
This device is not suitable for use on humans.
It is designed for training purposes only and must be kept unmistakable separate from fully operative
Customers are responsible for ensuring that the relevant people within their organisation are provided with access to
the information and the general safety instructions and warnings provided in this manual.
is a simulation and training device for realistic Advanced Life Support training on a
defibrillator (AED). Due to the similarity of its design and operating elements, the Trainer 1000 is particularly
suitable for training and instructing peo
does not deliver energy through adhesive electrodes, which means that it does not pose a risk
experienced trainees. At the same time, the Trainer 1000 can be used to simulate changes
state by using the AED Trainer’s IR remote control and intervene if the Trainer is used incorrectly.
ple in the use of the Lifepak® 1000 series defibrillators.

Trainer 1000 AED Training System Quick Start Guide
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
..............................
................................
................................
................................
................................
............................
................................
................................
................................
................................
................................
.......11Operating elements, controls and connections
................................
................................
................................
Preparing the Trainer 1000 for use
................................
................................
................................
................................
.16Unpacking and checking components
................................
................................
................................
................................
................................
................................
................................
Starting a training session
................................
................................
................................
................................
Verifying or creating an AED Trainer configu
................................
................................
................................
....17Connecting the training electrodes to the Trainer 1000
................................
................................
............................
Turning the unit On / Off
................................
................................
................................
................................
Attaching the training electrodes to the mannequin
................................
................................
................................
.18Using the remote control
................................
................................
................................
................................
................................
................................
................................
................................
................................
.....21Accessing setup mode
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
Setting options: CPRmax menu
................................
................................
................................
................................
...26Configuring the energy protocol
................................
................................
................................
................................
.28Changing settings for CUSTOMER
................................
................................
................................
...............................
................................
................................
................................
................................
................................
................................
................................
................................

................................
................................
Cleaning and maintaining the
................................
................................
................................
................................
Potential errors and remedies
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
...30................................
................................
................................
.......30................................
................................
................................
..34................................

The Trainer 1000 is designed only for training people in the use a
defibrillator on adults and children over the age
of 8, weighing more than 25 kg, and using standard training electrodes (black connector).
This ensures that trainees are taught which electrodes to use on which patients right from the outset of their
Press the PAUSE button on the remote control.
Select the required shock protocol from the table below and press the corresponding button on the remote
Press PAUSE again in order to verify your choice. The AED Trainer will turn itself off automatically.
You can now start the training session. Use the remote control to chose a particular problem situation. Please
refer to the rear of the Quick Start Guide
Once the AED has been switched on, it will display the message CONNECT ELECTRODES. Connect the training
electrodes and press the ELECTRODES button on the remote control to then proceed with an analysis.
Please note: The AED Trainer w
ill automatically open the shock protocol that was used last. Please refer to the
training system user manual for further information.
AED Trainer shock protocol
specific settings can be changed in setup mode (see user manual).

Remote control buttons and functions
changes all shock decisions to NO SHOCK ADVISED. This function can only be used
when the message PRESS FOR ANALYSIS or ANALYSIS IN PROGRESS, STAND CLEAR! is being
By pressing this button you simulate the fitting of the training electrodes to the mannequin.
Pressing the button again simulates disconnection of the electrodes.
Turn the MOTION SIMULATION function On or Off. This functio
Press once to activate service alert and to display the message CALL CUSTOMER SERVICE. To
cancel, turn off the AED Trainer or press button again.
Press once to activate LOW BATTERY alert.
This button is disabled. Intended only for use on Trainers with ECG mode.
PAUSE. Can be pressed at any time to stop any
resume operation. Press F1, F2, F3, F4, F5 or C1 for changing the shock protocol while the
Trainer 1000 is set to PAUSE. Press PAUSE again to verify the change and to turn the AED
n is only available during analysis.
Press again to deactivate.
of the Trainer 1000’s functions. Press again to

Defibrillation is a recognised method for treating certain, potentially fatal, cardiac arrhythmias, i.e. abnormal heart
beating. Defibrillation consists of delivering a therapeutic dose of electrical energy to the affected heart with a
Service is a training and simulation device for the work
performed with an automated external d
efibrillator (AED). The Trainer 1000 is not capable of detecting cardiac
arrhythmias or of delivering a defibrillation shock, but is a traini
ng device only that allows potential users to be
trained in the use of a defibrillator by simulating corresponding sounds and visual indicators.
Defibrillation is only one of the medical treatment methods available for resuscitating patients with shockabl
cardiac arrhythmias. Depending on the particulars of a case, patients might also require other treatment, such as:
Cardiopulmonary Resuscitation (CPR)
The success of a resuscitation attempt also depends on the length of t
ime that elapses between the onset of a
cardiac arrest during which no blood is circulated (e.g. in ventricular fibrillation or pulseless ventricular
tachycardia) and defibrillation.
According to the American Heart Association (AHA), the following is criti
survival from sudden cardiac arrest:
Early access to patients and identification of the problem
Early CPR by first responders or bystanders
Early advanced treatment by emergency medical services (EMS)
physiological state may also affect the likelihood of successful defibrillation. Failure to resuscitate a
patient is therefore not a reliable indicator of the performance of the resources, equipment or rescuers involved in
the resuscitation attempt. Patien
ts will sometimes exhibit a muscular response (such as jumping or twitching)
during energy transfer. The absence of such a response, however, is not a reliable indicator of the actual energy
delivered or the defibrillator’s performance.
llation largely depends on first aider’s confidence in administering it
repeated theoretical and practical training in the use of defibrillators can save lives.
is designed to help you with.

The Trainer 1000 can be used to simulate the following operating modes:
AED mode (with activatable and de
electrode discharge derivation)
ECG monitoring mode (requires Upgrade Kit); see
As with real live automated external defibrillators, the
issues specific instructions when detecting shockable rhythms or non
efibrillation) to the simulated patient requires the active involvement of the person operating the defibrillator.
This function allows expert users to decide on administering defibrillation shocks on the basis of the ECG analysis
ECG mode simulates the monitoring of the electrical activity of a patient’s heart using conventional ECG electrodes
that are attached using snap fasteners. The readout provided on the screen represents the con
lead II. The Trainer 1000 does not issue any spoken instructions in this mode. If the CPPS (Continuous Patient
Surveillance System) has been activated (setup menu), the user will be prompted to check the patient and change
tion electrodes if a shockable cardiac rhythm is detected.
Setup mode is used to select the relevant settings for specifying how the
evaluates the selected ECG rhythms and
shockable rhythms. The delivery of shocks

make use of the following terms:
Immediate hazard that will result in serious injury or death.
Hazard or unsafe practice that could result in serious personal injury.
Hazard or unsafe practice that could result in minor pers
onal injury, product damage or property
is powered by electricity. If not used properly and as described in these instructions, this
electrical energy may cause serious injury and death. Users must be thoroughly familiar with these operating
instructions and all of the Trainer 1000’s operating
controls, indicators, connections and accessories’ functions
before operating the device.
. It contains no responder
serviceable components and may carry
dangerous voltages. Always contact Customer Services f
Fire and electric shock hazard.
Do not immerse any part of the
in water or other fluids. Avoid spilling any fluids onto the device
and its accessories. Do not clean with ketones or other flammable agents. Unless specified otherwise, do not
utoclave or sterilise the
. Clean only using a slightly damp cloth and mild detergent.
Fire and explosion hazard.
Do not use defibrillators in the presence of flammable gasses or anaesthetics. Particular care must be taken when
using a defibrillator in c
lose proximity to sources of oxygen (e.g. anaesthesia bags or respirator hoses). Turn off
gas source or remove source before defibrillation. Please instruct users to observe this safety precaution before
; observe whenever possible for training
Possible electrical interference.
The use of cables, electrodes or accessories not explicitly specified for use with this training unit can affect the
Trainer 1000’s performance. Do not use any of the
’s accessories with fully operational def
used to treat patients and vice a verse. The Trainer 1000 must only be used with the components and accessories
specified in this user manual.
Potential device failure.
Always carry the charger when taking the device to training event
s and courses held outside your premises in
order to prevent the battery from running out of power.

Potential device malfunction.
The use of cables, electrodes or batteries from other manufacturers can cause the Trainer 1000 to malfunction,
and could void your warranty. The Trainer 1000 must only be used with the accessories specified in this user
Only use the original charg
er with the original cable supplied with the Trainer 1000 for connecting it to the mains.
Safety risk and potential device damage.
Monitors, defibrillators, training devices and their accessories (including tra
ferromagnetic materials. As with all ferromagnetic equipment, this training device must not be used in the
presence of the high electromagnetic fields generated by Magnetic Resonance Imaging (MRI) devices. The high
etic field created by an MRI device will attract the device with a force sufficient to cause death or serious
injury to persons located between the two. This magnetic attraction may also cause damage to the device. Please
contact the MRI device’s manufactu
rer for more information. The device should, in general, never be exposed to
magnetic fields as, e.g. emanate from annual magnets commonly used in medicine. Any spaces designated ‘Do not
bring your credit card into this area’ are not suitable for setting u
Route the power cables in such a way that they are not at risk of being damaged. Never use damaged cables!
Damaged cables can cause fire and potentially fatal electric shocks.
mechanical use i.e. immersing into water or dropping, can damage the device. Do not
continue using the device following such an incident. Please contact Coretec
The Trainer 1000, the training
Please keep this manual for future reference.
ining electrodes and cables) contain
Service GmbH’s technical services in
electrodes and cables do not contain any latex.

The Trainer 1000 features the following symbols with the following meanings:
Read this user manual before using the Trainer
Observe the safety instructions.
Do not dispose of as household waste
Do not dispose of the Trainer 1000 and its packaging as
CE certification provided by the manufacturer
The number behind the symbol is the year of manufacture.
Electrode connector symbol
This is where the electrodes are connected
This is where the cable for the remote control is connected and where the infrared receiver for
remote control is located.
Socket for the charger. Caution! Only use the supplied charger.
Do not connect a power supply plug or a different charger.
This is where the fuse is located and where it can be
Available as a spare part.
The fuse can also be replaced with a standard micro fuse with the following specifications:
G fuse pack 1.25A slow blow (250V) 20 x 5 mm

This indicator comprises three symbols that indicate whether the
defibrillator is ready for operation or whether there is a problem.
meaning of each of these symbols and the place / time it will be shown
will be explained in the following section. Please note that the Trainer
1000 is equipped with a plastic indicat
or placeholder instead of a display.
This means that, for training purposes, all of these symbols will be visible
The spanner symbol indicates a condition that might or is preventing the
defibrillator from being used normally. The corres
audio signal can be activated by pressing the symbol key on the remote
The OK symbol indicates that the defibrillator is ready for use.
will only be shown when de defibrillator is turned off.
indicator will display the battery symbol only when the
defibrillator is turned off. If the battery symbol is shown with one bar inside
it, the battery is running low. If the symbol is empty, the battery is running
very low and the OK symbol will subsequen
tly not be displayed when the
defibrillator is turned off. The corresponding indicator and audio signal can
be activated by pressing the symbol key on the remote control.
Issues voice prompts and audio signals.
ON/OFF button turns the power supply on and off. The green LED
above this button will always light up when the Trainer 1000 is turned on.

Pressing this red button will cause the device to simulate a delivery of a
For selecting the operating mode (manual or AED).
The two softkeys work in conjunction with the screen and can be used to
make selections while using the
softkeys depends on the action undertaken at the time and will be indicated
by their corresponding labels as shown on the screen.
Displays pertinent information for use during all modes of
For connecting training defibrillation electrodes (black) and training ECG
The relevant function of the

The heard rate meter will display heart rates between 20 and 300 beats per minute. This meter will only be shown
during AED monitoring, in manual mode and when using the 3
The battery status symbol on the screen indicates the relative charge held by the battery when the Trainer 1000 is
turned on, and can be adjusted through the r
emote control. This symbol can be displayed in two ways: 1. With
four solid bars inside it, indicating that the battery is fully charged. 2. Blank, indicating that the battery is running
very low, and accompanied by the message REPLACE BATTERY.
CG readout shown on the screen is a non
diagnostic ECG that is generated with the aid of the training
electrodes or lead II of the ECG cable. The ECG rhythm is either specified by the instructor through the remote
control, or as part of one of the pre
rammed scenarios. Please notify trainees that the presence of an ECG is
not a reliable indicator that the patient has a pulse.
This label defines the respective function of the relevant softkey. Examples of those are ANALYSE and

This section describes how to prepare the AED Trainer for use. Please check whether the device has been supplied
with all of the components specified below in order to prevent later complications.
Remove the AED Trainer from its box. Check the device and all of its accessories for signs of transport damage.
Keep the box for potential later use.
The box the AED Trainer is supplied in should contain:
2 AAA batteries (for the remote control)
Soft or hard bag, depending on order
Declaration of conformity
The AED Trainer is fitted with a
rechargeable lithium ion battery. This battery requires little attention and can hold
sufficient charge for running the Trainer 1000 for approx. 4 hours of non
Once empty, this battery can be recharged from the mains usi
Only use the supplied charger for operating the Trainer 1000 and charging the battery. Other mains adapters
might damage the Trainer 1000.
When switched off, the Trainer 1000 will fully recharge within approx. 2.5 hours (using th
The Trainer 1000’s battery can also be charged while it is switched on, although it will then take longer to charge.
Batteries age. This aging process depends on such factors as time and charge/discharge cycles. We recommend
he Trainer 1000’s batteries replaced by our customer services once the battery no longer provides
sufficient power to operate the Trainer 1000 for a full two hours in battery mode.
stop simulation when used properly.

Verifying or creating an AED Trainer
To simulate a LIFEPAK 1000 S with Defi
electrode ECG monitoring and manual mode, select “YES” under the setup
options ECG READOUT and “YES” under MANUAL ACCESS.
Check the AED Trainer settings and configure as required. Please refer to the re
levant sections of this manual for
detailed information on the setup options and configurations.
Connecting the training electrodes to the Trainer 1000
Push the electrode connector into the relevant socket on the AED Trainer until it engages.
, take a firm hold of the connector and carefully, but firmly, pull out of the socket. Never try to unplug
the training electrodes by pulling on the cable! Do not use any tools to unplug the electrodes, since these could
damage the Trainer 1000’s housing a
Turning the unit On / Off
Press the ON / OFF button on the AED Trainer to turn it on/off.
Once turned on, the AED Trainer will run a four
second check routine. While running this routine, all of the
indicators will blink once,
and the following message will be briefly shown:
Once this check routine has been successfully run, the green ON/OFF indicator will light up permanently and the
AED Trainer will automatically start up.
Depending on the AED Trainer’s c
onfiguration and electrode contact, the screen will now display one of the
STAND CLEAR! ANALYSIS, STAND CLEAR
The AED Trainer cannot be turned on if the Li
ion battery is low on charge.
this case, connect it to the charger. It should take no more than a few minutes for the Trainer to be charged
sufficiently for it to be turned on. However, the AED Trainer must always remain connected to the charger until
the battery has been sufficiently
recharged. (The charge indictor LED will go out when the battery is fully

Attaching the training electrodes to the mannequin
The illustration below shows where to position the training electrodes on the mannequin.
Figure: Attaching the tra
ining electrodes to the mannequin
Once the electrodes have been attached to the mannequin, the instructor must signal the device accordingly by
pressing the ELECTRODES button on the remote control. This will automatically activate the selected scenario wi
This training scenario will only start running if the training electrode connector has been properly inserted into the
socket on the AED Trainer and the instructor has verified that the electrodes have been correctly attached thro
If, while using the AED Trainer:
the training electrodes are removed from the device, it will display the message CONNECT ELECTRODES
the training electrodes, on the other hand, come off the mannequin, it will not display this m
This message function can be turned on/off by the instructor by pressing the electrode button on the remote
Never use defibrillator electrodes intended for use on patients. The Trainer 1000 will not reco
electrodes and they could damage the outer skin of the mannequin or leave behind stubborn residues.

The remote control requires 2x AAA batteries.
To fit and replace batteries, proceed as follows:
battery compartment cover on the rear of the remote control with your thumb and push open.
Insert two AAA batteries; make sure that the batteries’ (+) and (
) poles are correctly aligned.
Replace the battery compartment cover and make sure it engages.
ring a training session, the remote control can be used to create realistic training scenarios, i.e. can be used by
stop the AED Trainer from operating
configured routines that involve the messages NO SHOCK ADVISED, CONNECT E
CALL SERVICE and REPLACE BATTERY
select one of the six existing shock protocols (only possible while the device in PAUSE MODE)
(if applicable) select the ECG readout to be displayed on the screen.
The AED Trainer and remote control can e
ither communicate with one another through a wireless connection,
using infrared signals, or a remote control cable.
When using the remote control with a wireless connection, the instructor will have to stand behind the AED
Trainer. The remote control shou
ld always be held horizontally rather than aimed straight at the AED Trainer.
When the AED Trainer is turned on, all of the remote control settings from the previous training session, with the
exception of the current shock protocol, which will be saved, w
If the remote control is used with the cable, insert the cable’s right
angled connector into the corresponding
socket on the remote control. Connect the other end of the cable to the corresponding socket on the AED Tr
side. To remove, take hold of the cable’s connectors on both sides and pull out carefully.

Remote control buttons and functions
This button changes all shock
when the message PRESS FOR ANALYSIS or ANALYSIS IN PROGRESS, STAND CLEAR! is being
By pressing this button you simulate the fitting of the training electrodes to the mannequin.
Pressing the button again simulates disconnection of the electrodes.
Turn the MOTION SIMULATION function On or Off. This functio
Press once to activate service alert and to display the message CALL CUSTOMER SERVICE. To
cancel, turn off the AED Trainer or press button again.
once to activate LOW BATTERY alert.
This button is disabled. Intended only for use on Trainers with ECG mode.
PAUSE. Can be pressed at any time to stop any of
resume operation. Press F1, F2, F3, F4, F5 or C1 for changing the shock protocol while the
Trainer 1000 is set to PAUSE. Press PAUSE again to verify the change and to turn the AED
decisions to NO SHOCK ADVISED. This function can only be used
n is only available during analysis.
Press again to deactivate.
the Trainer 1000’s functions. Press again to

section describes how to:
select or change the Energy Protocol, Auto Analyse and CPR
select the required language for the voice prompts and text messages, and to use other device options.
These settings will not be
deleted when the device is turned off and will even be saved if the battery runs
Make sure that the AED Trainer is turned off.
Press both softkeys at the same time and simultaneously press the AED Trainer ON/OFF b
The screen will now display the configuration menu (set up mode).
The Trainer 1000’s setup mode is not password protected, as this is not relevant to a training device.
You will now be presented with 4 different menus in which to adjust the
Trainer 1000’s settings to suit your
This menu can be used to configure the audio settings.
This menu can be used to select the settings that correspond to your CPR algorithm and the display options
available to users in sem
Chose between AED operation only or manual operation for specialist personnel. You can also active the analysis
function for use in manual mode as a useful tool for the user.
Service mode (not accessible for customer)
These settings are used to adapt the Trainer 1000’s functions to mimic those of your defibrillator. This is also
where the language for the voice prompts, text messages and training scenarios can be adjusted.

The Device ID is your Trainer
1000’s serial number and is
specific to your device, which
is why it cannot be changed.
defibrillators, the date and
time are only used for data
purposes. Since the Trainer
1000 is not equipped with
these functions, the date and
time setting has been included
in the menu for demonstration
purposes only and cannot be
This option can be used to set
up the audio signals issued by
used to turn the audio alert
that accompanies the delivery
of defibrillation shocks on or
maintenance intervals in fully
This is where you can find the

You can select the energy to
be delivered during the 1
equal or greater than that of
150,175,200,225,250,300,325
As a measure of care, the
energy delivered by a shock
will not be increased if the
last heart rhythm was not
recognised findings on the
effectiveness of consecutive
shocks of equal or higher
shocks has been excluded.
Confirmation (only for SW
versions later than 1.10)
Maximise the amount of CPR
being administered by using
the advanced CPR settings.
CPR Time 1: 15s, 30s, 45s,
CPR Time 2: 15s, 30s, 45s,
30s, 45s, 60s, 90s, 120s, 180s
The device will repeatedly
prompt the user to perform
Select the times the rescuer
After every NSA (No Shock
Advised, i.e. analysis result: No
Check Breathing, Check vital

The Service Mode menu of this device cannot be changed.
For specifying the frequency
Surveillance System) is to
your target rescuer group
display the ECG readout in
Specify to what extent the
an analysis, thus increasing
function for use in manual

Select the level of function
Defi-electrode monitoring
target group the training
Select a sequence of heart
are not shockable in order

Setting options: CPRmax menu
The cprMAX technology has the following setup options:
Only applies to the first analysis. Prompts for CPR
Prompts for CPR after a shockable ECG rhythm is detected and while the defibrillator is
charging. Only applies to second and subsequent analyses if a SHOCK ADVISED. The SHOCK button will only be
bled once the charging process has completed and the CPR Time has elapsed.
For an abbreviated rhythm analysis after Initial CPR Time or Pre
Eliminates analysis after each shock and inserts a CPR prompt
This eliminates the three
Eliminates Pulse Check prompts after shocks, but only if shock is advised.
The duration of CPR can be increased by selecting the following setup options:
Initial CPR Time: 15 seconds
Pre-Shock CPR Time: 15 seconds or longer
Confirmation Analysis: Off
Please note: It is not necessary to activate all options in order to maximise the quantity of CPR administered
during treatment with an AED
If the Initial CPR Time is set to 15 seconds or longer, a CPR will be prompted by the time the electrodes have been
attached to the patient and the first analysis has been completed.
Once the electrodes have been attached, the AED will prompt
The display will now show a CPR Countdown Timer. The CPR time displayed by the timer is determined by the time
that was chosen in the setup options.
If the AED detects a shockable ECG rhythm, it will prompt to start CPR immediately and will then prompt
WITNESSED THE ARREST, PUSH CANCEL.
If you witnessed the arrest, you should proceed with the defibrillation. If you did not witness the arrest, you
should perform CPR. To proceed with the defibrillation, press the CANCEL softkey. This will end the CPR time and
will be followed by the voice p
shock in accordance with your AED your training.
To proceed with CPR, do not press the CANCEL softkey. The Initial CPR Time will then correspond to the time
options, for example, 90 seconds. The elapsing of the CPR Time will be followed by the voice
Proceed with the delivery of the shock in accordance with your AED training.
shockable ECG rhythm, it will prompt
should proceed to perform CPR for the time shown by the countdown timer.
immediately after the first analysis.
AR, ANALYSING NOW, STAND CLEAR
and the charging tone. Proceed with the delivery of the
to start CPR. There will be no other prompt. You

Shock CPR Time is set to 15 seconds or longer, you will be prompted to start CPR immediately after a
shockable rhythm is detected, before the shock is delivered and while the AED is charging.
Shock CPR Time refers to the second and any subsequent shocks.
If the AED determines that a rhythm is shockable after completion of the analys
is, it will display the following
The CPR Time will then correspond to the time selected in the Pre
options, such as, for example, 15 seconds. The elapsing of the CPR Time will be followed by the voice prompt
Proceed with the delivery of the shock in accordance with your AED training.
Stacked shocks and Pulse Checks
If STACKED SHOCKS have been set to OFF, the amount of CPR that will be administered will increase because you
prompted to perform CPR after each shock as opposed every third shock. Once a shock has been delivered,
the AED will prompt you to check the pulse and start CPR as opposed to starting an analysis. Once the CPR Time
has elapsed, the AED will prompt you star
If the PULSE CHECK option is set to OFF, the time available for performing CPR increases since the PULSE CHECK
prompt after delivery of a shock will have been eliminated. If the PULSE CHECK option is set to OFF, the AED will
ompt you to immediately start CPR after delivery of three consecutive shocks without performing a pulse check.
Combining Stacked shocks and Pulse Checks
If both the STACKED SHOCKS and PULSE CHECK options are set to OFF, the AED will proceed as follows:
The AED will not perform an analysis after delivering a shock.
The AED will not prompt for a PULSE CHECK after delivering a shock.
The AED will prompt you to START CPR, as detailed above.
Following the elapsing of the CPR Time, the AED will
erform an analysis. If the subsequent analysis determines that NO SHOCK is ADVISED, the AED
will prompt you to perform a PULSE CHECK.
If the CONFIRMATION ANALYSIS option is set to ON, the AED will perform an abbreviated rhythm analys
immediately before the delivery of a shock to confirm that there is still a shockable rhythm. Confirmation analyses
are only performed if the Initial CPR Time or Pre
Shock CPR Time options are enabled.
If the Initial CPR Time or Pre
ons are enabled, the AED will start a confirmation analysis and
STAND CLEAR. ANALYSING NOW, STAND CLEAR
as soon as the countdown timer reaches 0.
If the rhythm has changed, the confirmation analysis will cancel the shock and the AED will pr
If the rhythm is still shockable, the previous decision of SHOCK ADVISED will be confirmed and the AED

Configuring the energy protocol
Open setup mode (described in preceding sections above).
Keep pressing NEXT to go through the different shock protocol options. You can select the energy to be delivered
shock individually, and that to be delivered during the 3
e set to either: 150,175,200,225,250,300,325 and 360 joule
Press SELECT to save the required energy setting when it comes up and move to the next setup screen or press
ON/OFF to close the setup menu and to turn off the Trainer 1000 with the new settings.
The AED’s default shock energy settings start with 200, escalate to 300, and are then followed by 360 joules for all
subsequent shocks, which, in studies conducted by Medtronic Physio
ptiv-Biphasic™ technology, have proven to be the most effective and safe.
when CUSTOMER1 is shown on the display.
The first S will then be shown with a cursor. While the cursor is on the first S:
to change between S and N (S = Shock, N = No Shock).
For all successive Ss’: Press
to change between S, N and the space character.
to move the cursor to the right.
to finish. Your scenario will be saved.
The CPR Time is the period of time for which the AED Trainer will prompt the user to
perform a CPR. There are two different CPR Times, each of which can be set to any of seven
different periods of time (in seconds): 15,
for both CPR Time 1 and CPR Time 2 are 120 second.
CPR Time after delivery of shocks.
CPR Time for CPRs if NO SHOCK is ADVISED.
In consideration of the latest, internationally
equal or higher intensity, the option of delivering consecutive shocks has been excluded. Please contact Coretec
Service GmbH if you would like to receive any additional information on this.
and all subsequent ones. These
Control with the Medtronic Impulse
30, 45, 60, 90, 120, and 180. The default settings
recognised findings on the effectiveness of consecutive shocks of

Turning AUTO ANALYSE on/off
Auto Analyse is a function that enables the input required from users to be reduced to a minimum. When this
function is enabled, the Trainer 1000 will automatically start an ECG analysis as soon as the adhesive electrodes
have been attached to the patient. This minimises the chances of the AED being used incorrectly when operated

Cleaning and maintaining the
Always disconnect the Trainer 1000 from the charging voltage supply by pulling ou
The Trainer 1000’s surfaces can be cleaned with a conventional, mild all
make sure before use that the cleaner will not damage the Trainer’s surfaces.
Do not allow any liquid to find its way into
For safety reasons, we recommend not to switch on the device and not to reconnect it to the mains until it has
Do not use any easily flammable solvents such as Acetone, ethyl alcohol or white spirit for
fire hazards and can corrode the Trainer 1000’s surfaces.
The Trainer 1000 cannot be sterilized.
Batteries age. This aging process depends on such factors as time and charge/discharge cycles.
having the Trainer 1000’s batteries replaced by our customer services once the battery no longer
provides sufficient power to operate the Trainer 1000 for a full two hours in battery mode.
Recycle the packaging and the Trainer 1000, when it ha
Do not dispose of the Trainer 1000 and its packaging as household waste.
Please contact our customer services from more information on environmentally friendly disposal.
purpose cleaner and a damp cloth.
the Trainer through slots or connectors.
s reached the end of its life, in accordance with the local

Potential errors and remedies
The Trainer doesn’t start
The battery charge is so low
that there is not sufficient
current to start the Trainer.
Plug in the charger and wait a few minutes.
Now switch on the Trainer and
leave the charger in. When the charging LED goes
out, the battery is full and the charger can be
Only use the Trainer with the supplied USB stick or
a software update that has been specific
configured for your Trainer.
Check whether the fuse holder on the bottom of the
Trainer contains a fuse and whether the fuse is
Once you have finished, start the Trainer 1000.
Replace the fuse using only
G fuse packs 1.25 A/250 V, slow blow DIN 415671
as replacements and then start the Trainer 1000.
To change the language setting when you do not
speak the language currently in place, pr
Press down both of the soft keys while pressing
You will now be in the setup menu.
Keep pressing the left soft key until the frame is on
Press the right soft key. A submenu will open.
t soft key. You are now in the second
line-Press the right soft key. The language menu
Keep pressing the left soft key until you get to the
Press the right soft key to confirm your selection
rmation, please refer to the
description of the setup menu

Batteries inserted incorrectly
Remote cable is connected
The training electrodes are
If none of the above remedies work, the Trainer 1000 might have a fault. In this case, please contact the
manufacturer and the manufacturer’s customer services.
Insert batteries (2x AAA).
The remote control only works with batteries, even
cable. The cable is only a signal
cable, not a power cable.
Check whether the batteries are inserted the right
way round. Insert as specified in the battery
Insert fully charged batteries.
Since infrared does not use radio but optical
signals, the infrared transmitter (glass ball at the
front of the remote control) must be within the
visual range of the receiver (behind black disk) on
e remote control to a cable,
there is no need to take into account the infrared
Move closer to the Trainer or use cable.
Either connect at both ends or disconnect from bo
When the remote control cable is connected, the
infrared function is deactivated, which is why the
remote control will not work if it is connected at one
With some buttons, pressing them once will switch
function and pressing them again will switch
the function off again, e.g. connect electrodes, call
customer services, change batteries, pause.
Press this button only once.
When the Trainer 1000 is switched on,
electrodes must be connected and the
corresponding remote control button pressed ONCE
in order for the message “Connect electrodes” to
All of the functions will become active only after a
Not all of the buttons have functions in all setups.

Spanish, French, Portuguese, US English
German, French, British English, Italian, Dutch,
Danish, Finish, British English, Norwegian,
Spanish, British English, Greek, Italian,
English, Greek, Croatian, Hungarian,
Czech, British English, Lithuanian, Polish,
When placing an order, please also specify:
The required bag (soft or hard bag).
If no bag is specified, the Trainer 1000 is supplied with a soft bag.
The required charger (please see under accessories for available models)
The Trainer 1000 is always supplied with an AED and a manual mode by default. If you also require
simulation mode, please place a separate order for an ECG upgrade kit. If ordered together with a new Trainer
1000, the Trainer will be supplied with ECG mode ready installed.

ECG Upgrade Kit without cable
USB stick with language upg
control, remote control cable and batteries
Also suitable for use with the training electrodes
USB stick with upgrade, ECG remote control with
remote control cable and batteries.
To configure the upgrade, please state the
Trainer’s serial number and language version.
USB stick with upgrade, ECG remote control with
remote control cable and batteries, 3
To configure the upgrade, please state the
Trainer’s serial number and language version.
To configure the upgrade, please state the
Trainer’s serial number and required language
ECG remote control, remote control cable and
3-pin ECG cable for use with a Trainer 1000 with

LIFEPAK 1000 Software status while the Trainer 1000
software version is shown under Device Data in
the setup menu in the Trainer
Change to the LIFEPAK 1000 software that was taken
over to the Trainer 1000:
Removal of the confirmation analysis.
During the initial CPR, it is possible to spec
user will be prompted to perform CPR before the first
“Check pulse” can now also be set to “After second
NSA” in addition to the former options.
There is no need to protect the Trainer’s setup menu
up menu responds significantly faster

and menu navigation of the LIFEPAK® 1000,
but without delivery of shocks. So displayed
energies are only simulated.
360 joules. Default energy output settings
are 200, 300, 360 joules.
Wireless infrared signal: Range 4 meters
8.4 cm (3.3 in) x 4.6 cm (1.8 in) x 1.6
3 volts supplied by two AAA
55 g (.12 lb) with batteries
7-key membrane switch pad
labled for the following functions: shock
decision, electrode pad contact, motion,
service required, replace battery, pause and
Provides operating capability for
Provides operating capability
Medically trained personnel
can obtain the simulation of the ECG lead II
Allows user to configure the
On/Off, Shock, Menu, Two (2)
lower energy was unsuccessful
to auto analyze, auto analyze after first
before each analysis period
(post shock or after no shock
(if option set in setup menu)
Turns display on/off for AED mode
Audio alarm if the device
with an ECG display may be set up to allow
user to initiate a charge and shock without
cprMAX Technology Settings:
after first analysis regardless of analysis
decision. Can be set to OFF, 15, 30, 45, 60,
• Pre-shock-CPR–Allows for CPR while
device is charging. Can be set to OFF, 15 or
shockable rhythm after completion of Inital
shock CPR periods and prior to
Push to Shock promt (default Off)
allows for provision of CPR after each shock
(Always, After Every NSA,
ows device to promt for a pulse
check either after each shock, after every
NSA, or never promt for a pulse check
the setup whether a LIFEPAK(R) 1000 with
training scenarios which can be selected in
the setup or alternatively by remote
The Trainer 1000 is delivered
in different language versions. There are
several languages contained
which can be changed in the setup menu.
The following language combinations are
America: Spanish, French, Portuguese, US
• Central Europe: German, French, British
English, Italian, Dutch, US English
English, Norwegian, Swedish
• Southern Europe: Spanish, British English,
Greek, Italian, Portuguese
• Southeast Europe: British English, Greek,
Croatian, Hungarian, Serbian, Slovenian
• Eastern Europe: Czech, British English,
an, Polish, Russian, Slovenian
• Can also potentially be provided in other
Backlit LCD displays number of shocks
delivered, elapsed time, text and graphics
of heart rhythm and optional ECG
120 mm (4,7 in) x 89 mm (3,5 in)
Display Resolution: 320 x 240
dispaly, Display „…“ if heart rate is less than
bpm. Heart symbol flashes for each QRS
ECG information is a simulated signal from
Fully charged batteries run the device for at
least 3 hours. There is a warnin
device is turned off at low battery.
in charger is delivered with the
Device designed for low impact, low stress
environments. For use in temperature
controlled areas protected from
precipitation. Capable of
heavy useage and prolonged periods of
Operating Temperature: 0°C to 50
Emission: EN 55011 Class B
With bag and accessories:
Flame Protection according