CPS TEMP-SEEKER TM360 Operation Manual

TM360 TEMP-SEEKER
PRECISION THERMO-PSYCHROMETER
®
OPERATION MANUAL
www. .com
information@itm.com1.800.561.8187
GENERAL INFORMATION
Table of Contents
General Information 2-3 Unit Operation 4-6 TMX3RH Method 7 Wet-Sock Method 8 Psychrometric Chart 9 Accessory Probes 10-12 Warranty & Contact Information 13
TM360 Overview
The TM360 TEMP-SEEKER® is a high resolution thermo-psychrometer adaptable to accommodate all your temperature, humidity, dry/wet bulb and dew point measurements. The instrument offers 4 separate temperature channels for high accuracy thermistor probes, and an auxiliary port to connect a state of the art solid state RH / DB probe. The TM360’s memory functions continuously store the MIN and MAX readings from all the ports. Press the SCAN key and the display will alternate between all 4 ports in three second intervals providing continuous monitoring of multiple probe points during critical system diagnosis. The TM360 has an illuminated, low power LCD display making it easy to take readings in poorly illuminated machine rooms as well as provisions to connect to an external power adaptor to conserve battery life during extended monitoring applications. Designed in a rugged polyethylene case the TM360 is engineered to provide laboratory accuracy in rough field conditions. A generous probe storage area is provided for housing all your job specific probes directly in the unit itself. Probes can remain plugged in, reducing life shortening stress on the components.
Features:
• Four discrete temperature channels
• High accuracy silicon humidity transducer probe port
• Large, 4-digit, easy-to-read, backlit display
• Min / Max and memory functions
• Temperature differential between any 2 of the 4 temperature ports
• Direct relative humidity and wet bulb measurements
• Calibrated to NIST traceable standards
• Auto-off after 10 minutes of non-use or continuous ON mode
• Operates 30 continuous hours on 1 9V alkaline battery
• Integrated battery eliminator jack
• Rugged polyethylene carrying case
• Large probe storage area
• Switchable ˚C / ˚F
• 1 year warranty
2
www. .com
information@itm.com1.800.561.8187
TM360 Unit Specifications
Resolution ±0.1˚F or ˚C
GENERAL INFORMATION
Temperature Accuracy
RH Range 0 to 100%RH
RH Accuracy (TMX3RH) ±1.8%RH from 10 to 90%RH; ± 3% elsewhere in the range
DB Accuracy (TMX3RH)
Power Source
Battery Life 30hrs. continuous use
Automatic Power Off
Backlight 15 seconds
Dimensions 12" X 10" X 3" (30.5cm x 25.4cm x 7.6cm)
Weight 3lb (1.4 kg) with probes
Warranty 1 year
±0.4˚F / ±0.2˚C from 14.0 to 158˚F / -10.0 to 70˚C; 0.6˚F /
0.3˚C elsewhere in the range
±1.2% of reading ± 1.0˚F from -40˚F to 200˚F (± 1.2% of reading ± 0.5˚C from -40˚C to 90˚C)
9V alkaline battery or optional TMX3PS AC adapter for 100 to 240VAC (40 to 60 Hz)
After 10 minutes of non-use or user selectable continuous ON mode
Calibrated to NIST traceable standards
Only qualified service personnel should operate this unit. Some countries, states, etc…may require the user to be licensed. Please check with your local government agency.
Always wear gloves
Always wear safety glasses
3
www. .com
information@itm.com1.800.561.8187
UNIT OPERATION
Battery CompartmentBattery
KeypadLCD Display
4 Temperature PortsRelative Humidity Port
(Indoor Air Quality)
Jack
Eliminator
TMX3PS - Optional Power Supply Default Wet Bulb Ports A & B (See Also Wet Sock Method on pg. 8)