Corning 317 Instruction Manual

Calibrating
1. Select O2% mode
2. 1-point calibration
Example
%
hold probe 10␣ mm above fresh water
Auto endpoint or
to return to sample measurement, or continue with step 3 for the second cal point
3. 2-point calibration - repeat step 2, placing the probe in zero oxygen standard
auto or manual endpoint
4. Rinse probe
H20
99.
25.0°c
cal1
ATC
100.
25.0°c
cal1
ATC
0.
25.0°c
cal 2
ATC
%
%
Measuring Samples
1. Select O2% or O2 mg/L mode
2. Measure sample
Example
%
89.
25.0°c
sample
Auto endpoint or
3. Rinse and store probe
If you press
measuring immediately, and you will not need to press
read
The decimal point flashes during measurement.
When␣ the probe output has stabilized the␣ stability␣ indicator␣ appears. Manual endpoint - press Auto endpoint - the 317 automatically endpoints
mode
to select % or mg/L, the 317 begins
to start the sample measurement.
read
ATC
89.
25.0°c
ATC
H20
to endpoint
%
Using the Memory
Entering a Reading into Memory
The 317 can store up to 10 endpointed results.
Press
M
when the measurement has endpointed.
M 1 is displayed (or M 2 to M 10 if readings have already been stored).
M indicates the memory is full.
Recalling Memory
You can only recall stored memories if the current measurement has endpointed.
R
- the last stored memory is displayed. Press
Press to scroll through memories. RM 1 to RM 10 indicates which memory is being displayed.
Clearing last Memory
Press R - the last stored memory is displayed. Press to clear this memory. Only the last entered measure­ment can be cleared, i.e. M 2 cannot be cleared without first clearing M 3.
M
Clearing all Memories
Press
R
then press until M C is displayed.
Press
M
to clear the memories (press
without clearing the memories).
read
to exit
Using Continuous Measurement␣ Mode
In normal mode, the 317 will auto-off 10 minutes after the last key press, whether the unit is measuring or has endpointed. If you select continuous measurement mode, the 317 will not auto-off.
To select continuous measurement mode press and
M
for 2 seconds.
hold
Using Continuous Data Transfer Mode
In continuous data transfer mode readings are sent to the data output approximately every second. If the measurement endpoints (manually, or using auto endpoint) data transfer stops. If you recall memories all stored readings are output (from the first to the last stored).
To select continuous data transfer mode press and hold
R
for 2 seconds.
If you want to use this mode to monitor continually,
you also need to select continuous measurement mode.
To maintain continuity of sample readings
calibration data is not output in continuous data transfer mode.
Interfacing
Serial Output
The 317 can interface with computers, printers and other RS232 compatible devices. The RS232 Interface Kit (Cat. 470445) converts data from the 6 pin socket to RS232 format.
Type of communication - uni directional, baud rate 1200, data format - 1 start bit : 8 data bits : 1␣ stop bit : no parity.
Data is sent to the output socket at
measurement endpoint (manual or automatic) and memory recall.
To use the output to monitor continually, you
need to select continuous measurement mode and continuous data transfer mode.
CAUTION: The 317 does not meet IP67 when the waterproof output socket seal has been removed.
Maintenance
General
The 317 requires very little maintenance. Occasionally wipe the meter with a damp cloth. The casework is made of ABS/PC which is known to be affected by some organic solvents, including toluene, xylene and methyl-ethyl-ketone. It is good practice to wipe away any spillages as soon as they occur.
Every six months inspect and lightly grease the socket seals, and connector and battery ‘O’ rings with the grease supplied. Replace the ‘O’ rings if damaged.
CAUTION: To prevent permanent damage care should be taken to prevent liquid entering the 317 when fitting probes or replacing batteries. If it does, remove the batteries and allow the meter to dry before using.
Replacing the Batteries
Replace batteries as soon as the low battery indicator␣ appears. To prevent data loss always turn the 317 off using the batteries. See the Installation section for details on installing batteries. Do not mix old and new batteries.
NOTE: Check the Program Menu settings after changing the batteries.
Probe Maintenance
Clean the probe tip and refill/replace the membrane cap as described in the dissolved oxygen probe product insert.
mode
key before replacing the
Problem Solving
+ Err 1 - measurement out of range
Check probe is connected and immersed in sample. Check wetting cap has been removed.
Err 2 - cal 2 out of range
Check fresh zero oxygen solution is used. Clean the probe tip and refill the membrane cap. Check the membrane for damage, replace as necessary. Check barometric pressure and salinity settings.
Err 3 - cal 1 out of range
Clean the probe tip and refill the membrane cap. Check the membrane for damage, replace as necessary. Check barometric pressure and salinity settings.
Data Entry Errors
Entered barometric pressure value changes to 500 or 1100 - the 317 will not accept barometric pressure values outside this range.
Entered salinity value changes to 0 or 70 - the 317 will not accept salinity values outside this range.
Operating Hints
Remove the wetting cap from the end of the
probe before use.
The probe should remain connected to the 315
to maintain polarization.
When measuring, the sample should be stirred
at a constant speed.
Rinse probe with distilled water between
measurements, and after use.
Do not use standards after the expiration date.
Zero oxygen standard absorbs oxygen if left
exposed to air and this will cause inaccurate calibrations. Always use fresh standard.
For greatest accuracy, standards and samples
should be at the same temperature.
Regular maintenance is important to ensure
optimum performance. Refer to probe insert for details.
317 Dissolved Oxygen
Corning Incorporated Science Products Division Corning, New York 14831 USA. Tel: 1-800-222-7740/607-974-0353
For Technical Assistance: US and Canada call 1-800-222-7740/607-974-0353
Specifications
Measurement Ranges
0 to 200%
O
2
0.0 to 20.0 mg/L
Temp. –5.0 to 105.0°C
Resolution
1%
O
2
0.1 mg/L
Temp. 0.1°C
Relative Accuracy
1%
O
2
Temp. ± 0.5°C
Memory
10 memories
Barometric Pressure
500 to 1100 mbar, selectable
Salinity
0 to 70 ppt, selectable
Temperature Compensation
0.0 to 40.0°C (ATC probe)
Display
LCD
Outputs
Serial
Auto Off
10 minutes from last keypress
Operating Conditions
Temperature 5 to 40°C Humidity 5 to 80% (non condensing)
Size/Weight
3 1/4 x 7 3/4 x 1 3/4 inches (85 x 200 x 45 mm) 1lb (0.45kg)
Batteries
4 x AA/LR6 alkaline1.5V␣ d.c.
Regulatory Compliance
Complies with Part 15 of␣ FCC Rules (Class A computing device). Waterproof to IP67*
*when used as directed with probes with waterproof covers.
470457 a595
Installation
Input Connectors
connector cap
’O‘ ring seal
data output socket
input socket
Connect the dissolved oxygen probe as follows:
input socket
alignment
‘O’ rings
connector cover
1. Make sure connector cap is loosened.
2. Connect the probe mini-DIN connector to the meter input socket.
3. Lightly grease the ‘O’ rings with the grease supplied, then slide the waterproof cover over the socket. Make sure the cover is lined up correctly with the socket.
NOTE: Lightly greasing the cables will help removal of the connector cover in future.
4. Finger tighten the connector cap.
5. Allow a minimum of 1 hour for polarization.
CAUTION: To avoid damaging the connector always loosen the connector cap before removing the cover.
Installing the Batteries
hand grip coin battery
1. Remove the hand grip, and unscrew the battery cover using a coin.
2. Insert the batteries, following the illustration.
3. Lightly grease the ‘O’ ring seal, and replace the battery cover and hand grip.
cover
Fitting the Wrist Strap
Program Menu
The Program Menu allows you to set barometric pressure and salinity. The correction will only be applied
mg/L mode. You can only enter the Program Menu
in O
2
if the current measurement has endpointed.
Press and hold the the Program Menu - appears on the display.
mode
Press exit the Program Menu at any time. If you press when a value is flashing that value will not be entered.
P1 Barometric pressure affects the partial pressure
of O barometric pressure can be set between 500 and 1100 mbar. The 317 has a preset barometric pressure value of 1013 mbar. Use and to change the value, press enter the value and move on.
P2 Salinity affects the solubility of O
set between 0 and 70 ppt (g/L). The 317 has a preset salinity value of 0 ppt. Use and to change the value, press and move on.
read
Press
Correction Tables
Barometric Pressure Salinity
Entered BP Correction Entered Salinity Correction
mbar Factor (%) ppt Factor (%)
500 49.4 0 100.0 550 54.3 5 97.2 600 59.2 10 94.5 650 64.2 15 91.8 700 69.1 20 89.2 750 74.0 25 86.7 800 79.0 30 84.3 850 83.9 35 81.9 900 88.8 40 79.6
950 93.8 45 77.4 1000 98.7 50 75.2 1013 100.0 55 73.1 1050 103.7 60 71.1 1100 108.6 65 69.1
mbar = mmHg x 1.333
inches Hg x 33.858
mode
key for 2 seconds to access
to scroll through the options. Press
, and therefore the values obtained. The
2
mode
, and can be
2
mode
to enter value
to exit the Program Menu.
70 67.1
read
read
to
to
Controls
Press Press & Hold & Release for 2 Seconds
starts/stops turns auto measurement endpoint on/off
starts calibration no action
selects O O
stores result turns continuous in memory (up to 10) measurement
recalls result turns continuous from memory data transfer
% or enters Program Menu
2
mg/L/off
2
on/off
on/off
Display
stability/auto endpoint indicator
units shown
low batt. indicator shown here
100.
25.0 °c 10
ATC
temperature
%
M
memory, cal and error number shown here
here
prog. menu, cont. read and data transfer icons shown here
Spares and Accessories
476622 D.O. Probe (6' cable) 476623 D.O. Probe (30' cable) 470463 Sinkable Stainless Steel Sleeve and Cage 470444 Rubber Meter Holder and Neck Strap 470443 Hard Carrying Case 470447 Wrist Strap 470448 Instrument Sealing Kit 470449 Neck Strap 470445 RS232 Interface Kit 470438 Batteries, AA, 4/pk 470439 Guide Book to Conductivity/Dissolved Oxygen 473736 Fill Solution and Membrane Caps, 3/pk 470466 D.O. Electrode Fill Solution, 30 mL 477656 D.O. Sensor Cleaning Compound 473625 Zero Oxygen Standard, 500 mL 473739 Zero Oxygen Standard, Sachets, 30/case