Calibrating
1. Select O2% mode
2. 1-point calibration
Example
%
hold probe
10␣ mm
above fresh
water
Auto endpoint or
to return to sample measurement, or
continue with step 3 for the second cal
point
3. 2-point calibration - repeat step 2, placing the probe
in zero oxygen standard
auto or manual endpoint
4. Rinse probe
H20
99.
25.0°c
cal1
ATC
100.
25.0°c
cal1
ATC
0.
25.0°c
cal 2
ATC
%
%
Measuring Samples
1. Select O2% or O2 mg/L mode
2. Measure sample
Example
%
89.
25.0°c
sample
Auto endpoint or
3. Rinse and store probe
■ If you press
measuring immediately, and you will not need to press
read
■ The decimal point flashes during measurement.
When␣ the probe output has stabilized
the␣ stability␣ indicator␣ appears.
Manual endpoint - press
Auto endpoint - the 317 automatically endpoints
mode
to select % or mg/L, the 317 begins
to start the sample measurement.
read
ATC
89.
25.0°c
ATC
H20
to endpoint
%
Using the Memory
Entering a Reading into Memory
The 317 can store up to 10 endpointed results.
Press
M
when the measurement has endpointed.
M 1 is displayed (or M 2 to M 10 if readings have
already been stored).
M indicates the memory is full.
Recalling Memory
You can only recall stored memories if the current
measurement has endpointed.
R
- the last stored memory is displayed. Press ▼
Press
to scroll through memories. RM 1 to RM 10 indicates
which memory is being displayed.
Clearing last Memory
Press R - the last stored memory is displayed. Press
to clear this memory. Only the last entered measurement can be cleared, i.e. M 2 cannot be cleared without
first clearing M 3.
M
Clearing all Memories
Press
R
then press ▼ until M C is displayed.
Press
M
to clear the memories (press
without clearing the memories).
read
to exit
Using Continuous
Measurement␣ Mode
In normal mode, the 317 will auto-off 10 minutes after
the last key press, whether the unit is measuring or has
endpointed. If you select continuous measurement
mode, the 317 will not auto-off.
To select continuous measurement mode press and
M
for 2 seconds.
hold
Using Continuous
Data Transfer Mode
In continuous data transfer mode readings are sent to
the data output approximately every second. If the
measurement endpoints (manually, or using auto
endpoint) data transfer stops. If you recall memories all
stored readings are output (from the first to the last
stored).
To select continuous data transfer mode press and hold
R
for 2 seconds.
■ If you want to use this mode to monitor continually,
you also need to select continuous measurement
mode.
■ To maintain continuity of sample readings
calibration data is not output in continuous data
transfer mode.
Interfacing
Serial Output
The 317 can interface with computers, printers and
other RS232 compatible devices. The RS232
Interface Kit (Cat. 470445) converts data from the 6
pin socket to RS232 format.
Type of communication - uni directional, baud rate
1200, data format - 1 start bit : 8 data bits : 1␣ stop
bit : no parity.
■ Data is sent to the output socket at
measurement endpoint (manual or automatic)
and memory recall.
■ To use the output to monitor continually, you
need to select continuous measurement mode
and continuous data transfer mode.
CAUTION: The 317 does not meet IP67 when the
waterproof output socket seal has been removed.
Maintenance
General
The 317 requires very little maintenance.
Occasionally wipe the meter with a damp cloth. The
casework is made of ABS/PC which is known to be
affected by some organic solvents, including
toluene, xylene and methyl-ethyl-ketone. It is good
practice to wipe away any spillages as soon as they
occur.
Every six months inspect and lightly grease the
socket seals, and connector and battery ‘O’ rings
with the grease supplied. Replace the ‘O’ rings if
damaged.
CAUTION: To prevent permanent damage care
should be taken to prevent liquid entering the 317
when fitting probes or replacing batteries. If it does,
remove the batteries and allow the meter to dry
before using.
Replacing the Batteries
Replace batteries as soon as the low battery
indicator␣ appears. To prevent data loss always turn
the 317 off using the
batteries. See the Installation section for details on
installing batteries. Do not mix old and new
batteries.
NOTE: Check the Program Menu settings after
changing the batteries.
Probe Maintenance
Clean the probe tip and refill/replace the membrane
cap as described in the dissolved oxygen probe
product insert.
mode
key before replacing the
Problem Solving
+ Err 1 - measurement out of range
•
Check probe is connected and immersed in sample.
Check wetting cap has been removed.
Err 2 - cal 2 out of range
Check fresh zero oxygen solution is used. Clean the
probe tip and refill the membrane cap. Check the
membrane for damage, replace as necessary. Check
barometric pressure and salinity settings.
Err 3 - cal 1 out of range
Clean the probe tip and refill the membrane cap.
Check the membrane for damage, replace as
necessary. Check barometric pressure and salinity
settings.
Data Entry Errors
Entered barometric pressure value changes to 500
or 1100 - the 317 will not accept barometric pressure
values outside this range.
Entered salinity value changes to 0 or 70 - the 317
will not accept salinity values outside this range.
Operating Hints
■ Remove the wetting cap from the end of the
probe before use.
■ The probe should remain connected to the 315
to maintain polarization.
■ When measuring, the sample should be stirred
at a constant speed.
■ Rinse probe with distilled water between
measurements, and after use.
■ Do not use standards after the expiration date.
■ Zero oxygen standard absorbs oxygen if left
exposed to air and this will cause inaccurate
calibrations. Always use fresh standard.
■ For greatest accuracy, standards and samples
should be at the same temperature.
■ Regular maintenance is important to ensure
optimum performance. Refer to probe insert for
details.
317 Dissolved Oxygen
Corning Incorporated
Science Products Division
Corning, New York 14831
USA. Tel: 1-800-222-7740/607-974-0353
For Technical Assistance:
US and Canada call 1-800-222-7740/607-974-0353
Specifications
Measurement Ranges
0 to 200%
O
2
0.0 to 20.0 mg/L
Temp. –5.0 to 105.0°C
Resolution
1%
O
2
0.1 mg/L
Temp. 0.1°C
Relative Accuracy
1%
O
2
Temp. ± 0.5°C
Memory
10 memories
Barometric Pressure
500 to 1100 mbar,
selectable
Salinity
0 to 70 ppt, selectable
Temperature
Compensation
0.0 to 40.0°C (ATC probe)
Display
LCD
Outputs
Serial
Auto Off
10 minutes from last
keypress
Operating Conditions
Temperature 5 to 40°C
Humidity 5 to 80%
(non condensing)
Size/Weight
3 1/4 x 7 3/4 x 1 3/4 inches
(85 x 200 x 45 mm)
1lb (0.45kg)
Batteries
4 x AA/LR6
alkaline1.5V␣ d.c.
Regulatory Compliance
Complies with Part 15
of␣ FCC Rules (Class A
computing device).
Waterproof to IP67*
*when used as directed with
probes with waterproof
covers.
470457 a595
Installation
Input Connectors
connector
cap
’O‘ ring
seal
data output
socket
input
socket
Connect the dissolved oxygen probe as follows:
input socket
alignment
‘O’ rings
connector
cover
1. Make sure connector cap is loosened.
2. Connect the probe mini-DIN connector to the
meter input socket.
3. Lightly grease the ‘O’ rings with the grease
supplied, then slide the waterproof cover over
the socket. Make sure the cover is lined up
correctly with the socket.
NOTE: Lightly greasing the cables will help removal
of the connector cover in future.
4. Finger tighten the connector cap.
5. Allow a minimum of 1 hour for polarization.
CAUTION: To avoid damaging the connector always
loosen the connector cap before removing the
cover.
Installing the Batteries
hand grip coin battery
1. Remove the hand grip, and unscrew the battery
cover using a coin.
2. Insert the batteries, following the illustration.
3. Lightly grease the ‘O’ ring seal, and replace the
battery cover and hand grip.
cover
Fitting the Wrist Strap
Program Menu
The Program Menu allows you to set barometric
pressure and salinity. The correction will only be applied
mg/L mode. You can only enter the Program Menu
in O
2
if the current measurement has endpointed.
Press and hold the
the Program Menu - appears on the display.
mode
Press
exit the Program Menu at any time. If you press
when a value is flashing that value will not be entered.
P1 Barometric pressure affects the partial pressure
of O
barometric pressure can be set between 500
and 1100 mbar. The 317 has a preset
barometric pressure value of 1013 mbar. Use ▲
and ▼ to change the value, press
enter the value and move on.
P2 Salinity affects the solubility of O
set between 0 and 70 ppt (g/L). The 317 has a
preset salinity value of 0 ppt. Use ▲ and ▼ to
change the value, press
and move on.
read
Press
Correction Tables
Barometric Pressure Salinity
Entered BP Correction Entered Salinity Correction
mbar Factor (%) ppt Factor (%)
500 49.4 0 100.0
550 54.3 5 97.2
600 59.2 10 94.5
650 64.2 15 91.8
700 69.1 20 89.2
750 74.0 25 86.7
800 79.0 30 84.3
850 83.9 35 81.9
900 88.8 40 79.6
950 93.8 45 77.4
1000 98.7 50 75.2
1013 100.0 55 73.1
1050 103.7 60 71.1
1100 108.6 65 69.1
mbar = mmHg x 1.333
inches Hg x 33.858
mode
key for 2 seconds to access
to scroll through the options. Press
, and therefore the values obtained. The
2
mode
, and can be
2
mode
to enter value
to exit the Program Menu.
70 67.1
read
read
to
to
Controls
Press Press & Hold
& Release for 2 Seconds
starts/stops turns auto
measurement endpoint on/off
starts calibration no action
selects O
O
stores result turns continuous
in memory (up to 10) measurement
recalls result turns continuous
from memory data transfer
% or enters Program Menu
2
mg/L/off
2
on/off
on/off
Display
stability/auto
endpoint indicator
units shown
low batt.
indicator
shown
here
100.
25.0 °c 10
ATC
temperature
%
M
memory, cal and
error number
shown here
here
prog. menu,
cont. read and
data transfer
icons shown
here
Spares and Accessories
476622 D.O. Probe (6' cable)
476623 D.O. Probe (30' cable)
470463 Sinkable Stainless Steel Sleeve and Cage
470444 Rubber Meter Holder and Neck Strap
470443 Hard Carrying Case
470447 Wrist Strap
470448 Instrument Sealing Kit
470449 Neck Strap
470445 RS232 Interface Kit
470438 Batteries, AA, 4/pk
470439 Guide Book to Conductivity/Dissolved Oxygen
473736 Fill Solution and Membrane Caps, 3/pk
470466 D.O. Electrode Fill Solution, 30 mL
477656 D.O. Sensor Cleaning Compound
473625 Zero Oxygen Standard, 500 mL
473739 Zero Oxygen Standard, Sachets, 30/case