CPS Products CPS Products AC750 Manual

Test Equipment Depot - 800.517.8431
-
99 Washington Street Melrose, MA 02176
s®


OWNER'S MANUAL (En
glish}
CONTENTS
1. Safety Information
1.1 Preliminary
1.2 Usage
1.3 Symbols
1.4 General Use
2. Description
2.1 AC750 Layout ............................................................................5
2.2 Switch, Key and Input Jack Description
2.3 LCD Display ........................................................................... 7-8
3. Specifications
3.1 General .......................................................................................9
echnical Data .................................................................... 10-15
3.2 T
4. Operating Instructions
4.1 Reading Hold ...........................................................................16
4.2 Manual Measuring Range
4.3 Frequency Duty/Ratio Switch ...................................................16
4.4 Maximum / Minimum Measurement ........................................17
4.5 Function Key ............................................................................17
4.6 INRUSH Measurement .............................................................17
4.7 Back Light And Clamp Head Light ............................................18
4.8 Automatic Power-Off ................................................................18
4.9 Measurement Preparation ........................................................18
................................................................................1
.........................................................................................2
.....................................................................................3
................................................................................4
.....................................6
........................................................16
2
CONTENTS
4.10 Current Measurement ............................................................19
4.11 Voltage Measurement
4.12 Frequency and Duty Ratio Measurement
4.13 Resistance Measurement .......................................................21
4.14 Diode Test ..............................................................................21
cuit Continuity Test
4.15 Cir
4.16 Capacitance Measurement
4.17 Surge Current Measurement
4.18 Non-Contact Voltage Detection
5. Maintenance
5.1 Calibration
5.2 Battery Replacement ................................................................24
5.3 Probe Replacement ..................................................................24
6. Accessories ...........................................................28
7. Warranty ..............................................................29
................................................................................24
............................................................20
...............................20
............................................................22
.....................................................22
..................................................22
..............................................23
3
1. SAFETY INFORMATION
1.2 USAGE
1.2.1 DO NOT measure voltage greater than 750 V AC, 1000 V DC.
1.2.2 DO NOT measure resistor, capacitor, diode and circuit while
connected to power.
1.2.3 DO NOT measure capacitance before capacitor is discharged
completely.
1.2.4 DO NOT use meter in explosive gas, vapor or dusty environments.
1.2.5
DO NOT store or use meter in direct sunlight, high ambient
temperature or high humidity.
1.2.6 During test of currents, resistors, capacitors, diodes and circuit
connections, DO NOT connect meter to a voltage source.
1.2.7 DO NOT use unless meter bottom case and battery cover are
completely fastened .
1.2.8 During use, select the right function and measuring range.
During measurement, DO NOT
1.2.9 each range.
When measuring a circuit with the meter connected,
1.2.10 contact with probe tip (metal part).
During measurement, if voltage to be measured is more than
1.2.11 60 V DC or 30 V AC (RMS),
protection device
1.2.12 In the manual measuring range mode, when measuring an unknown value, select the highest measuring range first.
1.2.13 Before rotating Selector Switch to change measuring function, remove probe from circuit to be measured.
1.2.14 If you find any abnormal phenomena or failure of meter, stop using immediately.
exceed the indicated value for
DO NOT
always keep fingers behind finger
5
1. SAFETY INFORMATION
1.4 GENERAL USE
1.4.1 Do not open case to adjust or repair (only for technicians who fully
understand this meter and electrical shock hazard).
1.4.2 Before opening case or battery cover, remove probe from circuit to be
measured.
1.4.3 To avoid wrong readings causing electric shock, when “ “ appears on
meter display, replace battery immediately.
1.4.4 Clean meter with damp cloth and mild detergent. Do not use abrasives
or solvents.
1.4.5 Set Selector Switch to OFF position when meter not in use.
1.4.6 If meter is not used for an extended time, remove battery.
2. DESCRIPTION
2. DESCRIPTION
• This meter is a portable, professional multifunction measuring
instrument with LCD display and back light. Selector Switch is operated with one hand for easy operation.
• This meter is for measuring AC current, DC current, AC voltage,
DC voltage, frequency, duty ratio, resistance, capacitance measurement, circuit connection, diode test and non-contact voltage detection.
Other features include:
Automatic measuring range and manual measuring range
• Overload protection
• Reading hold function
• Max and Min measuring function
• Clamp head frequency measurement function
• Auto power-off function
• Relative measuring function
• Low battery indicator
7
2. DESCRIPTION
2.2 SWITCH, KEY AND INPUT JACK DESCRIPTION
BL/HOLD Back light or reading hold key
FUNC Function key (use with Selector Switch)
RANGE For switching manual measuring range state
REL For entering relative measurement state
Hz/% For frequency and duty ratio measurement function switch
MAX/MIN For maximum/minimum measurement function switch
OFF For turning power OFF
INPUT Jack
COM Jack
Selector Switch
99 Washington Street Melrose, MA 02176
-
Voltage, resistance, frequency, duty ratio, capacitance, diode, circuit connection input wire connecting terminal
Voltage, resistance, frequency, duty ratio, capacitance, diode, circuit connection common wire connecting terminal
For selecting function and measuring range
Test Equipment Depot - 800.517.8431
TestEquipmentD
epot.com
9
3. SPECIFICATIONS
3.1 GENERAL
Maximum Input Current 1000A AC; 1000A DC Maximum Input Voltage 750V AC (RMS); 1000V DC
AC Volts
DC Volts
AC Amps 60/600/1000 (+/- 2% reading) DC Amps 60/600/1000 (+/- 2% reading)
OHMs
Frequency
(Through Jaw Clamp)
Frequency
(Through Mode V)
Frequency
(Through Mode HZ/DUTY)
Measuring Range Overload Protection Yes (Full Range) Sampling Rate Approx. 3 times/sec. LCD Display Auto-Off After 15 minutes (if no operation) Test Leads (Red, Black) 39" (1 m), Included Battery Undervoltage Indication Polarity Indication Automatic ("-" = negative polarity) Display Indication
(Exceeds Max Range)
Temperature Coefficient <0.1 X accuracy/°C Clamp, Max Opening 1.6" (42 mm) Full measuring range overload
protection Battery One 9V DC (Included) Operation Temperature 32°F~122°F (0°C ~ 50°C) Storage Temperature -14°F~122°F (-10°C ~ 50°C) Size (H x W x D) 9.4˝ x 1.9˝ x 3.6˝ (238 x 50 x 92 mm) Weight (Including Battery) 0.92 lb (420 g)
6mV/600mV/6V/60V/600V/750V (+/-
0.6% reading); 750V (+/-0.8% reading) 6mV/600mV/6V/60V/600V/1000V (+/-
0.5% reading); 1000V (+/-0.8% reading)
600/6K/60K/600K/6M/60M (+/- 2% reading)
99.99 Hz to 999.9 Hz
99.99 Hz to 9.999kHz
9.999 Hz to 9.999 MHz
Display shows
"OL" or "-OL"
Yes
11
3. SPECIFICATIONS
3.2.1 TRUE RMS ZERO INPUT CHARACTERISTIC
3.2.1.1 For measuring non-sinusoidal wave signals, use true RMS measuring method, which has less error than traditional average response measuring method.
This true RMS meter can accurately measure non-sinusoi-
3.2.1.2 dal wave signals, but in AC function mode, when there is no signal to be measured (input terminal short circuit in AC voltage mode), clamp meter may show a reading from 1 to
50. These deviating readings are normal. In the designated measurement range, they will not affect the accuracy for multimeter measuring AC.
3.2.1.3 True RMS can be measured only when input signal reaches a certain level. Therefore, the measuring range of AC voltage and current should be specified at 2% ~100% of full range.
3.2.2 AC CURRENT
Measuring range Resolution Accuracy
60A 0.01A
1000A 1A
Maximum input current: 1000A AC
• Maximum input current: 0~600A: 40~400Hz; 600A~1000A: 40~60Hz
±(2.5% reading + 8 digits)600A 0.1A
12
3. SPECIFICATIONS
3.2.3 DC CURRENT
Measuring range Resolution Accuracy
60A 0.01A
1000A 1A
Maximum input current: 1000A DC
3.2.4 SURGE CURRENT
Measuring range Resolution Accuracy
60A 0.01A < 60A for reference only
600A 0.1A ±(5% reading + 60 digits)
1000A 1A
Time of integration: Measurement range: 20-1000A Frequency range: 40-400Hz
3.2.5 DC VOLTAGE
Measuring range Resolution Accuracy
600mV 0.1mV
6V 0.001V
60V 0.01V
600V 0.1V
1000V 1V ±(0.8% reading + 4 digits)
• Input impedance: 10M
• Maximum input voltage: 750V AC (RMS) or 1000V DC
Note: In the small voltage measuring range, the probe is not connected with the circuit to be tested, and the meter may have fluctuating readings, which is normal and caused by the meter’s high sensitivity. This does not affect actual measurement results.
100ms;
±(2.0% reading + 8 digits)600A 0.1A
±(0.5% reading + 5 digits)
13
3. SPECIFICATIONS
3.2.6 AC VOLTAGE
Measuring range Resolution Accuracy
600mV 0.1mV
6V 0.001V
60V 0.01V
600V 0.1V 750V 1V ±(0.8% reading + 4 digits)
Input impedance: 10M
• Maximum input voltage: 750V AC (RMS) or 1000V DC
• Frequency range: 40 - 400Hz
Note: In the small voltage measuring range,the probe is not connected with the circuit to be tested, and the meter may have fluctuating readings, which is normal and caused by the meter’s high sensitivity. This does not affect actual measurement results.
3.2.7 FREQUENCY
3.2.7.1 Clamp head measuring frequency (through range A):
Measuring range Resolution Accuracy
99.99Hz 0.01Hz
999.9Hz 0.1Hz
• Measuring scope: 10Hz-1kHz
• Input signal range: 20A AC (RMS) (input current will increase
when the frequency to be measured increases)
• Maximum input current: 1000A (RMS)
3.2.7.2 Through mode V:
Measuring range Resolution Accuracy
99.99Hz 0.01Hz
9.999Hz 0.001kH
• Measuring scope: 10Hz - 10kHz
• Input signal range: 20A AC (RMS) (input current will increase when
the frequency to be measured increases)
• Input impedance: 10M
14
• Maximum input voltage :750V AC (RMS)
±(0.8% reading + 5 digits)
±(1.5% reading + 5 digits)
±(1.5% reading + 5 digits)999.9Hz 0.1Hz
3. SPECIFICATIONS
3.2.7.3 Through mode HZ/DUTY:
Measuring range Resolution Accuracy
9.999Hz 0.01Hz
99.99Hz 0.1Hz
999.9Hz 0.001kHz
9.999kHz 0.01kHz
99.99kHz 0.1kHz
999.9kHz 0.001MHz
9.999MHz 0.001kH
• Overload protection: 250V DC or AC (RMS)
• Input voltage range: 2V (input voltage will increase when the frequency to be measured increases)
3.2.8 Duty Ratio
Measuring range Resolution Accuracy
0.1-99.9% 0.1% ±3.0%
3.2.8.1 Through mode A (from clamp head):
• Frequency response:10-1kHz
• Input current range: 20A AC (RMS)
• Maximum input current: AC 1000A
3.2.8.2 Through mode V:
• Frequency response: 10-10kHz
• Input voltage range: 60mV AC
• Input impedance: 10M
• Maximum input voltage: 750V AC (RMS)
3.2.8.2 Through mode HZ/DUTY:
• Frequency response: 10 - 10MHz
• Input voltage range: 2V AC (RMS) (input voltage will increase when the frequency to be measured increases)
• Maximum input voltage: 250V AC (RMS)
±(0.3% reading + 5 digits)
15
3. SPECIFICATIONS
3.2.9 RESISTANCE Measuring range Resolution Accuracy
600Ω 0.1Ω
6kΩ 0.001kΩ
60kΩ 0.01kΩ
600kΩ 0.1kΩ
6MΩ 0.001MΩ
60MΩ 0.1MΩ
• Open circuit voltage: about 0.4 V
• Overload protection: 250 V DC or AC (RMS)
3.2.10 CIRCUIT CONTINUITY TEST Measuring range Resolution Function
0.1Ω
• Overload protection: 250 V DC or AC (RMS)
3.2.11 CAPACITANCE
Measuring range Resolution Accuracy
9.999nF 0.001nF
99.99nF 0.01nF
999.9nF 0.1nF
9.999μF 0.001μF
99.99μF 0.01μF
999.9μF 0.1μF
9.999mF 0.001mF
99.99mF 0.01mF
• Overload protection: 250 V DC or AC (RMS)
±(0.8% reading + 3 digits)
±(2.0% reading + 5 digits)
If the resistance of circuit to be measured is less than 50Ω, the meter’s built-in buzzer may sound.
±(3.0% reading + 5 digits)
16
3. SPECIFICATIONS
3.2.12 DIODE TEST
Measuring
range
• Forward DC current is about 1mA
• Backward DC voltage is about 3.3V
• Overload protection: 250V DC or AC (RMS)
4.1 READING HOLD
1. During measurement, if reading hold is required, press “HOLD/BL” key and the value on the display will be locked. Press “HOLD/BL” key again to cancel reading hold state.
4.2 MANUAL MEASURING RANGE
1. RANGE key is automatic/manual measuring range key to trigger mode. The preset range is automatic measuring range. Press to switch to manual measuring range. In manual measuring range mode, click once to change to upper range. Continue to the top range, then continue to press this key to change to the bottom range, followed by recycling. If this key is pressed more than 2 seconds, it will switch back to automatic measuring range state.
Note: In capacitance and frequency measurement state, rhw manual measuring range key is invalid.
4.3 FREQUENCY/DUTY RATIO SWITCH
1. When meter is in AC voltage mode, if “Hz/%” key is pressed,
the meter will measure Hz, and measure AC voltage, AC current signal frequency. Click “Hz/%” key again and meter will measure DUTY cycle, voltage and current signal duty ratio. If in HZ/DUTY position, pressing HZ % key will switch between HZ and DUTY by recycling.
2. If “Hz/%” key is pressed again, meter will revert to voltage,
current measurement state.
Resolution Function
0.001V
Display approximate diode
forward voltage value
17
4. OPERATING GUIDANCE
Note: If meter is in maximum/minimum value measurement state, it can not switch to frequency, duty ratio measurement mode.
4.4 MAX/MIN MEASUREMENT CHOICE
1. Press “MAX/MIN” key to enter MAX mode, and always keep measurement maximum value; press “MAX/MIN” key again, to enter minimum value measurement state; press “MAX/MIN” key for the third time and the meter will display the difference of maximum and minimum value; press “MAX/MIN” key to repeat the above operations by recycling.
2. After entering MAX or MIN mode, meter will automatically save the measured maximum or minimum value.
3.
If the user presses “MAX/MIN” key more than 2 sec, meter will
restore normal measuring range.
Note: a) When meter is in maximum/minimum value measurement state, it is in manual measuring range mode. b) When meter is in frequency can not switch to maximum/ minimum value measurement mode.
4.5 FUNCTION SWITCH
1.
In resistance mode, press
among resistance, diode and continuity detection by recycling.
2.
In voltage and current mode, press
between AC and DC.
4.6 REL/INRUSH MEASUREMENT
1.
REL/INRUSH key is relative value measurement key
this key will enter relative value measurement mode. The current display value can be stored in the memory as reference value. When the user measures later, the display value is the difference for input value minus reference value. ie. REL (current reading)= Input value - Reference value.
2.
The relative value measurement can only be performed in
manual mode.
18
, duty ratio measurement state, it
FUNC key and meter will switch
FUNC key to switch
. Pressing
4. OPERATING GUIDANCE
3. In AC current measurement state, press REL/INRUSH more than
2 seconds to enter surge measuring state.
4.7 BACKLIGHT AND CLAMP HEAD LIGHT
1. During measurement, if ambient light is too dark to read the display, press and hold “BL/HOLD” key to activate backlight. Backlight will automatically turn off after 30 seconds.
2. During this period, pressing “BL/ HOLD” key more than two seconds will turn OFF backlight.
3. In the current mode, meter will turn backlight and clamp head light on. Backlight uses an LED with high current draw. Backlight will turn off after 30 seconds. Using backlight often will shorten battery life.
Note: When battery voltage 7.2 V, the LCD displays “ ” under­voltage symbol. If using backlight and battery voltage drops below
7.2 V due to high working current, the symbol may appear and
measurement accuracy is not guaranteed. Continue to use meter normally without using the backlight. Do not replace battery until
symbol shows under normal conditions.
4.8 AUTOMATIC POWER-OFF
1. After powering on, and if meter isn’t used for 15 minutes, meter will enter suspended state, automatically powering off to save the battery. Within 1 minute before shutdown, buzzer will sound five times. Meter will then enter a dormant state.
2. After automatic power-off, press FUNC key, to turn meter on again.
3. Holding FUNC key when powering on cancels automatic power­off function.
19
4. OPERATING GUIDANCE
4.9 MEASUREMENT PREPARATION
1. Rotate the Selector Switch to turn on power. When battery voltage is low (about 7.2 V), LCD displays symbol. Replace battery.
2. means that voltage or current should not be more than the specified value to protect internal line from damage.
3. Set Selector Switch to required measuring function and range.
4. When connecting line, first connect common test line, then charged test line. When removing the line, remove charged test line first.
4.10 CURRENT MEASUREMENT
WARNING
ELECTRIC SHOCK HAZARD. REMOVE PROBE FROM METER BEFORE MEASURING WITH CURRENT CLAMP.
1. Set Selector Switch to position A. At this time, meter is in AC current measurement state. Choose appropriate measuring range.
2. If you want to measure DC current, press FUNC key to enter direct current measurement state.
3. Hold trigger, open clamp head, clip one lead of measurement circuit to be tested in the clamp.
4. Read current value on LCD display.
Note:
1. Clamping two or more leads of circuit to be tested simultaneously will not give correct measuring results.
2. To get accurate readings, connect lead to be tested at center of current clamp.
3. indicates that maximum input AC current is 1000 A
4. To improve measurement precision, in DC current measurement state, if LCD display is not zero, press REL to return to zero, then measure.
20
4. OPERATING GUIDANCE
4.11 VOLTAGE MEASUREMENT
WARNING
ELECTRIC SHOCK HAZARD. PAY SPECIAL ATTENTION TO AVOID SHOCK WHEN MEASURING HIGH VOLTAGE. DO NOT INPUT VOLTAGE MORE THAN AC750 RMS.
1. Insert black probe to COM jack, insert red probe to INPUT jack,
choose appropriate measuring range.
2. Place Selector Switch to AC voltage V or mV position. At this
time, meter is in the DC voltage measurement state. To measure AC voltage, press FUNC key to enter AC voltage measurement state. Connect probe with voltage source or both ends of load in
3. parallel for measurement.
4. Read the voltage on the LCD.
Note:
1. In small voltage measuring range, probe is not connected with
circuit to be tested and meter may have fluctuating readings, which is normal and caused by the meter’s high sensitivity When meter is connected with circuit to be tested, you will get actual measured value. In relative measurement mode, automatic measuring range is
2.
invalid.
3. Indicates maximum input AC current is 750V AC or 1000V
DC Maximum input voltage is mode mV is 600 mV DC or AC.
4. If readings measured by meter are more than 750C RMS AC, a
“beep” alarm will sound.
.
21
4. OPERATING GUIDANCE
4.12 FREQUENCY AND DUTY RATIO MEASUREMENT
A) Clamp head measuring frequency (through AC current):
WARNING
ELECTRIC SHOCK HAZARD. REMOVE PROBE FROM METER BEFORE MEASURING WITH CURRENT CLAMP.
1. Set Selector Switch to position A .
2. Hold trigger, open clamp head, clip one lead of measurement circuit to be tested in the clamp.
3. Press Hz/% key to switch to frequency measuring state.
4.
Read current value on the LCD display.
5. Pressing Hz/% again enters duty ratio measuring state.
Note:
1. Clamping two or more leads of circuit to be tested simultaneously will not get correct measuring results.
2. Frequency measurement range is 10Hz~1kHz the frequency to be tested is less than 10Hz, or if frequency is higher than 10 kHz, accuracy is not guarantee
3. Duty ratio measuring range is 10 ~ 95%.
4. “ ”means that maximum input current is 1000A AC (RMS).
B) In Voltage Measurement Mode:
WARNING
ELECTRIC SHOCK HAZARD. PAY SPECIAL ATTENTION TO AVOID SHOCK WHEN MEASURING HIGH VOLTAGE. DO NOT INPUT VOLTAGE MORE THAN AC 750 RMS.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Place Selector Switch to V or mV position, press FUNC to enter
AC voltage measurement state.
3. Press Hz/% key to switch to frequency measuring state.
22
4. OPERATING GUIDANCE
4. Connect probe with signal or both ends of load in parallel for measurement.
5. VRead on LCD.
6. Pressing Hz/% again enters duty ratio measuring state.
Note:
1. Frequency measurement range is 10Hz~1kHz When the frequency to be tested is less than 10Hz,the LCD will show 00.0 Measuring frequency higher than 10kHz is possible, but accuracy is not guaranteed.
2. Duty ratio measuring range is 10 ~ 95%.
3.
means maximum input voltage is 750V AC (RMS).
C) In HZ/DUTY Measurement Mode:
WARNING
PAY SPECIAL ATTENTION TO AVOID SHOCK WHEN MEASURING HIGH VOLTAGE. DO NOT INPUT VOLTAGE MORE THAN AC 250V
RMS.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Set Selector Switch to position HZ.
3. Connect probe with signal or both ends of load in parallel for measurement.
4. Read LCD.
5. Press Hz/% again to enter duty ratio measuring state.
Note:
1. Frequency measurement range is 10Hz~1kHz When frequency to be tested is more than 10Hz, LCD will show 00.0. Measuring frequency higher than 10 kHz is possible, but accuracy is not guaranteed.
23
4. OPERATING GUIDANCE
4.13 RESISTANCE TEST
WARNING
ELECTRIC SHOCK HAZARD. WHEN MEASURING CIRCUIT IMPEDANCE, DETERMINE THAT THE POWER SUPPLY IS DISCONNECTED AND THE CAPACITOR IN THE CIRCUIT IS COMPLETELY DISCHARGED.
1. Insert black probe to COM jack, insert red probe to INPUT
jack.
2. Place Selector Switch to Ω position. At this time, meter
is in the measurement state. Connect probe to the both ends of resistor or circuit to be
3. tested for measurement.
4. LCD will show readings.
Note:
1. When input end is open, LCD shows “0L” out-of-range state.
2. When resistance to be tested > 1MΩ, the meter reading will be stable after a few seconds, which is normal for high resistance readings.
4.14 DIODE TEST
1. Insert black probe to COM jack, insert red probe to
jack.
Set Selector Switch to
2.
3. Press FUNC key to switch to measuring state.
4. Connect red probe to diode anode and connect black probe
to diode cathode.
5. Read the LCD.
Note:
1. Meter will show approximation of diode forward voltage
drop.
2. If probe has reverse connection or probe is open, LCD will
show “0L”.
24
Ω
INPUT
4. OPERATING GUIDANCE
4.15 CIRCUIT CONTINUITY TEST
WARNING
ELECTRIC SHOCK HAZARD. WHEN MEASURING CIRCUIT CONTINUITY, DETERMINE THAT THE POWER SUPPLY IS DISCONNECTED AND THE CAPACITOR IN THE CIRCUIT IS COMPLETELY DISCHARGED.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Set Selector Switch to Ω
3. Press FUNC key to switch to circuit continuity measuring state.
4. Connect probe to both ends of circuit to be tested. If resistance of circuit to be measured is less than 30Ω
5. may sound. Read circuit resistance value on the LCD.
6.
Note: If probe is open or circuits resistance to be tested is more than 600
Ω, display will show “0L”.
4.16 CAPACITANCE MEASUREMENT
, a buzzer
WARNING
ELECTRIC SHOCK HAZARD. TO AVOID ELECTRIC SHOCK, BEFORE MEASURING CAPACITANCE, DISCHARGE CAPACITANCE COMPLETELY.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Set Selector Switch to .
3. After discharging capacitance completely, connect probe to both
ends of capacitor to be tested.
4. Read capacitance on the LCD.
Note: To improve accuracy below 10nF measuring value, subtract the distributed capacitance of meter and cable.
25
4. OPERATING GUIDANCE
4.17 SURGE CURRENT MEASUREMENT
WARNING
ELECTRIC SHOCK HAZARD. REMOVE THE PROBE FROM THE METER BEFORE MEASURING WITH CURRENT CLAMP.
4.10.1 Set Selector Switch to position A.
4.10.2 Hold trigger, open clamp head, clip one lead of
measurement circuit to be tested in the clamp.
4.10.3 Press REL/INRUSH key more than 2 sec. to enter surge
current measurement mode. LCD will show “- - - -”, until motor start is detected. Meter shows and keeps surge current value.
Read current surge value on LCD display.
4.10.4
Note:
1. Clamping two or more leads of circuit to be tested simultaneously will not get correct measuring results.
2. To get accurate reading, connect lead to be tested at center of current clamp.
3. In manual measuring range mode, if LCD shows “OL”, which indicates over-range, choose a higher measuring range.
4. In manual measuring range mode, if you do not know value to be measured in advance, choose highest measuring range.
5. “ ”means maximum input current is 1000A AC (RMS).
4.18 NCV MEASUREMENT
1. Turn Selector Switch to NCV.
2. Place meter top close to the conductor. When test voltage is greater than 110 VAC (RMS) when meter is close to conductor, meter induction voltage indicator will turn on and buzzer will sound.
26
5. MAINTENANCE
Note:
1. Even if no indication, voltage may exist still. Do not use non-contact voltage detector to judge whether there is voltage in the wire. Detection operation could be affected by socket design, insulation thickness, type and other factors.
2. When inputting voltage on meter input terminal, due to existence of induced voltage, voltage induction indicator also may light.
3. External sources of interference (such as flashlight, motor, etc.) may incorrectly trigger non-contact voltage detection.
5.1 CALIBRATION
This meter should be recalibrated annually( within a temperature range of 65°F to 83°F (18°C to 28°C) and relative humidity less than 75%) .
5.2 BATTERY REPLACEMENT
WARNING
TO AVOID ELECTRIC SHOCK, MAKE SURE THAT THE TEST LEADS HAVE BEEN CLEARLY MOVE AWAY FROM THE CIRCUIT UNDER MEASUREMENT BEFORE OPENING THE BATTERY COVER.
1. When appears, battery should be replaced immediately.
2. Remove screw from battery cover.
3. Replace battery.
4. Replace battery cover and screw.
Note: Do not reverse battery polarity.
5.3 PROBE REPLACEMENT
Replace test leads if damaged or worn.
WARNING
USE MEETS EN 61010-031 STANDARD, RATED CAT III 1000V, OR BETTER TEST LEADS.
WARNING
TO AVOID ELECTRIC SHOCK, MAKE SURE THE PROBES ARE DIS­CONNECTED FROM THE MEASURED CIRCUIT BEFORE REMOVING THE REAR COVER. MAKE SURE THE REAR COVER IS TIGHTLY SCREWED BEFORE USING THE INSTRUMENT.
27
6. ACCESSORIES
6. ACCESSORIES
Description Part # Qty
Test Probe Set DMXTL 1 pair
WARRANTY
CPS Products, Inc. guarantees that all products are free of manufacturing and material defects to the original owner for one year from the date of purchase. If the equipment should fail during the guarantee period it will be repaired or replaced (at our option) at no charge. This guarantee does not apply to equipment that has been altered, misused or solely in need of field service maintenance. All repaired equipment will carry an independent 90-day warranty. This repair policy does not include equipment that is determined to be beyond economical repair.
Test Equipment Depot - 800.517.8431
- 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
28