Cornelius MJ30-4 PC, MJ32-4 PC, MJ30-4 PB, MJ31-4 PC, MJ31-4 PB User Manual

...
MILLENNIUM (MJ30-4 PB)
(MJ31-4 PB) (MJ32-4 PB) (MJ30-4 PC) (MJ31-4 PC) (MJ32-4 PC)
INSTALLATION & SERVICE
IMI CORNELIUS INC.
One Cornelius Place, Anoka, Minnesota 55303
Phone (800) 238-3600 or (612) 421-6120 Fax (612) 422-3255
Part No. 85318
Rev. B 2/99
Table of Contents
Description ...............................................................................................................................Page 2
Specifications ...........................................................................................................................Page 3
Receiving & Unpacking...........................................................................................................Page 4
Installation................................................................................................................................Page 4-6
Connect Drain
Connect Water Supply
Fill Ice Water Bath
Connect Power Supply/Start Refrigeration
Purge Air from System
Connect B-I-B Concentrate
Priming the Concentrate Pumps
Sanitizing prior to Initial Use
Apply Sealant to Base & Counter Top
Programming Instructions........................................................................................................Page 6
Brixing Concentrate .................................................................................................................Page 6
Cleaning & Sanitizing the System............................................................................................Page 7-8
Flushing the Concentrate Lines & Valves
Cleaning & Sanitizing the Concentrate Lines & Valves
Daily Cleaning & Maintenance................................................................................................Page 8
Maintenance of the Refrigeration System................................................................................Page 8-9
Cleaning of the Refrigeration Components
Ice Water Bath Cleaning
Troubleshooting Guide.............................................................................................................Page 10-17
Wiring Diagram........................................................................................................................Page 18
Exploded View/Parts Identification .........................................................................................Page 19-29
Recommended Spare Parts (Service Agents Only)..................................................................Page 32
Optional Accessories................................................................................................................Page 33
Warranty...................................................................................................................................Page 34
1
General Description
The Millennium™ is a state of the art four product B-I-B juice dispenser. It’s integral high performance, dual speed peristaltic mini pumps eliminate the need to purchase separate B-I-B pumps. The Millennium™ can draw product directly from the B-I-B’s up to 50 feet away (15 feet vertically) allowing for greater installation flexibility! Additionally, the Millennium™ incorporates a unique valve assembly which mixes the concentrates more thoroughly than similar models.
Wilshire realizes that counter space is at a premium in any operation. The Millennium™ maximizes the use of counter space by providing four product capability in a 15 inch wide chassis. Because the air circulation is through the top, the Millennium™ can be installed with zero clearance in the back. For improved serviceability, all components can be accessed through the top or front without disturbing equipment placed on either side.
With it’s innovative compact design, reduced operating expense, ease of use and service we’re certain the Millennium will meet the most demanding conditions for many years to come.
2
9-1/2 in.
30 in. (76.2 cm)
(24.13 cm)
Specifications
15 in. (38.1 cm)
28.25 in. (71.75 cm)
Model Number .........................................................................................................................MJ30-4
Electrical Ratings .......................................................................... 115VAC / 1 Phase / 60 Hz, 10 Amps
Electrical Connection .................................................................... 3 Prong Plug w/ Ground (supplied)
....................................................................................................... 2 Prong Plug w/ Ground (supplied with
....................................................................................................... export modles)
Power Supply ................................................................................ 15 Amp @ 120 volts
....................................................................................................... 10 Amp @ 220 volts
Water Requirements...................................................................... 3/8 in. (9.5 mm.) SAE male flare inlet
....................................................................................................... 20-100psi (1.4-7bar) max. static
pressure
....................................................................................................... 3.0 oz. (88.7 ml.) water flow per second
Refrigeration.................................................................................. 8.5 oz. R-134a Refrigerant
....................................................................................................... 1/3 hp. Compressor
....................................................................................................... Test Pressures:
....................................................................................................... High 460 psi (31.2 bar)
....................................................................................................... Low 200 psi (13.5 bar)
....................................................................................................... Air Cooled Condenser
Recommended Clearance..............................................................18 in. (45.72 cm.) on top, 4 in. on sides
Weight........................................................................................... Shipping, 150 lbs (68.1 kg.)
....................................................................................................... Operation, 241 lbs (109.4 kg.)
Approvals ...................................................................................... U.L., C.U.L., N.S.F.
* Low speed rating is 2/3 of the high speed
3
Receiving & Unpacking
1. Inspect the carton and note any damage,
regardless if it appears minor. If the carton is damaged, note on the consignee copy of the freight invoice “exterior carton damage ­concealed damage possible.”
4. Remove the four 9/16” (15 mm) retaining
bolts from the underside of the shipping pallet by carefully tilting the dispenser from side to side.
5. With the assistance of another person, lift
the dispenser from under the base and place it on the counter.
2. Cut the plastic banding strap and remove the
exterior carton sleeve, internal fillers and plastic bag around the dispenser. Carefully inspect for damage.
Note: IMI Cornelius is not responsible for damaged freight. If damage is found, you must save all packaging material and contact the freight carrier. Failure to contact the carrier within 48 hours of receipt may void your claim.
3. Confirm receipt of the installation kit
#720202300 packaged with the dispenser. If any item is missing, contact our customer service department at 1-800-238-3600 Monday - Friday between the hours of 8:00 AM and 4:30 PM CST.
Kit Contents:
Qty Description 4 3/8” barb
4 3/8” swivel nut 5 3/8”x3/8” elbow barb 6 3/8” nylon washer 10 #170 clamps 10 ft. 3/8” braided tubing 4 ft. 1/2” I.D. clear tubing 1 1/2”x1/2” 90o elbow 4 Generic B-I-B connectors 2 Flavor strip kits 1 Installation template
The dispenser is extremely heavy when operational. Make certain the counter can support a minimum of 300 lbs directly below the dispenser.
Installation
Typically the dispenser is placed directly on the counter and a food grade silicone sealant applied around the base. However, an optional leg kit is available. The following instructions assume the optional leg kit will not be used.
Depending on the type of counter, it may be
necessary to provide access through the counter at the rear of the dispenser for the drain, power, water and concentrate connections. Refer to the
template enclosed in the installation kit.
Connect Water Supply:
Water pipe connections and fixtures directly connected to a potable water supply shall be sized, installed, and maintained according to federal, state, and local laws.
Wilshire recommends that a ½” O.D. copper supply line with a shut-off valve and water filter be located within 3-6 feet (0.91-1.83 m) of the dispenser.
a. Assemble the 3/8” braided water supply line
furnished in the Installation Kit by attaching the straight barb on one end and the elbow barb on the other. Secure with clamps provided.
4
b. Insert a 3/8” nylon washer into the elbow
barb of the supply line. Connect the assembled supply line to the water inlet fitting located on the rear of the dispenser.
c. Connect the water supply line to the shut
off. Turn on the water and check for leaks.
NOTE: The dynamic water pressure must be 20 PSIG (1.4 bar) minimum to ensure correct valve flow control and must not exceed 100 PSIG (7 bar) to avoid valve damage.
Connect B-I-B Concentrate
a. Attach the supplied 3/8” (9.5mm) elbow
barbs to the end of each 3/8” I.D. B-I-B concentrate supply line (not supplied). Secure with clamps provided. Insert a 3/8” nylon washer into the elbow barb and connect the supply lines to the rear of the dispenser noting the valve numbering (numbered 1-4 from left to right as viewed from the front of the dispenser).
b. Route the concentrate supply lines to the
B-I-B location and attach the gray plastic B-I-B connector (supplied) to the end of each line. Secure with clamps provided.
Fill Ice Water Bath:
a. Remove the top cover b. Remove the large red plastic plug from the
filler hole in the top of the refrigeration deck.
c. Fill the water bath with cool water until it
begins to trickle out of the overflow tube.
d. Reinstall the red plastic plug and top cover.
Connect Power Supply / Start Refrigeration:
a. Start the refrigeration system by plugging
the power cord into a 115 VAC (+ 10%), 15 amp grounded receptacle.
The Millennium forms an ice bank of approximately 30 lbs (13.61 kg) in 4-5 hours at a room temperature of 75oF (24oC). Once the ice bank has grown to the proper size, the ice bank control will shut down the refrigeration circuit.
NOTE: It is normal to see water trickle from the over flow as the ice bank forms.
Connect Drain:
a. Attach the clear drain hose to the barb fitting
on the drip tray. Route the drain hose out from the rear of the dispenser and connect it to a suitable drain source, ensuring compliance with all federal, state, and local plumbing codes. Slide the drip tray into place.
Purge Air from the Water Coil
Prior to initial use, purge all air from the valves by pushing the dispensing switch repeatedly. Continue until a steady flow of water is observed. Repeat the above on the remaining valves.
Priming the Concentrate Pumps
a. Turn off the water supply. b. Connect the concentrate lines to their
respective B-I-B.
c. Depress and hold each start button until
concentrate is observed flowing from the dispense nozzle.
d. Turn on the water supply.
5
Sanitizing Prior to Initial Use
L.E.D.
CANCEL/POUR
The beverage system must be cleaned and sanitized after installation is completed to safe guard against any possible contaminants which may have entered the system during transport or installation. Refer to the “Cleaning & Sanitizing the System” section of this manual for procedures.
Apply Sealant to Base & Countertop
NOTE: The portion control has a full memory retention in case of a power failure.
SMALL
S
MEDIUM
M
LARGE
L
▲▼
EXTRA-LARGE
XL
If the optional legs are not used to raise the dispenser off the counter, the entire perimeter of the base must be sealed at the counter top with silicone (or other food grade approved sealant) in order to comply with N.S.F. standards. Refer to the template supplied in the Installation Kit.
Programming Portion
Control (optional)
a. Simultaneously, press and hold “S” (small)
and “XL” (extra large) push button switches on the Coded Autoset Portion Control Module until the LED light in the lower right hand corner of the module starts blinking. Release the switches. The blinking LED indicates the programming mode is active.
b. Place the cup under the white mixing valve
nozzle and push the selected size button (small, medium, large, or extra large). Hold the button in until the cup fills to the desired portion, then release the button. Repeat the above procedure for the remaining sizes.
CANCEL/POUR ▲▼:
a. Push and release cancel/pour button to stop
the valve from dispensing.
b. Push and hold for a continuous pour.
Brixing Concentrate
The following procedures describe how to adjust the water to concentrate ratio (brix) according to taste. Contact your concentrate supplier for recommended brix ratios.
a. Remove the flavor strip cover above the
dispense valves.
b. Sample the finished drink. Increase or
decrease the water ratio by turning the water flow control screw clockwise (more water) or counterclockwise (less water) until the desired water to concentrate ratio is achieved. Repeat procedure on remaining valves.
c. Replace the flavor strip cover.
c. After programming all the drink sizes, press
and release the “cancel/pour” ▲▼ switch to return the Coded Autoset Portion Control to the operational mode. The blinking LED light will go out.
d. If at a future date, it is decided to change the
portion size of the drinks, the individual sizes can be adjusted by the above procedure. It is not necessary to reprogram every size.
NOTE: If the flow control does not respond there may be debris caught between the internal ceramic spool and sleeve. Try dislodging the debris by pressing the dispense switch several times or turning the flow control adjustment screw all the way in and out several times.
6
Cleaning & Sanitizing the
System
c. Connect the bag valve to the gray bag
connector and submerse the parts in the bucket of hot water.
The dispenser must be cleaned and sanitized after installation and as required by state and local health departments, or every 3 months minimum. Your state or local health codes may require more frequent and extensive sanitizing procedures.
Cleaning and Sanitizing Equipment and Supplies:
Recommended sanitizer/cleaner:
Stera-Sheen® Green Label prepared to ensure 200 ppm of available chlorine (one-2 oz. packet/1 gallon (3.8 L) of water = 200 ppm). Solution temperature should be between 80°F - 100°F (26.7°C - 37.8°C).
One clean 5 gallon (19 L) bucket
One clean 1 gallon (3.9 L) container or
bucket
Clean non-abrasive clothes
A small brush
Three empty bag in the box bags (to cut
valve fittings off)
Flushing the Concentrate Lines & Valves
Cleaning and sanitizing is not required for potable water circuits. Potable water lines should remain connected and operational during the cleaning and sanitizing procedures for juice circuits.
a. Fill the 5 gallon (19 L) bucket with clean
extremely hot tap water, approximately 140oF (60oC).
b. Take the three empty bag in the boxes and
remove the bags from the carton. With scissors, cut the bag valve off of the bag and clean the valve by rinsing it under hot tap water.
d. Depress and hold the dispense button until
the concentrate has been fully purged from the product lines and valves.
IMPORTANT
Do Not flush more than three valve circuits
simultaneously.
e. Once the concentrate is purged, pulse each
valve for 15 seconds on and then release the button. Repeat this pulsing for 15 cycles for each circuit being cleaned. Once the 15 cycles have been completed, allow each valve to dispense for 3 continuous minutes.
f. Remove the nozzles and static mixers from
the dispenser and rinse them under hot tap water to remove any remnant of excess concentrate. Repeat for each circuit to be cleaned.
g. Replace the nozzles and static mixers into
their proper location.
h. Discard any remaining hot water left in the
bucket.
Cleaning & Sanitizing the Concentrate Lines and Valves
a. Prepare 5 gallons (19 L) of Stera-Sheen®
Green Label cleaning and sanitizing solution
by mixing one 2 oz. packet/1 gallon (3.8 L) of potable water. This will provide enough sanitizing solution to clean and sanitize all 4 concentrate circuits for most installations. Installations that have 50 or more feet of concentrate line may require more sanitizing solution.
IMPORTANT
Use potable water at 80°F-100°F (26.7°C-
26.7°C) to create solution. Water temperatures above this range will breakdown the chlorine count and minimize sanitation.
7
b. Submerse the bag connector and bag valve
assembly into the bucket of sanitizing solution.
Daily Cleaning and
Maintenance
c. Depress and hold the dispense button until
sanitizing solution is dispensed through the mixing valves nozzle.
d. Pulse each valve for 15 seconds on and then
release the button. Repeat this pulsing for 15 cycles for each circuit being sanitized. Once the 15 cycles have been completed, allow the sanitizing solution to stand in the product lines and valves for 30 minutes.
e. While waiting, remove the nozzles and static
mixers and place them into a separate container with 2 qts. (1.9 L) of sanitizing solution. Agitate vigorously using the small brush to remove any excess concentrate. Allow the parts to soak for 30 minutes.
f. Clean the dispensing valves mixing chamber
with the brush and sanitizing solution (this is the cavity from which the nozzle is removed).
a. On a daily basis, clean the external cabinet
splash areas using a clean damp cloth. Remove and wash the cup rest, dispensing nozzles, and static mixers with clean water. Wipe dry with a clean soft cloth.
b. Wipe the drip tray in place on the unit, wash
the tray out with a mild soap solution, then rinse the tray by pouring water down the drip tray’s drain.
c. Clean all external surfaces of the dispenser
with a sponge and a mild soap solution. Rinse the sponge out with clean water, then wring the excess water from the sponge and wipe all external surfaces of the dispenser.
d. Wipe the dispenser dry with a clean soft
cloth. DO NOT USE ABRASIVE TYPE CLEANERS. Install the cup rest,
dispensing nozzles, and static mixers on the dispenser.
g. Replace the nozzles and static mixers into
their proper locations and discard the sanitizing solution used to soak them in.
Note: Do not reuse the sanitizing solution used to clean the nozzles, static mixers, etc.
h. Activate the dispensing valves for two more
cycles (15 seconds on then off) with sanitizing solution then run solution continuously through the dispensing valves for 2 minutes.
i. Disconnect the bag valves from the gray bag
connectors and re-connect the bag connectors to their appropriate product bags.
j. Depress and hold each dispense button until
juice appears at the outlet of the dispensing nozzle. Dispense and discard two 8 oz. (237 ml) cups of juice and verify that there is no chlorine off taste
Maintenance of the
Refrigeration System
Cleaning of the refrigeration components should be performed by a qualified service technician.
a. Continuous maintenance of this dispenser is
a basic requirement for proper operation and sanitation, including all support equipment utilized in the daily operation of this equipment.
b. Cleaning of the Refrigeration Components: c. Disconnect the power before removing the
top bonnet of the dispenser.
d. The dispenser’s ventilation grilles and the
condenser fins should be cleaned periodically to maintain efficient refrigeration and to avoid compressor failure. The condenser fins can be cleaned
8
with a vacuum cleaner or a soft bristle brush.
e. Clean the exterior surfaces of the
compressor, agitator motor, fan motor, and fan blade with a damp cloth to wipe off the accumulated dust.
Ice Water Bath Cleaning
The water bath should be cleaned to obtain maximum cooling efficiency. It is recommended that the water bath be cleaned two to four times annually, depending upon local conditions and/or as required by state and local health departments.
A convenient time to do this is at the time the dispenser is being sanitized. To save time, the water bath can be drained while the dispenser is being sanitized.
a. Remove the cup rest and splash panel to
access the water bath’s drain hose. Extend the drain hose to a suitable waste receptacle and allow the water bath to drain.
b. Remove the dispenser’s top bonnet by
removing the two screws located on the top of the bonnet. While lifting the bonnet up and away from the dispenser, unplug the merchandiser electrical cord from the power strip located on the refrigeration deck.
c. In order to remove the refrigeration deck for
the water bath and evaporator cleaning, remove the two screws that secure the deck to the cabinet structure. Unplug the refrigeration deck’s main power strip cord from the dispenser’s power cord. Disconnect the red, white and blue 12 pin plugs from the electrical box. Disconnect the key switch a by unplugging the white 2 pin plug at the transformer.
Note: It will be necessary to melt the ice bank to be able to pull the refrigeration deck up and away from the dispenser. Warm water may be used to accelerate the melting. In order to prevent ice bank control damage, do not direct the warm water stream on to the sensing bulb
of the ice bank control. Never use an ice pick or other instrument to remove ice from the evaporator. Such practice can result in a punctured refrigerant circuit or damage to the water bath tank.
d. Once the ice bank is sufficiently melted, lift
the entire refrigeration deck up and away from the water bath by lifting the deck using it’s service handle strap.
e. Prepare 1 gallon (3.9 L) of cleaning and
sanitizing solution (see “Sanitizing the Concentrate Lines & Valves”). Pour the cleaning and sanitizing solution into the water bath and clean the sides and bottom of the tank, the product coils, and associated brackets with a fiber brush.
f. Utilizing the cleaning and sanitizing solution
in the water bath, the fiber brush, and a clean cloth soaked with sanitizing solution, clean the refrigeration deck’s evaporator coils, agitator motor shaft and blade, and the ice bank sensing bulb.
g. Drain the cleaning and sanitizing solution
from the water bath and rinse/flush all the components with clean water.
h. Re-install the refrigeration deck into the
water bath and secure with two screws.
i. Connect the refrigeration deck’s power strip
to the dispenser’s power cord. Connect the red, white and blue 12 pin plugs to the electrical box. Connect the key switch to the transformer.
j. Fill the water bath with cool potable water
until water begins to flow from the overflow tube.
k. Plug the merchandisers power cord into
power strip located on the refrigeration deck and re-install the top bonnet. Secure the bonnet assembly with two screws.
l. Install the splash panel and cup rest and plug
the dispenser’s power cord into the electrical outlet.
9
Troubleshooting Guide
Problem Probable Cause Remedy
DISPENSER IS TOTALLY INOPERATIVE
DISPENSER WILL NOT COOL
Drink temperature should be 35-45°F (2-7°C)
No power present at the source.
Power is present at the source...dispenser is still inoperative.
Refrigeration shut off switch is in the off position.
Low water bath level.
Depleted ice bank ( not applicable after period of heavy usage.)
Check the power circuit for a blown fuse or tripped circuit breaker. Check the supply line for the proper amperage and voltage
ratings.....dedicated 15 amps /115
volts.
Check the power cord connection located behind the front splash panel.
Refrigeration shut off switch is located on the top of the dispenser . This switch will disable the compressor and the condenser fan motor. Be sure the switch is in the on position.
Check water bath level. If 2 gallons or more are added, cause of problem has been confirmed.
Check compressor. If running, allow the system time to recover (approximately 1 hour). If compressor is not running or the system is not recovering see the Refrigeration System Troubleshooting section.
WITH POWER PRESENT, UNIT WILL NOT DISPENSE ALL VALVES.
Defective Ice bank control.
Key lock switch in the OFF position.
Transformer not connected or defective.
Fuse blown between the transformer and voltage regulator boards.
10
See Refrigeration System Troubleshooting section.
Check position of the key lock switch. Check for loose wire connections. Run continuity check, replace as needed.
Check the connection between the power strip and the transformer. Check transformer output at terminal junction on control box for 26 VAC.
Check the fuse located in the electrical box, left side. This box is located on the refrigeration deck and will require the removal of the upper bonnet. Replace fuse as needed.
Troubleshooting Guide
Problem Probable Cause Remedy
UNIT DISPENSING CONCENTRATE ONLY NO WATER TO ALL VALVES
FROM SINGLE VALVE UNIT WILL NOT DISPENSE WATER OR CONCENTRATE.
No water reaching dispenser from the water supply line.
Water pressure over 100 psi or 7 bar. High pressure will not allow the water solenoids to open.
Freeze up of water coil.
Inoperative or defective push button switch, or portion control board.
Defective Voltage Regulator (VRB) board.
Check line to see if the water is shut off at the source.
Install regulator onto incoming water line to reduce the pressure to between 30 and 80 psi. (2-5- bar).
Refer to the Refrigeration System Troubleshooting section.
Check for loose or broken wire connections at the switch. Check the continuity of switch. Replace as needed.
The VRB boards are contained in the electrical box mounted on the refrigeration deck.. There is one for each valve.
CAUTION: Access to this panel requires the removal of the upper bonnet while the dispenser is energized and operational. Check the power input for 26 VAC and the output for 28 VDC.
FROM SINGLE VALVE, UNIT IS DISPENSING CONCENTRATE ONLY.
Inoperative water valve solenoid coil.
Solenoid makes clicking sound when energized, but still does, not function.
Flow control binding or defective.
11
Check to see all wire connections are secured at the coil. Check power for 26 VDC. If power is present, coil maybe defective , replace as needed. Armature maybe stuck, disassemble and check function. Return spring could also be defective replace as needed.
Disassemble the coil assembly and check plunger seal for defects. Replace as required. Inspect valve seat for restrictions.
Disassemble flow control assembly and check the spool and sleeve assembly . Be sure the spool moves freely within the sleeve and there are no restrictions.
Troubleshooting Guide
Problem Probable Cause Remedy
FROM SINGLE VALVE, UNIT IS DISPENSING WATER ONLY.
Concentrate container empty.
BIB connector is not engaged or improperly installed.
Air leak in product suction line.
No power at the pump motor.
Defective gear motor assembly.
Replace.
Check BIB connector for proper installation. Plunger should be completely inserted with concentrate filling the line to the dispenser.
Check the 3/8 flare connection where the product line attaches to the unit. Tighten fitting using a flare gasket. If line is drawing air, pump will not prime.
Check for broken or loose wire connection going to pump motor. Check for 26 VDC at pump motor when the dispense button is energized.
Check motor function by energizing circuit with the dispense button. If motor runs, but pump does not turn, a broken gear box is indicated. The pump motor and gear box are replaced as a unit.
DISPENSER IS BRIXING OUT TOO HIGH.
Pump running backwards.
Defective concentrate pump tubing.
Supply water pressure too low.
Defective flow control.
Check the direction of pump rotation...correct rotation is clockwise as viewed from the front of the unit. If pump is turning in a counter clockwise direction, this indicates the pump wires are reversed.
Check the pump tubing for ruptures, wear, etc. Replace as needed following the instructions printed in this manual.
Water pressure must be 20 psi flowing or above.
Disassemble and check flow control assembly for binding, defective compensating spring , trapped debris. Be sure the spool moves freely within the sleeve, replace parts as needed.
12
Troubleshooting Guide
Problem Probable Cause Remedy
DISPENSER IS BRIXING OUT TOO HIGH. (Cont.)
DISPENSER IS BRIXING OUT TOO LOW.
Low viscosity or highly concentrated concentrate. Example 11 to 1 ratio.
Supply water pressure is too high.
Defective or improperly regulating flow control .
Locate the pump switches grouped behind the front splash panel. They are marked SPEED SWITCH with the corresponding valve number. Placing the switch in the low position slows down the concentrate pumping rate requiring less water to brix.
Water pressure has a high limit of 100 psi/ 7 bar or over. Place regulator in supply line to bring pressure down to 4 bar/ 50 psi flowing.
High water pressure will cause the flow control not to regulate properly. If the problem exists after the pressure has been reduced to 50 psig/ 4 bar, the flow control maybe defective.. Disassemble and check the spool and sleeve function and the compensating spring . Replace parts as required.
Speed switch in the low position.
Air leak in the concentrate suction line.
Worn concentrate pump tubing.
Defective poppet valve and or spring.
Pump speed switch is in the low position. Move switch to high increasing the pumping rate of the concentrate.
Check that the concentrate line connection is secured tightly to the dispenser with the proper gaskets. Check the connector at the concentrate source to be sure it is properly attached.
Remove the front half of the concentrate pump and inspect the tubing . Worn or stretched tubing will decrease the amount of concentrate pumped. After 1 year tubing should be replaced.
Disassemble the right side of the valve which contains the syrup poppet assembly. Be sure the valve is not binding, free of debris, return spring is actuating and the diaphragm is not torn or punctured.
13
Troubleshooting Guide
Problem Probable Cause Remedy
WATER LEAKS FROM BOTTOM OF DISPENSING VALVE.
WATER CONTINUOUSLY DRIPS FROM NOZZLE.
UNIT CONTINUES TO DISPENSE AFTER BUTTON HAS BEEN RELEASED.
Nozzle pieces improperly assembled.
Water solenoid valve not shutting off tightly.
Inoperative or defective push button switch, or portion control board.
Relay on voltage regulator board hangs up or is stuck open.
Static mixer inserted in nozzle body upside down. Cone shaped end must be visible when remounting the nozzle assembly to the dispenser.
Disassemble water solenoid located at the valve and check for foreign material in the valve seat. Check the armature for binding, check the plunger and the seat for defects. Replace parts as needed.
Check the operation of the push button for sticking. Check continuity of switch. Replace as needed.
CAUTION: Access to this panel requires removal of the bonnet while the dispenser is plugged in and operational. *Voltage regulator board is found in the electrical box mounted to the refrigeration deck. There is one board for each valve. Gently tap the relay. If dispensing stops, relay was hung up . If tapping the relay solves the problem, it should be noted that if a relay sticks once , it will probably stick again. Replace as needed.
TWO OR MORE PUMPS OPERATE WHEN ONLY ONE BUTTON IS PUSHED.
Defective Voltage Regulator (VRB) Board.
Insulator pad missing or not properly positioned between the transistor on VRB board and mounting bracket.
14
Pull the connector marked DOOR , lower right hand connector on the VRB board. If the unit still continues to dispense the VRB board is defective and needs to be replaced.
On the valve that is dispensing in error, check the insulator pad between the transistor on the top of the VRB board and the mounting bracket. If this pad is missing or improperly aligned, the VRB board will make contact with the bracket. Current will run through the bracket and set off the valve. Refer to drawing found in this manual.
Troubleshooting Guide
Problem Probable Cause Remedy
TWO OR MORE PUMPS OPERATE WHEN ONLY ONE BUTTON IS PUSHED.
(Cont.)
WITH POWER PRESENT DISPENSER WILL NOT COOL
Drink temperature should be 35-45°F 2-7°C
VRB Board mounting screw insulator is missing or improperly installed.
Refrigeration shut off switch is in the off position.
Depleted ice bank (not applicable after period of heavy usage.)
Check the corresponding VRB board to the valve that is dispensing in error. Where the screw mounts the transistor to the bracket, be sure that the spacer bushing is present and the screw is not over tightened. Screw contact with the bracket will cause a short and the firing of the valve when other valves are energized.
Refrigeration shut off switch is located on the top of the dispenser. This switch will disable the compressor and condenser fan motor. Be sure the switch is in the on position.
Remove rear bonnet covering the refrigeration deck. CAUTION should be observed when working in this area as the dispenser is energized and operational. Check to see if the compressor is running . If so allow the system to recover (approximately 1 hour).
WITH POWER PRESENT COMPRESSOR WILL NOT RUN. (Compressor shell is cold to touch.)
COMPRESSOR RUNS BUT NO COOLING IS TAKING PLACE.
Loose or broken wire connections.
Defective compressor start component.
Ice bank control not connected or defective.
Low refrigerant charge.
Refrigerant restriction.
Check the wire connections at the compressor. Be sure they are tightly secured.
Check the compressor start relay. Replace as required.
Check the ice bank control connections. To test, place jumper between the 2 leads. If the compressor starts ,the ice bank control is defective.
Contact a refrigeration service technician. They will check for leaks (sign of oil at joints and fitting) and will recharge if necessary.
Refrigeration technician will examine the system for, kinks in lines, restrictions, etc. Heat exchanger may need to be replaced.
15
Troubleshooting Guide
Problem Probable Cause Remedy
COMPRESSOR RUNS BUT NO COOLING. TRIPPING OVERLOAD. (Compressor shell is hot to touch).
Condenser fan not running.
Inadequate air flow through the dispenser.
Blocked condenser fins.
Ambient air over 100°F 38°C.
Check that the condenser fan is running. If the fan is binding or turning slowly, an overload condition will occur.
Air flow is top front in, exhaust out the rear. NEVER PLACE ANYTHING ON TOP OF THE DISPENSER. This will block the airflow over heating the compressor and causing it to trip the overload.
Remove the rear bonnet and check the condenser fins. If unit is run in a dusty environment, condenser fins can become clogged, stopping the air flow. Brush out and clean with vacuum.
If the dispenser is working in an environment over 100° ambient or is drawing in discharged air, during a heavy period, the unit may trip the overload.
UNIT IS DISPENSING CONCENTRATE ONLY REFRIGERATION SYSTEM HAS BEEN RUNNING FOR EXTENDED PERIOD WITHOUT SHUTTING OFF FROZEN WATER COIL.
Defective overload.
Defective compressor.
Insufficient water level in the water bath.
Agitator motor not operating.
Improperly located ice bank control sensor.
Overload works on shell temperature and amperage draw and can be defective. Replace as needed.
Refrigeration technician will verify if the compressor is defective and needs to be replaced.
See the Start-up and Operation Section of this manual. Be sure the water bath is filled to fill line.
Check the agitator motor located on the refrigeration deck. Be sure it is plugged in and receiving power. Check that the impeller is not obstructed. Replace if needed.
Melt the ice bank and check the location of the sensing bulb. Be sure it is all the way down to the bottom of the tube.
16
Troubleshooting Guide
Problem Probable Cause Remedy
UNIT IS DISPENSING CONCENTRATE ONLY REFRIGERATION SYSTEM HAS BEEN RUNNING FOR EXTENDED PERIOD WITHOUT SHUTTING OFF FROZEN WATER COIL.
Compressor not cycling off. Check the ice bank control. Be sure
that the control has not overheated and fused the contacts together. Replace as required.
17
Loading...
+ 41 hidden pages