This instruction manual represents an integral part of the equipment and must be kept readily
available for use. Read the warnings contained herein carefully before installing and using this
equipment. In addition to offering information concerning routine maintenance for the ice slush
drinks machine and technical back-up for troubleshooting, this manual aims to help the user
make the most of the machine’s potential, adapting it to suit the specific needs of the various
countries it will be used in. Modifications or attempts to modify the equipment will not only
result in the forfeiture of the guarantee, but are also extremely dangerous.
The maintenance operations must be carried out by qualified professionals. Never attempt to
repair the machine yourselves as the intervention of non-qualified persons, as well as being
hazardous, could also lead to serious damage to the machine.
ATTENTION: Before starting up the machine it’s necessary to perform all the cleaning
and sanitation procedures.
TECHNICAL DATA
PLATE DATA
The voltage and the frequency are indicated on the serial number plate, located behind the
drip tray and the right hand side near the controls.
Machine Bowl n° Volt Hertz Ampere
230 50 7.3 VFCB POST MIX 2
115 60 13.8
230 50 7.3 VFCB PRE MIX 2
115 60 13.8
VFCB POST MIX 3 115 60 21.5
IMPORTANT: This machine is intended for use at a maximum ambient temperature of
32°C (90°F).
TRANSPORTATION INDICATIONS
To prevent the oil contained in the compressor from flowing out into the cooling circuit, the
equipment must be transported, stored, and handled in a vertical position, as per the
indications given on the packing. The wooden pallet, equipped with housing for the lifting
forks, allows the packed equipment to be moved using normal handling and hoisting means.
1. Release the equipment from the packing, then slide it off
upwards (see Fig. 1). The machine has to be installed in level
with horizontal plane;
2. Checking the machine identification after removing the
packing, you must check that the equipment you have
received is exactly as you ordered, making sure the
specifications indicated on the invoice or the delivery note
are identical to those on the data plate.
Fig. 1
3. Equipment accessories
On opening the packing you will find the following accessories inside the bowls:
• this instruction manual;
• 1 tube of Vaseline to be used for the maintenance machine the requires;
• 2 bags, each one containing a drip tray
• 2 evaporator frontal seals
4. Positioning - make sure the machine’s bodywork is well ventilated (20cm (8”) all around at
least ) and do not install it near heat sources.
We recommend you keep the room temperature at between 59°F (15°C) and 90°F (32°C).
IMPORTANT: All the pieces of packing must be kept out of reach of children as they
represent potential hazards.
Before inserting the plug into the mains socket for your own safety you must take careful note
of the following precautions.
• The machine’s electrical system can only be considered
safe when it is connected correctly to an earthing system, as
provided for by the national safety regulations. The manufacturer
cannot be held responsible for any damages that may be caused
by failure to earth the system.
• For the system to be installed correctly and safely, it is
essential to provide a suitable socket which complies with the
national safety standards in force (see Fig. 2).
• Check the length of the power supply cable to make sure it is not being crushed, do not use
extension cables and, to remove the plug, first turn off the switch, then hold the plug tightly
and pull gently.
• Do not obstruct the ventilation and the heat dissipation grids as bad airing, in addition to
reducing the output and causing bad functioning, could also lead to serious damage to the
equipment.
•Plug:
Machine VFCB post mix 115-60
2 bowl
VFCB pre mix 230-50
2 bowl
Fig. 2
VFCB post mix 115-60
3 bowl
VFCB post mix 230-50
2 bowl
VFCB pre mix 230-50
2 bowl
Plug Nema 5-20 Nema L5-30 cs-sd or cs-sa
IMPORTANT: If the power supply cable is damaged, it must be replaced by qualified
If necessary apply the calibrator/separator relative measuring cup (H1) and push the button
(or the micro switch) on the valve (l 1) until both the soda water and the syrup run into the
measuring cup separately (Fig. 9).
•
Adjust the water flow to the minimum level in order to reduce the inlet pressure. Turn the
water adjustment screw on the flow control counter clockwise until the minimum position to
decrease water flow (clockwise to increase it).
•
Reconnect the syrup.
•
Dispense product into a cup or into the
measuring cup by depressing the micro switch
on the valve. Check brix using a refractometer or
the measuring cup. The brix should be set to
13% ± 1%. To increase the brix adjust the syrup
adjustment screw on the valve clockwise to
increase syrup flow or counter clockwise to
decrease syrup flow.
CAUTION: Do not turn the syrup flow
control counter clockwise to far .
•
When the brix has been set to specification,
reposition the valve to its original setting &
tighten retaining nuts.
•
Repeat the operation for the second bowl.
•
Once the adjustments have been completed for each bowl, place the switch (L1) in the “ON”
position.
•
Both bowls will fill until the level sensor is activated.
•
Turn mixing motors “ON” (switch located on the control panel). The dispenser has built in
delays for compressor & bowl fill switch.
•
NOTE: During initial filling it is normal for some foaming to take place. The bowl level
management will adjust as the product freezes & the foam will dissipate.
t the time you want the machine to turn to freezing mode by following steps above.
NOTE: Once the settings have been saved, the unit will keep the settings
.
Enter Time Programming on Initial Installation or in the Event of a Time Change
hour clock the time is P.M when the dot at the bottom
LEFT BOWL
•
Turn the power switch OFF.
•
Press and hold the left “Select function"
button and turn ON the power switch.
Release the “Select Function" button
when the hour digits start blinking.
•
Set the hour by pressing the "Auto
Timer" clock button until the appropriate
hour is shown.
NOTE: When using a 12
right corner of the LED display is lit; A.M. when dot is not lit.
•
Press the left “Select function" button to set the minutes, then press the "Auto Timer“
clock button until the appropriate minutes are set.
•
Press the “Select function" button one more time to save your settings.
Turn the power switch ON. Make sure the "Auto Timer" is OFF.
•
Press the left "Auger ON/OFF" button ON.
•
Press and hold the “Select function" button until you hear a long beep and the LED,
"Cold" and the "Auto Timer" clock light begins to blink.
•
Press the "Auto Timer" clock button to set the hour you want it to turn to refrigeration
mode, (defrost mode) and then press the “Select function" button to save the hour
setting.
•
Press the "Auto Timer" clock button to set the minutes to complete time setting that
you want it to turn to refrigeration mode (defrost mode). Then press the “Select
function" button to save the minute setting. The "Cold" light will turn off and the
"Freeze" light and "Auto Time" light will begin blinking.
•
Se
Then press the “Select function" button to save the time settings for freeze mode.
NOTE: When the light on the "Auto Timer" clock button is "ON", the defrost timer
is activated. To turn OFF the defrost timer, press the "Auto Timer" clock button
(s) until the light (s) on the clock button (s) turns off.
Operate in Automatic Mode (with Defrost Timer Activated)
Operate in Manual Mode (without Defrost Timer Activated)
product
•
Turn power switch ON and wait for LED
display to light up.
•
Press the left hand "Auger ON/OFF" button ON.
•
To operate in defrost mode press the "Auto Timer"
button until it is illuminated.
•
When setting automatic times, please keep in mind it will take time for the frozen
product to become liquid or vice versa.
NOTE: As regard the RIGTH BOWL repeat the same operations using the buttons
on the right side of the touch pad.
•
Turn the power switch ON and wait for
LED display to light up.
•
Make sure the clock button is OFF (LED
light on clock button should not be lit up).
•
First turn auger on by pressing the "Auger ON/OFF"
button until it beeps.
NOTE: The auger must be on before the unit will allow the cooling or freezing
mode to be activated.
•
Then select refrigeration or freezing mode by pressing the “Select function"
button until the light under the selection you desire is lit up.
NOTE: In the cooling mode, the LED will read the actual temperature of the
(the temperature setting is preset to NSF standards and is not adjustable). In the
"Freeze" or "Off" mode the LED will read the current time.
•
Turn the power switch OFF.
•
Press and hold the left "Auger On/Off" button and turn
the power switch ON. Release the "Auger On/Off" button
when either "12" or "24" are shown (indicates the current
hour view).
Press the "Auto Timer" button until the desired hour display type is shown (12 to 24).
•
Press the "Select Function" button until the desired temperature display type is
shown (°F or °C).
•
Press the "Select Function" button until the current time is displayed to store the
changes. The unit is now ready for use.
Press and hold the left "Auger On/Off" button and turn ON the power switch. Release the
"Auger On/Off" button when either "12" or "24" are shown (indicates the current hour
view).
Press the "Select Function" button until either °F or °C is shown on the display.
Press the "Auto Timer" button until the desired temperature display type is shown.
•
Store the change by pressing the "Select Function" button until the current time
is displayed. The unit is now ready for use.
•
Press the "Select Function" button until the "Cold" LED is lit. The display will now
show the current bowl temperature in either °F or °C depending on which was
selected in the Setting the °F or °C Temperature Display section.
•
Turn On the auger on the side that you want to display the bowl temperature (press the
“FILTER CLEANING” Alarm
A filter cleaning alarm will activate when the unit is running hot due to insufficient internal air
circulation. When this occurs a “Filt” message will appear on the touch pad LED display readout
and an intermittent audible tone will also sound to alert the operator of this condition.
The “Filt” message will appear when the alarm activates (a beeping sound every 4-5 seconds).
To determine the condition that caused the alarm, see list of conditions below:
• Condition: The filter is dirty and needs to be cleaned.
Corrective Action: Clean and replace filter following instructions(Removing and Cleaning Filter).
• Condition: The unit is positioned too close to a wall or other object restricting air flow and
causing the machine to run at a higher temperature.
Corrective Action: Reposition unit to maximize ventilation space (Installation Instructions).
• Condition: The filter is not properly installed.
Corrective Action: Properly install filter (Removing and cleaning filter).
• Condition: The unit has been installed near a heat source, such as a coffee machine, ice
maker or cold beverage machine which expels hot air from its vents, causing the machine to
run at a high temperature (installation near a heat source should be avoided)
Corrective Action: Reposition unit to maximize ventilation space.
“SYSTEM OVER TEMPERATURE” Alarm
A system over temperature alarm will activate as a safety precaution when the unit has
overheated to protect the compressor.
• The system automatically goes to “OFF” status where the compressor’s operations is
stopped, while augers will keep working to avoid forming ice blocks.
• When this occurs an “Err” message will appear on the touch pad LED readout accompanied
by a continuous buzzer sound to alert the operator of this condition.
• When this alarm activates, turn off all switches. Then determine the condition. (See “Filter
Cleaning” Alarm Section for Conditions and Corrective Actions)
NOTE: Once the settings have been saved, the unit will keep the settings
.
Enter Time Programming on Initial Installatio
n or in the Event of a Time Change
NOTE: When using a 12 hour clock the time is P.M when the dot at the bottom
•
Turn the power switch OFF.
•
Press and hold the left “MODE” button and
turn the power switch ON.
Release the “MODE” button when the hour
digits start blinking.
•
Set the hour by pressing the “Auto
Timer” clock button until the appropriate
hour is shown.
right corner of the LED display is lit; A.M. when dot is not lit.
•
Press the left “MODE” button to set the minutes, then press the “Auto Timer” clock
button until the appropriate minutes are set.
•
Press the left “MODE” button one more time to save your settings.
Turn the power switch ON. Make sure the “Auto Timer” is OFF.
•
Press the “Auger ON/OFF” button ON (one at least) .
•
Press and hold the left “MODE” button until you hear a long beep and the LED,
"Cold" and the "Auto Timer" clock light begins to blink.
•
Press the “Auto Timer” clock button to set the hour you want it to turn to refrigeration
mode (defrost mode) and then press the “MODE” button to save the hour setting.
•
Press the “Auto Timer” clock button to set the minutes to complete time setting that
you want it to turn to refrigeration mode (defrost mode). Then press the left “MODE”
button to save the minute setting. The “Cold” light will turn off and the "Freeze" light
and “Auto Time” light will begin blinking.
•
Set the time you want the machine to turn to freezing mode by following steps above.
Then press the left “MODE” button to save the time settings for freeze mode.
NOTE: When the light on the “Auto Timer” clock button is “ON”, the defrost timer
is activated. To turn OFF the defrost timer, press the "Auto Timer" clock button
until the light on the clock button turns off.
Operate in Automatic Mode (with Defrost Timer Activated)
Operate in Manual Mode (without Defrost Timer Activated)
and is not adjustable).
•
Turn power switch ON and wait for LED
display to light up.
•
Press "Auger ON/OFF" buttons ON.
•
To operate in defrost mode press the “Auto Timer”
button until it is illuminated.
•
When setting automatic times, please keep in mind it will take time for the frozen
product to become liquid or vice versa.
NOTE: The defrost timer is the same for all the bowls.
The defrost mode can be activated only if one auger button at least is on.
•
Turn the power switch ON and wait for
LED display to light up.
•
Make sure the clock button is OFF (LED
light on clock button should not be lit up).
•
First turn auger on by pressing the “Auger ON/OFF”
button until it beeps.
NOTE: The auger must be on before the unit will allow the cooling or freezing
mode to be activated.
•
Then select refrigeration or freezing mode by pressing the “MODE” button until the light
under the selection you desire is lit up.
NOTE: In the cooling mode, the LED display will read the actual temperature of the
product (the temperature setting is preset to NSF standards
Pushing the thermometer button it is possible to see the temperature into the left
bowl (led 1 light up) or into the right one (led 2 light up).
In the “Freeze” or “Off” mode the LED will read the current time.
Press and hold the left "Auger On/Off" button and
turn the power switch ON. Release the "Auger On/Off"
button when either “12” or “24” are shown (indicates
the current hour view).
Press the “Auto Timer” button until the desired hour display type is shown (12 to 24).
•
Press the left “MODE” button on the left until the desired temperature display type is
shown (°F or °C).
•
Press the left “MODE” button on the left until the current time is displayed to store the
changes. The unit is now ready for use.
•
Press the left “MODE” button until the “Cold” LED is lit, than press the “bowl
Temperature “ button choosing the bowl. The display will now show the current bowl
temperature in either °F or °C depending on which was selected in the Setting the
°F or °C Temperature Display section. Turn On the auger on the side that you want
to display the bowl temperature (press the “Auger On/ Off” button).
“FILTER CLEANING” Alarm
A filter cleaning alarm will activate when the unit is running hot due to insufficient internal air
circulation. When this occurs a “Filt” message will appear on the touch pad LED display readout
and an intermittent audible tone will also sound to alert the operator of this condition.
The “Filt” message will appear when the alarm activates (a beeping sound every 4-5 seconds).
To determine the condition that caused the alarm, see list of conditions below:
• Condition: The filter is dirty and needs to be cleaned.
Corrective Action: Clean and replace filter following instructions(Removing and Cleaning Filter).
• Condition: The unit is positioned too close to a wall or other object restricting air flow and
causing the machine to run at a higher temperature.
Corrective Action: Reposition unit to maximize ventilation space (Installation Instructions).
• Condition: The filter is not properly installed.
Corrective Action: Properly install filter (Removing and cleaning filter).
• Condition: The unit has been installed near a heat source, such as a coffee machine, ice
maker or cold beverage machine which expels hot air from its vents, causing the
machine to run at a high temperature (installation near a heat source should be
avoided)
Corrective Action: Reposition unit to maximize ventilation space.
“SYSTEM OVER TEMPERATURE” Alarm
A system over temperature alarm will activate as a safety precaution when the unit has
overheated to protect the compressor.
• The system automatically goes to “OFF” status where the compressor’s operations is
stopped, while augers will keep working to avoid forming ice blocks.
• When this occurs an “Err” message will appear on the touch pad LED readout accompanied
by a continuous buzzer sound to alert the operator of this condition.
• When this alarm activates, turn off all switches. Then determine the condition. (See “Filter
Cleaning” Alarm Section for Conditions and Corrective Actions)
beneath the tap (Q) and lower the lever (R)
very gently (Fig.12).
2. Adjusting the consistency:
To vary the consistency of the ice slush, turn
the knob (S) as shown in Fig. 13. Turning the
knob counter-clockwise will increase viscosity
(make the product denser). Turning the knob
clockwise will decrease viscosity (make the
product less dense).
ATTENTION: This device only changes the
consistency of the ice slush
dispensed, it does not effect the
cooling temperature.
ATTENTION: When the level of the ice slush
inside the bowl is below the
minimum, to stop the product
becoming too dense you must
switch off the cooling system
(OFF position), or top up the
bowl,
An indicator gauge for reference is located on the
back of the unit approximately 6 inches below the
adjustment knob.
5. Pull the lock pin/safety pin (Y) upwards, if the machine version is
without pin, unscrew one of the two locking knobs (Y1) – Fig.18.
Y1
Fig. 18
6. Slide the locking bar (Z) – Fig. 19 - outwards so that the bowls are
fully released. Unlock relief valve on the bowl lid cap and remove it
(Fig. 20).
Fig. 19
7. Lift the bowl lid cap lever (Fig. 20), remove the cap, unscrew the
knobs (A1) so the bowls can be lowered slightly, open the tap to
remove any remaining liquids and then extract the bowl from its
seating, pulling it outwards (Fig. 20).
Fig. 20
8. Remove the tap from its seating, pressing the two clamping wings at
the same time (Fig. 21) and pushing upwards.
Fig. 21
9. Dismount the tap, holding the body (R) pressed downwards, then
slide the lever (L) out of its seating (Fig. 22). Wash all the parts
thoroughly with hot water and washing-up liquid, rinse them well, and
proceed to remount them.
10. Separate the bowl from its cover, releasing the fastening (D1)
by pulling them upwards as shown in Fig.23.
11. Wash the bowl lid cap (Fig. 20), the bowl and its cover
carefully with water and washing-up liquid, rinse them well, and
proceed to remount them, making sure the sealing strip (E1) is
positioned correctly between the cover and the bowl. To
guarantee the seal, the rounded part of the said sealing strip (as
shown in Fig. 24) must be facing the cover.
12. Unscrew the fastening knob (F1) clockwise as the
direction indicated by the arrow (threading to the left) and
proceed by extracting the spiral scraper (G1) and the seals (H1)
and (I1). Clean the individual parts thoroughly (Fig. 25).
13. Clean the drip tray (J1) and the evaporator (K1) well (Fig. 26).
CAUTION: Do not wash any components in a dishwasher.
14. Proceed to remount the mixing unit as follows:
• dampen the sealing strip (l1) and insert it into its seating in
the correct position (Fig. 26A);
• apply a copious amount of Vaseline (supplied with the
machine) to the suction cup seal (H1) (on the part that comes
into contact with the evaporator K1) and insert it in its spiral
seating (Fig. 25 , Fig. 26);
• remount the spiral scraper (G1) (Fig.25);
• fasten all the components in place by screwing the knob F1
counter-clockwise (Fig.25).
15. Remount the bowl by pushing it into its seating (Fig. 27).
16. Proceed to fasten the bowl in place by tightening the
knobs (A1) – Fig. 26 and the locking bar (Z) – Fig. 19.
Fig. 24
Fig. 25
Fig. 26
Fig. 26A
Fig. 27
17. Remount the tap, remembering to smear both is seating in the bowl
and the seals (J) with Vaseline (Figure 28).
18. Dismount the drip tray (L1) by rotating it slightly and pulling
it outwards (Fig. 29). Wash all the parts carefully and
proceed to reassemble it by following the aforesaid
operations, remembering to reinsert the condensation
discharge pipe (M1) in its seating.
Fig. 29
19.If the drip tray is outfitted with coupling T0(Fig. 30) it is
possible to discharge the water in the drip tray without
removing it. In order to do that proceed as follow:
• remove the tap T1 (Fig. 30);
• connect a tube with the coupling T0 (Fig. 30).
NOTE:
“Six monthly” cleaning procedure is the standard time and it is suitable for the most of the
products, but the cleaning procedure depends on the kind of product used. If the product used
is a made of milk or 100% fruit plus sugar, for example, “Three monthly” cleaning procedure is
request. In case of doubt please contact the assistance or do the “Three monthly” cleaning
procedure.
ATTENTION: To guarantee the cooling system a good performance level, the routine
cleaning of the condenser filter is essential.
As explained earlier, an acoustic signal, accompanied by the word
FILT appearing on the control panel display, will warn the operator
when the filter is clogged and so must be cleaned before the
machine comes to an automatic stop.
Proceed as follows:
• disconnect the machine from the power supply;
• remove the filter as showed in Fig. 31 and clean it;
• dismount the cover (B) – Fig. 32;
• disconnect the plug (D) – Fig. 33;
• un-thread the back (E) – Fig. 33, pulling it high.
• clean the condenser with a brush if necessary (Fig.34).
You are reminded that all the operations described above
must be carried out with the machine switched off and the
power cable disconnected.
INSUFFICIENT AIR FLOW ALARM
To protect the entire cooling system, the ice slush drinks
maker is fitted with an electronic system which issues an
acoustic warning signal in the event of insufficient
ventilation or if the condenser filter is clogged with dust. In
addition to this, the words Filt and then Err will appear on the
control panel display.
If the operator does not intervene when the signal is issued,
restoring normal working conditions (i.e. ensuring better
ventilation or cleaning the filter), this electronic system will stop
the machine automatically.
SPINDLE BUSHING: replace every 6 to 12 months
depending on conditions of use and level of maintenance.
Replacement should ONLY be done by a qualified service
technician
•Remove the Rulon bushing using a flathead screwdriver as
shown
•Insert the brass guide tool on the drive shaft and slip the
bushing over the guide tool and firmly press into place.
•Using the special tool C and rubber mallet, gently seat the
bushing into place as shown
• Lubricate the bell shaped seal and reassemble onto
evaporator/auger shaft.
BELL SHAPED SEAL: replace every 1 to 6 months depending on conditions of use,
level of maintenance and lubrication. This part should be lubricated during
reassembly after every cleaning
BOWL GASKET: it’s the red gasket located at the rear of the bowl, replace as
necessary depending on conditions of use and level of maintenance; this part should
be lubricated during the re-assembling after every cleaning.
O-RING: the dispense valve o-ring should be replaced every 6 to 12 months or as
necessary when wear is apparent. Lubricate the o-ring each time they are replaced
or the dispense handle is removed for cleaning
The “sold out” pressure switches ,set on the electro-valves, detect the presence (or not) of
product in the bag in box. The “sold out” pressure switches are not related to ANY control
board. They work INDEPENDENTLY one from the other (left bowl sold out works
independently from right bowl sold out). When the “sold out” pressure switches detect the lack
of product, they stop the mix (product and water) dispensing. In managing bowls level
condition they are active only during a bowl loading phase
Level bowl =Level probes
The “level bowl” probes function is to control and manage the product level in the bowl. They
work INDEPENDENTLY one from the other (left bowl level probes work independently from
right bowl level probes).
Cut out level bowl = overflow probes
The “cut out level bowl” probes are safety probes. Their function is to avoid bowl’s overflow if
the “level bowl” probes are out of function (both of them or just one of them). If they are
active they inhibit the “pressure and level control board”.
If the “level bowl” probes are uncovered ,the mix is introduced in the bowl. If the “sold out”
During the first filling condition it necessary that both “level bowl” probes (the left bowl and
“Sold out” Pressure switch
Active
only
during the
bowl loading phase
“Sold out” Pressure switch
Pressure means product, no pressure means no product
First Filling Condition
First Filling Condition
Each time the unit is turned off and then turned on
Each time the there is a lack of power supply to the unit
First Filling Condition
pressure switch has detected product ,the bowl filling goes on until the “level bowl” probes
are covered (4 minutes max); otherwise the unit begins to buzz after 4 minutes because
the “level bowl” probes are not covered yet and the led (out of product) on the frontal
panel is lit. So that it is necessary to turn the unit off, remove the empty bag in box,
replace it with a new one and turn the unit on. The unit will work as a first filling condition.
the right one) are reached otherwise the unit begins to buzz and it does not work
“Sold out” pressure switch begins to work only if it detects product
4 minutes is the maximum time for filling the bowls
Managing bowl Level
“Level bowl” probes
uncovered
Product introduced in the bowl until the “level bowl”
probes are covered
The “pressure and level control board” INDIRECTLY detects the lack of product because
when the “sold out” pressure switch , set on the electro-valve, has detected the product
missing, it stops the mix dispensing. In this situation the “level bowl” probes will never
be covered, so after 1 minute (set time) the “pressure and level control board” detects
the lack of product and the led (out of product) on the frontal panel is lit.
If the “level bowl” probes are not covered by a minute ( i.e. the unit dispenses the mix
continuously so the mix introduced in the bowl does not counterbalance the mix
dispensed) the bowl loading phase stops and it is necessary to turn the unit off and then
turn the unit on. The unit works as a first filling condition.
only if the “level
bowl” probes are out
of service (both of
them or just one)
The overflow condition
If just one “cut out
level bowl” is reached
it is enough to stop
in both the bowls.
When the product
level in the bowl
reaches the “cut
out level bowl”
probes the unit
interrupts the bowl
loading phase.
Until both “cut out
level bowl” probes are
uncovered the product
can not be introduced
in any bowl ( i.e. until
left bowl is not under
the LT “cut out level
bowl” ,the unit do not
fill the right bowl).
Case Study
Some events that can happen during unit working are following:
1. If a bag in box is empty during the first filling condition, the “sold out” pressure switch,
set on the electro-valve, detects the lack of product and stops mix dispensing. After 4
minutes the “pressure and level control board” turns the frontal led (out of product)
on. So that it is necessary to turn the unit off, remove the empty bag in box , replace it
with a new one and turn the unit on. The un
2. If the unit is switched off and then switched on without removing the empty bag in
box, the unit works as a first filling condition and it happens the same situation
described at point 1.
3. If the “level bowl” probes are covered and the unit is switched off and then switched on,
the unit ignores the first filling condition because the “level bowl” probes are already
covered.
4. If the product inside a bag in box finishes during a managing bowl level, it is necessary
to switch the unit off, remove the empty bag in box and replace it with a new one and
then turn the unit on. The unit works as a first filling condition and, when the “level
bowl” probes are covered, the unit stops the mix dispensing in the bowl.
5. If the product inside a bag in box finishes as soon as the “level bowl” probes are
covered, the unit does not detect the end of product. As soon as the “level bowl” probes
are uncovered the “sold out” pressure switch detects the end of product and stops the
mix dispensing. After 1 minute the “pressure and level control board” lights the relative
led (out of product) on. So that it is necessary to turn the unit off, remove the empty
bag in box , replace it with a new one and turn the unit on. The unit will work as a first
filling condition.
The pressure inside the bowls is managed by the “pressure and level control board”.
First filling condition
In first filling condition the pressure inside the bowls is not checked until product reaches the
“level bowl” probes in BOTH the bowls. When the “level bowl” probes are satisfied in BOTH the
bowls the “pressure and level control board” begins to work and maintains the pressure inside
each bowl between 120mbar (on ) e 160mbar (off). After 1 minute and 30 seconds, if the
pressure inside the bowls does not reach 160mbar, there are 2 possibilities:
•there is no more gas in the CO2 external tank; so the “out gas” frontal led is lit and it is
necessary to turn the unit off, remove the empty CO2 tank and replace it with a new
one
•there is a lack of CO2; so the CO2 dispensing is interrupted in order to avoid the CO2
external tank run out. It is important to find out the origin of this lack.
Pressure managing
The pressure inside the bowl (between 120mbar e 160mbar) is managed INDEPENDENTLY
from “level bowl” probes. For example if the “level bowl” probes (in both bowls or just in one)
are uncovered and a bag in box is empty, the led (out of product) on the frontal panel is lit
(after 1 minute) but the pressure inside the bowl remains between 120mbar e 160mbar even if
the bag in box is not removed and replaced with a new one.
VFCB PRE MIX
Safety Probes
Level Probes
Product
Left Bowl Right Bowl
Description
Sold out = Product presence probes
The “sold out” probes detect the presence (or not) of product in the external tank/s.
They work independently one from the other (left bowl sold out works independently from
right bowl sold out). In first filling condition they are working after 30 seconds from the unit
switched on . In managing bowls level condition they are active only during a bowl loading
phase and they are working with an hysteresis of 3 seconds.
Level bowl =Level probes
The “level bowl” probes function is to control and manage the product level in the bowl. They
work independently one from the other (left bowl level probes work independently from right
bowl level probes).
Cut out level bowl = overflow probes
The “cut out level bowl” probes are safety probes. Their function is to avoid bowl’s overflow if
the “level bowl” probes are out of function (both of them or just one of them). If they are
active they inhibit the “pressure and level control board”.
If the “level bowl” probes are uncovered, the product is introduced in the bowl. After
30 seconds if the “sold out” probes have detected product ,the bowl filling goes on
until the “level bowl” probes are covered (4 minutes max); otherwise the unit begins
to buzz because the external tank is empty (or both tanks or just one as showed by
the led on the frontal panel). So that it is necessary to turn the unit off, remove the
empty tank (or the empty tanks) , replace it (them) with a new one/s and turn the
unit on. The unit work as a first filling condition. During the first filling condition it
necessary that both “level bowl” probes (the left bowl and the right one) are reached
otherwise the unit begins to buzz and it does not work.
Each time the unit is turned off and then turned on
Each time the there is a lack of power supply to the unit
“Sold out” probes begin to work with delay of 30 seconds
from the unit switched on
4 minutes is the maximum time for filling the bowls
Managing bowl Level
“Sold out” probes
Active only during the bowl loading phase
“Sold out” probes
“Level bowl”
probes uncovered
The “sold out” probes detect the product presence (or not) in the external tanks with an
hysteresis of 3 seconds. 3 seconds software hysteresis are introduced to prevent the
probes recognize the tank bowl conduit as empty when there is residual gas in it.
If the “level bowl” probes are not covered by a minute ( i.e. the unit dispenses the mix
continuously so the mix introduced in the bowl does not counterbalance the mix dispensed)
the bowl loading phase stops and it is necessary to turn the unit off and then turn the unit
on. The unit works as a first filling condition.
Detect the lack of product with an hysteresis of 3 second.
Product introduced in the bowl until “level bowl” probes are covered
VFCB Operator’s Manual
The overflow condition
pha
se.
Overflow Condition
When the product
level in the bowl
reaches the “cut out
level bowl” probes
only if the “level
bowl” probes are out
of service (both of
them or just one)
The overflow condition
If just one “cut out
level bowl” is reached
it is enough to stop the
bowl loading phase in
both the bowls.
the unit interrupts
the bowl loading
Until both “cut out
level bowl” probes
are uncovered the
product can not be
introduced in any
bowl ( i.e. until the
product level in the
left bowl is not under
the LT “cut out level
bowl” the unit do not
fill the right bowl).
Case Study
It follows some events that can happen during unit working:
1. If external tank (or tanks) is (are) empty during the first filling condition, the “sold
out” probes , after 30 seconds , detect the lack of product and stop the unit. So it is
necessary to switch the unit off, remove the empty tank and replace it with a new
one and then turn the unit on.
2. If the unit is switched off and then switched on without removing the external tank
empty (or tanks), the unit works as a first filling condition and it happens the same
situation described at point 1.
3. If the “level bowl” probes are covered and the unit is switched off and then switched
on, the unit ignores the first filling condition because the “level bowl” probes are
already covered.
4. If the product inside the external tank finishes during a managing bowl level (the
relative led must be lit), it is necessary to switch the unit off, remove the empty
tank and replace it with a new one and then turn the unit on. The unit works as a
first filling condition and, when the “level bowl” probes are covered, the unit stops
the product dispensing in the bowl.
5. If product inside the external tank finishes as soon as the “level bowl” probes are
covered, the unit does not detect the end of product because the monitoring of the
“sold out” probes (covered / uncovered) is active only during the bowl loading
phase. As soon as the “level bowl” probes are uncovered the “sold out” probes
(after 3 second hysteresis ) detect the end of product; so it is necessary to turn the
unit off, remove the empty tank and replace it with a new one and then turn the
unit on.
The pressure inside the bowls is managed by the “pressure and level control board”
First filling condition
In first filling condition the pressure inside the bowls is not checked until product reaches the
“level bowl” probes in BOTH the bowls. When the “level bowl” probes are satisfied in BOTH
the bowls the “pressure and level control board” begins to work and maintains the pressure
inside each bowl between 120mbar (on ) e 160mbar (off). After 1 minute and 30 seconds, if
the pressure inside the bowls does not reach 160mbar, there are 2 possibilities:
•there is no more gas in the CO2 external tank; so the “out gas” frontal led is lit and it is
necessary to turn the unit off, remove the empty CO2 tank and replace it with a new
one
•there is a lack of CO2; so the CO2 dispensing is interrupted in order to avoid the CO2
external tank run out. It is important to find out the origin of this lack.
Pressure managing
The pressure inside the bowl (between 120mbar e 160mbar) is managed INDEPENDENTLY
from “level bowl” probes. For example if the “level bowl” probes (in both bowls or just in one)
are uncovered and the external tank (product tank) is empty, the led is lit in the frontal panel
(after 3 second hysteresis) but the pressure inside the bowl remains between 120mbar e
160mbar even if the external tank is not removed and replaced with new one (product tank).
NOTE: The following procedures must be performed by a qualified service technician.
Problem Possible Cause Solution
“Filt” or “Err” message
appears on the touchpad LED
readout
The machine does not cool, or
cools only partially, but the
compressor/s is/are running
• The filter is dirty and needs
to be cleaned
• The unit is positioned too
close to a wall or other object
restricting air flow and
causing the machine to run at
a higher temperature
• The filter is not properly
installed
• The unit has been installed
near a heat source, such as a
coffee machine, ice maker or
cold beverage machine which
expels hot air from its vents,
causing the machine to run at
a high temperature.
(Installation near a heat
source should be avoid)
• The space around the
machine is inadequate for
ventilation
• Freezer is in defrost
• The condenser fins are
clogged with airborne
particles
• Fan motor is not running
• Refrigerant is leaking
• Clean and replace filter
following instructions
(Removing and Cleaning
Filter)
• Reposition unit to maximize
ventilation space (see
installation figures)
• Properly install filter see
“Removing and cleaning filter”
• Reposition unit to maximize
ventilation space (see
installation figures)
• Allow at least 8” (20cm)
between the machine and
anything next to it; keep
away from heat sources
• Return to freeze mode
• Remove the side panels.
Using a brush or compressed
air clean the condenser
• Check the fan motor’s
electrical connections and, if
disconnected, reconnect. If
still not operating, replace the
motor