![](/html/75/75fa/75fa6f54815033fb711f0a80d41bbf8b06305a208c46f21e15767f3e90eb0bea/bg1.png)
Instruction Manual
Corning Model 311 Portable Conductivity Meter
This manual contains complete instructions for
setting up and using the Corning Model 311
Portable Conductivity Meter.
The information contained in this manual was
correct at the time of going to print. However, we
continue to improve products and reserve the right
to change specifications, equipment and
maintenance procedures at any time.
This manual is copyrighted, and all rights
are reserved. No part of this manual may be
reproduced by any means or in any form without
prior consent in writing.
CB.014.98 ©1998 Corning Incorporated
1
![](/html/75/75fa/75fa6f54815033fb711f0a80d41bbf8b06305a208c46f21e15767f3e90eb0bea/bg2.png)
Instruction Manual
Corning Model 311 Portable Conductivity Meter
Contents
1 Set Up
1.1 Unpacking and Installation 3
1.2 Product Description 3
1.3 Functional Description 4
2 Operation
2.1 Calibration 5
2.2 Measurements 6
3 Support Information
3.1 Probe Maintenance and Cleaning 7
3.2 Low Battery Indicator 7
3.3 Specifications 7
Consumables/Accessories 8
Warranty 8
2
![](/html/75/75fa/75fa6f54815033fb711f0a80d41bbf8b06305a208c46f21e15767f3e90eb0bea/bg3.png)
Instruction Manual
Corning Model 311 Portable Conductivity Meter
1.1 Unpacking and Installation
1. Unpack the instrument and examine it
carefully to make sure that no damage has
occurred during shipping. If there is any
noticeable damage, notify your distributor
immediately.
2. In addition to this manual, you should find the
following items:
Meter w/Wrist Strap 1
Conductivity Probe (476612) 1
Carry Case (470442) 1
1413 µS Cal Std Sachets (473738) or
100ml Sample Bottle 1
12.88 mS Cal Std Sachets (473737) or
100ml Sample Bottle
Warranty Card 1
Guide to Conductivity and
Dissolved Oxygen (470439)
Calibration Trimmer Screwdriver 1
9V Battery (471027) 1
3. Save all packing materials until you are sure
that the instrument functions correctly. Any
defective or damaged items must be returned
in the original carton together with the
supplied accessories.
1.2 Product Description
The Corning Model 311 is a multi-range, water
resistant, Conductivity Meter with automatic
temperature compensation. The sensors in the
meters probe utilize the latest technology in
conductivity measurement. The four-ring
potentiometric method incorporated into the probe
has been proven to provide higher accuracy than
the more common amperometric method. The
Model 311 has a range of 0 to 199,900 µS/cm and
can be used in any sample from treated water to
brine. The temperature of all conductivity
measurements are automatically compensated for
by the meters ATC circuitry. The probe is easily
detached and can be changed without having to
be returned to the factory.
1
1
The meter is constructed with a water-resistant,
rugged ABS plastic housing which allows the
versatility of both laboratory and field use.
Other features include a Low Battery Indicator
warning which prevents erroneous readings due
to low battery voltage.
4. Fill in and return the warranty card. For your
own record make a note of the serial number,
date of purchase and supplier on page 8 of
this manual.
3