4
Micro+ Operating Manual
English
Warnings
● If an unexpectedly high CO reading
is displayed, this could be due to CO
poisoning. Seek further medical advice.
● Never use alcohol or cleaning agents
containing alcohol or other organic
solvents as these vapours will damage
the CO sensor inside.
● Under no circumstances should the
instrument be immersed in liquid or
splashed with liquid.
● People with lung disease or chest
ailments may not be able to achieve
the breath-hold. In such cases, the
user should inhale and hold their
breath when the breath test is started,
and exhale, if necessary, before the
countdown has completed. This may
result in a lower reading, but may still
provide an indication of a smoking habit.
● During start-up, if the
icon is
displayed then calibration is due.
Please refer to page 13 for instructions.
The unit may give false readings if not
calibrated.
● During start-up if the
icon
is displayed then the D-piece requires
changing. It is recommended that the
D-piece is changed every month or
earlier if visibly soiled.
● The battery life is indicated by the
icon. When the
icon is displayed
the batteries should be changed.
● The disposable cardboard mouthpieces
are single-use only as re-use can
increase the risk of cross-infection.
● See Bedfont’s Infection Control and
Maintenance Guidelines for further
information on infection control.
● Changing prole settings does not affect
the smoker’s CO reading, only the trafc
light colour classication displayed.
● Care should be taken not to cover the
exhaust port of the instrument during
a breath test. Labels should never be
placed over this outlet.
● The USB port should be used to
connect to a PC, in order to run the
COdata+ PC software please ensure
that the monitor is only connected to
a computer that is manufactured in
accordance with EN60950.
● Keep the PC out of reach of the patient
when the Micro
+
is connected.
Contraindications
● The sensor has a cross-sensitivity to
hydrogen which could affect the CO
result. Hydrogen could be present on
the breath due to certain gastrointestinal
conditions. The sensor also has a crosssensitivity to nitric oxide and ethylene.
● Temperature may affect the accuracy of
the Micro
+
. The instrument should be
calibrated at the temperature at which
it is expected to be used. If the Micro
+
is used at a lower temperature than
when it was calibrated the readings
may be lower. When used at a higher
temperature than calibrated the
readings may be higher.
Caution
Federal law restricts this device to sale by
or on the order of a physician.
www.covita.net | tel (800) 707-5751