CPS Products MD100WHE, MD100WVHE, MDXCP, MD100W, MD50WHE User manual

...
For use with free BLACKMAX® app featuring CPS®Link
US Patent No. 9,043,161. For Android or iOS mobile devices.
OWNER’S MANUAL (English)
(Software Version 1.1 or Higher with CPS® Link™ App)
2 Valve Series: MD50W, MD50WHE, MD50WVHE
4 Valve Series: MD100W, MD100WHE, MD100WVHE
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CONTENTS
Overview ...................................................................................................................................2
General Safety Instructions Controls and Features Manifold Accessories Quick Start Instructions (For Use With BLACKMAX Mobile APP) BLACKMAX App Home Screen Testing: Pressure or Vaccum Hold Data Logging Help Screen Quick Start Instructions (Wireless Manifold without APP) Initial Operation Setting To Zero Pressure Testing: Pressure or Vaccum Hold Programming/Setting Units Setting APO On or Off Connecting to an A/C system Recovering Refrigerant Evacuating Procedure Charging Procedure - Low Side Charging Procedure - High Side Appendix A (Specifications: Manifold) Appendix B (Specifications: MDXVG Vacuum Gauge Accessory) Appendix C (Specifications: MDXBK Charger Accessory) Appendix D (Battery Selection and Replacement) Appendix E (Cleaning Sensor in MDXVG Vacuum Gauge) Appendix F (Limited Warranty) Appendix G (Refrigerants Stored in MD50W and MD100W series) Appendix H (Error Indicator Screens)
...........................................................................................................................12
.............................................................................................................................13
.......................................................................................................................15
.......................................................................................................3
........................................................................................................... 4-5
................................................................................................................6
.............................................7
............................................................................................ 8-10
..........................................................................................11
.....................................................14
.........................................................................................................16
.................................................................................... 16-19
............................................................................................. 19-22
.............................................................................................................22
..................................................................................................23
...........................................................................................................23
.............................................................................................................24
..............................................................................................24
..............................................................................................25
....................................................................................25
...........................................25
.......................................................26
...................................................................26
......................................................26
...............................................................................................27
........................................27
.....................................................................................28
OVERVIEW
New Wireless BLACKMAX digital manifolds can be used as conventional, stand-alone manifolds or can also be used with a free APP allowing a mobile device (running on iOS or Android platforms) to communicate directly with and control select functions on these models remotely. The resulting APP provides the HVAC/R technician with the convenience of running and monitoring a BLACKMAX digital manifold (from up to 300 feet away), while also providing additional capabilities not found in non-wireless versions (such as comprehensive views of system measurements, graphical depictions of test functions, data logging, and more). If your mobile device is NOT available, your wireless manifold may be operated as a stand alone device using the yellow keypad.
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
GENERAL SAFETY INSTRUCTIONS
Please read, follow and understand the contents of this entire manual, with special attention given to Danger, Warning and Caution statements.
FOR USE BY PROFESSIONALLY TRAINED AND CERTIFIED OPERATORS ONLY.
MOST STATES, COUNTRIES, ETC., MAY REQUIRE THE USER TO BE LICENSED. PLEASE CHECK WITH YOUR LOCAL GOVERNMENT AGENCY.
DANGER: EXPLOSION/RUPTURE RISK!!! Use extreme caution when working with
flammable refrigerants. DO NOT use manifold with unregulated nitrogen, carbon dioxide or other very high pressure gasses. NEVER USE MANIFOLD WITH OXYGEN. Unregulated gasses can cause components in a refrigeration system to rupture. Severe Injury or Death can occur.
DANGER: ELECTRICAL SHOCK HAZARD: Always disconnect power source when servicing
equipment.
WARNING: DO NOT APPLY MORE THAN 900 PSI TO ANY PORT ON MANIFOLD.
FOLLOW ALL EQUIPMENT MANUFACTURER’S TEST PROCEDURES WHEN SERVICING THOSE SPECIFIC TYPES OF EQUIPMENT.
WARNING:
WARNING: Avoid breathing refrigerant vapors and lubricant vapor or mist. Breathing high
NOTE: It may be unlawful to discharge or exhaust refrigerants into the
Remove all hoses with care as they may contain liquid refrigerant under pressure.
Contact with refrigerant may cause frostbite or other related injuries. Wear proper personal protective equipment such as safety goggles and gloves. When disconnecting any hose, please use extreme caution.
concentration levels may cause heart arrhythmia, loss of consciousness, or even cause suffocation. Exposure may irritate eyes, nose, throat and skin. Please read manufacturer’s Material Safety Data Sheet for further safety information on refrigerants and lubricants.
atmosphere. CPS strongly recommends the practice of recovery or recycling. Check your local government agencies for refrigerant handling regulations.
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CONTROLS AND FEATURES - BLACKMAX WIRELESS DIGITAL MANIFOLD
1. [ON/OFF] - Press and hold to turn On or Off
2. [SET] - For entering programed values
3. [REF] (Refrigerant) - Scrolls through 5 pre-selected “favorite” or commonly used
refrigerants
4. [MODE] - Programming (change English vs. Metric units; Turn Auto-Power OFF)
5. [SH/SC] (Superheat, Subcool)- Controls display of Superheat/ Subcool, Saturation
temperatures, and Delta T.
6. [TEST] - Press to obtain either “Pressure Hold” or “Vacuum Hold” tests.
7. [CLR] (Clear)- Sets pressure to zero point . Also ends any test in process.
8. [BACKLIGHT] - Press once to turn light On or Off
9. HIGH SIDE VALVE- Controls flow to right side ¼” SAE male port
10. REFRIGERANT VALVE (MD100 Series Only)- Controls flow through RIGHT MIDDLE
¼” SAE Male Port.
11. SIGHTGLASS
12. VACUUM VALVE (MD100 Series Only)- Controls flow through LEFT MIDDLE 3/8”
SAE PORT.
13. LOW SIDE VALVE- Controls flow to LEFT SIDE ¼” SAE Port.
14. CENTER PORT (MD50W Series Only)- A ¼” SAE Male “T” port used as common
connection to vacuum pump, recovery unit or refrigerant supply tank. The addition
of a ¼” SAE Male “T” fitting with valve core allows for a dual hook up between
vacuum pump and refrigerant supply tank.
15. 9V DC Charger Port - Connection for Accessory MDXBK
16. Vacuum Gauge Accessory Port - Connection for accessory MDXVG
17. Thermistor Sensors - Ideal for taking Superheat and Subcool true temperatures on
round pipes, at any angle
18. Swivel Hook -
19. Battery Compartment - Uses 4 AA batteries
20. Hose Hangers - For securing ends of unused hoses
21. Connection Points- For (2) thermistor sensors
22. Storage Posts - For thermistor sensors
23. SAE Connection Ports - For SAE Hoses
- Monitor refrigerant flow
Heavy Duty 360°
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CONTROLS AND FEATURES - BLACKMAX WIRELESS DIGITAL MANIFOLD
21.) Connection Points For Thermistor Sensors
4 Valve Manifold (Front)
4.) Mode
3.) Refrigerant
2.) Set
1.) On/Off
5.) Superheat/ Subcool
6.) Test
7.) Clear
8.) Backlight
13.) Low Side Valve
4 Valve Manifold (Back)
16.) Vacuum Gauge Accessory Port
15.) 9V DC Charger Port
Valve
11.) Sightglass12.) Vacuum
22.) Storage Posts
9.) High Side Valve
10.) Refrigerant Valve
18.) Swivel Hook
19.) Battery Compartment
20.) Hose Hangers
22.) SAE Hose Connection Ports
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
ACCESSORIES
Vacuum Gauge Accessory Kit MDXVG
AVT45 Brass
Tee Connector
(Included with MDXVG Kit)
Universal Charger MDXBK
4 (AA) Rechargeable NiMH Batteries
Euro Plug
Australia/New Zealand/China Plug
Thermistor Sensor MDXCP Soft Case MDXSC
Singapore/ Hong Kong Plug
Included (2)
U.S. Plug
Vacuum Gauge Accessory
(only available with MDXVG Kit)
5 Ft. CordUK/Ireland/
Optional MDXSC Soft Case
• Padded sides
MDXCP Thermistor Sensor with 6 Ft. lead for taking Superheat and Subcool true temperatures on pipes
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
• Rear storage pocket with zipper
• Protective skirt with drawstring
• Holds 7” tablet, instruction manual, thermistor sensors, etc.
QUICK START INSTRUCTIONS (For Use With BLACKMAX Mobile APP)
If you plan to use the wireless manifold without the free app, go to page 14 for quick start instructions - manifold only.
1. Download and install the FREE APP for your wireless BLACKMAX Digital Manifold;
• For iOS mobile devices - Navigate to the iTunes APP Store and
download the free BLACKMAX Wireless Digital Manifold APP.
• For Android mobile devices - Navigate to Google Play and
download the BLACKMAX Wireless Digital Manifold APP.
2. Install 4 AA Alkaline (included) or rechargeable batteries in back of manifold.
3. Press and hold [ON/OFF] (green button) to turn manifold ON.
4. Open the “BLACKMAX CPS Link” icon from your mobile device.
Indicates
wireless manifold has been
detected
Scroll Up/
Down to
refresh the
scanning
process
5. The application will appear on your mobile device screen in landscape mode.
6. Select one BLACKMAX Wireless Digital Manifold to operate.
7. Review and adhere to all safety instructions.
8. Connect the BLACKMAX manifold to an AC system. (see pages 24-26)
9. You are now ready to work.
Press “>” to go to Home
Screen
Or Swipe LEFT
here to open
Side Screen
(Help Or Exit)
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BLACKMAX APP HOME SCREEN
View 1
Working Temperature(s)
View 2
Touch arrows
to toggle between
Home Screens
(View 1
Or View 2)
Home
Screen
Current Refrigerant
Hold
Vacuum
Hold
Presure
Logging
High SideLow Side
Analog Scale
Digital Scale
Touch icons here to
select any of these
functions
Settings Data
CHANGE OR CONFIRM SETTINGS
Side Menu
OR
Left Swipe
Touch
Settings Icon
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
From Right
Side Of Screen
And Touch
Settings
BLACKMAX APP SETTINGS SCREEN
Settings
Select One
Change Units
Target Superheat
Zero Offset
Open manifold to atmosphere before each use especially if
!
manifold has been dropped or pressurized beyond 800 psig (55 Bar).
Change Language
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CHANGE OR CONFIRM REFRIGERANTS
Move screen Up/Down then select a new active refrigerant.
become the default (but not affect FAVORITE FIVE. Changing FAVORITE FIVE refrigerants permanently must be done using the manifold. See page 21 for more information)
10
Note: Refrigerant selected will
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
TESTING: Pressure Hold or Vacuum Hold
Pressure Hold Test
1. Touch “Start” or “Stop” at any time.
2. Default duration = 5 minutes (Max time = 100 mins)
3. You may navigate to other screens while this test runs
gray line is
starting point
Final graph will show initial, current and final pressures.
Vacuum Hold Test
Normally, systems with vacuum levels within 250-500 microns are deemed a good seal and are ready for charging.
1. Connect BLACKMAX #MDXVG Vacuum Gauge Accessory to Digital Manifold & initiate vacuum to enable test. (Note: See Appendix E for cleaning the sensor).
2. Touch “Start” or “Stop” at any time.
3. Default duration = 5 minutes (Max time = 100 mins)
4. You may navigate to other screens while this test runs
gray line is
starting point
Final graph will show initial, current and final vacuum.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
DATA LOGGING
Mobile Device Logs
1. Shows logs stored in the mobile device’s local storage.
2. For Geolocation, Delete, View or Export logs, touch icons.
All stored files have
option of showing
Geolocation, Delete,
View Or Export
New Log Tab
1. For starting new data logging in the Manifold
2. Select frequency in which the manifold will record values
3. View maximum time remaining
4. Start/Stop logging at any time (or let system run until no more Manifold memory).
Touch
to select
interval
time
Logs stored in
mobile device
Maximum
time
remaining
Touch
icon to
download
a file
Manifold Logs
1. Shows logs stored in manifold
2. Can only erase entire memory
3. Download any file to mobile device
12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
HELP SCREEN
APP HELP
Select for general description of any item;
1. General
2. Home
3. Pressure Hold
4. Vacuum Hold
5. Data Logging
6. Settings
7. Side Menu
Video
Watch instructional videos on how to use this app
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Main Screen
OPERATING THE WIRELESS MANIFOLD WITHOUT THE APP
If your mobile device is NOT available, Wireless BLACKMAX Digital Manifolds may also be operated as stand-alone devices. Simply follow these instructions.
QUICK START INSTRUCTIONS (WIRELESS DIGITAL MANIFOLD)
1. Install 4 AA Alkaline (included) or rechargeable batteries into back of manifold.
2. Press and hold [ON/OFF] (green button) to turn manifold ON.
3. WELCOME SCREEN appears briefly (displays current version of software).
4. “LOC” and “SEL” will appear on the display.
5. Intermittently press [CLR] to scroll among regions:
• AUSTRALIA (Australia, Asian Pacific)
• EUROPE (All of continental Europe, plus Great Britain, Russia)
• USA (Includes USA, Canada, Caribbean, Central or South America)
6. Press [SET] to select desired region.
7. To change the region selected:
A. Press and hold [SET], then [ON/OFF] until display appears, then release BOTH keys. B. This will take you back to step 5 above. Select another region.
8. The next screen will be the “MAIN” screen [FIG.1] showing default status for the following:
Battery charge status/APO (On or Off)
High and low side working pressures
High and low side temperatures
Current working refrigerant
Battery Level
Indicator,
(and APO Status)
APO
Psig Psig
FIG.1
R
Current Working Refrigerant
High and Low Side Working Pressures
°F
High and Low Side Working Temperatures
T LOW
°F
T HIGH
9. Review and adhere to all safety instructions.
10. Connect MD50W or MD100W series to an AC system.
11. You are now ready to work.
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Main Screen
INITIAL OPERATION
Saturation Temperature Screen
Super Heat, Sub Cool Screen
1. To Select A Refrigerant Other Than the One Shown- From Main Screen [FIG.1], press [REF] to scroll through a list of 5 common factory programmed refrigerants.
Note: To change default refrigerants, go to page 20. Note: Thermistor sensors must be connected to get a T Low and / or T High temperature reading.
2. To View Vapor (VSAT) and Liquid (LSAT) Saturation Temperatures, press [SH/SC]
FIG.2
R
APO
FIG.1
R
Refrigerants
APO
Psig Psig
°F
T LOW
T HIGH
°F
VSAT
Psig Psig
°F
VSAT
°F
LSAT
LSAT
3. Note: In this mode, and to prevent errors, refrigerant cannot be changed using [REF]
key. Press [CLR] to return to the Main Screen [FIG.1] from this view.
4. To View Superheat (SH) and Sub-cool (SC) Temperatures- Press [SH/SC] again.
5. Note: In this mode, and to prevent errors, refrigerant cannot be changed using the
APO
FIG.3
R
[REF] key. Press [CLR] to return to Main Screen [FIG.1] from this view.
Thermistor Sensors - The
Super Heat
Psig Psig
°F
SH
Sub
°F
SC
Cool
enclosed clamps can be used to take temperatures of pipes or other round surfaces within its jaw range (1/8” to 1-1/8”). This greatly improves the calculation of Superheat and Subcool temperatures, as the sensor is virtually isolated from ambient conditions, allowing sensors to read true pipe temperatures while attached.
Operation - Plug one lead into the LEFT (Blue) socket and one lead into the RIGHT
Blue Red
(Red) socket. Apply Thermistor Sensors to system tubing and read the resulting display temperatures.
6. To Display Difference Between High and
Low Side Temperatures- Press [SH/SC] again. Press [CLR] to return to Main Screen [FIG.1].
7. To turn manifold OFF, press and hold the
[ON/OFF] until the screen shows “OFF”.
15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
SETTING TO ZERO PRESSURE
Pressure Zero Screen
Test Pressure Set Screen
Test Time Pressure Set Screen
1. It is recommended the manifold be opened to
the atmosphere before every working session. If display reads other than zero in the selected
APO
FIG.4
pressure units, zero can be re-set by pressing and holding [CLR] at main screen. The Pressure Zero Screen [FIG.4] appears and will then return user to Main Screen [FIG.1], where both Pressure Displays should now read zero. This operation should be performed especially if manifold has been dropped or pressurized beyond 800 PSIG (55 bar) full scale.
PRESSURE HOLD OR VACUUM HOLD TESTING
The BLACKMAX MD100W and MD50W series manifold can perform Pressure Hold and Vacuum Hold tests. Test pressures and durations are fully settable by the user. Any test can be aborted at any time by pressing [CLR].
PRESSURE HOLD TEST
A pressure hold test should be used to determine if a large leak exists (not to determine if no leak or small leaks exist). Electronic Leak Detectors are the only instrument to determine leak rates less than 1 oz./year. In practice there should be no pressure hold detected if the system is deemed leak proof enough to add refrigerant. Large leaks should be found before adding refrigerant. Low cost tracer gas mixtures such as 95% N2 and 5% H2 are popular and can be sensed by a CPS E_MOS leak detector.
1. At Main Screen [FIG.1] press [TEST]. The Test Pressure Set Screen will appear
[FIG.5]. [Note: The MDXVG Vacuum Gauge accessory should not be plugged in at this time].
2. Press [SET] to advance to Test Time (Pressure) Set Screen [FIG.6].
FIG.5
APO
16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
APO
hh : mm
FIG.6
Adjust Test Duration In 5 Minute Increments
PRESSURE HOLD TEST (CONTINUED)
Pressure Set Screen
Start Vs. Final Pressures
MDXVG Vacuum Gauge Accessory Connection
Vacuum Display Screen
3. Use [UP/DOWN ARROWS] to adjust test duration in 5 minute increments to
desired setting (Note: Minimum time is pre-set at 5 minutes. Maximum time is 95 minutes).
4. Press [SET] to store Test Duration. Display will advance to Pressure Set Screen [FIG.8].
FIG.7
APO
FIG.8
Psig
hh : mm
Input Test Pressure
Original Set System Pressure Value
Psig Psig
Final System Pressure Value
5. Connect manifold Low and High Side Service hoses to the system to be tested.
Pressurize the system. Regulated Nitrogen or Nitrogen/Hydrogen trace gas is recommended.
6. The pressure in the system will be displayed in the UPPER LEFT.
7. Press [SET] to start test. Starting test pressure is indicated in upper left display [FIG.7]. Current system pressure is shown in upper right display. Timer (lower left display) decrements and when at zero, a bell will sound indicating test completion and Start Vs. Final Pressures will be displayed [FIG.8] to determine if system leaks.
VACUUM HOLD TEST
Normally, systems with vacuum levels within 250-500 microns are deemed as having a good seal and are ready for charging. As with any recharged system, a good electronic leak detector should be used to test for leaks after the charge has been completed. MD50W or MD100W manifold series will function as accurate vacrometers when used with the MDXVG Remote Vacuum Accessory (page 6).
1. Plug the MDXVG Vacuum Accessory into the upper right side of the manifold [FIG.9].
2. The Vacuum Display Screen [FIG.10] appears.
FIG.10
Connection port
FIG.9
mic
Vac
17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
VACUUM HOLD TEST
Test Vacuum Set Screen
Vacuum Hold Test Time Set Screen
3. Assemble the BLACKMAX MDXVG vacuum gauge accessory to the Suction Port on an AC system as shown [FIG.11].
Suction
Suction
port
port
FIG.11
Discharge
Discharge
port
port
Vacuum Gauge
Accessory
4. In this mode, the manifold will show current vacuum in the system. The display begins to show vacuum levels at 25000 microns or less and is accurate to
AVT45 Brass
Tee Connector
APO
FIG.12
+/- 10 microns. Continue to run vacuum pump until the desired vacuum level is reached. Once reached, proceed to Vacuum Hold Test.
5. Vacuum Hold Test - To perform this test, press
Vac
[CLR] to go to the Main Screen [FIG.1] Vacuum Gauge Accessory should still be plugged into manifold.
6. At Main Screen [FIG.1] press [TEST]. The Test Vacuum Set Screen appears [FIG.12]
7. Press [SET] to advance to Vacuum Hold Test –
APO
FIG.13
Time Set Screen [FIG.13].
8. Use [UP/DOWN ARROWS] to adjust test
duration in 5 minutes increments (minimum time is pre-programmed at 5 minutes).
A) Press [SET] to store Vacuum Test duration
time. Vacuum Hold Test Screen [FIG.14]
Vac
hh : mm
Set test duration (adjust in 5 minute increments
will now be displayed.
NOTE: The Vacuum Reading in the upper Left display is still active. The Reading in the upper Right display is not active unit steps 2 and 3 are completed.
18
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
VACUUM HOLD TEST
Vacuum Display Screen
Vacuum Display Screen
FIG.14
Current Vacuum Level
Psig
Leak test duration
B) Continue to Evacuate the system to the desired level which is indicated in the
upper left display once the desired vacuum level is reached: a. On the MD100W, close the vacuum valve b. On the MD50W, close the High and Low side valves. C) Press [SET] to start test. Starting test vacuum level is indicated in upper left
display [FIG.15]. Current system vacuum level is shown in upper right of
display and timer is shown in lower left. The timer decrements and when zero, a
bells sounds and test ends. At end of test, Start vs. Final Vacuum levels will be
displayed to enable determination of any vacuum loss over a set period of time.
hh : mm
Inactive
FIG.15
Desired Test Level
Timer counts backwards
Current System Vacuum Level
D) Press [CLR] to return to Main Screen [FIG.1].
NOTE: If it is required to rerun the vacuum pump after STEP 3, disconnect and reconnect the MDXVG to reset it in the vacrometer mode.
PROGRAMMING/SETTING UNITS
According to your geographic location, MD50W and MD100W manifold series are fac­tory programmed with 5 common refrigerants in those regions (see chart below), and units of measure (Temperature, Pressure and Vacuum) for specific regions. See chart on
next page for details. Any of these units can be changed via the Program Mode.
19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
PROGRAMMING/SETTING UNITS
Program Screen
Refrigerant Selection Screen
There are 2 major options. In each, a blinking symbol represents a choice the user can make by pressing [SET].
1. PrO (Programming) a. Change Refrigerants: Change any or all of 5 common refrigerants accessible at
Main Screen.
b. Change Temperature Units: Change °F or °C. c. Change Pressure Units: Change Psig, kPa, Mpa, bar and KgCm2 pressure units. d. Change Vacuum Units: Change Microns vs. Hg vacuum units.
2. APO (Set Automatic Power OFF) - Manifolds come factory pre-set to ON.
Pre-Programmed Refrigerants
The following 5 common refrigerants have been programmed for easy access from the
Main Screen using the [REF] key:
PRE-PROGRAMMED REFRIGERANTS and UNITS OF MEASURE
USA AUSTRALIA EUROPE
R-134a R-134a R-134a R-22 R-22 R-404A R-404A R-404A R-744 (CO2) R-407A R-744 (CO2) R-410A R-410A R-410A R-407c
PSIG, F, Micron kPA, C, Micron Bar, C, Micron
PROGRAMMING/SETTING REFRIGERANTS
The recommended way to change refrigerants involves making a new list of 5 refrigerants, and change all five of them in the same session. Example: If only R-22 needs to be changed to R-1234yf (and if using USA defaults), program the following items: R-410A, R-1234yf, R-404A, R-134a and R-407A.
1. At Main Screen [FIG.1] press and hold [MODE] until “PrO” (Program) Screen [FIG.16] appears. “PrO” will blink.
2. Press and hold [SET] to enter the Refrigerant Selection Screen [FIG.17].
APO
FIG.16 FIG.17
APO
R
R11 is initial default refrigerant
Counter
3. The Initial Refrigerant (blinking in upper right) should be R-11 and Counter (Lower Right) should start at 000.
4. Use [UP/DOWN ARROWS] to scroll through 93 refrigerants. Press and hold either arrow to scroll through refrigerants progressively faster. Release to stop scrolling. Last refrigerant displayed will blink.
20
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Temperature Units Screen
PROGRAMMING/SETTING REFRIGERANTS
5. To select refrigerant displayed, press [REF]. A tone will sound and counter will
Pressure Units Screen
increment by one indicating that the first refrigerant in the sequence of 5 favorites has been replaced with one selected.
6. To select another refrigerant, repeat steps 3 and 4.
7. To exit this menu and go to the next (Temperature Units Screen), press [SET], or
8. To return to Main Screen [FIG.1], press and hold [CLR] after step 6.
9. If all 5 refrigerants are changed, manifold automatically advances to next menu
without having to press [SET]. Main Screen [FIG.1] can then be accessed by pressing and holding [CLR].
PROGRAMMING/SETTING TEMPERATURE UNITS (°F VS. °C)
1. At Main Screen [FIG.1] press and hold [MODE] until “PrO” screen [FIG.16] appears. “PrO” will blink.
2. Press [SET] to enter “PrO” menu. The Refrigerant Selection Screen appears [FIG.17].
3. Press [SET]. The Temperature Units Screen [FIG.18] will appear (blinking).
4. Press [SET] to select this menu. The symbol for the current units blinks.
5. Use [UP/DOWN ARROWS] to toggle between °F /°C
6. Press [SET] to select °F vs. °C and go to next menu (Pressure Units), or
7. To return to Main Screen [FIG.1], press and hold [CLR].
APO
Temperature
Units
( °F or °C)
°C °F
FIG.18
°C °F
PROGRAMMING/SETTING PRESSURE UNITS (PSIG, KGCM2, KPA,BAR, MPA)
1. At Main Screen [FIG.1] press and hold [MODE] until “PrOScreen appears FIG.16]. “PrO” will blink.
2. Press [SET]. The Refrigerant Selection Screen appears [FIG.17].
3. Press [SET]. The Temperature Units Screen (blinking) appears [FIG.18].
4. Press UP ARROW to advance to Pressure Units Screen [FIG.19]. See blinking symbols.
5. Press [SET] to select this menu. Current units blink.
6. Use [UP/DOWN ARROWS] to scroll through units (Psig, KgCm2, kPa,bar, Mpa)
7. Press [SET] to select pressure units desired and go to next menu (Vacuum Units), or
8. To return to Main Screen [FIG.1], press and hold [CLR].
APO
Psig k Pa Mpa bar Psig k Pa Mpa bar KgCm2 KgCm2
FIG.19
Pressure Units (Psig, KgCm2, kPa,bar, Mpa)
21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
PROGRAMMING/SETTING VACUUM UNITS (MIC VS. INHG)
Vacuum Units Screen
Units Summary Screen
Set APO Screen
1. At main screen [FIG.1] press and hold [MODE] until “PrO” screen [FIG.16] appears. “PrO” blinks.
2. Press [SET] to enter “ PrO” Screen. The Refrigerant Selection Screen appears [FIG.17].
3. Press [SET] to bypass Refrigerants Screen. The Temperature Units Screen (blinking) appears [FIG.18].
4. Press [MODE] to advance to the Pressure Units Screen [FIG.19].
5. Press [SET] to advance to Vacuum Units Screen (blinking) [FIG.20].
6. Press [SET] to select this menu.
Current units will blink.
7. Use [UP/DOWN ARROWS] to select
either microns (mic) or inches of mercury (inHg).
8. Press [SET] to select desired units
APO
FIG.20
mic inHg
Vacuum Units (“mic” or “inHg”)
and go to next menu (Units Summary Screen), or
9. To return to the Main Screen [FIG.1] press and hold [CLR].
ENTER PROGRAMMING CHANGES (MAKE CHANGES CURRENT)
1. The next menu item after selecting Vacuum Units is the Units Summary Screen [FIG.21].
Blinking symbols represents units chosen in previous steps.
2. Press [SET] to select units shown on this screen-
making them the current working units.
3. Or after pressing [SET], use [UP/DOWN ARROWS] to display previous units. These can be restored by pressing SET Menu. In either case, units become current units and manifold returns to Main Screen [FIG.1]
APO
Psig
°F
FIG.21
Blinking symbols indicate
mic
Pressure, Temperature and Vacuum Units chosen in previous steps
SETTING APO (AUTOMATIC POWER) ON OR OFF
When APO is ON and there has been no user activity (usually 10-15 minutes), manifolds turn OFF to conserve battery power. About 1 minute before turning OFF, an alarm will sound to signify that override is possible by pressing any key, except the [ON/OFF] key. This will re-start the APO timer. APO is suspended when performing either the vacuum or pressure hold test or when the unit is used as a vacuum gauge. When APO is OFF, the user must turn a manifold off manually.
1. From Main Screen [FIG.1], press and hold [MODE]. The
PrO icon will blink [FIG.16].
2. Press DOWN ARROW to go to (blinking) Set APO screen.
[FIG.22].
3. At SET APO Screen briefly press [SET]. “APO On” will
appear and blink [FIG.23].
4. Use [UP/DOWN ARROW] to change display from APO On
[FIG.23] to APO OFF as desired.
5. Press [SET] to select desired APO state and return to Main
Screen [FIG.1]
22
APO
FIG.22
FIG.23
APO
From either screen, select either Automatic Power ON or OFF
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CONNECTING TO AN A/C SYSTEM
Various service and test procedures below can be performed after your manifold has been properly connected to an AC system as shown:
FIG.24 FIG.25
(L) (H) (L) (H)
RECOVERING REFRIGERANT
MD50W 2 Valve Manifold [FIG.24] MD100W 4 Valve Manifold [FIG.25]
1.) Connect Low (A) and High (B) Service hoses to
service ports (E) and (D)
2.) Connect hose from to open port on the tee(C) to Refrigerant Recovery Unit inlet port
3.) Start Recovery Unit 3.) Open low side valve (A)
4.) Open low side valve (A) 4.) Open high side valve (B)
5.) Open high side valve (B) 5.) Start Recovery Unit
6.) Complete refrigerant recovery process. 6.) Open valve (D)
7.) Disconnect manifold from system.
8.) Proceed to the repair system, then go to
Evacuation Procedure
1.) Connect Low (1) and High (2) Side Service hoses to equipment’s service ports (L) and (H)
2.) Connect hose from port (4) to Refrigerant Recovery unit inlet port
7.) Complete refrigerant recovery process. Disconnect manifold from system.
8.) Proceed to the repair system, then go to Evacuation Procedure
Reference [Fig.24] and [Fig.25]
23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
EVACUATING PROCEDURES
MD50W 2 Valve Manifold [FIG.24] MD100W 4 Valve Manifold [FIG.25]
1.) Connect Low (A) and High (B) Service hoses to
service ports (E) and (D)
2.) Connect hose from tee (C) to Inlet port of Vacuum Pump. Use the fitting on the tee with the valve core.
3.) Connect another hose from the open port of tee (C) to the valve of a refrigerant supply tank. Supply tank valve must be off.
4.) Check Low and High Side pressure readings, if zero PSIG or lower continue. If pressure reads higher than 0 PSIG, run Refrigerant RECOVERY
5.) Once it is determine no refrigerant is present in system, start vacuum pump.
6.) Open low side valve (A)
7.) Open high side valve (B)
8.) Note: A Vacuum Gauge (VG200) can be
connected via service tee to equipment to monitor vacuum pump’s progress.
9.) Evacuate system according to manufacturer’s specifications; close high and low side hand knobs.
10.) Disconnect Vacuum Pump hose from valved port on Tee (C). The valve core will seal when hose is removed, thus preserving the vacuum.
The unit is now ready for Vacuum Leak test or High Side Charging Procedure.
1.) Connect Low (1) and High (2) Service hoses to service ports (L) and (H)
2.) Connect 3/8” hose from port (4) to Vacuum pump inlet port
3.) Connect another hose from port (3) to refrigerant supply tank’s valve. Supply tank valve must remain off.
6.) Open low side valve (A) Open high side valve (B)
7.) Open valve (D). Open valve (C), but do not open supply tank valve.
8.) Note: Optional Vacuum Gauge Accessory (MDXVG) can be connected via service tee to the equipment to monitor vacuum pump’s progress.
10.) Close valve (D) once the required vacuum level is reached. Turn off vacuum pump.
CHARGING PROCEDURE - LOW SIDE
MD50W 2 Valve Manifold [FIG.24] MD100W 4 Valve Manifold [FIG.25]
1.) The following instructions are to add charge to a pressurized system, known as “topping off”.
2.) Connect Low (A) and High (B) Service hoses to
service ports (E) and (D). All valves should be closed.
3.) Connect a hose from the open port on tee (C) to valve fitting on supply tank.
4.) Open supply tank valve. For blends, turn tank upside down or make sure liquid refrigerant is being charged.
5.) Push valve core on tee (C) until refrigerant comes out. This will purge the hose from the supply tank.
6.) Turn refrigerant system on. Monitor pressure gauges. Use superheat and subcooling features to help determine when charge amount is optimal.
7.) Slightly open Low Side Valve (A) to allow liquid refrigerant to flow into low side of equipment. Caution should be used when charging liquid refrigerant into low side of a refrigeration system.
8.) Close Low Side valve (A) when desired amount of refrigerant has been added. Note the use of a scale is required.
9.) Let equipment temperatures and pressures stabilize. Check superheat and/or subcooling levels. Repeat charge if necessary.
10.) Once done, close off supply tank valve and disconnect all hoses. Use Caution as all hose connections are under pressure.
Once charge is complete, an electronic refrigerant leak detector should be used to confirm leaks do not exist.
24
2.) Connect Low (1) and High (2) Service hoses to service ports (L) and (H)
3.) Connect a hose from port (3) to refrigerant supply tank’s valve.
5.) Open valve (C). With hose disconnect from port (4), slightly open valve (D) until refrigerant comes out. This will purge the hose from the supply tank.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CHARGING PROCEDURE - HIGH SIDE
High Side Charging should be done on a completely evacuated system.
MD50W 2 Valve Manifold [FIG.24] MD100W 4 Valve Manifold [FIG.25]
1.) A scale should be used to weigh in the charge. Place tank on scale so liquid refrigerant is supplied.
2.) Open Supply tank valve to fill manifold with liquid refrigerant.
3.) Set scale to record charge amount.
4.) Open High Side Valve (B). Monitor weight reading.
5.) Once desired charge amount has been reached, close High Side Valve (B) and supply tank valve.
6.) Start refrigeration system. Open Low Side valve (A) to allow all remaining refrigerant in manifold and
service hoses to charge into system.
7.) Close Low Side Valve (A). 7.) Close Low Side Valve (A) and Valve (C).
8.) Monitor Low and High Side pressures. Superheat and subcooling features of manifold can be used at this time.
9.) Disconnect service hoses from equipment. Use Caution as all hoses are under pressure.
Once charge is complete, an electronic refrigerant leak detector should be used to confirm leaks do not exist.
APPENDIX A (MANIFOLD SPECIFICATIONS)
High side pressure range -14.7 to 800.0 PSIG (-1.0 to 55 bar) Low side pressure range -14.7 to 800.0 PSIG (-1.0 to 55 bar) Pressure reading accuracy +/- 1% of reading or +/- 0.5 PSIG (+/- 0.05 bar) Pressure reading resolution 0.5 PSIG (0.05 bar) High and low side burst pressure 1500.0 PSIG (103.4 bar) High side temperature range -58.0 °F to 302 °F (-50 °C to 150°C) Low side temperature range -58.0 °F to 302 °F (-50 °C to 150°C) Temperature reading accuracy +/- 0.36 °F (+/- 0.2°C) Vacuum indication accuracy +/- 10 microns (requires MDXVG accessory) SH, SC and saturation temperature accuracy +/- 0.5°F (+/- 0.3°C) Instrument working temperature -4.0 °F to 122°F (-20°C to 50°C) Instrument working RH range 0% to 100% non-condensing Total number of stored refrigerants 93 APO (Automatic Power Off) 10 minutes. Can be disabled by user. MD100W size (L x H x W) 8.1” x 3.5” x 7.0” (20.5 cm x 8.9 cm x 17.8cm) MD50W size (L x H x W) 8.1” x 2.5” x 7.0” (20.5 cm x 6.4 cm x 17.8cm) MD100W weight 3.3 lb. (1.5 kg) MD50W weight 3.0 lb. (1.4 kg) Display Life (Using 4 industrial alkaline AA batteries (included) Display Life Using 4 Rechargeable NiMH AA batteries (not included):
Backlight full ON: 25+ hrs. continuous duty
Backlight full OFF: 60+ hrs. continuous duty
Backlight full ON: 20+ hrs. continuous duty from full charge
Backlight full OFF: 50+ hrs. continuous duty from full charge
APPENDIX B (MDXVG VACUUM GAUGE ACCESSORY SPECIFICATIONS)
Sensor Type
Display Pressure Range
Accuracy +/- 10% of reading or +/- 10 microns, whichever is greater
Operating Temperature Range
Operating and Storage Humidity 0 to 95%, non-condensing Storage Temperature -40 to 85°C (-40 to 185°F)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Self - heated thermistor bridge with integral temperature compensation from 0 to 50 °C (32 to 122 °F).
0 to 99 microns 1 micron resolution 100 to 990 microns 10 micron resolution 1000 to 9900 microns 20 micron resolution 10000 to 25000 microns 50 micron resolution
above 25000 microns
Compensated (accuracy as stated): 0°C TO 50°C (32°F TO 122°F) Non-compensated (add +/- 0.5% of reading error for every °C outside
compensated range): -10 to 0°C (14 to 32°F)
AUTO-RANGING SCALES:
Will display low side pressure reading, in selected unit.
25
APPENDIX C (MDXBK CHARGER ACCESSORY SPECIFICATIONS)
Operating Temperature Range 32°F to 122°F (0°C to 50°C)
Operating and Storage Humidity 0-95%
Storage Temperature -40°F to 185°F (-40 to 85°C) Power Source 100-240VAC input, 9VDC 2.0 amp maximum output
Battery Type
Rechargeable Battery Cycle Life
AA size, NiMH rechargeable, approximate battery life is 20,000 to 50,000 hours*
1000 recharges *based on usage with or without the LCD backlight energized.
APPENDIX D (BATTERY SELECTION AND REPLACEMENT)
For optimum life and economic value, CPS Products, Inc. recommends that MD50W and MD100W manifold series be powered by rechargeable NiMH (nickel metal hydride) batteries. CPS Products offers an optional “Charger Kit” (model MDXBK) that comes with a power adapter and 4 NiMH batteries. Do not connect the battery charger if there are no rechargeable batteries in the manifolds.
NOTE: Alkaline batteries should not, and cannot be charged in the BLACKMAX MD50W or MD100W Series. Battery charge status is shown by the battery indicator symbol in the upper left [FIG.1] Main Screen and all other screens. NOTE: A slight increase in temperature near the charging socket is normal. Remove battery door and install 4 AA size batteries, observing
battery polarity. When using rechargeable batteries for the first time, turn manifold off and connect one end of charger to Manifold 9V DC plug [FIG.26] and the other end to an appropriate wall outlet socket. Charge NiMH batteries for 2 hours. The Battery Symbol on the display will blink during charging. MD50W and MD100W manifold series can be used while charging the batteries, but charging time will be considerably extended. The preferred way to use the Rapid Charging Feature is to charge while the manifold is OFF.
FIG.26
Connection for 9V DC Charger
APPENDIX E (CLEANING SENSOR IN MDXVG)
Sensor may become contaminated with dust, oil, etc. from pulling vacuums.
1. Never use a cotton swab or similar object to clean the sensor. This may damage the
sensor.
2. Open all knobs, valves, and cap all the ports except for the VAC port.
3. Invert the MDXVG and add about 1 tablespoon of 100% denatured alcohol into the VAC port (see page 2) so as to flush out contaminants.
4. Gently shake the unit to produce a swirling effect of the alcohol inside the housing.
5. Turn right side up, open a port to pour out the rubbing alcohol.
6. Repeat the above procedure until the alcohol appears to be clean and clear of contaminants.
7. Check the gasket located inside the VAC port. If worn or deformed, replace with a new CPS gasket (PN: HXG)
8. Open all ports to allow sensors to air dry (usually about 1 hour). DO NOT BLOW DRY.
26
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
APPENDIX F (LIMITED WARRANTY)
BLACKMAX digital manifolds are covered by a CPS Products, Inc. LIMITED WARRANTY. CPS Products, Inc. warrants, to the original purchaser, that its BLACKMAX line of digital manifolds and accessories are free from defects in assembly, materials and workmanship for 1 year from the date of purchase, or 15 months from the date of manufacture, whichever is sooner. If any BLACKMAX product is believed to be defective it must be returned to CPS Products. If the product is found to be defective, CPS will repair or replace the item(s), at our option, at no charge.
This warranty does not apply to products or accessories that have been altered, misused, submerged, exposed to extreme heat or cold, lightning strikes, or electrical voltage exceeding published product specifications. This warranty also does not apply to products that are simply in need of routine field service maintenance.
To assure your BLACKMAX purchase is on file with CPS Products, please fill in a Product Registration form and mail to CPS within 15 days from the date of purchase, along with a copy of the original sales receipt. Return of the registration form is not required to be covered by the Limited Product Warranty.
To process a warranty claim, please contact CPS USA at 1-800-277-3808 or CPS Products N.V. (Belgium) at (32) 3 281 30 40; or CPS Australia at 08 8340 7055, within 15 days of discovering defects in assembly, materials, or workmanship. CPS will provide a return authorization number (RGA #). Do not attempt to repair BLACKMAX products. CPS has the sole right to determine if a claim is valid and if compensation is justified.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER EXPRESS WARRANTIES. TO THE EXTENT ALLOWED BY LAW, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED TO THE DURATION OF THE LIMITED PRODUCT WARRANTY. IN NO CASE SHALL CPS PRODUCTS INC. HAVE ANY LIABILITY FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND, INCLUDING BUT NOT LIMITED TO: THE LOSS OF THE USE OF THE PRODUCT OR ANY ASSOCIATED EQUIPMENT, COST OF ANY SUBSTITUTED EQUIPMENT, FACILITIES OR SERVICES, DOWNTIME, THE CLAIMS OF THIRD PARTIES, DAMAGE TO PROPERTY OR PERSONAL INJURY. SOME JURISDICTIONS DO NOT ALLOW LIMITS ON WARRANTIES OR ON REMEDIES FOR BREACH IN CERTAIN TRANSACTIONS. IN SUCH JURISDICTIONS, THE LIMITS IN THIS PARAGRAPH MAY NOT APPLY.
THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS, AND YOU MAY ALSO HAVE OTHER RIGHTS WHICH VARY FROM JURISDICTION TO JURISDICTION.
APPENDIX G (REFRIGERANTS STORED IN MEMORY)
MD50W and MD100W manifold series are pre-loaded with 93 of the most common refrigerants.
R11 R13B1 R290 R407D R416A R424A R504 R113 R141B R32 R407F R114 R142B R401A R408A R426A R508A R115 R143 R401B R409A R418A R116 R143A R401C R41 R419A
R12 R152A R402A R410A R420A R123 R176 R402B R410B R421A R428A R600
R123A R21 R403A R411A R421B R434A R600A
R1234YF R218 R403B R411B
R124 R22 R404A R412A R601A
R125 R23 R405A
R1270 R236ea R406A R422C R500
R13 R236FA R407A R414A R134 R245ca R407B R414B R502
R134A R245fa R407C R415A R423A / 39TC R503
R413A /
M049
R417A / M059
R422A / M079
R422B
R422D / M029
R425A R507A
R427A /
M089 /
FX100
R437A /
M049PLUS
R438A /
M099
R501
(Sub-critical
range only)
R508B
R509A
R601
R744
27
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
APPENDIX H (ERROR INDICATOR SCREENS)
If certain malfunctions occur, MD50W and MD100W series will display error messages:
oP -Appears in display areas reserved for temperature readings when thermistor sensor connected to that port is open circuited or not connected. Replace sensor with CPS model number TMX3C.
SP –Appears in display areas reserved for temperature readings when thermistor sensor connected to that port is short circuited. Replace sensor with CPS model number TMX3C.
Err (Upper Left Display) - Appears in upper left display when T function is selected and one or both thermistor sensors are defective as indicated by oP or SP indicators mentioned above.
Err (Lower Left Display) -Appears in display areas reserved for temperature readings when user tries to read saturation temperatures and calibration of pressure channels is invalid. This defect requires manifold be returned to factory.
#73-156 RevB
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...