•You introduced software that is not authorized by Coulter into your computer. Only operate your|
system’s computer with software authorized by Coulter.|
Coulter Corporation urges its customers to comply with all national health and safety standards such as
the use of barrier protection. This may include, but it is not limited to, protective eye wear, gloves, and
suitable laboratory attire when operating or maintaining this or any other automated laboratory analyzer.
In the USA, for Service call Coulter Customer Operations 1-800-526-7694.
1GENERAL SERVICE AND MAINTENANCE, 1
1.1LASER MAINTENANCE, 1
1.2GENERAL PROCEDURES, 1
Power Down, 3
Power Up, 3
Optimize the Disk, 3
Daily, 4
Weekly, 4
Monthly, 4
Set Screen Saver, 5
Position Rocker Bed, 6
Diluter F Key Functions, 6
All service and maintenance of the laser module must be done by trained
Coulter personnel. Removal of the module must be performed only by a
Coulter Representative.
1.2GENERAL PROCEDURES
GENERAL SERVICE AND MAINTENANCE 1
1
GENERAL
WARNING
Do not try to remove the laser module from the Diluter.
IMPORTANT
If a power failure or reset occurs during a cycle, after power up, press
PRIME APERT before cycling samples.
This section details the cleaning, replacement, and adjustment
procedures that are your responsibility. General operations that may
apply to numerous procedures are described below. They include:
Power Down
Power Up
Optimize the Disk
Set Screen Saver
Position Rocker Bed
Diluter
Coulter recommends a Preventive Maintenance Inspection (PMI) once a
year by a Coulter Service Representative.
Figure 1 illustrates component location on the Diluter front panel.
F Key Functions
PN 4237187A (November 1993)1
GENERAL SERVICE AND MAINTENANCE
A
GENERAL
67
1
8
9
VL30 VL31 VL32 VL33 VL34 VL35 VL36 VL37
VL19
10
11
12
GA1
VC15
RG2
PM8
VL43
VL38
VL40
VL44
PM7
VL20 VL18 VL22
VL47
VL46
VL48
VL50
SOL47
VC14
VL49
13
7187002
PM10
PM9
14
17. HGB CUVETTE HOLDER
18. BATH OVERFLOW CHAMBER (VC10)
19. NEEDLE-RINSE PUMP (PM1)
20. WBC DILUENT DISPENSER (PM3)
21. RINSE MIX CHAMBER PUMP (PM11)
22. 1mL BACKWASH PUMP (PM6)
23. RBC DILUENT DISPENSER (PM2)
171615
VL14VL15VL16
VL13
VL4VL7
12345
VL3
VL2
VL1
PM4
VL6
VL5
9. ELECTRONIC MANOMETER WINDOW
10. MANOMETER RANGE SCALE
11. CBC LYTIC REAGENT PUMP (PM8)
12. SHEATH TANK (VC15)
13. 5mL CLEANING AGENT PUMP (PM10)
14. 5mL HGB PUMP (PM9)
15. FOAM TRAP (VC14)
16. WASTE CHAMBER (VC11)
SOL1
SOL5
PM11
1. ERYTHROLYSE II PUMP (PM4)
2. RBC BATH (VC2)
3. VACUUM ISOLATOR (VC1)
4. WBC BATH (VC3)
5. STABILYSE PUMP (PM12)
6. CBC LYTIC REAGENT PUMP (PM7)
7. MANOMETER REGULATOR
8. SHEATH FLOW RESTRICTOR
PM6
2221201918
PM = PUMP
PM1
VL11VL10
PM3
VL12
PM2
Figure 1 Diluter Front Panel
VC = VACUUM CHAMBER
23
2PN 4237187A (November 1993)
GENERAL PROCEDURES
Power Down
1.Turn off the Graphic Printer.
2.Turn off the Ticket Printer.
3.Turn off the DMS.
4.Press
5.Turn off the MAIN POWER circuit breaker on the Power Supply.
6.Unplug the primary power cord from the wall outlet.
7.Verify that the power is disconnected:
8.Turn off the MAIN POWER circuit breaker.
POWER OFF on the Diluter keypad.
a.Turn on the MAIN POWER circuit breaker.
b.Check the front panel of the Power Supply and be sure there
are no lamps lighted. If any lamp lights, the power is still
connected; you must repeat steps 5, 6 and 7.
1
GENERAL
Power Up
Optimize the Disk
1.Plug the primary power cord into the wall outlet.
2.Turn on the MAIN POWER circuit breaker on the Power Supply.
3.Press
4.Turn on the DMS.
5.Turn on the Ticket Printer.
6.Turn on the Graphic Printer.
This system is equipped with the OPTune™ utility, a software utility
that optimizes the hard drive on your DMS. Optimizing organizes files
on your hard drive so that the DMS is faster and more efficient. The
OPTune utility runs when you turn on the DMS and when you reset the
DMS.
POWER ON on the Diluter keypad.
PN 4237187A (November 1993)3
GENERAL SERVICE AND MAINTENANCE
GENERAL
1
The OPTune utility performs three types of optimization: daily, weekly
and monthly. The type it uses depends on how long it has been since the
last optimization.
Daily
The fastest method, this type leaves each file
•100% defragmented
•sorted in ascending order by name
•in contiguous order
Weekly
Takes about 50% longer than the daily method. Does everything the
daily method does, plus it optimizes in the packed mode. It arranges the
files on your hard drive so that they are end-to-end, with no space
between them. Thus, new files are likely to be written to disk without
being fragmented.
Monthly
The most thorough method. Does everything the weekly method does,
plus it physically arranges files on the disk in the same order as the
sorted directory entries. This method takes longer, but increases
efficiency when accessing many files in sequential order.
The OPTune utility initially takes about 45 minutes to an hour to
optimize a 240MB hard drive. After that, daily optimization takes about
30 to 40 seconds. The OPTune utility reorganizes only those files that
need it; it does not reoptimize unnecessarily.
You can bypass the OPTune utility process by pressingwhile the
OPTune utility screen is in view. When the message Do you want to stop?[Y/N] appears, press. This message appears: User terminated Verify.
The optimizing process stops and the system goes into the DMS
program.
If there is a hard disk failure during the optimization process, for
example lost clusters or cross-linked files, the OPTune utility prompts
you to continue. Answer Yes to go on with the process. However, if this
happens more than once a week, record the incident in your DMS
maintenance log or your logbook and call Coulter Customer Operations.
4PN 4237187A (November 1993)
GENERAL PROCEDURES
If you always leave your system turned on, we recommend that you run
the OPTune utility at least once a week, using one of these two options:
1
Option A
1.Turn off the Graphic Printer.
2.Turn off the DMS.
3.Wait at least 15 seconds.
4.Turn on the DMS.
5.Turn on the Graphic Printer.
Option B
Select
Special Functions
Sample Analysis Set up
Reset DMS
When you reset the DMS or turn it off then back on, these functions
default to these conditions:
GENERAL
Set Screen Saver
•Data Base storage: ON (DB↑)
•XB: ON (XB↑)
•AutoPrint: NONE (PR↓)
•AutoTransmit: OFF (HC↓)
Reset these options as needed according to your laboratory’s protocol.
The screen saver function makes the screen go blank when it is not being
used. Figure 2 illustrates the Setup window. Select
Special Functions
Set Up
System Set Up
Screen Saver Time Out
PN 4237187A (November 1993)5
GENERAL SERVICE AND MAINTENANCE
GENERAL
1
Set the time, in minutes, for the length of time to elapse before the screen
goes blank. You can enter from 0 to 99 minutes.
•A number from 1 to 99 means that, if there is no keystroke in that
many minutes, the screen goes blank. The screen also goes blank
when the compressor times out.
The screen turns back on under these conditions:
-The DMS receives a sample run.
-You press a key on the DMS keyboard. The DMS executes the
keystroke as it applied to the screen before it went blank.
•Zero means the screen saver will not be activated.
Sample AnalysisControlsStart UpSpecial Functions
DMS 2ADiagnostics
Control set up
Sample analysis set up
C System set up|
Screen Saver Setupanguage
Screen Saver Delay: 0 Minutes QAP ID #|
Change Supervisor Password
Screen Saver Time Out|
(931951) COPYRIGHT 1989-1993 COULTER CORPORATION, ALL RIGHTS RESERVED
F1-Help F3-Error Log F5-Other F9-Main menu F10-Previous menu
07/14/93 17:17 OPR DMS↑ PR↓ TK↓ HC↓ DB↑ XB↓ WL↑ HWL↓ QC↑59
CBC/Diff/Retic
Set Up|
Shift
Reagents
Institution
Communication def
Electronic reference
atabase
Figure 2 Screen Saver Setup
Position Rocker Bed
When the compressor is off, the rocker bed is not locked in position and
can be tilted easily.
Diluter F Key Functions
The Diluter F key functions are listed in Table 1. To select these
functions, press
6PN 4237187A (November 1993)
F X X ENTER where XX is the function number.
GENERAL PROCEDURES
During these functions, a related message appears on the Diluter keypad;
when the routine is finished, the message returns to the function
number.
For example, when you select F01, the burn circuit activates and the
message appears;
CLEAN APERTURES
after which the message returns to:
FUNCTION = 01
1
GENERAL
To repeat the function, press
Table 1 Diluter F Key Functions
FunctionMessageDescription
F01
F02
F03
F04
F05
F06
CLEAN APERTURES
PRIME CBC LYSE
RBC AND WBC MIX
DILUENT DISPENSE
SOLENOID XX
RELEASE B.S.V.
The burn circuit activates. This function applies current across
the apertures to "burn" protein or debris off the apertures.
The following sequence occurs twice:
•Drains both baths then dispenses CBC lytic reagent into the
WBC bath three times to prime the lyse line.
•The WBC bath drains and the cycle repeats.
•It then fills both baths with diluent and primes the apertures.
Causes mixing bubbles to enter both the RBC and WBC baths.
Dispenses 1 mL of diluent through the Secondary mode
aspirator tip.
Use to manually activate individual solenoids (see Table 7).
Enter the desired solenoid number (for numbers less than 10,
enter a leading zero), and press ENTER.
•Drains the baths
•Bleeds off pressure
•Rotates the BSV back and forth five times
ENTER. To exit the function, press STOP.
Use to check BSV function, free a stuck valve, or when cleaning
the BSV to ease removal. Press STOP twice to exit the function.
This function does not refill the baths; use RINSE to refill them
with diluent.
continued
PN 4237187A (November 1993)7
GENERAL SERVICE AND MAINTENANCE
Table 1 Diluter F Key Functions
FunctionMessageDescription
continued
F07
GENERAL
BARCODE READ RATE
TEST
Checks the bar-code read rate on the cassette bar-code label,
and on the four types of tube bar-code labels.
1
F08
F09
PRIME SWEEP FLOW
ZAP APERTURES
1.Enable the bar-code reader.
2.Remove the cap-piercing needle so that extensive piercings
do not cause a plugged needle.
3.Place a cassette of bar-code labeled tubes on the rocker
bed.
4.Select the function, then the following appears on the
Diluter display:
A00B00C00D00E00
A represents Interleaved 2-of-5
B represents Code 39®Bar code
C represents Codabar
D represents the cassette label
E represents code 128
The cassette automatically moves to the piercing station
and the bar-code reader scans both the cassette and tube
labels. On the display, the appropriate category increments
by one each time a label from that category is read. Press
STOP twice to exit the function.
Primes the sweep-flow lines.
•Applies pressure behind the baths
•Drains and rinses the baths four times with cleaning agent
•Draws the cleaning agent behind the apertures
•Zaps the apertures 10 times
After this function is complete, you must execute the startup
cycle to purge the system of the cleaning agent.
F10
F11
F12
F13
RAISE NEEDLE
EXTENDED PRIME
EXTENDED CLEAR
PURGE MODE
Causes the piercing needle to rise and the rocker bed to lock in
the backward position. Press STOP twice to exit.
Pulls fluid through the apertures to the vacuum isolator chamber
for 60 seconds. Aperture current turns on and aperture lamp
intensity increases for viewing the aperture screen.
Clears the apertures for 60 seconds. Use to clear bleach.
Use to clear air bubbles or debris from the flow cell.
continued
8PN 4237187A (November 1993)
GENERAL PROCEDURES
Table 1 Diluter F Key Functions
FunctionMessageDescription
continued
F14
F15
F16
F17
F20
F44
F45
PRIME PAK LYSE
PRIME PAK PRESRV
PRIME DILUENT
PRIME SCATTER PAK
FILL SHEATH TANK
CLEAR FLOW CELL 1
CLEAR FLOW CELL 2
Used to prime the diff lyse lines for troubleshooting. For Service
use only.
Used to prime the diff preservative reagent lines for
troubleshooting. For Service use only.
Use to prime diluent lines when you replace the diluent
container.
Use to prime the diff reagents when you replace the SCATTER
PAK.
Used to fill the sheath tank as needed. For Service use only.
What it does to flow cell
•purges it with cleaning agent
•turns on its aperture current
•applies low pressure to it
•alternately applies vacuum to the bottom
•checks it for a clog
•leaves it in diluent
See the Clear Flow Cell procedure for instructions on the use of
this function.
What it does to flow cell
•purges it with cleaning agent
•turns on its aperture current
•applies 30 psi to top of it
•alternately applies vacuum to the bottom
•checks it for a clog
•leaves it in diluent
1
GENERAL
See the Clear Flow Cell procedure for instructions on the use of
this function.
F46
CLEAR FLOW CELL 3
Use only if F44 and F45 fail to clear flow cell.
What it does to flow cell
•purges it with cleaning agent
•turns on its aperture current
•applies low pressure to it
•alternately applies vacuum to the bottom
•leaves it in cleaning agent
See the Clear Flow Cell procedure for instructions on the use of
this function.
continued
PN 4237187A (November 1993)9
GENERAL SERVICE AND MAINTENANCE
Table 1 Diluter F Key Functions
FunctionMessageDescription
continued
F55
F56
GENERAL
F57
1
F95
F96
LATRON CONTROL
F55 = DIFF
F56 = RETIC
F57 = DIFF + RETIC
SOLENOIDS FREE
TUBE ADVANCE
Access this function before you aspirate LATRON primer and
control.
Do not use this function to aspirate anything except
LATRON control or LATRON primer. Other material can
cause damage to the system.
This function causes approximately 1.5 mL of primer or control to
be aspirated through the Secondary mode aspirate tip directly
into the flow cell.
Frees the pressure from all solenoids. Press STOP twice to exit.
Use when you adjust the tube detector; causes a cassette to
move through the piercing station one tube position at a time.
Press STOP to halt the advance and make adjustments, ENTER
to continue the advance and check alignment. Press STOP twice
to exit the function.
CAUTION
1.3COMPONENTS
This section includes
List of Cylinders
List of Manifolds
List of Pumps
List of Vacuum Chambers
Pinch Valve Functions
Solenoid Functions
Summary of Pinch Valve/Solenoid/Component Relationships
10PN 4237187A (November 1993)
COMPONENTS
Cylinders (CL)
Manifolds (MF)
Table 2 lists the cylinders and the components they operate.
Table 2 Cylinders
CylinderOperates
CL1BSV front
CL2BSV back
CL3needle
CL4left lift
CL5right lift
CL6tube ram
CL7bar-code reader
CL8bed rocker
CL9belt advance
PM1needle rinse, 0.77 mL
PM2RBC diluent dispense, 10 mL
PM3WBC diluent dispense, 6 mL
PM45 diff lyse, 0.536 mL
PM6backwash, 1 mL
PM73 diff lyse, 0.52 mL
PM83 diff lyse, 0.52 mL
PM9Hgb blank, 5 mL
PM10cleaning agent, 5 mL
PM11rinse, 1 mL
PM12leukocyte preservative, 0.193 mL
PM13Primary mode aspiration, 0.250 mL
PM14Secondary mode aspiration, 0.150 mL
Vacuum Chambers (VC)
Table 5 lists the vacuum chambers and the components they control.
Table 6 lists the pinch valves sequentially, shows which solenoid (SL)
controls the pinch valve, and gives each valve’s function. See Figure 3 for
most locations.
Legend: NC = Normally Closed
NO = Normally Open
NA = Normally Activated
Table 6 Pinch Valves
Pinch
ValveSolenoidPinch Valve Function
VL1SL18NC path from RBC diluent dispenser to BSV rear section, port 9
NO path from RBC diluent dispenser to BSV rear section, port 10
VL2SL18NC path from WBC diluent dispenser to BSV rear section, port 10
NO path from WBC diluent dispenser to BSV center section, port 8
VL3SL9NC 5 psi path to bath overflow chamber to drain it
NO vent to bath overflow chamber
continued
PN 4237187A (November 1993)13
GENERAL SERVICE AND MAINTENANCE
Pinch
ValveSolenoidPinch Valve Function
continued
GENERAL
1
VL4SL63NC diff lyse pump dispense
VL5SL8NC RBC diluent dispenser dispense
VL6SL7NC WBC diluent dispenser dispense
VL7--Not used
VL10SL13NC backwash pump dispense for Secondary mode
VL11SL42NC mix chamber rinse pump dispense
VL12SL42NC pressure to rinse pump for mix chamber
VL13SL11NC RBC count
VL14SL18NC path for rinse trough drain to VIC, primary mode
VL15SL12NC pressure for WBC bath drain
VL16SL10NC WBC count
VL18SL6NC pressure for Hgb cuvette drain
VL19SL16NC •VIC drain
VL20SL57NC Primary mode drain to diff waste chamber
VL22SL12NC vacuum to main waste chamber
Table 6 Pinch Valves
NO diff lyse pump fill
NO RBC diluent dispenser fill
NO WBC diluent dispenser fill
NO •backwash pump dispense for Primary mode
•backwash pump fill for both modes
NO mix chamber rinse pump fill
NC vacuum to rinse pump for mix chamber
NO not used
NO not used
NO vent to bath overflow chamber
NO not used
NO vent for Hgb cuvette
•5 psi to VIC
•Primary and Secondary mode backwash drain to waste
chamber
NO •not used
•vacuum to VIC
•not used
NO Secondary mode drain to diff waste chamber
NO pressure to main waste chamber
continued
14PN 4237187A (November 1993)
COMPONENTS
Table 6 Pinch Valves
Pinch
ValveSolenoidPinch Valve Function
continued
VL23SL12NC RBC bath drain
NO output to main waste chamber
VL24SL12NC WBC bath drain
NO output to main waste chamber
VL26SL50NO rinse from PM1 into mix chamber when deactivated
VL28SL28NC Hgb cuvette drain to waste chamber
NO vent
VL29SL49NC vent diff waste chamber
NO not used
VL30SL51NC sample pressure to mix chamber
NO vent for mix chamber
VL31SL54NC sheath flow from tank to bottom half of flow cell
NO not used
VL32SL54NC sheath flow from tank to top half of flow cell
NO not used
VL33SL49NC sample drain from flow cell
NO not used
VL34SL55NC forward flush drain to waste
NO not used
VL35SL12NO high vacuum to low vacuum regulator
NC not used
VL36SL61NC diff rinse pump - cleaner
NO diff rinse pump - diluent
VL37SL46NC path for flush upper section of flow cell
NO not used
VL38SL17NC high vacuum to VIC
NO low vacuum to VIC
VL40SL57NC vacuum for secondary backwash drain to diff waste chamber
NO not used
VL43SL2NC dispense lyse to WBC bath
NO lyse pumps fill
VL44SL60NC diff preservative pump dispense
NO diff preservative pump fill
VL45--Not used
1
GENERAL
continued
PN 4237187A (November 1993)15
GENERAL SERVICE AND MAINTENANCE
Pinch
ValveSolenoidPinch Valve Function
continued
GENERAL
1
VL46SL3NC pressure to lyse pumps
VL47SL61NC cleaner pump dispense
VL48SL2NC Hgb blank pump dispense
VL49SL48NC vacuum supply to drain mix chamber to diff waste chamber
VL50SL56NC •vacuum
VL52SL44NO blood and diff lyse delivery to mix chamber, Secondary mode
VL53SL45NO diff preservative delivery path to mix chamber
VL54SL65NO blood and diff lyse delivery to mix chamber, Primary mode
VL55SL52NO sample flow from mix chamber to flow cell
VL56SL50NC vacuum on mix chamber
VL57SL48NC drain path not used
VL58SL46NC flush flow cell lower section
VL59SL41NC reverse flush drain from flow cell to waste chamber
VL68 NASL24NC Primary mode aspiration disabled
VL69SL24NC from secondary aspiration pump to main waste chamber
THREE WAY PINCH VALVEDUAL ACTION PINCH VALVESOLENOIDMINI PINCH VALVE
7187001A
19
1613
ROCKER BED
DILUTER BASE (IN FRONT OF THE ROCKER BED)
MF13
46
MF11
363534333231
30
43384044
474648
49
29
28
24
23
221820
1514
7
4
32
1
5
6
1012 11
22
25
24
58
69
68
77
88
59
54
52
26
82
55
21
58 57
56
53
13
51
15
52
17 53
41
54
44 55
45
56
57
48
60
49
61
50
65
BSV
7985
81 76
78
83
74
73
34 64 27 26 36 23
9
30 62 63 37 59 28
29
86
TOP
87
BOTTOM
Table 7 Solenoids
SolenoidFunctionAction
continued
GENERAL
1
SL81Level senseMF1
Figure 3 illustrates solenoid and pinch valve locations.
20PN 4237187A (November 1993)
Figure 3 Pinch Valve and Solenoid Locations
2.1GENERAL INFORMATION
For optimum performance of your COULTER®STKS with Reticulocyte
Analysis, you must calibrate the CBC parameters. The WBC differential
parameters are calibrated at the factory; they do not require calibration in
the laboratory.
CALIBRATION 2
Your laboratory is responsible for the final calibration of the CBC
parameters, for recording the calibration factors, and for establishing
reference values for ramp-pulse and precision-pulse tests. Coulter
recommends S-CAL
alternative to whole-blood calibration.
Calibrate the CBC parameters:
•At installation
•After the replacement of any component that involves the dilution
characteristics (such as the BSV), or the primary measurements
(such as the apertures)
•When advised to do so by your Coulter Service representative
Daily quality assurance programs monitor ongoing instrument and
system performance. However, we recommend that you verify the status
of instrument calibration quarterly, and/or when controls begin to show
evidence of unusual trends, or when controls exceed the manufacturer’s
defined acceptable limits.
®
calibrator, or an exact equivalent, as an acceptable
2
CALIBRATION
•Although the instrument is relatively insensitive to minor room
temperature changes, perform calibration when the room
temperature is stable and within the normal ambient temperature
range. If the average ambient room temperature changes more than
10°F from the calibrating temperature, it is advisable to verify
calibration and to recalibrate if necessary.
•In the normal process of tracking data for an extended period of
time, your laboratory can make a specific decision to recalibrate a
given parameter. Never adjust to a specific value for an individual
sample.
PN 4237187A (November 1993)21
CALIBRATION
2.2PRECALIBRATION STEPS
CALIBRATION
•If a problem or malfunction occurs while performing these
procedures, see the Troubleshooting procedures later in this
manual. If the corrective procedures fail to eliminate the problem,
call Coulter Customer Operations immediately.
Before calibrating, perform the following procedures in the order given.
1.Follow a, b or c as it applies to your instrument.
a.If the instrument is in the shutdown condition, and you
routinely shut down the instrument for a minimum of 30
minutes every 24 hours in COULTER CLENZ
go to step 2.
®
cleaning agent,
2
b.If you routinely use COULTER CLENZ cleaning agent, but you
are beginning calibration after processing patient samples, shut
down the instrument in the cleaning agent for 30 minutes to
assure optimum aperture cleanliness before proceeding.
CAUTION
Do not aspirate bleach; bleach can damage the flow cell.
c.If you do not routinely use COULTER CLENZ cleaning agent,
perform the Bleach the Apertures procedure in Chapter 3.
2.Check the reagent containers for:
a.Sufficient quantity
b.Not beyond expiration date
c.No precipitates, turbidity, particulate matter, or unusual color
d.Proper connections between the Diluter and the reagent
containers
22PN 4237187A (November 1993)
PRECALIBRATION STEPS
WARNING
The contents of the waste container and its associated tubing can
include residual biological material and must be handled with care.
Avoid skin contact and clean up spills immediately. Dispose of the
contents of the waste container in accordance with acceptable
laboratory procedures.
3.Check the waste container for:
a.Sufficient capacity
b.Proper connections
4.Perform the startup procedures, including a quality-control check,
as described in the Operator’s Guide, PN 4237188. Verify that the
following are acceptable: background count, reproducibility,
carryover, and control recovery.
Reproducibility Check, CBC/Diff
Before you calibrate the STKS with either the S-CAL Kit or whole blood,
perform a 10-sample reproducibility study on the CBC parameters in the
Primary mode. You can also perform this study if you need to check
instrument precision for the CBC or diff parameters.
In the Secondary mode, when you perform a Reproducibility check, do
not cycle more than eight samples per tube. If you are running 31
samples, use four separate vials.
2
CALIBRATION
Sample Requirements
Collect enough blood for 10 cycles from a single donor who:
•Is receiving no medication.
•Has normal hematologic parameters, with a WBC count of 10,000
±1,000.
•Has normal erythrocyte, leukocyte and platelet morphology and
PN 4237187A (November 1993)23
CALIBRATION
•If you want to check the diff parameters, with diff values of
Neutrophils40 to 72%
Lymphocytes17 to 45%
Monocytes4 to 12%
Eosinophils0 to 10%
Basophils0 to 1%
Procedure, Primary Mode
CALIBRATION
2
1.At the Diluter, press
PRIME APERT to activate the pneumatics.
2.At the Analyzer:
a.In SYSTEM CONFIGURATION, be sure that the BLOOD
DETECTOR is set to ENABLED.
b.In SYSTEM RUN, verify that the # aspirations/tube is 1.
3.At the DMS, select
Special Functions
Calibration
Reproducibility
If data remains from a previous study, pressDelete Table.
4.Cycle one sample of normal whole blood in the Primary mode.
CAUTION
Do not pierce a specimen tube more than five times; additional piercing
can obstruct or clog the needle.
5.At the Analyzer, set the number of aspirations per tube to 5.
6.Aliquot the well-mixed normal whole-blood sample into two redtopped tubes.
a.Place the tubes into consecutive positions in a cassette, place
the cassette in the loading bay, and press
START/CONT.
b.Monitor the system for normal sample flow.
24PN 4237187A (November 1993)
Loading...
+ 98 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.