General Safety Instructions
Controls and Features
Manifold Accessories
Quick Start Instructions (For Use With BLACKMAX Mobile APP)
BLACKMAX App Home Screen
Testing: Pressure or Vaccum Hold
Data Logging
Help Screen
Quick Start Instructions (Wireless Manifold without APP)
Initial Operation
Setting To Zero Pressure
Testing: Pressure or Vaccum Hold
Programming/Setting Units
Setting APO On or Off
Connecting to an A/C system
Recovering Refrigerant
Evacuating Procedure
Charging Procedure - Low Side
Charging Procedure - High Side
Appendix A (Specifications: Manifold)
Appendix B (Specifications: MDXVG Vacuum Gauge Accessory)
Appendix C (Specifications: MDXBK Charger Accessory)
Appendix D (Battery Selection and Replacement)
Appendix E (Cleaning Sensor in MDXVG Vacuum Gauge)
Appendix F (Limited Warranty)
Appendix G (Refrigerants Stored in MD50W and MD100W series)
Appendix H (Error Indicator Screens)
New Wireless BLACKMAX digital manifolds can be used as conventional, stand-alone
manifolds or can also be used with a free APP allowing a mobile device (running on
iOS or Android platforms) to communicate directly with and control select functions
on these models remotely. The resulting APP provides the HVAC/R technician with the
convenience of running and monitoring a BLACKMAX digital manifold (from up to 300
feet away), while also providing additional capabilities not found in non-wireless versions
(such as comprehensive views of system measurements, graphical depictions of test
functions, data logging, and more). If your mobile device is NOT available, your wireless
manifold may be operated as a stand alone device using the yellow keypad.
Please read, follow and understand the contents of this entire manual, with special attention given
to Danger, Warning and Caution statements.
FOR USE BY PROFESSIONALLY TRAINED AND CERTIFIED OPERATORS ONLY.
MOST STATES, COUNTRIES, ETC., MAY REQUIRE THE USER TO BE LICENSED.
PLEASE CHECK WITH YOUR LOCAL GOVERNMENT AGENCY.
DANGER: EXPLOSION/RUPTURE RISK!!! Use extreme caution when working with
flammable refrigerants. DO NOT use manifold with unregulated nitrogen,
carbon dioxide or other very high pressure gasses. NEVER USE MANIFOLD WITH OXYGEN. Unregulated gasses can cause components in a refrigeration
system to rupture. Severe Injury or Death can occur.
DANGER: ELECTRICAL SHOCK HAZARD: Always disconnect power source when servicing
equipment.
WARNING: DO NOT APPLY MORE THAN 900 PSI TO ANY PORT ON MANIFOLD.
FOLLOW ALL EQUIPMENT MANUFACTURER’S TEST PROCEDURES WHEN
SERVICING THOSE SPECIFIC TYPES OF EQUIPMENT.
WARNING:
WARNING: Avoid breathing refrigerant vapors and lubricant vapor or mist. Breathing high
NOTE: It may be unlawful to discharge or exhaust refrigerants into the
Remove all hoses with care as they may contain liquid refrigerant under pressure.
Contact with refrigerant may cause frostbite or other related injuries. Wear proper
personal protective equipment such as safety goggles and gloves. When disconnecting
any hose, please use extreme caution.
concentration levels may cause heart arrhythmia, loss of consciousness, or even cause
suffocation. Exposure may irritate eyes, nose, throat and skin. Please read manufacturer’s
Material Safety Data Sheet for further safety information on refrigerants and lubricants.
atmosphere. CPS strongly recommends the practice of recovery or
recycling. Check your local government agencies for refrigerant handling
regulations.
Move screen Up/Down then select a new
active refrigerant.
become the default (but not affect FAVORITE FIVE.
Changing FAVORITE FIVE refrigerants permanently
must be done using the manifold. See page 21 for
more information)
3. You may navigate to other screens while this test runs
gray line is
starting point
Final graph will show initial, current and final
pressures.
Vacuum Hold Test
Normally, systems with vacuum levels within 250-500 microns are deemed a good seal and
are ready for charging.
1. Connect BLACKMAX #MDXVG Vacuum Gauge Accessory to Digital Manifold & initiate
vacuum to enable test. (Note: See Appendix E for cleaning the sensor).
If your mobile device is NOT available, Wireless BLACKMAX Digital Manifolds may also
be operated as stand-alone devices. Simply follow these instructions.
QUICK START INSTRUCTIONS (WIRELESS DIGITAL MANIFOLD)
1. Install 4 AA Alkaline (included) or rechargeable batteries into back of manifold.
2. Press and hold [ON/OFF] (green button) to turn manifold ON.
3. WELCOME SCREEN appears briefly (displays current version of software).
4. “LOC” and “SEL” will appear on the display.
5. Intermittently press [CLR] to scroll among regions:
• AUSTRALIA (Australia, Asian Pacific)
• EUROPE (All of continental Europe, plus Great Britain, Russia)
• USA (Includes USA, Canada, Caribbean, Central or South America)
6. Press [SET] to select desired region.
7. To change the region selected:
A. Press and hold [SET], then [ON/OFF] until display appears, then release BOTH keys.
B. This will take you back to step 5 above. Select another region.
8. The next screen will be the “MAIN” screen [FIG.1] showing default status for the following:
• Battery charge status/APO (On or Off)
• High and low side working pressures
• High and low side temperatures
• Current working refrigerant
Battery Level
Indicator,
(and APO Status)
APO
Psig Psig
FIG.1
R
Current
Working
Refrigerant
High and Low
Side Working
Pressures
°F
High and Low
Side Working
Temperatures
T LOW
°F
T HIGH
9. Review and adhere to all safety instructions.
10. Connect MD50W or MD100W series to an AC system.
1. To Select A Refrigerant Other Than the One Shown- From Main Screen [FIG.1], press
[REF] to scroll through a list of 5 common factory programmed refrigerants.
Note: To change default refrigerants, go to page 20.
Note: Thermistor sensors must be connected to get a T Low and / or T High temperature reading.
2. To View Vapor (VSAT) and Liquid (LSAT) Saturation Temperatures, press [SH/SC]
FIG.2
R
APO
FIG.1
R
Refrigerants
APO
Psig Psig
°F
T LOW
T HIGH
°F
VSAT
Psig Psig
°F
VSAT
°F
LSAT
LSAT
3. Note: In this mode, and to prevent errors, refrigerant cannot be changed using [REF]
key. Press [CLR] to return to the Main Screen [FIG.1] from this view.
4. To View Superheat (SH) and Sub-cool (SC) Temperatures- Press [SH/SC] again.
5. Note: In this mode, and to
prevent errors, refrigerant
cannot be changed using the
APO
FIG.3
R
[REF] key. Press [CLR] to return
to Main Screen [FIG.1] from
this view.
Thermistor Sensors - The
Super
Heat
Psig Psig
°F
SH
Sub
°F
SC
Cool
enclosed clamps can be used to
take temperatures of pipes or
other round surfaces within its
jaw range (1/8” to 1-1/8”). This greatly improves the calculation of Superheat and Subcool
temperatures, as the sensor is virtually isolated from ambient conditions, allowing
sensors to read true pipe temperatures while attached.
Operation - Plug one lead into the LEFT
(Blue) socket and one lead into the RIGHT
BlueRed
(Red) socket. Apply Thermistor Sensors to
system tubing and read the resulting display
temperatures.
6. To Display Difference Between High and
Low Side Temperatures- Press [SH/SC]
again. Press [CLR] to return to Main Screen [FIG.1].
the atmosphere before every working session.
If display reads other than zero in the selected
APO
FIG.4
pressure units, zero can be re-set by pressing
and holding [CLR] at main screen. The Pressure
Zero Screen [FIG.4] appears and will then
return user to Main Screen [FIG.1], where
both Pressure Displays should now read zero.
This operation should be performed especially
if manifold has been dropped or pressurized
beyond 800 PSIG (55 bar) full scale.
PRESSURE HOLD OR VACUUM HOLD TESTING
The BLACKMAX MD100W and MD50W series manifold can perform Pressure Hold
and Vacuum Hold tests. Test pressures and durations are fully settable by the user. Any test can be aborted at any time by pressing [CLR].
PRESSURE HOLD TEST
A pressure hold test should be used to determine if a large leak exists (not to determine
if no leak or small leaks exist). Electronic Leak Detectors are the only instrument to
determine leak rates less than 1 oz./year. In practice there should be no pressure hold
detected if the system is deemed leak proof enough to add refrigerant. Large leaks
should be found before adding refrigerant. Low cost tracer gas mixtures such as 95%
N2 and 5% H2 are popular and can be sensed by a CPS E_MOS leak detector.
1. At Main Screen [FIG.1] press [TEST]. The Test Pressure Set Screen will appear
[FIG.5]. [Note: The MDXVG Vacuum Gauge accessory should not be plugged in
at this time].
2. Press [SET] to advance to Test Time (Pressure) Set Screen [FIG.6].
3. Use [UP/DOWN ARROWS] to adjust test duration in 5 minute increments to
desired setting (Note: Minimum time is pre-set at 5 minutes. Maximum time is
95 minutes).
4. Press [SET] to store Test Duration. Display will advance to Pressure Set Screen [FIG.8].
FIG.7
APO
FIG.8
Psig
hh : mm
Input Test
Pressure
Original Set
System Pressure
Value
Psig Psig
Final System
Pressure
Value
5. Connect manifold Low and High Side Service hoses to the system to be tested.
Pressurize the system. Regulated Nitrogen or Nitrogen/Hydrogen trace gas is
recommended.
6. The pressure in the system will be displayed in the UPPER LEFT.
7. Press [SET] to start test. Starting test pressure is indicated in upper left display
[FIG.7]. Current system pressure is shown in upper right display. Timer (lower left
display) decrements and when at zero, a bell will sound indicating test completion
and Start Vs. Final Pressures will be displayed [FIG.8] to determine if system
leaks.
VACUUM HOLD TEST
Normally, systems with vacuum levels within 250-500 microns are deemed as having a
good seal and are ready for charging. As with any recharged system, a good electronic leak
detector should be used to test for leaks after the charge has been completed. MD50W
or MD100W manifold series will function as accurate vacrometers when used with the
MDXVG Remote Vacuum Accessory (page 6).
1. Plug the MDXVG Vacuum Accessory into the upper right side of the manifold [FIG.9].
3. Assemble the BLACKMAX
MDXVG vacuum gauge
accessory to the Suction
Port on an AC system as
shown [FIG.11].
Suction
Suction
port
port
FIG.11
Discharge
Discharge
port
port
Vacuum Gauge
Accessory
4. In this mode, the manifold will show current vacuum
in the system. The display begins to show vacuum
levels at 25000 microns or less and is accurate to
AVT45 Brass
Tee Connector
APO
FIG.12
+/- 10 microns. Continue to run vacuum pump until
the desired vacuum level is reached. Once reached,
proceed to Vacuum Hold Test.
5. Vacuum Hold Test - To perform this test, press
Vac
[CLR] to go to the Main Screen [FIG.1] Vacuum
Gauge Accessory should still be plugged into
manifold.
6. At Main Screen [FIG.1] press [TEST]. The Test Vacuum Set Screen appears [FIG.12]
7. Press [SET] to advance to Vacuum Hold Test –
APO
FIG.13
Time Set Screen [FIG.13].
8. Use [UP/DOWN ARROWS] to adjust test
duration in 5 minutes increments (minimum
time is pre-programmed at 5 minutes).
A) Press [SET] to store Vacuum Test duration
time. Vacuum Hold Test Screen [FIG.14]
Vac
hh : mm
Set test
duration (adjust
in 5 minute
increments
will now be displayed.
NOTE: The Vacuum Reading in the upper Left display is still active. The Reading in the
upper Right display is not active unit steps 2 and 3 are completed.
B) Continue to Evacuate the system to the desired level which is indicated in the
upper left display once the desired vacuum level is reached:
a. On the MD100W, close the vacuum valve
b. On the MD50W, close the High and Low side valves.
C) Press [SET] to start test. Starting test vacuum level is indicated in upper left
display [FIG.15]. Current system vacuum level is shown in upper right of
display and timer is shown in lower left. The timer decrements and when zero, a
bells sounds and test ends. At end of test, Start vs. Final Vacuum levels will be
displayed to enable determination of any vacuum loss over a set period of time.
hh : mm
Inactive
FIG.15
Desired Test
Level
Timer counts
backwards
Current System
Vacuum Level
D) Press [CLR] to return to Main Screen [FIG.1].
NOTE: If it is required to rerun the vacuum pump after STEP 3, disconnect and reconnect the
MDXVG to reset it in the vacrometer mode.
PROGRAMMING/SETTING UNITS
According to your geographic location, MD50W and MD100W manifold series are factory programmed with 5 common refrigerants in those regions (see chart below), and units of measure (Temperature, Pressure and Vacuum) for specific regions.See chart on
next page for details. Any of these units can be changed via the Program Mode.
There are 2 major options. In each, a blinking symbol represents a choice the user can
make by pressing [SET].
1. PrO (Programming)
a. Change Refrigerants: Change any or all of 5 common refrigerants accessible at
Main Screen.
b. Change Temperature Units: Change °F or °C.
c. Change Pressure Units: Change Psig, kPa, Mpa, bar and KgCm2 pressure units.
d. Change Vacuum Units: Change Microns vs. Hg vacuum units.
2. APO (Set Automatic Power OFF) - Manifolds come factory pre-set to ON.
Pre-Programmed Refrigerants
The following 5 common refrigerants have been programmed for easy access from the
The recommended way to change refrigerants involves making a new list of 5
refrigerants, and change all five of them in the same session. Example: If only R-22
needs to be changed to R-1234yf (and if using USA defaults), program the following
items: R-410A, R-1234yf, R-404A, R-134a and R-407A.
1. At Main Screen [FIG.1] press and hold [MODE] until “PrO” (Program) Screen
[FIG.16] appears. “PrO” will blink.
2. Press and hold [SET] to enter the Refrigerant Selection Screen [FIG.17].
APO
FIG.16FIG.17
APO
R
R11 is initial
default
refrigerant
Counter
3. The Initial Refrigerant (blinking in upper right) should be R-11 and Counter (Lower
Right) should start at 000.
4. Use [UP/DOWN ARROWS] to scroll through 93 refrigerants. Press and hold either
arrow to scroll through refrigerants progressively faster. Release to stop scrolling.
Last refrigerant displayed will blink.
1. At main screen [FIG.1] press and hold [MODE] until “PrO” screen [FIG.16] appears. “PrO” blinks.
2. Press [SET] to enter “ PrO” Screen. The Refrigerant Selection Screen appears [FIG.17].
3. Press [SET] to bypass Refrigerants Screen. The Temperature Units Screen (blinking) appears [FIG.18].
4. Press [MODE] to advance to the Pressure Units Screen [FIG.19].
5. Press [SET] to advance to Vacuum Units Screen (blinking) [FIG.20].
6. Press [SET] to select this menu.
Current units will blink.
7. Use [UP/DOWN ARROWS] to select
either microns (mic) or inches of
mercury (inHg).
8. Press [SET] to select desired units
APO
FIG.20
mic inHg
Vacuum
Units
(“mic” or
“inHg”)
and go to next menu (Units Summary
Screen), or
9. To return to the Main Screen [FIG.1]
press and hold [CLR].
ENTER PROGRAMMING CHANGES (MAKE CHANGES CURRENT)
1. The next menu item after selecting Vacuum Units is the Units Summary Screen [FIG.21].
Blinking symbols represents units chosen in
previous steps.
2. Press [SET] to select units shown on this screen-
making them the current working units.
3. Or after pressing [SET], use [UP/DOWN ARROWS] to display previous units. These can
be restored by pressing SET Menu. In either
case, units become current units and manifold
returns to Main Screen [FIG.1]
APO
Psig
°F
FIG.21
Blinking symbols indicate
mic
Pressure, Temperature and
Vacuum Units chosen in
previous steps
SETTING APO (AUTOMATIC POWER) ON OR OFF
When APO is ON and there has been no user activity (usually
10-15 minutes), manifolds turn OFF to conserve battery power.
About 1 minute before turning OFF, an alarm will sound to
signify that override is possible by pressing any key, except the
[ON/OFF] key. This will re-start the APO timer. APO is suspended
when performing either the vacuum or pressure hold test or
when the unit is used as a vacuum gauge.
When APO is OFF, the user must turn a manifold off manually.
1. From Main Screen [FIG.1], press and hold [MODE]. The
PrO icon will blink [FIG.16].
2. Press DOWN ARROW to go to (blinking) Set APO screen.
[FIG.22].
3. At SET APO Screen briefly press [SET]. “APO On” will
appear and blink [FIG.23].
4. Use [UP/DOWN ARROW] to change display from APO On
[FIG.23] to APO OFF as desired.
5. Press [SET] to select desired APO state and return to Main
Screen [FIG.1]
22
APO
FIG.22
FIG.23
APO
From either
screen,
select either
Automatic
Power ON
or OFF
1.) The following instructions are to add charge to a pressurized system, known as “topping off”.
2.) Connect Low (A) and High (B) Service hoses to
service ports (E) and (D). All valves should be closed.
3.) Connect a hose from the open port on tee (C) to
valve fitting on supply tank.
4.) Open supply tank valve. For blends, turn tank upside down or make sure liquid refrigerant is being
charged.
5.) Push valve core on tee (C) until refrigerant comes
out. This will purge the hose from the supply tank.
6.) Turn refrigerant system on. Monitor pressure gauges. Use superheat and subcooling features to help
determine when charge amount is optimal.
7.) Slightly open Low Side Valve (A) to allow liquid refrigerant to flow into low side of equipment. Caution
should be used when charging liquid refrigerant into low side of a refrigeration system.
8.) Close Low Side valve (A) when desired amount of refrigerant has been added. Note the use of a scale
is required.
9.) Let equipment temperatures and pressures stabilize. Check superheat and/or subcooling levels.
Repeat charge if necessary.
10.) Once done, close off supply tank valve and disconnect all hoses. Use Caution as all hose connections
are under pressure.
Once charge is complete, an electronic refrigerant leak detector should be used to confirm leaks do not exist.
24
2.) Connect Low (1) and High (2) Service hoses
to service ports (L) and (H)
3.) Connect a hose from port (3) to refrigerant
supply tank’s valve.
5.) Open valve (C). With hose disconnect from
port (4), slightly open valve (D) until
refrigerant comes out. This will purge the hose
from the supply tank.
1.) A scale should be used to weigh in the charge. Place tank on scale so liquid refrigerant is supplied.
2.) Open Supply tank valve to fill manifold with liquid refrigerant.
3.) Set scale to record charge amount.
4.) Open High Side Valve (B). Monitor weight reading.
5.) Once desired charge amount has been reached, close High Side Valve (B) and supply tank valve.
6.) Start refrigeration system. Open Low Side valve (A) to allow all remaining refrigerant in manifold and
service hoses to charge into system.
7.) Close Low Side Valve (A).7.) Close Low Side Valve (A) and Valve (C).
8.) Monitor Low and High Side pressures. Superheat and subcooling features of manifold can be used at this time.
9.) Disconnect service hoses from equipment. Use Caution as all hoses are under pressure.
Once charge is complete, an electronic refrigerant leak detector should be used to confirm leaks do not exist.
APPENDIX A (MANIFOLD SPECIFICATIONS)
High side pressure range -14.7 to 800.0 PSIG (-1.0 to 55 bar)
Low side pressure range -14.7 to 800.0 PSIG (-1.0 to 55 bar)
Pressure reading accuracy +/- 1% of reading or +/- 0.5 PSIG (+/- 0.05 bar)
Pressure reading resolution 0.5 PSIG (0.05 bar)
High and low side burst pressure 1500.0 PSIG (103.4 bar)
High side temperature range -58.0 °F to 302 °F (-50 °C to 150°C)
Low side temperature range -58.0 °F to 302 °F (-50 °C to 150°C)
Temperature reading accuracy +/- 0.36 °F (+/- 0.2°C)
Vacuum indication accuracy +/- 10 microns (requires MDXVG accessory)
SH, SC and saturation temperature accuracy +/- 0.5°F (+/- 0.3°C)
Instrument working temperature -4.0 °F to 122°F (-20°C to 50°C)
Instrument working RH range 0% to 100% non-condensing
Total number of stored refrigerants 93
APO (Automatic Power Off) 10 minutes. Can be disabled by user.
MD100W size (L x H x W) 8.1” x 3.5” x 7.0” (20.5 cm x 8.9 cm x 17.8cm)
MD50W size (L x H x W) 8.1” x 2.5” x 7.0” (20.5 cm x 6.4 cm x 17.8cm)
MD100W weight 3.3 lb. (1.5 kg)
MD50W weight 3.0 lb. (1.4 kg)
Display Life (Using 4
industrial alkaline AA
batteries (included)
Display Life Using 4
Rechargeable NiMH AA batteries
(not included):
Backlight full ON: 25+ hrs. continuous duty
Backlight full OFF:60+ hrs. continuous duty
Backlight full ON: 20+ hrs. continuous duty from full charge
Backlight full OFF:50+ hrs. continuous duty from full charge
APPENDIX B (MDXVG VACUUM GAUGE ACCESSORY SPECIFICATIONS)
Sensor Type
Display Pressure Range
Accuracy+/- 10% of reading or +/- 10 microns, whichever is greater
Operating Temperature Range
Operating and Storage Humidity 0 to 95%, non-condensing
Storage Temperature-40 to 85°C (-40 to 185°F)
Self - heated thermistor bridge with integral temperature compensation from
0 to 50 °C (32 to 122 °F).
0 to 99 microns 1 micron resolution
100 to 990 microns 10 micron resolution
1000 to 9900 microns20 micron resolution
10000 to 25000 microns 50 micron resolution
above 25000 microns
Compensated (accuracy as stated): 0°C TO 50°C (32°F TO 122°F)
Non-compensated (add +/- 0.5% of reading error for every °C outside
compensated range): -10 to 0°C (14 to 32°F)
AUTO-RANGING SCALES:
Will display low side pressure reading, in
selected unit.
25
APPENDIX C (MDXBK CHARGER ACCESSORY SPECIFICATIONS)
Operating Temperature Range32°F to 122°F (0°C to 50°C)
Operating and Storage Humidity 0-95%
Storage Temperature-40°F to 185°F (-40 to 85°C)
Power Source100-240VAC input, 9VDC 2.0 amp maximum output
Battery Type
Rechargeable Battery Cycle Life
AA size, NiMH rechargeable, approximate battery life is 20,000 to
50,000 hours*
1000 recharges *based on usage with or without the LCD backlight
energized.
APPENDIX D (BATTERY SELECTION AND REPLACEMENT)
For optimum life and economic value, CPS Products, Inc. recommends that MD50W and MD100W
manifold series be powered by rechargeable NiMH (nickel metal hydride) batteries. CPS Products
offers an optional “Charger Kit” (model MDXBK) that comes with a power adapter and 4 NiMH
batteries. Do not connect the battery charger if there are no rechargeable batteries in the manifolds.
NOTE: Alkaline batteries should not, and cannot be charged in the BLACKMAX MD50W or
MD100W Series. Battery charge status is shown by the battery indicator symbol in the upper left [FIG.1] Main Screen and all other screens. NOTE: A slight increase in temperature near
the charging socket is normal. Remove battery door and install 4 AA size batteries, observing
battery polarity. When using rechargeable
batteries for the first time, turn manifold off
and connect one end of charger to Manifold
9V DC plug [FIG.26] and the other end to an
appropriate wall outlet socket. Charge NiMH
batteries for 2 hours. The Battery Symbol on
the display will blink during charging. MD50W
and MD100W manifold series can be used
while charging the batteries, but charging time
will be considerably extended. The preferred
way to use the Rapid Charging Feature is to
charge while the manifold is OFF.
FIG.26
Connection
for 9V DC
Charger
APPENDIX E (CLEANING SENSOR IN MDXVG)
Sensor may become contaminated with dust, oil, etc. from pulling vacuums.
1. Never use a cotton swab or similar object to clean the sensor. This may damage the
sensor.
2. Open all knobs, valves, and cap all the ports except for the VAC port.
3. Invert the MDXVG and add about 1 tablespoon of 100% denatured alcohol into the VAC port (see page 2) so as to flush out contaminants.
4. Gently shake the unit to produce a swirling effect of the alcohol inside the housing.
5. Turn right side up, open a port to pour out the rubbing alcohol.
6. Repeat the above procedure until the alcohol appears to be clean and clear of contaminants.
7. Check the gasket located inside the VAC port. If worn or deformed, replace with a new CPS
gasket (PN: HXG)
8. Open all ports to allow sensors to air dry (usually about 1 hour). DO NOT BLOW DRY.
BLACKMAX digital manifolds are covered by a CPS Products, Inc. LIMITED WARRANTY.
CPS Products, Inc. warrants, to the original purchaser, that its BLACKMAX line of digital
manifolds and accessories are free from defects in assembly, materials and workmanship for
1 year from the date of purchase, or 15 months from the date of manufacture, whichever is sooner. If any
BLACKMAX product is believed to be defective it must be returned to CPS Products. If the product is found to be
defective, CPS will repair or replace the item(s), at our option, at no charge.
This warranty does not apply to products or accessories that have been altered, misused, submerged, exposed
to extreme heat or cold, lightning strikes, or electrical voltage exceeding published product specifications. This
warranty also does not apply to products that are simply in need of routine field service maintenance.
To assure your BLACKMAX purchase is on file with CPS Products, please fill in a
Product Registration form and mail to CPS within 15 days from the date of purchase, along with a copy of the
original sales receipt. Return of the registration form is not required to be covered by the Limited Product Warranty.
To process a warranty claim, please contact CPS USA at 1-800-277-3808 or CPS Products
N.V. (Belgium) at (32) 3 281 30 40; or CPS Australia at 08 8340 7055, within 15 days of
discovering defects in assembly, materials, or workmanship. CPS will provide a return
authorization number (RGA #). Do not attempt to repair BLACKMAX products. CPS has the sole right to determine
if a claim is valid and if compensation is justified.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER EXPRESS WARRANTIES. TO THE EXTENT ALLOWED
BY LAW, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
ARE LIMITED TO THE DURATION OF THE LIMITED PRODUCT WARRANTY. IN NO CASE SHALL CPS
PRODUCTS INC. HAVE ANY LIABILITY FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY
KIND, INCLUDING BUT NOT LIMITED TO: THE LOSS OF THE USE OF THE PRODUCT OR ANY ASSOCIATED
EQUIPMENT, COST OF ANY SUBSTITUTED EQUIPMENT, FACILITIES OR SERVICES, DOWNTIME, THE
CLAIMS OF THIRD PARTIES, DAMAGE TO PROPERTY OR PERSONAL INJURY. SOME JURISDICTIONS DO
NOT ALLOW LIMITS ON WARRANTIES OR ON REMEDIES FOR BREACH IN CERTAIN TRANSACTIONS. IN
SUCH JURISDICTIONS, THE LIMITS IN THIS PARAGRAPH MAY NOT APPLY.
THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS, AND YOU MAY ALSO HAVE OTHER RIGHTS WHICH
VARY FROM JURISDICTION TO JURISDICTION.
APPENDIX G (REFRIGERANTS STORED IN MEMORY)
MD50W and MD100W manifold series are pre-loaded with 93 of the most common refrigerants.
If certain malfunctions occur, MD50W and MD100W series will display error messages:
oP -Appears in display areas reserved for temperature readings when thermistor sensor
connected to that port is open circuited or not connected. Replace sensor with CPS model
number TMX3C.
SP –Appears in display areas reserved for temperature readings when thermistor sensor
connected to that port is short circuited. Replace sensor with CPS model number TMX3C.
Err (Upper Left Display) - Appears in upper left display when T function is selected and
one or both thermistor sensors are defective as indicated by oP or SP indicators mentioned
above.
Err (Lower Left Display) -Appears in display areas reserved for temperature readings when
user tries to read saturation temperatures and calibration of pressure channels is invalid.
This defect requires manifold be returned to factory.