Craftsman 82005 Owner's Manual

Owner's Manual
Autoranging Industrial Multimeter
Model No. 82005
Sears, Roebuck and Co., Hoffman Estates, IL 60179
www.craftsman.com 061510
CAUTION
: Read, understand and
follow Safety Rules and Operating Instructions in this manual before using this product.
2
TABLE OF CONTENTS
Page Warranty 2 Safety Instructions 3 Safety Symbols 4 Control, Jacks, & Display 5 Specifications 6 Battery Installation 9 Fuse Installation 10 Operating Instructions 11 DC Voltage Measurements 11 AC Voltage Measurements 12 DC Current Measurements 13 AC Current Measurements 14 Resistance and Continuity Measurements 15 Diode tests 16 Capacitance Measurements 17 Frequency/Duty Cycle Measurements 17 Autoranging 18 Relative, Backlight, Hold, Auto Off, Low Battery 19 Maintenance 20 Troubleshooting 22 Service and Parts 22
ONE YEAR FULL WARRANTY
ONE YEAR Full WARRANTY ON CRAFTSMAN MULTIMETER If this CRAFTSMAN MULTIMETER fails to give complete satisfaction within one year from the date of purchase, RETURN IT TO THE NEAREST SEARS STORE OR OTHER CRAFTSMAN OUTLET IN THE UNITED STATES, and Sears will replace it, free of charge. If this CRAFTSMAN MULTIMETER is used for commercial or rental purposes, this warranty applies for 90 days from the date of purchase. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state
Sears, Roebuck and Co., Dept. 817WA, Hoffman Estates, IL 60179
For Customer AssistanceCall 9am-5 PM (EST) Monday through Friday 1-888-326-1006
3
WARNING: USE EXTREME CAUTION IN THE USE OF THIS DEVICE. Improper use of this device can result in
injury or death. Follow all safeguards suggested in this manual in addition to the normal safety precautions used in working with electrical circuits. DO NOT service this device if you are not qualified to do so.
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that
exceeds the specified maximum:
Input Limits
Function Maximum Input V DC, V AC 1000V DC/AC rms mAAC/DC 400mAAC/DC A AC/DC 10AAC (20A 30secs.) Frequency, Resistance,
Capacitance, Diode, Continuity
600V DC/ACrms
2. USE EXTREME CAUTION working with high voltages.
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above earth ground.
4. DO NOT measure current of circuits whose voltage is greater than 1000V.
5. NEVER connect the meter leads across a voltagesource while the function switch is in the current, resistance, or diode mode. Doing so can damage the meter.
6. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance or diode tests.
7. ALWAYS turn off the power and disconnect the test leads before opening the covers to replace the fuse or batteries.
8. NEVER operate the meter unless the back cover is in place and fastened securely.
4
SAFETY SYMBOLS
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal injury or damage to the meter.
This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result in damage to the meter.
This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage, with respect to earth ground, exceeds (in this case) 1000 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.
WARNING
CAUTION
MAX
1000V
5
CONTROLS, JACKS, AND DISPLAY
1. 4,000 count LCD display
2. RANGE button
3. Hz and % button
4. Mode button
5. Function switch
6. mA, µA and 10A input jacks
7. COM input jack
8. Positive input jack
9. Backlight button
10. RELATIVE button
11. HOLD button
Note: Tilt stand and battery compartment are on rear of unit.
•))) Continuity Diode test Battery status
n nano (10-9) (capacitance) µ micro (10-6) (amps) m milli (10-3) (volts, amps) A Amps k kilo (103) (ohms) F Farads (capacitance) M mega (106) (ohms)  Ohms Hz Hertz (frequency) V Volts % Percent(dutyratio) REL Relative AC Alternatingcurrent Auto Autoranging DC Direct current HOLD Display hold
6
SPECIFICATIONS
Function Range
Resolution
Accuracy
DC Voltage 400mV 0.1mV
(0.5% reading + 2 digits)
4V 0.001V
(1.2% reading + 2 digits)
40V 0.01V 400V 0.1V 1000V 1V
(1.5% reading + 2 digits)
AC Voltage 400mV 0.1mV
(2.0% reading + 10 digits)
4V 0.001V
(2.0% reading + 5 digits)
40V 0.01V 400V 0.1V 1000V 1V
(2.5% reading + 5 digits)
DC Current
400A 0.1A
(1.0% reading + 5 digits)
4000A 1A
(1.5% reading + 3 digits)
40mA 0.01mA 400mA 0.1mA 10A 0.01A
(2.5% reading + 5 digits)
AC Current
400A 0.1A
(2.5% reading + 10 digits)
4000A 1A
(2.5% reading + 5 digits)
40mA 0.01mA 400mA 0.1mA
10A 0.01A
(3.0% reading + 7 digits)
Resistance
400 0.1
(1.2% reading + 4 digits)
4k 0.001k
(1.0% reading + 2 digits)
40k 0.01k
(1.2% reading + 2 digits)
400k 0.1k 4M 0.001M 40M 0.01M
(2.0% reading + 3 digits)
Capacitance
40nF 0.01nF
(5.0% reading + 7 digits)
400nF 0.1nF
(3.0% reading + 5 digits)
4F 0.001F 40F 0.01F
7
100F 0.1F
(5% reading + 5 digits)
Frequency 5.999Hz 0.001Hz
(1.5% reading + 1 digits)
59.99Hz 0.01Hz
599.9Hz 0.1Hz
(1.2% reading + 3 digits)
5.999kHz 0.001kHz
59.99kHz 0.01kHz
599.9kHz 0.1kHz
5.999MHz 0.001MHz
(1.5% reading + 4 digits)
9.999MHz 0.001MHz
Sensitivity: 0.5V rms <500kHz; 3V rms >500kHz
Duty Cycle
0.1 to
99.9%
0.1%
(1.2% reading + 2 digits)
Pulse width: 100µs to 100ms, Frequency: 5Hz to 150kHz
NOTE: Accuracy is stated at 65oF to 83oF (18oC to 28oC) and
less than 75% RH.
Note: Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measure circuit.(+ digits) – This is the accuracy of the A/D converter.
Enclosure Double molded, waterproof (IP67) Shock (Drop Test) 6.5 feet (2 meters) Diode Test Test current of 0.3mA typical, open
circuit voltage 1.5V DC typical
Continuity Check Audible signal will sound if the
resistance is less than 100 (approx.), test current <0.3mA
Input Impedance 7.8MΩ AC Response Average responding ACV Bandwidth 50Hz to 400Hz Crest Factor ≤3 at full scale up to 500V,
decreasing linearly to ≤1.5 at
1000V
Display 4,000 count backlit liquid crystal Overrange indication OL”isdisplayed Auto Power Off 30 minutes (approximately)