.
NOTE: Accuracy is stated at 65
75% RH.
200mV 0.1mV
%(0.5% reading + 2 digits) 2000mV 1mV 20V 0.01V 200V 0.1V 600V 1V 200V 0.1V AC Voltage 600V 1V
2000"A 1"A 20mA 200mA
10"A 100"A %(1.2% reading + 2 digits)
10A 10mA 200$ 0.1$
%(1.2% reading + 10 digits
(50/60Hz)
%(1.0% reading + 2 digits)
%(2.0% reading + 2 digits)
%(0.8% reading + 2 digits)
2000$ 1$ 20k$ 0.01k$ 200k$ 0.1k$ 2000k$ 1k$ %(1.0% reading + 2 digits) 9V 10mV Battery Test
%(1.0% reading + 2 digits)
1.5V 10mV
o
F to 83oF (18oC to 28oC) and less than
7