Corning 430 Instruction Manual

430 pH Meter
Instruction Manual
This manual contains complete instructions for setting up and using the 430 pH meter. Applications information is also available.
The information contained in this manual was correct at the time of going to print. However, we continue to improve products and reserve the right to change specifications, equipment and maintenance procedures at any time.
This manual is copyrighted, and all rights are reserved. No part of this manual may be reproduced by any means or in any form without prior consent in writing.
The power supply unit is classed as IEC Class II equipment (equipment providing an adequate degree of protection against electric shocks, in which additional safety precautions, for example, double or reinforced insulation, are included). The 430 is intended for use by persons knowledgeable in safe laboratory practices. If the 430 is not used in accordance with these instructions for use, the protection provided by the equipment may be impaired.
The 430 is suitable for direct current only. This equipment generates, uses and can radiate radio frequency
energy and if not installed and used in accordance with the instruction manual, may cause interference with radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment. Operation of this equipment in a residential area is likely to cause interference in which case the user at his own expense will be required to take whatever measures may be required to correct the interference.
There are no user replaceable parts in the 430 or power supply unit. Do not remove the covers.
109117-1 Rev. A, 10/96
1
Contents
1 Set Up
1.1 Unpacking and Installation ................................................ 3
1.2 Input and Output Connections........................................... 4
1.3 Display and Controls ......................................................... 4
2 Operation
2.1 Calibrating a pH Electrode ................................................ 6
2.2 Measuring Samples........................................................... 7
2.3 Meter Reset and Buffer Sets ............................................. 7
2.4 mV Measurements ............................................................ 8
2.5 Temperature Measurements ............................................. 8
2.6 Chart Recorder .................................................................. 9
3 Support Information
3.1 Basic Theory.................................................................... 10
3.2 Operating Hints................................................................ 11
3.3 Maintenance .................................................................... 12
3.3.1 Meter Maintenance.......................................................... 12
3.3.2 Electrode Maintenance.................................................... 12
3.4 Problem Solving .............................................................. 14
3.4.1 Meter Reset ..................................................................... 14
3.4.2 Meter Test ....................................................................... 14
3.4.3 Meter Error Codes ........................................................... 15
3.4.4 Electrode Problem Solving .............................................. 16
3.5 Specifications .................................................................. 18
Consumables and Accessories ............................... inside rear cover
Warranty .................................................................. inside rear cover
2
109117-1 Rev. A, 10/96
1.1 Unpacking and Installation
1. The main carton will contain the following: Meter with a shorting (test) clip in the pH socket
Power Supply This Instruction Manual Memo Cards, English, French, German, Italian, Spanish Warranty Card Calibration Certificate
2. If you ordered the meter with an electrode, electrode arm
and accessories, the top tray of the main carton will contain these items. Refer to the accessories packing sheet for a list of the additional components.
3. To attach the electrode arm to the pH meter:
a. The electrode arm can be attached to the left or right-
hand side. Using the screwdriver supplied remove the appropriate end cap from the pH meter.
b. Slide the arm base into the recess and tighten the fixing
screw. Replace the end cap.
c. Fit the electrode arm onto the post. Adjust the tension
screw as required.
1
electrode arm
arm base
tension screw
fixing screw
4. Fill in and return the warranty card. For your own record
make a note of the serial number, date of purchase and supplier on the inside front cover of this manual.
109117-1 Rev. A, 10/96
end cap
screwdriver
3
1.2 Input and Output Connections
1
DC - power
supply unit socket
REC - recorder output socket, see Section 2.6 for details
1. Disconnect the shorting clip from the pH socket and retain
it by clipping it over the socket. Connect the electrode. If you are using an electrode incorporating ATC connect the other lead to the ATC socket.
NOTE The 430 will not accept separate reference and pH
electrodes.
2. If you are using a separate ATC probe connect it to the
ATC socket.
3. Connect the power supply unit to the DC socket.
ATC - ATC (automatic temperature compensation) probe socket
pH - combination electrode socket
1.3 Display and Controls
4
109117-1 Rev. A, 10/96
1.3 Display and Controls (cont)
mode
cal
on/off
read
430 Display
mode
slope value displayed
pH trend indicator
Selects pH, mV or temperature mode.
Starts a calibration sequence in pH mode.
Turns the display off, and places the meter in standby mode. Wakes meter up.
Starts sample measurement in all modes. Press again to freeze display at endpoint.
1
calibration buffer indicator
automatic temperature compensation
Example: pH Reading, with ATC
109117-1 Rev. A, 10/96
5
2.1 Calibrating a pH Electrode
1 point calibration
Place the tip of the electrode in the calibration buffer and
2
press It does not matter which buffer you do first.
The meter will automatically endpoint when the reading is stable and the appropriate buffer symbol will appear on the display.
Rinse the tip of the electrode with distilled or deionized water and goto Measuring Samples (Section 2.2) or 2 point calibration.
NOTE If you are using an ATC probe (or an electrode
2 point calibration
Place the tip of the electrode in the second buffer and press
The meter will automatically endpoint when the reading is stable and the appropriate buffer symbol will appear on the display. The meter will also briefly display the electrode slope value.
Rinse the tip of the electrode with distilled or deionized water and goto Measuring Samples (Section 2.2).
cal
incorporating ATC), temperature is being measured and the buffer value is being corrected for temperature (see Section 2.5). If you are not using an ATC in the system, the Model 430 assumes a temperature of 25°C.
cal
NOTE For maximum accuracy, we recommend using an ATC
probe or the 3-in-1 Combination electrode which includes an ATC probe. If you do not have an ATC probe you should make sure that all buffers and samples are at 25°C.
6
109117-1 Rev. A, 10/96
2.2 Measuring Samples
NOTE Good laboratory practice dictates that electrodes should
be calibrated at least daily.
Place the tip of the electrode in the sample and press to start the measurement. The decimal point will be flashing whenever the electrode is reading.
When the reading is stable, press to freeze the display.
NOTE The display of the Model 430 will show both a digital
and an analog pH reading.
2.3 Meter Reset and Buffer Sets
This process resets the meter electronics and places the selected buffer set in the meter memory. When calibrating you can use any two of the three buffers in the selected set.
Buffer Set 1 (b=1): 4.00, 7.00, 10.01 pH Buffer Set 2 (b=2): 4.01, 7.00, 9.21 pH Buffer Set 3 (b=3): 4.01, 6.86, 9.18 pH
To reset the meter and buffer set:
Turn the meter off
on/off
read
read
2
Press and hold and turn meter on
Release key.
The display will show b=1, b=2, or b=3.
Press until display shows the buffer set you want.
Press and the meter will reset and put the selected buffer
109117-1 Rev. A, 10/96
mode
cal
read
mode
set into the meter memory.
on/off
7
Loading...
+ 14 hidden pages