TAKING PICTURES
IR Imaging
Controls the IR imaging and display modes.
• IMAGE TYPE: sets the IR d ispla y modes .
HIGHLIGHT: Over lays h ighes t temp erat ures o nly f or hot s top lo cati ng.
ATB: O ver lays A dapt ive T herm al Ble nding ar eas de ter mined b y ATB hig h
and ATB l ow set tin gs.
THERMAL ONLY: IR image o nly is d ispla yed.
VISIBLE ONLY: V ISI BLE im age onl y is dis playe d.
• COLOURISATION: Configures the IR image colour palette (as display ed on the
colour bar for reference). Industry Standard colour palettes are available:
Rainbo w, Amb er, Sepia , Gre y posi tiv e, Gre y nega tiv e, Iro n, Hot m etal .
• ATB HIGH: S elec ts the t empe ratu re thr esho ld abo ve whi ch the AT B ove rlay
is act ive .
ATB HIGH ACTIVATE: Enable or Disable.
TEMPERATURE: selects the temperature threshold.
• ATB L OW: S elec ts the t emper atu re thr esho ld belo w whi ch the AT B
over lay is a cti ve.
ATB LOW ACTIVATE: Enable or Disable.
TEMPERATURE: selects the temperature threshold.
• RANGE: Sele cts IR t empe ratu re ran ge.
Low : low ra nge -1 0°C t o 140 °C
High: h igh ran ge -10 °C to 3 80° C
• UNITS: Selects the global temperature units.
TEMPERATURE UNITS: Celsius, Fahrenheit, Kelvin.
THERMAL ANALYSIS
SPOT Configuration
Confi gures t he activ e tempe rature mea sure ment SPOT S & BOX ope rat ion.
• SPOTS/BOX : Config ures t he num ber of ac tiv e SPO TS in t he disp lay/B OX disp lay.
Options selectable: 1-SPOT, 2-SPOT, 3-SPOT, BOX.
• MOVE SPOT/BOX: Selec ts the A CTI VE SP OT or B OX cor ner to be m ove d by the
joys tick i n the pr evie w disp lay.
Opti ons sel ect able: 1 ,2, 3 to sel ect th e act ive SP OT (dis play ed in bol d in the
preview display).
• SPOT 1/B OX EMIS SIVI TY: Con figur es emi ssiv it y of SP OT1 a nd also t he BO X area .
Opti ons sel ect able: 0 .5 to 1 .0 in st eps on 0. 05.
• SPOT2 EMISSIVITY: Config ures e missi vit y of SPOT2 .
Opti ons sel ect able: 0 .5 to 1 .0 in st eps on 0. 05.
• SPOT3 EMISSIVITY: Config ures em issi vit y of SPOT3.
Opti ons sel ect able: 0 .5 to 1 .0 in st eps on 0. 05.
Global IR Parameter
Confi gures t he GL OBAL I R param eter s used b y the IR r adiom etr ic anal ysis .
Thes e parameter s are shared b y all th e SPO TS and t he BOX a rea ar e displ ayed
on the pr eview dis play.
• REFLECTED T: Configures the GLOBAL reflect ed temperature.
Options selectable: -10C to 380C.
• ATMOSPHERIC T: C onfigures the GLOBAL atmospher ic temperature.
Options selectable: -10C to 380C.
• HUMIDITY: Configures the GLOBA L humidity.
Opti ons sel ect able: 0 % to 10 0% in st eps of 1% .
• DISTANCE: Configu res t he GLO BAL di stan ce of th e objec t fr om the c amera .
Opti ons sel ect able: 1 m to 10 0m in st eps of 1 m.
www.GlobalTestSupply.com