Maximum Input Current 1000A AC; 1000A DC
Maximum Input Voltage750V AC (RMS); 1000V DC
AC Volts
DC Volts
AC Amps60/600/1000 (+/- 2% reading)
DC Amps60/600/1000 (+/- 2% reading)
OHMs
Frequency
(Through Jaw Clamp)
Frequency
(Through Mode V)
Frequency
(Through Mode HZ/DUTY)
Measuring Range Overload Protection Yes (Full Range)
Sampling RateApprox. 3 times/sec.
LCD Display Auto-OffAfter 15 minutes (if no operation)
Test Leads (Red, Black)39" (1 m), Included
Battery Undervoltage Indication
Polarity IndicationAutomatic ("-" = negative polarity)
Display Indication
(Exceeds Max Range)
Temperature Coefficient<0.1 X accuracy/°C
Clamp, Max Opening1.6" (42 mm)
Full measuring range overload
protection
BatteryOne 9V DC (Included)
Operation Temperature32°F~122°F (0°C ~ 50°C)
Storage Temperature-14°F~122°F (-10°C ~ 50°C)
Size (H x W x D)9.4˝ x 1.9˝ x 3.6˝ (238 x 50 x 92 mm)
Weight (Including Battery)0.92 lb (420 g)
3.2.1.1 For measuring non-sinusoidal wave signals, use true RMS
measuring method, which has less error than traditional
average response measuring method.
This true RMS meter can accurately measure non-sinusoi-
3.2.1.2
dal wave signals, but in AC function mode, when there is
no signal to be measured (input terminal short circuit in AC
voltage mode), clamp meter may show a reading from 1 to
50. These deviating readings are normal. In the designated
measurement range, they will not affect the accuracy for
multimeter measuring AC.
3.2.1.3 True RMS can be measured only when input signal reaches
a certain level. Therefore, the measuring range of AC
voltage and current should be specified at 2% ~100% of
full range.
3.2.2 AC CURRENT
Measuring rangeResolutionAccuracy
60A0.01A
1000A1A
Maximum input current: 1000A AC
•
• Maximum input current: 0~600A: 40~400Hz;
600A~1000A: 40~60Hz
±(2.5% reading + 8 digits)600A0.1A
12
3. SPECIFICATIONS
3.2.3 DC CURRENT
Measuring rangeResolutionAccuracy
60A0.01A
1000A1A
•
Maximum input current: 1000A DC
3.2.4 SURGE CURRENT
Measuring rangeResolutionAccuracy
60A0.01A< 60A for reference only
600A0.1A±(5% reading + 60 digits)
1000A1A
Time of integration:
Measurement range: 20-1000A
Frequency range: 40-400Hz
3.2.5 DC VOLTAGE
Measuring rangeResolutionAccuracy
600mV0.1mV
6V0.001V
60V0.01V
600V0.1V
1000V1V±(0.8% reading + 4 digits)
• Input impedance: 10M
• Maximum input voltage: 750V AC (RMS) or 1000V DC
Note: In the small voltage measuring range, the probe is not
connected with the circuit to be tested, and the meter may have
fluctuating readings, which is normal and caused by the meter’s
high sensitivity. This does not affect actual measurement results.
100ms;
±(2.0% reading + 8 digits)600A0.1A
±(0.5% reading + 5 digits)
13
3. SPECIFICATIONS
3.2.6 AC VOLTAGE
Measuring rangeResolutionAccuracy
600mV0.1mV
6V0.001V
60V0.01V
600V0.1V
750V1V±(0.8% reading + 4 digits)
•
Input impedance: 10M
• Maximum input voltage: 750V AC (RMS) or 1000V DC
• Frequency range: 40 - 400Hz
Note: In the small voltage measuring range,the probe is not connected with the circuit to
be tested, and the meter may have fluctuating readings, which is normal and caused by
the meter’s high sensitivity. This does not affect actual measurement results.
3.2.7 FREQUENCY
3.2.7.1 Clamp head measuring frequency (through range A):
Measuring rangeResolutionAccuracy
99.99Hz0.01Hz
999.9Hz0.1Hz
• Measuring scope: 10Hz-1kHz
• Input signal range: 20A AC (RMS) (input current will increase
when the frequency to be measured increases)
• Maximum input current: 1000A (RMS)
3.2.7.2 Through mode V:
Measuring rangeResolutionAccuracy
99.99Hz0.01Hz
9.999Hz0.001kH
• Measuring scope: 10Hz - 10kHz
• Input signal range: 20A AC (RMS) (input current will increase when
the frequency to be measured increases)
• Input impedance: 10M
14
• Maximum input voltage :750V AC (RMS)
±(0.8% reading + 5 digits)
±(1.5% reading + 5 digits)
±(1.5% reading + 5 digits)999.9Hz0.1Hz
3. SPECIFICATIONS
3.2.7.3 Through mode HZ/DUTY:
Measuring rangeResolutionAccuracy
9.999Hz0.01Hz
99.99Hz0.1Hz
999.9Hz0.001kHz
9.999kHz0.01kHz
99.99kHz0.1kHz
999.9kHz0.001MHz
9.999MHz0.001kH
• Overload protection: 250V DC or AC (RMS)
• Input voltage range: 2V (input voltage will increase when the
frequency to be measured increases)
3.2.8 Duty Ratio
Measuring rangeResolutionAccuracy
0.1-99.9%0.1%±3.0%
3.2.8.1 Through mode A (from clamp head):
• Frequency response:10-1kHz
• Input current range: ≥ 20A AC (RMS)
• Maximum input current: AC 1000A
3.2.8.2 Through mode V:
• Frequency response: 10-10kHz
• Input voltage range: ≥ 60mV AC
• Input impedance: 10M
• Maximum input voltage: 750V AC (RMS)
3.2.8.2 Through mode HZ/DUTY:
• Frequency response: 10 - 10MHz
• Input voltage range: ≥ 2V AC (RMS) (input voltage will increase
when the frequency to be measured increases)
3.2.10 CIRCUIT CONTINUITY TEST
Measuring rangeResolutionFunction
0.1Ω
• Overload protection: 250 V DC or AC (RMS)
3.2.11 CAPACITANCE
Measuring rangeResolutionAccuracy
9.999nF0.001nF
99.99nF0.01nF
999.9nF0.1nF
9.999μF0.001μF
99.99μF0.01μF
999.9μF0.1μF
9.999mF0.001mF
99.99mF0.01mF
• Overload protection: 250 V DC or AC (RMS)
±(0.8% reading + 3 digits)
±(2.0% reading + 5 digits)
If the resistance of circuit to
be measured is less than 50Ω,
the meter’s built-in buzzer may
sound.
±(3.0% reading + 5 digits)
16
3. SPECIFICATIONS
3.2.12 DIODE TEST
Measuring
range
• Forward DC current is about 1mA
• Backward DC voltage is about 3.3V
• Overload protection: 250V DC or AC (RMS)
4.1 READING HOLD
1. During measurement, if reading hold is required, press “HOLD/BL”
key and the value on the display will be locked. Press “HOLD/BL”
key again to cancel reading hold state.
4.2 MANUAL MEASURING RANGE
1. RANGE key is automatic/manual measuring range key to trigger
mode. The preset range is automatic measuring range. Press to
switch to manual measuring range. In manual measuring range
mode, click once to change to upper range. Continue to the top
range, then continue to press this key to change to the bottom
range, followed by recycling. If this key is pressed more than 2
seconds, it will switch back to automatic measuring range state.
Note: In capacitance and frequency measurement state, rhw manual
measuring range key is invalid.
4.3 FREQUENCY/DUTY RATIO SWITCH
1. When meter is in AC voltage mode, if “Hz/%” key is pressed,
the meter will measure Hz, and measure AC voltage, AC current
signal frequency. Click “Hz/%” key again and meter will measure
DUTY cycle, voltage and current signal duty ratio. If in HZ/DUTY
position, pressing HZ % key will switch between HZ and DUTY
by recycling.
2. If “Hz/%” key is pressed again, meter will revert to voltage,
current measurement state.
ResolutionFunction
0.001V
Display approximate diode
forward voltage value
17
4. OPERATING GUIDANCE
Note: If meter is in maximum/minimum value measurement state,
it can not switch to frequency, duty ratio measurement mode.
4.4 MAX/MIN MEASUREMENT CHOICE
1. Press “MAX/MIN” key to enter MAX mode, and always keep
measurement maximum value; press “MAX/MIN” key again, to
enter minimum value measurement state; press “MAX/MIN”
key for the third time and the meter will display the difference of
maximum and minimum value; press “MAX/MIN” key to repeat
the above operations by recycling.
2. After entering MAX or MIN mode, meter will automatically save
the measured maximum or minimum value.
3.
If the user presses “MAX/MIN” key more than 2 sec, meter will
restore normal measuring range.
Note:
a) When meter is in maximum/minimum value measurement
state, it is in manual measuring range mode.
b) When meter is in frequency
can not switch to maximum/ minimum value measurement mode.
4.5 FUNCTION SWITCH
1.
In resistance mode, press
among resistance, diode and continuity detection by recycling.
2.
In voltage and current mode, press
between AC and DC.
4.6 REL/INRUSH MEASUREMENT
1.
REL/INRUSH key is relative value measurement key
this key will enter relative value measurement mode. The
current display value can be stored in the memory as reference
value. When the user measures later, the display value is
the difference for input value minus reference value. ie.
REL (current reading)= Input value - Reference value.
2.
The relative value measurement can only be performed in
manual mode.
18
, duty ratio measurement state, it
FUNC key and meter will switch
FUNC key to switch
. Pressing
4. OPERATING GUIDANCE
3. In AC current measurement state, press REL/INRUSH more than
2 seconds to enter surge measuring state.
4.7 BACKLIGHT AND CLAMP HEAD LIGHT
1. During measurement, if ambient light is too dark to read the
display, press and hold “BL/HOLD” key to activate backlight.
Backlight will automatically turn off after 30 seconds.
2. During this period, pressing “BL/ HOLD” key more than two
seconds will turn OFF backlight.
3. In the current mode, meter will turn backlight and clamp head
light on. Backlight uses an LED with high current draw. Backlight
will turn off after 30 seconds. Using backlight often will shorten
battery life.
Note: When battery voltage ≤ 7.2 V, the LCD displays “” undervoltage symbol. If using backlight and battery voltage drops below
7.2 V due to high working current, the symbol may appear and
measurement accuracy is not guaranteed. Continue to use meter
normally without using the backlight. Do not replace battery until
symbol shows under normal conditions.
4.8 AUTOMATIC POWER-OFF
1. After powering on, and if meter isn’t used for 15 minutes, meter
will enter suspended state, automatically powering off to save the
battery. Within 1 minute before shutdown, buzzer will sound five
times. Meter will then enter a dormant state.
2. After automatic power-off, press FUNC key, to turn meter on
again.
3. Holding FUNC key when powering on cancels automatic poweroff function.
19
4. OPERATING GUIDANCE
4.9 MEASUREMENT PREPARATION
1. Rotate the Selector Switch to turn on power. When battery voltage
is low (about ≤ 7.2 V), LCD displays symbol. Replace
battery.
2.means that voltage or current should not be more than the
specified value to protect internal line from damage.
3. Set Selector Switch to required measuring function and range.
4. When connecting line, first connect common test line, then
charged test line. When removing the line, remove charged test
line first.
4.10 CURRENT MEASUREMENT
WARNING
ELECTRIC SHOCK HAZARD. REMOVE PROBE FROM METER BEFORE
MEASURING WITH CURRENT CLAMP.
1. Set Selector Switch to position A. At this time, meter is in AC current
measurement state. Choose appropriate measuring range.
2. If you want to measure DC current, press FUNC key to enter direct
current measurement state.
3. Hold trigger, open clamp head, clip one lead of measurement circuit to
be tested in the clamp.
4. Read current value on LCD display.
Note:
1. Clamping two or more leads of circuit to be tested simultaneously will
not give correct measuring results.
2. To get accurate readings, connect lead to be tested at center of current
clamp.
3. indicates that maximum input AC current is 1000 A
4. To improve measurement precision, in DC current measurement state,
if LCD display is not zero, press REL to return to zero, then measure.
20
4. OPERATING GUIDANCE
4.11 VOLTAGE MEASUREMENT
WARNING
ELECTRIC SHOCK HAZARD. PAY SPECIAL ATTENTION TO AVOID
SHOCK WHEN MEASURING HIGH VOLTAGE. DO NOT INPUT
VOLTAGE MORE THAN AC750 RMS.
1. Insert black probe to COM jack, insert red probe to INPUT jack,
choose appropriate measuring range.
2. Place Selector Switch to AC voltage V or mV position. At this
time, meter is in the DC voltage measurement state. To measure
AC voltage, press FUNC key to enter AC voltage measurement
state.
Connect probe with voltage source or both ends of load in
3.
parallel for measurement.
4. Read the voltage on the LCD.
Note:
1. In small voltage measuring range, probe is not connected with
circuit to be tested and meter may have fluctuating readings,
which is normal and caused by the meter’s high sensitivity
When meter is connected with circuit to be tested, you will get
actual measured value.
In relative measurement mode, automatic measuring range is
2.
invalid.
3.Indicates maximum input AC current is 750V AC or 1000V
DC Maximum input voltage is mode mV is 600 mV DC or AC.
4. If readings measured by meter are more than 750C RMS AC, a
“beep” alarm will sound.
.
21
4. OPERATING GUIDANCE
4.12 FREQUENCY AND DUTY RATIO MEASUREMENT
A) Clamp head measuring frequency (through AC current):
WARNING
ELECTRIC SHOCK HAZARD. REMOVE PROBE FROM METER
BEFORE MEASURING WITH CURRENT CLAMP.
1. Set Selector Switch to position A .
2. Hold trigger, open clamp head, clip one lead of measurement
circuit to be tested in the clamp.
3. Press Hz/% key to switch to frequency measuring state.
4.
Read current value on the LCD display.
5. Pressing Hz/% again enters duty ratio measuring state.
Note:
1. Clamping two or more leads of circuit to be tested simultaneously
will not get correct measuring results.
2. Frequency measurement range is 10Hz~1kHz the frequency to
be tested is less than 10Hz, or if frequency is higher than 10 kHz,
accuracy is not guarantee
3. Duty ratio measuring range is 10 ~ 95%.
4. “ ”means that maximum input current is 1000A AC (RMS).
B) In Voltage Measurement Mode:
WARNING
ELECTRIC SHOCK HAZARD. PAY SPECIAL ATTENTION TO AVOID
SHOCK WHEN MEASURING HIGH VOLTAGE. DO NOT INPUT
VOLTAGE MORE THAN AC 750 RMS.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Place Selector Switch to V or mV position, press FUNC to enter
AC voltage measurement state.
3. Press Hz/% key to switch to frequency measuring state.
22
4. OPERATING GUIDANCE
4. Connect probe with signal or both ends of load in parallel for
measurement.
5. VRead on LCD.
6. Pressing Hz/% again enters duty ratio measuring state.
Note:
1. Frequency measurement range is 10Hz~1kHz When the frequency
to be tested is less than 10Hz,the LCD will show 00.0 Measuring
frequency higher than 10kHz is possible, but accuracy is not
guaranteed.
2. Duty ratio measuring range is 10 ~ 95%.
3.
means maximum input voltage is 750V AC (RMS).
C) In HZ/DUTY Measurement Mode:
WARNING
PAY SPECIAL ATTENTION TO AVOID SHOCK WHEN MEASURING
HIGH VOLTAGE. DO NOT INPUT VOLTAGE MORE THAN AC 250V
RMS.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Set Selector Switch to position HZ.
3. Connect probe with signal or both ends of load in parallel for
measurement.
4. Read LCD.
5. Press Hz/% again to enter duty ratio measuring state.
Note:
1. Frequency measurement range is 10Hz~1kHz When frequency
to be tested is more than 10Hz, LCD will show 00.0. Measuring
frequency higher than 10 kHz is possible, but accuracy is not
guaranteed.
23
4. OPERATING GUIDANCE
4.13 RESISTANCE TEST
WARNING
ELECTRIC SHOCK HAZARD. WHEN MEASURING CIRCUIT
IMPEDANCE, DETERMINE THAT THE POWER SUPPLY IS
DISCONNECTED AND THE CAPACITOR IN THE CIRCUIT IS
COMPLETELY DISCHARGED.
1. Insert black probe to COM jack, insert red probe to INPUT
jack.
2. Place Selector Switch to Ω position. At this time, meter
is in the measurement state.
Connect probe to the both ends of resistor or circuit to be
3.
tested for measurement.
4. LCD will show readings.
Note:
1. When input end is open, LCD shows “0L” out-of-range state.
2. When resistance to be tested > 1MΩ, the meter reading
will be stable after a few seconds, which is normal for high
resistance readings.
4.14 DIODE TEST
1. Insert black probe to COM jack, insert red probe to
jack.
Set Selector Switch to
2.
3. Press FUNC key to switch to measuring state.
4. Connect red probe to diode anode and connect black probe
to diode cathode.
5. Read the LCD.
Note:
1. Meter will show approximation of diode forward voltage
drop.
2. If probe has reverse connection or probe is open, LCD will
show “0L”.
24
Ω
INPUT
4. OPERATING GUIDANCE
4.15 CIRCUIT CONTINUITY TEST
WARNING
ELECTRIC SHOCK HAZARD. WHEN MEASURING CIRCUIT
CONTINUITY, DETERMINE THAT THE POWER SUPPLY IS
DISCONNECTED AND THE CAPACITOR IN THE CIRCUIT IS
COMPLETELY DISCHARGED.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Set Selector Switch to Ω
3. Press FUNC key to switch to circuit continuity measuring state.
4. Connect probe to both ends of circuit to be tested.
If resistance of circuit to be measured is less than 30Ω
5.
may sound.
Read circuit resistance value on the LCD.
6.
Note: If probe is open or circuits resistance to be tested is more than
600
Ω, display will show “0L”.
4.16 CAPACITANCE MEASUREMENT
, a buzzer
WARNING
ELECTRIC SHOCK HAZARD. TO AVOID ELECTRIC SHOCK,
BEFORE MEASURING CAPACITANCE, DISCHARGE CAPACITANCE
COMPLETELY.
1. Insert black probe to COM jack, insert red probe to INPUT jack.
2. Set Selector Switch to .
3. After discharging capacitance completely, connect probe to both
ends of capacitor to be tested.
4. Read capacitance on the LCD.
Note: To improve accuracy below 10nF measuring value, subtract the
distributed capacitance of meter and cable.
25
4. OPERATING GUIDANCE
4.17 SURGE CURRENT MEASUREMENT
WARNING
ELECTRIC SHOCK HAZARD. REMOVE THE PROBE FROM THE
METER BEFORE MEASURING WITH CURRENT CLAMP.
4.10.1 Set Selector Switch to position A.
4.10.2 Hold trigger, open clamp head, clip one lead of
measurement circuit to be tested in the clamp.
4.10.3 Press REL/INRUSH key more than 2 sec. to enter surge
current measurement mode. LCD will show “- - - -”, until
motor start is detected. Meter shows and keeps surge
current value.
Read current surge value on LCD display.
4.10.4
Note:
1. Clamping two or more leads of circuit to be tested
simultaneously will not get correct measuring results.
2. To get accurate reading, connect lead to be tested at center of
current clamp.
3. In manual measuring range mode, if LCD shows “OL”, which
indicates over-range, choose a higher measuring range.
4. In manual measuring range mode, if you do not know value to
be measured in advance, choose highest measuring range.
5. “ ”means maximum input current is 1000A AC (RMS).
4.18 NCV MEASUREMENT
1. Turn Selector Switch to NCV.
2. Place meter top close to the conductor. When test voltage is
greater than 110 VAC (RMS) when meter is close to conductor,
meter induction voltage indicator will turn on and buzzer will
sound.
26
5. MAINTENANCE
Note:
1. Even if no indication, voltage may exist still. Do not use non-contact voltage
detector to judge whether there is voltage in the wire. Detection operation could
be affected by socket design, insulation thickness, type and other factors.
2. When inputting voltage on meter input terminal, due to existence of induced
voltage, voltage induction indicator also may light.
3. External sources of interference (such as flashlight, motor, etc.) may incorrectly
trigger non-contact voltage detection.
5.1 CALIBRATION
This meter should be recalibrated annually( within a temperature range
of 65°F to 83°F (18°C to 28°C) and relative humidity less than 75%) .
5.2 BATTERY REPLACEMENT
WARNING
TO AVOID ELECTRIC SHOCK, MAKE SURE THAT THE TEST LEADS
HAVE BEEN CLEARLY MOVE AWAY FROM THE CIRCUIT UNDER
MEASUREMENT BEFORE OPENING THE BATTERY COVER.
1. When appears, battery should be replaced immediately.
2. Remove screw from battery cover.
3. Replace battery.
4. Replace battery cover and screw.
Note: Do not reverse battery polarity.
5.3 PROBE REPLACEMENT
Replace test leads if damaged or worn.
WARNING
USE MEETS EN 61010-031 STANDARD, RATED CAT III 1000V, OR
BETTER TEST LEADS.
WARNING
TO AVOID ELECTRIC SHOCK, MAKE SURE THE PROBES ARE DISCONNECTED FROM THE MEASURED CIRCUIT BEFORE REMOVING
THE REAR COVER. MAKE SURE THE REAR COVER IS TIGHTLY
SCREWED BEFORE USING THE INSTRUMENT.
27
6. ACCESSORIES
6. ACCESSORIES
DescriptionPart #Qty
Test Probe SetDMXTL1 pair
WARRANTY
CPS Products, Inc. guarantees that all products are free of manufacturing
and material defects to the original owner for one year from the date
of purchase. If the equipment should fail during the guarantee period it
will be repaired or replaced (at our option) at no charge. This guarantee
does not apply to equipment that has been altered, misused or solely
in need of field service maintenance. All repaired equipment will carry
an independent 90-day warranty. This repair policy does not include
equipment that is determined to be beyond economical repair.
Test Equipment Depot - 800.517.8431
- 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com