Coulter STKS Troubleshooting

COULTER®STKS Analyzer with Reticulocyte Analysis
Special Procedures and Troubleshooting
COULTER
STKS
PN 4237187B (March 1995)
READ ALL PRODUCT MANUALS AND CONSULT WITH COULTER-TRAINED PERSONNEL BEFORE
ATTEMPTING TO OPERATE INSTRUMENT
HAZARDS AND OPERATIONAL PRECAUTIONS AND LIMITATIONS
WARNINGS, CAUTIONS, and IMPORTANTS alert you as follows:
WARNING - Might cause injury. CAUTION - Might cause damage to the instrument. IMPORTANT - Might cause misleading results.
CAUTION|
|
System integrity might be compromised and operational errors might occur if:|
|
Thisequipmentisusedinamannerotherthanspecified.Operatetheinstrumentasinstructedinthe| Product Manuals.|
|
You introduced software that is not authorized by Coulter into your computer. Only operate your| system’s computer with software authorized by Coulter.|
Coulter Corporation urges its customers to comply with all national health and safety standards such as the use of barrier protection. This may include, but it is not limited to, protective eye wear, gloves, and suitable laboratory attire when operating or maintaining this or any other automated laboratory analyzer.
In the USA, for Service call Coulter Customer Operations 1-800-526-7694.
1 GENERAL SERVICE AND MAINTENANCE, 1
1.1 LASER MAINTENANCE, 1
1.2 GENERAL PROCEDURES, 1 Power Down, 3 Power Up, 3 Optimize the Disk, 3
Daily, 4 Weekly, 4
Monthly, 4 Set Screen Saver, 5 Position Rocker Bed, 6 Diluter F Key Functions, 6
1.3 COMPONENTS, 10 Cylinders (CL), 11 Manifolds (MF), 11 Pumps (PM), 12 Vacuum Chambers (VC), 12 Pinch Valve (VL) Functions, 13 Solenoid (SL) Functions, 17
CONTENTS
2 CALIBRATION, 21
2.1 GENERAL INFORMATION, 21
2.2 PRECALIBRATION STEPS, 22 Reproducibility Check, CBC/Diff, 23
Sample Requirements, 23 Procedure, Primary Mode, 24
Carryover Check, 26
Sample Requirements, 26 Procedure, 26
2.3 CBC CALIBRATION WITH S-CAL CALIBRATOR, 28 Calibration Summary, 28
2.4 CALIBRATE CBC PARAMETERS WITH WHOLE BLOOD, 37
3 CLEANING PROCEDURES, 39
3.1 CLEAN BLOOD SAMPLING VALVE (BSV), 39 Procedure A, Rinse the BSV Outside Sections, 39
PN 4237187A (November 1993) i
CONTENTS
Procedure B, Remove and Clean the BSV Sections, 40
3.2 CLEAN APERTURES, 44 Procedure A, Zap the Apertures, 44 Procedure B, Bleach the Apertures, 44 Procedure C, Clean the Aperture Baths, 49 Procedure D, Clogged Apertures, 52 Procedure E, Clear Flow Cell, 53
3.3 TRANSPORT SYSTEM, 54 Clean Rinse Trough, 54 Clean Cassettes, 56 Clean Stripper Plate, 56 Clean Rocker Bed Belt, 56
3.4 OTHER COMPONENTS, 56 Clean Air Filters, 57 Clean Vacuum Isolator Chamber (VIC), 58 Clean Waste Chamber, 60 Clean Vacuum Trap Bottle, 61 Bleach RBC Internal Electrode, 62 Drain Overflow Chamber, 64 Clean Drip Plate, 64
4 REPLACE/ADJUST PROCEDURES, 67
4.1 REPLACE REAGENTS, 67
4.2 REPLACE WASTE CONTAINER, 68
4.3 REPLACE FUSES, 69
4.4 REPLACE APERTURE BATHS, 70
4.5 REPLACE O-RINGS, 71
4.6 REPLACE CHECK VALVES, 72
4.7 REPLACE AIR FILTERS, 73
4.8 REPLACE CHOKES, 73
4.9 ATTACH TUBING TO A FITTING, 73
4.10 REPLACE PIERCING NEEDLE, 73 Decontaminate Needle, 73
ii PN 4237187A (November 1993)
Replace Needle, 74
4.11 REPLACE CASSETTE SPRING CLIP, 76
4.12 REPLACE CASSETTE LABELS, 77
4.13 CHANGE RINSE TROUGH PAD, 77
4.14 ADJUST HEMOGLOBIN LAMP VOLTAGE, 78
4.15 ADJUST TUBE DETECTOR, 79
5 TROUBLESHOOTING, 81
5.1 ANALYZER RESET PROCEDURE, 81
5.2 DMS RESET PROCEDURE, 83 DMS Not Working, 83 Printer Not Ready, 84 Lost Clusters, 84
CONTENTS
5.3 RAMP AND PRECISION TESTS, 85
5.4 RETIC REPRODUCIBILITY CHECK, 85
5.5 PROLONGED SHUTDOWN PROCEDURE, 87
5.6 ANALYZER CLOCK STOPPED, 87
5.7 ANALYZER ALERT MESSAGES, 89
5.8 DILUTER DIAGNOSTIC MESSAGES, 100
5.9 DMS ERROR MESSAGES, 107
5.10 GRAPHIC/LASER PRINTER PRECAUTIONS, 113
INDEX, 119
ILLUSTRATIONS
1 Diluter Front Panel, 2 2 Screen Saver Setup, 6 3 Pinch Valve and Solenoid Locations, 20 4 Reproducibility, 25
PN 4237187B (March 1995) iii
CONTENTS
5 Carryover Acceptable Screen, 28 6 CBC Calibration Screen, 30 7 Calibration Criteria Flowchart, 33 8 Blood Sampling Valve (BSV), 41 9 RBC and WBC Aperture Baths, 45 10 Aperture Bath, Blocks and Housings, 51 11 Rinse Trough, 55 12 Location of Air Filters, 57 13 Clean VIC, 59 14 Drip Plate, 64 15 Replace Piercing Needle, 75 16 Tube Detector Adjustment, 80 17 Reset Button Location, 82 18 Retic Reproducibility, 86
TABLES
1 Diluter F Key Functions, 7 2 Cylinders, 11 3 Manifolds, 11 4 Pumps, 12 5 Vacuum Chambers, 12 6 Pinch Valves, 13 7 Solenoids, 17 8 Open-Fuse Indications, 69 9 Analyzer ALERT Messages, 89 10 Diluter Diagnostic Messages, 101 11 DMS Error Messages, 107
iv PN 4237187B (March 1995)
1.1 LASER MAINTENANCE
All service and maintenance of the laser module must be done by trained Coulter personnel. Removal of the module must be performed only by a Coulter Representative.
1.2 GENERAL PROCEDURES
GENERAL SERVICE AND MAINTENANCE 1
1
GENERAL
WARNING
Do not try to remove the laser module from the Diluter.
IMPORTANT
If a power failure or reset occurs during a cycle, after power up, press PRIME APERT before cycling samples.
This section details the cleaning, replacement, and adjustment procedures that are your responsibility. General operations that may apply to numerous procedures are described below. They include:
Power Down Power Up Optimize the Disk Set Screen Saver Position Rocker Bed Diluter
Coulter recommends a Preventive Maintenance Inspection (PMI) once a year by a Coulter Service Representative.
Figure 1 illustrates component location on the Diluter front panel.
F Key Functions
PN 4237187A (November 1993) 1
GENERAL SERVICE AND MAINTENANCE
A
GENERAL
67
1
8
9
VL30 VL31 VL32 VL33 VL34 VL35 VL36 VL37
VL19
10
11
12
GA1
VC15
RG2
PM8
VL43
VL38
VL40
VL44
PM7
VL20 VL18 VL22
VL47
VL46
VL48
VL50
SOL47
VC14
VL49
13
7187002
PM10
PM9
14
17. HGB CUVETTE HOLDER
18. BATH OVERFLOW CHAMBER (VC10)
19. NEEDLE-RINSE PUMP (PM1)
20. WBC DILUENT DISPENSER (PM3)
21. RINSE MIX CHAMBER PUMP (PM11)
22. 1mL BACKWASH PUMP (PM6)
23. RBC DILUENT DISPENSER (PM2)
17 16 15
VL14 VL15 VL16
VL13
VL4 VL7
12345
VL3
VL2
VL1
PM4
VL6
VL5
9. ELECTRONIC MANOMETER WINDOW
10. MANOMETER RANGE SCALE
11. CBC LYTIC REAGENT PUMP (PM8)
12. SHEATH TANK (VC15)
13. 5mL CLEANING AGENT PUMP (PM10)
14. 5mL HGB PUMP (PM9)
15. FOAM TRAP (VC14)
16. WASTE CHAMBER (VC11)
SOL1
SOL5
PM11
1. ERYTHROLYSE II PUMP (PM4)
2. RBC BATH (VC2)
3. VACUUM ISOLATOR (VC1)
4. WBC BATH (VC3)
5. STABILYSE PUMP (PM12)
6. CBC LYTIC REAGENT PUMP (PM7)
7. MANOMETER REGULATOR
8. SHEATH FLOW RESTRICTOR
PM6
22 21 20 19 18
PM = PUMP
PM1
VL11VL10
PM3
VL12
PM2
Figure 1 Diluter Front Panel
VC = VACUUM CHAMBER
23
2 PN 4237187A (November 1993)
GENERAL PROCEDURES
Power Down
1. Turn off the Graphic Printer.
2. Turn off the Ticket Printer.
3. Turn off the DMS.
4. Press
5. Turn off the MAIN POWER circuit breaker on the Power Supply.
6. Unplug the primary power cord from the wall outlet.
7. Verify that the power is disconnected:
8. Turn off the MAIN POWER circuit breaker.
POWER OFF on the Diluter keypad.
a. Turn on the MAIN POWER circuit breaker.
b. Check the front panel of the Power Supply and be sure there
are no lamps lighted. If any lamp lights, the power is still connected; you must repeat steps 5, 6 and 7.
1
GENERAL
Power Up
Optimize the Disk
1. Plug the primary power cord into the wall outlet.
2. Turn on the MAIN POWER circuit breaker on the Power Supply.
3. Press
4. Turn on the DMS.
5. Turn on the Ticket Printer.
6. Turn on the Graphic Printer.
This system is equipped with the OPTune™ utility, a software utility that optimizes the hard drive on your DMS. Optimizing organizes files on your hard drive so that the DMS is faster and more efficient. The OPTune utility runs when you turn on the DMS and when you reset the DMS.
POWER ON on the Diluter keypad.
PN 4237187A (November 1993) 3
GENERAL SERVICE AND MAINTENANCE
GENERAL
1
The OPTune utility performs three types of optimization: daily, weekly and monthly. The type it uses depends on how long it has been since the last optimization.
Daily
The fastest method, this type leaves each file
100% defragmented
sorted in ascending order by name
in contiguous order
Weekly
Takes about 50% longer than the daily method. Does everything the daily method does, plus it optimizes in the packed mode. It arranges the files on your hard drive so that they are end-to-end, with no space between them. Thus, new files are likely to be written to disk without being fragmented.
Monthly
The most thorough method. Does everything the weekly method does, plus it physically arranges files on the disk in the same order as the sorted directory entries. This method takes longer, but increases efficiency when accessing many files in sequential order.
The OPTune utility initially takes about 45 minutes to an hour to optimize a 240MB hard drive. After that, daily optimization takes about 30 to 40 seconds. The OPTune utility reorganizes only those files that need it; it does not reoptimize unnecessarily.
You can bypass the OPTune utility process by pressing while the OPTune utility screen is in view. When the message Do you want to stop? [Y/N] appears, press . This message appears: User terminated Verify. The optimizing process stops and the system goes into the DMS program.
If there is a hard disk failure during the optimization process, for example lost clusters or cross-linked files, the OPTune utility prompts you to continue. Answer Yes to go on with the process. However, if this happens more than once a week, record the incident in your DMS maintenance log or your logbook and call Coulter Customer Operations.
4 PN 4237187A (November 1993)
GENERAL PROCEDURES
If you always leave your system turned on, we recommend that you run the OPTune utility at least once a week, using one of these two options:
1
Option A
1. Turn off the Graphic Printer.
2. Turn off the DMS.
3. Wait at least 15 seconds.
4. Turn on the DMS.
5. Turn on the Graphic Printer.
Option B
Select
Special Functions
Sample Analysis Set up
Reset DMS
When you reset the DMS or turn it off then back on, these functions default to these conditions:
GENERAL
Set Screen Saver
Data Base storage: ON (DB)
XB: ON (XB)
AutoPrint: NONE (PR)
AutoTransmit: OFF (HC)
Reset these options as needed according to your laboratory’s protocol.
The screen saver function makes the screen go blank when it is not being used. Figure 2 illustrates the Setup window. Select
Special Functions
Set Up
System Set Up
Screen Saver Time Out
PN 4237187A (November 1993) 5
GENERAL SERVICE AND MAINTENANCE
GENERAL
1
Set the time, in minutes, for the length of time to elapse before the screen goes blank. You can enter from 0 to 99 minutes.
A number from 1 to 99 means that, if there is no keystroke in that many minutes, the screen goes blank. The screen also goes blank when the compressor times out.
The screen turns back on under these conditions:
- The DMS receives a sample run.
- You press a key on the DMS keyboard. The DMS executes the
keystroke as it applied to the screen before it went blank.
Zero means the screen saver will not be activated.
Sample Analysis Controls Start Up Special Functions
DMS 2A Diagnostics
Control set up Sample analysis set up C System set up|
Screen Saver Setup anguage
Screen Saver Delay: 0 Minutes QAP ID #|
Change Supervisor Password Screen Saver Time Out|
(931951) COPYRIGHT 1989-1993 COULTER CORPORATION, ALL RIGHTS RESERVED
F1-Help F3-Error Log F5-Other F9-Main menu F10-Previous menu 07/14/93 17:17 OPR DMSPRTKHCDBXBWLHWLQC 59
CBC/Diff/Retic
Set Up|
Shift Reagents Institution Communication def Electronic reference
atabase
Figure 2 Screen Saver Setup
Position Rocker Bed
When the compressor is off, the rocker bed is not locked in position and can be tilted easily.
Diluter F Key Functions
The Diluter F key functions are listed in Table 1. To select these functions, press
6 PN 4237187A (November 1993)
F X X ENTER where XX is the function number.
GENERAL PROCEDURES
During these functions, a related message appears on the Diluter keypad; when the routine is finished, the message returns to the function number.
For example, when you select F01, the burn circuit activates and the message appears;
CLEAN APERTURES
after which the message returns to:
FUNCTION = 01
1
GENERAL
To repeat the function, press
Table 1 Diluter F Key Functions
Function Message Description
F01
F02
F03 F04
F05
F06
CLEAN APERTURES
PRIME CBC LYSE
RBC AND WBC MIX DILUENT DISPENSE
SOLENOID XX
RELEASE B.S.V.
The burn circuit activates. This function applies current across the apertures to "burn" protein or debris off the apertures.
The following sequence occurs twice:
Drains both baths then dispenses CBC lytic reagent into the WBC bath three times to prime the lyse line.
The WBC bath drains and the cycle repeats.
It then fills both baths with diluent and primes the apertures.
Causes mixing bubbles to enter both the RBC and WBC baths. Dispenses 1 mL of diluent through the Secondary mode
aspirator tip. Use to manually activate individual solenoids (see Table 7).
Enter the desired solenoid number (for numbers less than 10, enter a leading zero), and press ENTER.
Drains the baths
Bleeds off pressure
Rotates the BSV back and forth five times
ENTER. To exit the function, press STOP.
Use to check BSV function, free a stuck valve, or when cleaning the BSV to ease removal. Press STOP twice to exit the function. This function does not refill the baths; use RINSE to refill them with diluent.
continued
PN 4237187A (November 1993) 7
GENERAL SERVICE AND MAINTENANCE
Table 1 Diluter F Key Functions
Function Message Description
continued
F07
GENERAL
BARCODE READ RATE TEST
Checks the bar-code read rate on the cassette bar-code label, and on the four types of tube bar-code labels.
1
F08 F09
PRIME SWEEP FLOW ZAP APERTURES
1. Enable the bar-code reader.
2. Remove the cap-piercing needle so that extensive piercings do not cause a plugged needle.
3. Place a cassette of bar-code labeled tubes on the rocker bed.
4. Select the function, then the following appears on the Diluter display:
A00B00C00D00E00
A represents Interleaved 2-of-5 B represents Code 39®Bar code C represents Codabar D represents the cassette label E represents code 128
The cassette automatically moves to the piercing station and the bar-code reader scans both the cassette and tube labels. On the display, the appropriate category increments by one each time a label from that category is read. Press STOP twice to exit the function.
Primes the sweep-flow lines.
Applies pressure behind the baths
Drains and rinses the baths four times with cleaning agent
Draws the cleaning agent behind the apertures
Zaps the apertures 10 times
After this function is complete, you must execute the startup cycle to purge the system of the cleaning agent.
F10
F11
F12 F13
RAISE NEEDLE
EXTENDED PRIME
EXTENDED CLEAR PURGE MODE
Causes the piercing needle to rise and the rocker bed to lock in the backward position. Press STOP twice to exit.
Pulls fluid through the apertures to the vacuum isolator chamber for 60 seconds. Aperture current turns on and aperture lamp intensity increases for viewing the aperture screen.
Clears the apertures for 60 seconds. Use to clear bleach. Use to clear air bubbles or debris from the flow cell.
continued
8 PN 4237187A (November 1993)
GENERAL PROCEDURES
Table 1 Diluter F Key Functions
Function Message Description
continued
F14
F15
F16
F17
F20 F44
F45
PRIME PAK LYSE
PRIME PAK PRESRV
PRIME DILUENT
PRIME SCATTER PAK
FILL SHEATH TANK CLEAR FLOW CELL 1
CLEAR FLOW CELL 2
Used to prime the diff lyse lines for troubleshooting. For Service use only.
Used to prime the diff preservative reagent lines for troubleshooting. For Service use only.
Use to prime diluent lines when you replace the diluent container.
Use to prime the diff reagents when you replace the SCATTER PAK.
Used to fill the sheath tank as needed. For Service use only. What it does to flow cell
purges it with cleaning agent
turns on its aperture current
applies low pressure to it
alternately applies vacuum to the bottom
checks it for a clog
leaves it in diluent See the Clear Flow Cell procedure for instructions on the use of
this function. What it does to flow cell
purges it with cleaning agent
turns on its aperture current
applies 30 psi to top of it
alternately applies vacuum to the bottom
checks it for a clog
leaves it in diluent
1
GENERAL
See the Clear Flow Cell procedure for instructions on the use of this function.
F46
CLEAR FLOW CELL 3
Use only if F44 and F45 fail to clear flow cell. What it does to flow cell
purges it with cleaning agent
turns on its aperture current
applies low pressure to it
alternately applies vacuum to the bottom
leaves it in cleaning agent See the Clear Flow Cell procedure for instructions on the use of
this function.
continued
PN 4237187A (November 1993) 9
GENERAL SERVICE AND MAINTENANCE
Table 1 Diluter F Key Functions
Function Message Description
continued
F55 F56
GENERAL
F57
1
F95 F96
LATRON CONTROL
F55 = DIFF F56 = RETIC F57 = DIFF + RETIC
SOLENOIDS FREE TUBE ADVANCE
Access this function before you aspirate LATRON primer and control.
Do not use this function to aspirate anything except LATRON control or LATRON primer. Other material can cause damage to the system.
This function causes approximately 1.5 mL of primer or control to be aspirated through the Secondary mode aspirate tip directly into the flow cell.
Frees the pressure from all solenoids. Press STOP twice to exit. Use when you adjust the tube detector; causes a cassette to
move through the piercing station one tube position at a time. Press STOP to halt the advance and make adjustments, ENTER to continue the advance and check alignment. Press STOP twice to exit the function.
CAUTION
1.3 COMPONENTS
This section includes
List of Cylinders List of Manifolds List of Pumps List of Vacuum Chambers Pinch Valve Functions Solenoid Functions Summary of Pinch Valve/Solenoid/Component Relationships
10 PN 4237187A (November 1993)
COMPONENTS
Cylinders (CL)
Manifolds (MF)
Table 2 lists the cylinders and the components they operate.
Table 2 Cylinders
Cylinder Operates
CL1 BSV front CL2 BSV back CL3 needle CL4 left lift CL5 right lift CL6 tube ram CL7 bar-code reader CL8 bed rocker CL9 belt advance
CL10 stripper plate
1
GENERAL
Table 3 lists the Manifolds and their output.
Table 3 Manifolds
Manifold Output
MF1 5 psi, level sense switches MF2 pneumatic/hydraulic source input to Diluter, waste MF3 30 psi MF4 vacuum MF6 30 psi, solenoid (SL) manifold MF7 5 psi MF8 5 psi, mixing bubbles
MF9 30 psi MF10 vacuum MF11 30 psi, SL manifold MF12 vacuum MF13 30 psi, SL manifold MF15 30 psi
PN 4237187A (November 1993) 11
GENERAL SERVICE AND MAINTENANCE
Pumps (PM)
Table 4 lists the pumps and what they operate.
GENERAL
1
Table 4 Pumps
Pump Operates
PM1 needle rinse, 0.77 mL PM2 RBC diluent dispense, 10 mL PM3 WBC diluent dispense, 6 mL PM4 5 diff lyse, 0.536 mL PM6 backwash, 1 mL PM7 3 diff lyse, 0.52 mL PM8 3 diff lyse, 0.52 mL PM9 Hgb blank, 5 mL
PM10 cleaning agent, 5 mL PM11 rinse, 1 mL PM12 leukocyte preservative, 0.193 mL PM13 Primary mode aspiration, 0.250 mL PM14 Secondary mode aspiration, 0.150 mL
Vacuum Chambers (VC)
Table 5 lists the vacuum chambers and the components they control.
Table 5 Vacuum Chambers
Vacuum Chamber Controls
VC1 vacuum isolator (VIC) VC2 RBC bath VC3 WBC bath VC4 RBC bath aperture 1 VC5 RBC bath aperture 2 VC6 RBC bath aperture 3 VC7 WBC bath aperture 1
continued
12 PN 4237187A (November 1993)
COMPONENTS
Table 5 Vacuum Chambers
Vacuum Chamber Controls
continued
VC8 WBC bath aperture 2
VC9 WBC bath aperture 3 VC10 bath overflow VC11 waste chamber (Diluter) VC13 waste chamber (Diff) VC14 foam trap VC15 sheath tank VC16 sheath flow VC17 vacuum tank VC18 flow cell VC19 mixing chamber VC20 foam trap
1
GENERAL
Pinch Valve (VL) Functions
Table 6 lists the pinch valves sequentially, shows which solenoid (SL) controls the pinch valve, and gives each valve’s function. See Figure 3 for most locations.
Legend: NC = Normally Closed
NO = Normally Open NA = Normally Activated
Table 6 Pinch Valves
Pinch
Valve Solenoid Pinch Valve Function
VL1 SL18 NC path from RBC diluent dispenser to BSV rear section, port 9
NO path from RBC diluent dispenser to BSV rear section, port 10
VL2 SL18 NC path from WBC diluent dispenser to BSV rear section, port 10
NO path from WBC diluent dispenser to BSV center section, port 8
VL3 SL9 NC 5 psi path to bath overflow chamber to drain it
NO vent to bath overflow chamber
continued
PN 4237187A (November 1993) 13
GENERAL SERVICE AND MAINTENANCE
Pinch
Valve Solenoid Pinch Valve Function
continued
GENERAL
1
VL4 SL63 NC diff lyse pump dispense
VL5 SL8 NC RBC diluent dispenser dispense
VL6 SL7 NC WBC diluent dispenser dispense
VL7 -- Not used
VL10 SL13 NC backwash pump dispense for Secondary mode
VL11 SL42 NC mix chamber rinse pump dispense
VL12 SL42 NC pressure to rinse pump for mix chamber
VL13 SL11 NC RBC count
VL14 SL18 NC path for rinse trough drain to VIC, primary mode
VL15 SL12 NC pressure for WBC bath drain
VL16 SL10 NC WBC count
VL18 SL6 NC pressure for Hgb cuvette drain
VL19 SL16 NC • VIC drain
VL20 SL57 NC Primary mode drain to diff waste chamber
VL22 SL12 NC vacuum to main waste chamber
Table 6 Pinch Valves
NO diff lyse pump fill
NO RBC diluent dispenser fill
NO WBC diluent dispenser fill
NO • backwash pump dispense for Primary mode
backwash pump fill for both modes
NO mix chamber rinse pump fill
NC vacuum to rinse pump for mix chamber
NO not used
NO not used
NO vent to bath overflow chamber
NO not used
NO vent for Hgb cuvette
5 psi to VIC
Primary and Secondary mode backwash drain to waste chamber
NO • not used
vacuum to VIC
not used
NO Secondary mode drain to diff waste chamber
NO pressure to main waste chamber
continued
14 PN 4237187A (November 1993)
COMPONENTS
Table 6 Pinch Valves
Pinch
Valve Solenoid Pinch Valve Function
continued
VL23 SL12 NC RBC bath drain
NO output to main waste chamber
VL24 SL12 NC WBC bath drain
NO output to main waste chamber VL26 SL50 NO rinse from PM1 into mix chamber when deactivated VL28 SL28 NC Hgb cuvette drain to waste chamber
NO vent VL29 SL49 NC vent diff waste chamber
NO not used VL30 SL51 NC sample pressure to mix chamber
NO vent for mix chamber VL31 SL54 NC sheath flow from tank to bottom half of flow cell
NO not used VL32 SL54 NC sheath flow from tank to top half of flow cell
NO not used VL33 SL49 NC sample drain from flow cell
NO not used VL34 SL55 NC forward flush drain to waste
NO not used VL35 SL12 NO high vacuum to low vacuum regulator
NC not used VL36 SL61 NC diff rinse pump - cleaner
NO diff rinse pump - diluent VL37 SL46 NC path for flush upper section of flow cell
NO not used VL38 SL17 NC high vacuum to VIC
NO low vacuum to VIC VL40 SL57 NC vacuum for secondary backwash drain to diff waste chamber
NO not used VL43 SL2 NC dispense lyse to WBC bath
NO lyse pumps fill VL44 SL60 NC diff preservative pump dispense
NO diff preservative pump fill VL45 -- Not used
1
GENERAL
continued
PN 4237187A (November 1993) 15
GENERAL SERVICE AND MAINTENANCE
Pinch
Valve Solenoid Pinch Valve Function
continued
GENERAL
1
VL46 SL3 NC pressure to lyse pumps
VL47 SL61 NC cleaner pump dispense
VL48 SL2 NC Hgb blank pump dispense
VL49 SL48 NC vacuum supply to drain mix chamber to diff waste chamber
VL50 SL56 NC • vacuum
VL52 SL44 NO blood and diff lyse delivery to mix chamber, Secondary mode
VL53 SL45 NO diff preservative delivery path to mix chamber
VL54 SL65 NO blood and diff lyse delivery to mix chamber, Primary mode
VL55 SL52 NO sample flow from mix chamber to flow cell VL56 SL50 NC vacuum on mix chamber
VL57 SL48 NC drain path not used
VL58 SL46 NC flush flow cell lower section
VL59 SL41 NC reverse flush drain from flow cell to waste chamber
VL68 NA SL24 NC Primary mode aspiration disabled
VL69 SL24 NC from secondary aspiration pump to main waste chamber
VL73 NA SL58 NC Secondary mode aspiration and backwash disabled VL74 NA SL58 NC Primary mode aspiration and backwash disabled VL76 NA SL64 NC Primary mode aspiration disabled
Table 6 Pinch Valves
NO vacuum to lyse pumps and Hgb blank pump
NO cleaner pump fill
NO Hgb blank pump fill
NO drain path for diff waste chamber
sheath fill
not used
NO • sheath flow
sheath pressure
flow cell rinse to bottom of flow cell
NC not used
NC not used
NC not used
NO pressure on mix chamber
NO not used
NO not used
NO not used
NO from primary aspiration pump to main waste chamber
NO Secondary mode aspiration enabled
continued
16 PN 4237187A (November 1993)
COMPONENTS
Table 6 Pinch Valves
Pinch
Valve Solenoid Pinch Valve Function
continued
VL77 SL62 NC Secondary mode backwash to waste chamber
NO not used VL78 SL58 NO diff lyse delivery enabled, Primary mode
VL79 NA SL64 NC Secondary mode aspiration disabled
VL81 SL59 NO Primary mode backwash enabled VL82 SL23 NO needle rinse backwash diluent VL83 SL58 NO diff lyse delivery enabled, Secondary mode VL84 -- Not used VL85 SL59 NO Secondary mode backwash enabled VL86 SL59 NC vacuum via vacuum tank to aspiration pumps
NO pressure to aspiration pumps VL87 SL13 NC Secondary mode rinse trough drain
NO Primary mode rinse trough drain VL88 SL62 NC Primary mode backwash to waste chamber
NO not used
1
GENERAL
Solenoid (SL) Functions
Table 7 lists the solenoid functions and actions. See Figure 3 for most locations.
Table 7 Solenoids
Solenoid Function Action
Diluter
SL1 Rinse needle PM1 SL2 Lyse/Hgb valve VL43, VL48 SL3 Lyse/Hgb pump VL46 SL4 Hgb drain valve VL28 SL5 Backwash pump PM6 SL6 Hgb/waste drain VL18 SL7 WBC dispense VL6, PM3
continued
PN 4237187A (November 1993) 17
GENERAL SERVICE AND MAINTENANCE
GENERAL
1
Table 7 Solenoids
Solenoid Function Action
continued
SL8 RBC dispense VL5, PM2 SL9 Whole blood aspiration/
Overflow bath drain SL10 WBC count VL16 SL11 RBC count VL13 SL12 Bath drain (VL 14) VL15, VL22, VL23,
SL13 Manual backwash VL10, VL87 SL14 Cleaner VL27 SL15 Sample valve return CL2 SL16 Vacuum isolator drain VL19 SL17 Segment sample VL38, VL68, VL69,
SL18 WBC dispense to RBC bath VL1, VL2, VL14 SL19 WBC sample mix WBC bath SL20 RBC sample mix RBC bath
Transport System
SL21 Right lift CL5 SL22 Left lift CL4 SL23 Needle up VL82, CL3 SL24 Manual probe in VL69, VL68, CL1 SL25 Ram tube CL6 SL26 Bar-code reader VL66, CL7 SL27 Backward bed rock CL8 SL28 Bed lock L1 SL29 Stripper plate CL10 SL30 Forward bed rock CL8 SL31 Flipper retract L41, L42 SL32 Spare SL33 Needle vent pressure QD1-11 to needle SL34 Manual probe out - NO CL1 SL35 Spare
VL3, VL86
VL24, VL35
CL2
continued
18 PN 4237187A (November 1993)
COMPONENTS
Table 7 Solenoids
Solenoid Function Action
continued
SL36 Belt advance CL9 SL37 Needle down CL3 SL38 Spare SL39 Spare SL40 Spare
Diff
SL41 Rev flush VL59 SL42 Rinse - diff VL11, VL12 SL43 Lyse - diff PM1 SL44 Lyse disable - diff, manual VL52 SL45 Quench VL53 SL46 Flush diff VL37, VL58 SL47 Cleaner - diff PM10 SL48 Drain - diff VL57, VL49 SL49 Exit - diff VL29, VL33 SL50 LATRON deliver VL26, VL56 SL51 Sample pressure - diff VL30 SL52 Sample disable - diff VL55 SL53 Waste - diff VC13 SL54 Sheath flow VL31, VL32 SL55 Forward flush VL34 SL56 Sheath refill VL50 SL57 Probe/needle drain VL20, VL40 SL58 Aspirate enable/
Lyse disable SL59 Backwash disable VL81, VL85 SL60 Quench enable VL44 SL61 Cleaner enable VL36, VL47 SL62 Rinse metering tube VL77, VL88 SL63 Lyse - diff enable VL4 SL64 Segment diff - NO VL76, VL79 SL65 Lyse disable - diff auto VL54
VL78, VL83 (VL74, L73)
1
GENERAL
continued
PN 4237187A (November 1993) 19
GENERAL SERVICE AND MAINTENANCE
37
1
5
50
47
ON REAR
THREE WAY PINCH VALVEDUAL ACTION PINCH VALVESOLENOIDMINI PINCH VALVE
7187001A
19
1613
ROCKER BED
DILUTER BASE (IN FRONT OF THE ROCKER BED)
MF13
46
MF11
363534333231
30
43384044
474648
49
29
28
24
23
221820
1514
7
4
32
1
5
6
10 12 11
22
25
24
58
69
68
77
88
59
54
52
26
82
55
21
58 57
56 53
13
51
15
52
17 53
41
54 44 55 45
56
57 48
60 49
61 50
65
BSV
7985
81 76
78
83
74
73
34 64 27 26 36 23
9
30 62 63 37 59 28
29
86
TOP
87
BOTTOM
Table 7 Solenoids
Solenoid Function Action
continued
GENERAL
1
SL81 Level sense MF1
Figure 3 illustrates solenoid and pinch valve locations.
20 PN 4237187A (November 1993)
Figure 3 Pinch Valve and Solenoid Locations
2.1 GENERAL INFORMATION
For optimum performance of your COULTER®STKS with Reticulocyte Analysis, you must calibrate the CBC parameters. The WBC differential parameters are calibrated at the factory; they do not require calibration in the laboratory.
CALIBRATION 2
Your laboratory is responsible for the final calibration of the CBC parameters, for recording the calibration factors, and for establishing reference values for ramp-pulse and precision-pulse tests. Coulter recommends S-CAL alternative to whole-blood calibration.
Calibrate the CBC parameters:
At installation
After the replacement of any component that involves the dilution characteristics (such as the BSV), or the primary measurements (such as the apertures)
When advised to do so by your Coulter Service representative
Daily quality assurance programs monitor ongoing instrument and system performance. However, we recommend that you verify the status of instrument calibration quarterly, and/or when controls begin to show evidence of unusual trends, or when controls exceed the manufacturer’s defined acceptable limits.
®
calibrator, or an exact equivalent, as an acceptable
2
CALIBRATION
Although the instrument is relatively insensitive to minor room temperature changes, perform calibration when the room temperature is stable and within the normal ambient temperature range. If the average ambient room temperature changes more than 10°F from the calibrating temperature, it is advisable to verify calibration and to recalibrate if necessary.
In the normal process of tracking data for an extended period of time, your laboratory can make a specific decision to recalibrate a given parameter. Never adjust to a specific value for an individual sample.
PN 4237187A (November 1993) 21
CALIBRATION
2.2 PRECALIBRATION STEPS
CALIBRATION
If a problem or malfunction occurs while performing these procedures, see the Troubleshooting procedures later in this manual. If the corrective procedures fail to eliminate the problem, call Coulter Customer Operations immediately.
Before calibrating, perform the following procedures in the order given.
1. Follow a, b or c as it applies to your instrument.
a. If the instrument is in the shutdown condition, and you
routinely shut down the instrument for a minimum of 30 minutes every 24 hours in COULTER CLENZ go to step 2.
®
cleaning agent,
2
b. If you routinely use COULTER CLENZ cleaning agent, but you
are beginning calibration after processing patient samples, shut down the instrument in the cleaning agent for 30 minutes to assure optimum aperture cleanliness before proceeding.
CAUTION
Do not aspirate bleach; bleach can damage the flow cell.
c. If you do not routinely use COULTER CLENZ cleaning agent,
perform the Bleach the Apertures procedure in Chapter 3.
2. Check the reagent containers for:
a. Sufficient quantity
b. Not beyond expiration date
c. No precipitates, turbidity, particulate matter, or unusual color
d. Proper connections between the Diluter and the reagent
containers
22 PN 4237187A (November 1993)
PRECALIBRATION STEPS
WARNING
The contents of the waste container and its associated tubing can include residual biological material and must be handled with care. Avoid skin contact and clean up spills immediately. Dispose of the contents of the waste container in accordance with acceptable laboratory procedures.
3. Check the waste container for:
a. Sufficient capacity
b. Proper connections
4. Perform the startup procedures, including a quality-control check, as described in the Operator’s Guide, PN 4237188. Verify that the following are acceptable: background count, reproducibility, carryover, and control recovery.
Reproducibility Check, CBC/Diff
Before you calibrate the STKS with either the S-CAL Kit or whole blood, perform a 10-sample reproducibility study on the CBC parameters in the Primary mode. You can also perform this study if you need to check instrument precision for the CBC or diff parameters.
In the Secondary mode, when you perform a Reproducibility check, do not cycle more than eight samples per tube. If you are running 31 samples, use four separate vials.
2
CALIBRATION
Sample Requirements
Collect enough blood for 10 cycles from a single donor who:
Is receiving no medication.
Has normal hematologic parameters, with a WBC count of 10,000 ±1,000.
Has normal erythrocyte, leukocyte and platelet morphology and
PN 4237187A (November 1993) 23
CALIBRATION
If you want to check the diff parameters, with diff values of
Neutrophils 40 to 72% Lymphocytes 17 to 45% Monocytes 4 to 12% Eosinophils 0 to 10% Basophils 0 to 1%
Procedure, Primary Mode
CALIBRATION
2
1. At the Diluter, press
PRIME APERT to activate the pneumatics.
2. At the Analyzer:
a. In SYSTEM CONFIGURATION, be sure that the BLOOD
DETECTOR is set to ENABLED.
b. In SYSTEM RUN, verify that the # aspirations/tube is 1.
3. At the DMS, select
Special Functions
Calibration
Reproducibility
If data remains from a previous study, press Delete Table.
4. Cycle one sample of normal whole blood in the Primary mode.
CAUTION
Do not pierce a specimen tube more than five times; additional piercing can obstruct or clog the needle.
5. At the Analyzer, set the number of aspirations per tube to 5.
6. Aliquot the well-mixed normal whole-blood sample into two red­topped tubes.
a. Place the tubes into consecutive positions in a cassette, place
the cassette in the loading bay, and press
START/CONT.
b. Monitor the system for normal sample flow.
24 PN 4237187A (November 1993)
Loading...
+ 98 hidden pages