Contents
1 Set Up
1.1 Unpacking and Installation 3
1.2 Product Description 3
1.3 Functional Description 4
2 Operation
2.1 Calibration 5
2.2 Measurements 6
3 Support Information
3.1 Probe Maintenance and Cleaning 7
3.2 Low Battery Indicator 7
3.3 Specifications 7
Consumables/Accessories 8
Warranty 8
2
Instruction Manual
Corning Model 311 Portable Conductivity Meter
1.1 Unpacking and Installation
1. Unpack the instrument and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your distributor immediately.
2. In addition to this manual, you should find the following items:
Meter w/Wrist Strap 1 Conductivity Probe (476612) 1 Carry Case (470442) 1
1413 µS Cal Std Sachets (473738) or 100ml Sample Bottle 1
12.88 mS Cal Std Sachets (473737) or 100ml Sample Bottle
Warranty Card 1 Guide to Conductivity and
Dissolved Oxygen (470439) Calibration Trimmer Screwdriver 1
9V Battery (471027) 1
3. Save all packing materials until you are sure
that the instrument functions correctly. Any defective or damaged items must be returned in the original carton together with the supplied accessories.
1.2 Product Description
The Corning Model 311 is a multi-range, water resistant, Conductivity Meter with automatic temperature compensation. The sensors in the meters probe utilize the latest technology in conductivity measurement. The four-ring potentiometric method incorporated into the probe has been proven to provide higher accuracy than the more common amperometric method. The Model 311 has a range of 0 to 199,900 µS/cm and can be used in any sample from treated water to brine. The temperature of all conductivity measurements are automatically compensated for by the meters ATC circuitry. The probe is easily detached and can be changed without having to be returned to the factory.
1
1
The meter is constructed with a water-resistant, rugged ABS plastic housing which allows the versatility of both laboratory and field use.
Other features include a Low Battery Indicator warning which prevents erroneous readings due to low battery voltage.
4. Fill in and return the warranty card. For your
own record make a note of the serial number, date of purchase and supplier on page 8 of this manual.