Copper Mountain Technologies R140, R54, R60, RP180, R180 Operating Manual

1-Port VNA Series
R54
R140
R60
R180/RP180
Operating Manual
Software version 18.3.0
T A B L E O F C O N T E N T S
INTRODUCTION ....................................................................................................................... 8
SAFETY INSTRUCTIONS .......................................................................................................... 9
1. GENERAL OVERVIEW ....................................................................................................... 11
1.1 Description ........................................................................................................................................ 11
1.2 Specifications ................................................................................................................................... 11
1.3 Measurement Capabilities ........................................................................................................... 11
1.4 Principle of Operation .................................................................................................................. 16
2. PREPARATION FOR USE .................................................................................................. 18
2.1 General Information ...................................................................................................................... 18
2.2 Software Installation ..................................................................................................................... 20
2.3 Top Panel ........................................................................................................................................... 20
2.4 Test Port ............................................................................................................................................. 23
2.5 Mini B USB Port ............................................................................................................................... 23
2.6 External Trigger Signal Input Connector (R140 model only) ........................................ 24
2.7 External Reference Frequency Input Connector (R140 model only) .......................... 24
2.8 Reference Frequency Input/Output Connector (R60 and R180 model only) 24
2.9 External Trigger Signal Input/Output Connector (R60 and R180 model only) 25
3. GETTING STARTED .......................................................................................................... 26
3.1 Analyzer Preparation for Reflection Measurement ........................................................... 27
3.2 Analyzer Presetting ........................................................................................................................ 27
3.3 Stimulus Setting .............................................................................................................................. 28
3.4 IF Bandwidth Setting .................................................................................................................... 29
3.5 Number of Traces, Measured Parameter and Display Format Setting ...................... 30
3.6 Trace Scale Setting ........................................................................................................................ 31
3.7 Analyzer Calibration for Reflection Coefficient Measurement ..................................... 31
3.8 SWR and Reflection Coefficient Phase Analysis Using Markers .................................. 34
4. MEASUREMENT CONDITIONS SETTING ......................................................................... 36
4.1 Screen Layout and Functions ..................................................................................................... 36
4.1.1 Left and Right Softkey Menu Bars .................................................................................... 36
4.1.2 Top Menu Bar ........................................................................................................................... 37
4.1.3 Instrument Status Bar ........................................................................................................... 40
4.2 Channel Window Layout and Functions ................................................................................ 42
4.2.1 Channel Title Bar .................................................................................................................... 43
4.2.2 Trace Status Field ................................................................................................................... 43
4.2.3 Graph Area ................................................................................................................................. 46
4.2.4 Markers ........................................................................................................................................ 47
4.2.5 Channel Status Bar ................................................................................................................. 48
4.3 Quick Channel Setting Using Mouse....................................................................................... 49
4.3.1 Active Channel Selection ..................................................................................................... 49
2
4.3.2 Active Trace Selection .......................................................................................................... 50
4.3.3 Display Format Setting ......................................................................................................... 50
4.3.4 Trace Scale Setting................................................................................................................. 50
4.3.5 Reference Level Setting ....................................................................................................... 51
4.3.6 Marker Stimulus Value Setting .......................................................................................... 52
4.3.7 Switching between Start/Center and Stop/Span Modes ......................................... 52
4.3.8 Start/Center Value Setting .................................................................................................. 53
4.3.9 Stop/Span Value Setting ...................................................................................................... 53
4.3.10 Sweep Points Number Setting ........................................................................................ 54
4.3.11 IF Bandwidth Setting .......................................................................................................... 54
4.3.12 Power Level Setting ............................................................................................................ 55
4.4 Channel and Trace Display Setting ......................................................................................... 55
4.4.1 Setting the Number of Channel Windows ..................................................................... 55
4.4.2 Channel Activating ................................................................................................................. 56
4.4.3 Active Channel Window Maximizing ............................................................................... 57
4.4.4 Number of Traces Setting .................................................................................................... 58
4.4.5 Active Trace Selection .......................................................................................................... 59
4.5 Measurement Parameters Setting ............................................................................................ 62
4.5.1 S-Parameters............................................................................................................................. 62
4.5.2 Trace Format ............................................................................................................................. 62
4.5.3 Rectangular Format ............................................................................................................... 63
4.5.4 Polar Format ............................................................................................................................. 65
4.5.5 Smith Chart Format ................................................................................................................ 66
4.5.6 Data Format Setting ............................................................................................................... 68
4.6 Trigger Setting ................................................................................................................................. 69
4.6.1 External Trigger (except R54) ............................................................................................ 71
4.6.1.1 Point Feature ..................................................................................................................... 71
4.6.1.2 External Trigger Polarity .............................................................................................. 71
4.6.1.3 External Trigger Position ............................................................................................. 72
4.6.1.4 External Trigger Delay .................................................................................................. 73
4.6.2 Trigger Output (except R54/R140) .................................................................................. 76
4.6.2.1 Switching ON/OFF Trigger Output ........................................................................... 77
4.6.2.2 Trigger Output Polarity ................................................................................................. 78
4.6.2.3 Trigger Output Function ............................................................................................... 79
4.7 Scale Setting ..................................................................................................................................... 80
4.7.1 Rectangular Scale ................................................................................................................... 80
4.7.2 Rectangular Scale Setting ................................................................................................... 80
4.7.3 Circular Scale ............................................................................................................................ 81
4.7.4 Circular Scale Setting ............................................................................................................ 81
4.7.5 Automatic Scaling ................................................................................................................... 82
4.7.6 Reference Level Automatic Selection ............................................................................. 83
4.7.7 Electrical Delay Setting ........................................................................................................ 83
4.7.8 Phase Offset Setting .............................................................................................................. 84
4.8 Stimulus Setting .............................................................................................................................. 86
4.8.1 Sweep Type Setting ............................................................................................................... 86
4.8.2 Sweep Span Setting ............................................................................................................... 86
3
4.8.3 Sweep Points Setting ............................................................................................................ 87
4.8.4 Stimulus Power Setting ........................................................................................................ 88
4.8.5 Segment Table Editing ......................................................................................................... 88
4.9 Trigger Setting ................................................................................................................................. 91
4.10 Measurement Optimizing ......................................................................................................... 92
4.10.1 IF Bandwidth Setting .......................................................................................................... 92
4.10.2 Averaging Setting ................................................................................................................. 92
4.10.3 Smoothing Setting ............................................................................................................... 93
4.10.4 Trace Hold Function ............................................................................................................ 95
4.11 Cable Specifications .................................................................................................................... 96
4.11.1 Selecting the type of cable ............................................................................................... 96
4.11.2 Manually specify Velocity Factor and Cable Loss .................................................... 97
4.11.3 Editing table of cables........................................................................................................ 98
5. CALIBRATION AND CALIBRATION KIT ......................................................................... 100
5.1 General Information .................................................................................................................... 100
5.1.1 Measurement Errors ............................................................................................................. 100
5.1.2 Systematic Errors .................................................................................................................. 100
5.1.2.1 Directivity Error .............................................................................................................. 101
5.1.2.2 Source Match Error ....................................................................................................... 101
5.1.2.3 Reflection Tracking Error ........................................................................................... 101
5.1.3 Error Modeling ....................................................................................................................... 101
5.1.3.1 One-Port Error Model .................................................................................................. 101
5.1.4 Analyzer Test Port Defining .............................................................................................. 102
5.1.5 Calibration Steps ................................................................................................................... 103
5.1.6 Calibration Methods ............................................................................................................. 104
5.1.6.1 Normalization ................................................................................................................. 104
5.1.6.2 Expanded Normalization ............................................................................................ 105
5.1.6.3 Full One-Port Calibration ........................................................................................... 105
5.1.6.4 Waveguide Calibration ................................................................................................ 106
5.1.7 Calibration Standards and Calibration Kits ................................................................ 107
5.1.7.1 Types of Calibration Standards ................................................................................ 107
5.1.7.2 Calibration Standard Model ...................................................................................... 107
5.2 Calibration Procedures ............................................................................................................... 110
5.2.1 Calibration Kit Selection .................................................................................................... 110
5.2.2 Reflection Normalization ................................................................................................... 112
5.2.3 Full One-Port Calibration ................................................................................................... 114
5.2.4 Error Correction Disabling ................................................................................................. 116
5.2.5 Error Correction Status ....................................................................................................... 116
5.2.6 System Impedance Z0 ......................................................................................................... 117
5.2.7 Port Extension ........................................................................................................................ 117
5.2.8 Auto Port Extension ............................................................................................................. 119
5.3 Calibration Kit Management .................................................................................................... 121
5.3.1 Calibration Kit Selection for Editing ............................................................................. 121
5.3.2 Calibration Kit Label Editing ............................................................................................ 121
5.3.3 Predefined Calibration Kit Restoration ........................................................................ 122
4
5.3.4 Calibration Standard Editing ............................................................................................ 124
5.3.5 Calibration Standard Defining by S-Parameter File ................................................ 126
5.4 Automatic Calibration Module ................................................................................................ 128
5.4.1 Automatic Calibration Module Features ...................................................................... 129
5.4.2 Automatic Calibration Procedure ................................................................................... 130
6. MEASUREMENT DATA ANALYSIS ................................................................................. 131
6.1 Markers ............................................................................................................................................. 131
6.1.1 Marker Adding ........................................................................................................................ 132
6.1.2 Marker Deleting ..................................................................................................................... 133
6.1.3 Marker Stimulus Value Setting ........................................................................................ 133
6.1.4 Marker Activating .................................................................................................................. 134
6.1.5 Reference Marker Feature ................................................................................................. 134
6.1.6 Marker Properties .................................................................................................................. 136
6.1.6.1 Marker Coupling Feature ............................................................................................ 136
6.1.6.2 Marker Value Indication Capacity ........................................................................... 138
6.1.6.3 Multi Marker Data Display ......................................................................................... 138
6.1.6.4 Marker Data Alignment ............................................................................................... 139
6.1.6.5 Memory trace value display ...................................................................................... 140
6.1.7 Marker Position Search Functions .................................................................................. 141
6.1.7.1 Search for Maximum and Minimum ....................................................................... 142
6.1.7.2 Search for Peak............................................................................................................... 143
6.1.7.3 Search for Target Level ............................................................................................... 145
6.1.7.4 Search Tracking .............................................................................................................. 147
6.1.7.5 Search Range .................................................................................................................. 147
6.1.8 Marker Math Functions ....................................................................................................... 148
6.1.8.1 Trace Statistics ............................................................................................................... 149
6.1.8.2 Bandwidth Search ......................................................................................................... 151
6.1.8.3 Flatness ............................................................................................................................. 155
6.1.8.4 RF Filter Statistics ......................................................................................................... 157
6.2 Memory Trace Function ............................................................................................................. 160
6.2.1 Saving Trace into Memory ................................................................................................. 161
6.2.2 Memory Trace Deleting ...................................................................................................... 161
6.2.3 Trace Display Setting .......................................................................................................... 162
6.2.4 Memory Trace Math ............................................................................................................. 163
6.3 Fixture Simulation ........................................................................................................................ 164
6.3.1 Port Z Conversion ................................................................................................................. 164
6.3.2 De-embedding ........................................................................................................................ 165
6.3.3 Embedding ............................................................................................................................... 167
6.4 Time Domain Transformation .................................................................................................. 168
6.4.1 Time Domain Transformation Activating .................................................................... 170
6.4.2 Time Domain Transformation Span ............................................................................... 172
6.4.3 Time Domain Transformation Type ............................................................................... 174
6.4.4 Time Domain Transformation Window Shape Setting ........................................... 174
6.4.5 Frequency Harmonic Grid Setting .................................................................................. 175
6.5 Time Domain Gating ................................................................................................................... 177
5
6.5.1 Time Domain Gate Activating .......................................................................................... 178
6.5.2 Time Domain Gate Span..................................................................................................... 178
6.5.3 Time Domain Gate Type ..................................................................................................... 179
6.5.4 Time Domain Gate Shape Setting .................................................................................. 179
6.6 S-Parameter Conversion ............................................................................................................ 180
6.7 Limit Test ......................................................................................................................................... 182
6.7.1 Limit Line Editing ................................................................................................................. 183
6.7.2 Limit Test Enabling/Disabling ......................................................................................... 185
6.7.3 Limit Test Display Management ..................................................................................... 185
6.7.4 Limit Line Offset .................................................................................................................... 186
6.8 Ripple Limit Test........................................................................................................................... 187
6.8.1 Ripple Limit Editing ............................................................................................................. 187
6.8.2 Ripple Limit Enabling/Disabling ..................................................................................... 189
6.8.3 Ripple Limit Test Display Management ....................................................................... 189
7. CABLE LOSS MEASUREMENT ........................................................................................ 191
7.1 Cable Loss Measurement Algorithm ..................................................................................... 191
8. ANALYZER DATA OUTPUT ............................................................................................ 193
8.1 Analyzer State ................................................................................................................................ 193
8.1.1 Analyzer State Saving ......................................................................................................... 193
8.1.2 Analyzer State Recalling .................................................................................................... 195
8.1.3 Autosave and Autorecall State of Analyzer ................................................................ 195
8.2 Channel State ................................................................................................................................. 196
8.2.1 Channel State Saving .......................................................................................................... 196
8.2.2 Channel State Recalling ..................................................................................................... 197
8.3 Trace Data CSV File ..................................................................................................................... 199
8.3.1 CSV File Saving ...................................................................................................................... 199
8.4 Trace Data Touchstone File ...................................................................................................... 201
8.4.1 Touchstone File Saving ...................................................................................................... 202
8.4.2 Touchstone File Recalling ................................................................................................. 203
8.5 Graph Printing ................................................................................................................................ 205
8.5.1 Graph Printing Procedure .................................................................................................. 205
8.5.2 Quick saving program screen shot ................................................................................. 206
9. SYSTEM SETTINGS ......................................................................................................... 207
9.1 Analyzer Presetting ...................................................................................................................... 207
9.2 Program Exit ................................................................................................................................... 207
9.3 Analyzer System Data ................................................................................................................. 208
9.4 System Correction Setting ........................................................................................................ 208
9.5 User Interface Setting ................................................................................................................. 210
10. SPECIFICS OF WORKING WITH TWO OR MORE DEVICES ........................................ 212
10.1 Installation of additional software...................................................................................... 212
10.2 Connecting devices to a USB port ....................................................................................... 212
10.3 Synchronizing the work of analyzers ................................................................................. 212
10.4 Adding / removing devices ..................................................................................................... 214
6
10.5 Frequency adjustment of the internal generators ........................................................ 214
10.5.1 Manual frequency adjustment ....................................................................................... 215
10.5.2 Automatic frequency adjustment ................................................................................. 215
10.6 Features of analyzers calibration ........................................................................................ 216
10.6.1 Calibration Type .................................................................................................................. 217
10.6.2 Scalar Transmission Normalization............................................................................. 218
10.6.3 Expanded Transmission Normalization ..................................................................... 218
10.7 Selection of the measured S-parameters ......................................................................... 221
11. MAINTENANCE AND STORAGE ................................................................................... 222
11.1 Maintenance Procedures ......................................................................................................... 222
11.2 Instrument Cleaning ................................................................................................................. 222
11.3 Factory Calibration .................................................................................................................... 223
11.4 Storage Instructions .................................................................................................................. 223
Appendix 1 .......................................................................................................................... 224
7

INTRODUCTION

This Operating Manual represents design, specifications, overview of functions, and detailed operation procedure for the Vector Network Analyzer, to ensure effective and safe use of the technical capabilities of the instrument by the user.
Vector Network Analyzer operation and maintenance should be performed by qualified engineers with initial experience in operating of microwave circuits and PC.
The following abbreviations are used in this Manual:
PC – Personal Computer
DUT – Device Under Test
IF – Intermediate Frequency
CW – Continuous Wave
SWR – Standing Wave Ratio
VNA – Vector Network Analyzer
8

SAFETY INSTRUCTIONS

Carefully read through the following safety instructions before putting the Analyzer into operation. Observe all the precautions and warnings provided in this Manual for all the phases of operation, service, and repair of the Analyzer.
The VNA must be used only by skilled and specialized staff or thoroughly trained personnel with the required skills and knowledge of safety precautions.
The Analyzer complies with INSTALLATION CATEGORY I as well as POLLUTION DEGREE 2 in IEC61010–1.
The Analyzer is a MEASUREMENT CATEGORY I (CAT I) device. Do not use as CAT II, III, or IV device.
The Analyzer is tested in stand-alone condition or in combination with the accessories supplied by Copper Mountain Technologies against the requirement of the standards described in the Declaration of Conformity. If it is used as a system component, compliance of related regulations and safety requirements are to be confirmed by the builder of the system.
Never operate the Analyzer in the environment containing inflammable gasses or fumes.
Operators must not remove the cover or part of the housing. The Analyzer must not be repaired by the operator. Component replacement or internal adjustment must be performed by qualified maintenance personnel only.
Electrostatic discharge can damage your Analyzer when connected or disconnected from the DUT. Static charge can build up on your body and damage the sensitive circuits of internal components of both the Analyzer and the DUT. To avoid damage from electric discharge, observe the following:
Always use a desktop anti static mat under the DUT. Always wear a grounding wrist strap connected to the desktop anti static
mat via daisy-chained 1 MΩ resistor.
Connect the PC and the body of the DUT to protective grounding before
you start operation.
9
SAFETY INSTRUCTIONS
CAUTION
This sign denotes a hazard. It calls attention to a procedure, practice or condition that, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the instrument.
Note
This sign denotes important information. It calls attention to a procedure, practice, or condition that is essential for the user to understand.
10
Measured parameters
S
11,
Cable loss.
Number of measurement channels
Up to 4 logical channels. Each logical channel is represented on the screen as an individual channel window. A logical channel is defined by such stimulus signal settings as frequency range, number of test points, etc.
Data traces
Up to 4 data traces can be displayed in each channel window. A data trace represents one of such parameters of the DUT as magnitude and phase of S11, DTF, Cable loss.
Memory traces
Each of the 4 data traces can be saved into memory for further comparison with the current values.
Data display formats
SWR, Return loss, Cable loss, Phase, Expanded phase, Smith chart diagram, DTF SWR, DTF Return loss, Group delay.
Sweep setup features
Sweep type
Linear frequency sweep, logarithmic frequency sweep, and segment frequency sweep.
Measured points per sweep
Set by user from 2 to 100,001.

1. GENERAL OVERVIEW

1.1 Description

The VNA is designed for use in the process of development, adjustment and testing of antenna-feeder devices in industrial and laboratory facilities, as well as in field, including operation as a component of an automated measurement system. The Analyzer is designed for operation with external PC, which is not supplied with it.

1.2 Specifications

The specifications of Analyzer model can be found in its corresponding datasheet.

1.3 Measurement Capabilities

11
GENERAL OVERVIEW
Segment sweep
A frequency sweep within several user-defined segments. Frequency range, number of sweep points, IF bandwidth and measurement delay should be set for each segment.
Power settings
Two modes of output power level.
Power levels depending on device.
Sweep trigger
Trigger modes: continuous, single, hold.
Trace display functions
Trace type
Data trace, memory trace.
Trace math
Data trace modification by math operations: addition, subtraction, multiplication or division of measured complex values and memory data.
Auto scaling
Automatic selection of scale division and reference level value to have the trace most effectively displayed.
Electrical delay
Calibration plane moving to compensate for the delay in the test setup. Compensation for electrical delay in a DUT during measurements of deviation from linear phase.
Phase offset
Phase offset defined in degrees.
Accuracy enhancement
Calibration
Calibration of a test setup (which includes the Analyzer and adapter) significantly increases the accuracy of measurements. Calibration allows to correct the errors caused by imperfections in the measurement system: system directivity, source match, and tracking.
Calibration methods
The following calibration methods are available:
reflection normalization; full one-port calibration.
Reflection normalization
The simplest calibration method.
Full one-port calibration
Method of calibration that ensures high accuracy.
12
GENERAL OVERVIEW
Factory calibration
The factory calibration of the Analyzer allows performing measurements without additional calibration and reduces the measurement error after reflection normalization.
Mechanical calibration kits
The user can select one of the predefined calibration kits of various manufacturers or define own calibration kits.
Electronic calibration modules
Electronic calibration modules manufactured by COPPER MOUNTAIN TECHNOLOGIES make the Analyzer calibration faster and easier than traditional mechanical calibration.
Defining of calibration standards
Different methods of calibration standard defining are available:
standard defining by polynomial model; standard defining by data (S-parameters).
Error correction interpolation
When the user changes such settings as start/stop frequencies and number of sweep points, compared to the settings of calibration, interpolation or extrapolation of the calibration coefficients will be applied.
Marker functions
Data markers
Up to 16 markers for each trace. A marker indicates stimulus value and the measured value in a given point of the trace.
Reference marker
Enables indication of any maker values as relative to the reference marker.
Marker search
Search for max, min, peak, or target values on a trace.
Marker search additional features
User-definable search range. Functions of specific condition tracking or single operation search.
Setting parameters by markers
Setting of start, stop and center frequencies by the stimulus value of the marker and setting of reference level by the response value of the marker.
Marker math functions
Statistics, bandwidth, flatness, RF filter.
13
GENERAL OVERVIEW
Statistics
Calculation and display of mean, standard deviation and peak-to-peak in a frequency range limited by two markers on a trace.
Bandwidth
Determines bandwidth between cutoff frequency points for an active marker or absolute maximum. The bandwidth value, center frequency, lower frequency, higher frequency, Q value, and insertion loss are displayed.
Flatness
Displays gain, slope, and flatness between two markers on a trace.
RF filter
Displays insertion loss and peak-to-peak ripple of the passband, and the maximum signal magnitude in the stopband. The passband and stopband are defined by two pairs of markers.
Data analysis
Port impedance conversion
The function of conversion of the S-parameters measured at 50 Ω port into the values, which could be determined if measured at a test port with arbitrary impedance.
De-embedding
The function allows to exclude mathematically the effect of the fixture circuit connected between the calibration plane and the DUT from the measurement result. This circuit should be described by an S­parameter matrix in a Touchstone file.
Embedding
The function allows to simulate mathematically the DUT parameters after virtual integration of a fixture circuit between the calibration plane and the DUT. This circuit should be described by an S-parameter matrix in a Touchstone file.
S-parameter conversion
The function allows conversion of the measured S­parameters to the following parameters: reflection impedance and admittance, transmission impedance and admittance, and inverse S-parameters.
14
GENERAL OVERVIEW
Time domain transformation
The function performs data transformation from frequency domain into response of the DUT to radiopulse in time domain. Time domain span is set by the user arbitrarily from zero to maximum, which is determined by the frequency step. Windows of various forms allow better tradeoff between resolution and level of spurious sidelobes.
Time domain gating
The function mathematically removes unwanted responses in time domain what allows obtaining frequency response without influence from the fixture elements. The function applies reverse transformation back to frequency domain from the user-defined span in time domain. Gating filter types are: bandpass or notch. For better tradeoff between gate resolution and level of spurious sidelobes the following filter shapes are available: maximum, wide, normal and minimum.
Other features
Analyzer control
Using external personal computer via USB interface.
Familiar graphical user interface
Graphical user interface based on Windows operating system ensures fast and easy Analyzer operation by the user.
The software interface of Analyzers is compatible with modern tablet PCs and laptops.
Saving trace data
Saving the traces in graphical format and saving the data in Touchstone and *.csv (comma separated values) formats on the hard drive are available.
Remote control
COM/DCOM
Remote control via COM/DCOM. COM automation runs the user program on an Analyzer PC. DCOM automation runs the user program on a LAN-networked PC. Automation of the instrument can be achieved in any COM/DCOM-compatible language or environment, including Python, C++, C#, VB.NET, LabVIEW, MATLAB, Octave, VEE, Visual Basic (Excel) and many others.
Socket
Data transfer between the PC user and the computer that is connected to the device, can be also performed via Socket (TCP, port 5025).
15
GENERAL OVERVIEW

1.4 Principle of Operation

The Analyzer consists of the Analyzer Unit, some supplementary accessories, and personal computer (which is not supplied with the package). The Analyzer Unit is powered and controlled by PC via USB-interface. The block diagram of the Analyzer is represented in Figure 1.1.
The Analyzer Unit consists of a source oscillator, a local oscillator, a source power attenuator, a directional coupler and other components which ensure the Analyzer operation. The test port is the source of the test signal. The incident and reflected signals from the directional coupler are supplied into the mixers, where they are converted into IF, and are transferred further to the 2-channel receiver. The 2-channel receiver, after filtration, digitally encodes the signals and supplies them for further processing (filtration, phase difference measurement, magnitude measurement) into the signal processor. The filters for the IF are digital and have passband from 10 Hz to 30(100) kHz. The combination of the assemblies of directional couplers, mixers, and 2­channel receiver forms two similar signal receivers.
An external PC controls the operation of the components of the Analyzer. To fulfill the S-parameter measurement, the Analyzer supplies the source signal of the assigned frequency from test port to the DUT, then measures magnitude and phase of the signal reflected by the DUT, and after that compares these results to the magnitude and phase of the source signal.
16
Figure 1.1The VNA block diagram
17

2. PREPARATION FOR USE

2.1 General Information

Unpack the VNA and other accessories.
Connect the Analyzer to the PC using the USB Cable supplied in the package. Install the software (supplied on the flash drive) onto your PC. The software installation procedure is described below.
Warm-up the Analyzer for the time stated in its specifications.
Assemble the test setup using cables, connectors, fixtures, etc, which allow DUT connection to the Analyzer.
Perform calibration of the Analyzer. Calibration procedure is described in section
5.
18
PREPARATION FOR USE
Attention!
To avoid motherboard damage you must use USB cables supplied in the package or similar ones according to the specifications shown in Figure 2.1 and Figure 2.2 (for R180/RP180 only)
Figure 2.1 USB TYPE C TO C 2.0, 3A
Figure 2.2 USB TYPE C TO USB 2.0 A MALE, 3A
19
PREPARATION FOR USE
Minimal system requirements for the PC
WINDOWS 2000/XP/VISTA/7/8
1.5 GHz Processor
2 GB RAM
USB 2.0 High Speed
Flash drive contents
Setup_RVNA_vX.X.X.exe installer file
(X.X.X – program version number);
Driver folder contains the driver;
Doc folder contains documentation.
Driver installation
Connect the Analyzer to your PC via the supplied USB cable.
When you connect the Analyzer to the PC for the first time, Windows will automatically detect the new USB device and will open the USB driver installation dialog (Windows 2000/XP/VISTA/7/8).
In the USB driver installation dialog, click on Browse and specify the path to the driver files, which are contained in the Driver folder on the USB flash drive.
Program and related files installation
Run the Setup_RVNA_vX.X.X.exe installer file from the supplied USB flash drive. Follow the instructions of the installation wizard.

2.2 Software Installation

The software is installed to the external PC running under Windows operating system. The Analyzer is connected to the external PC via USB interface.
The supplied USB flash drive contains the following software:
The procedure of the software installation is performed in two steps. The first one is the driver installation. The second step comprises installation of the program, documentation and other related files.

2.3 Top Panel

The top panel view of Analyzers is represented in the figures below. The top panel is equipped with the READY/STANDBY LED indicator running in the following modes:
20
PREPARATION FOR USE
green blinking light – standby mode. In this mode the current
consumption of the device from the USB port is minimum;
green glowing light – normal device operation.
Figure 2.3 R140 top panel
Figure 2.4 R54 top panel
Figure 2.5 R60 top panel
21
Figure 2.6 R180 top panel
PREPARATION FOR USE
22
PREPARATION FOR USE

2.4 Test Port

The test port (type-N male 50 Ω) is intended for DUT connection. It is also used as a source of the stimulus signal and as a receiver of the response signal from the DUT.

2.5 Mini B USB Port

The mini B USB port view is represented in Figure 2.7, Figure 2.8, Figure 2.9 and Figure 2.10. It is intended for connection to USB port of the personal computer via the supplied USB cable.
Figure 2.7 Mini B USB port R54
Figure 2.8 Mini B USB port R140
23
Figure 2.9 Mini B USB port R60
PREPARATION FOR USE
Figure 2.10 Mini B USB port R180

2.6 External Trigger Signal Input Connector (R140 model only)

This connector allows the user to connect an external trigger source. Connector type is SMA female. TTL compatible inputs of 3 V to 5 V magnitude have up to 1 us pulse width. Input impedance is at least 10 kΩ.

2.7 External Reference Frequency Input Connector (R140 model only)

External reference frequency - see in its specifications, input level is 2 dBm ± 2 dB, input impedance at «Ref In» is 50 Ω. Connector type is SMA female.
2.8 Reference Frequency Input/Output Connector (R60 and R180
model only)
External reference frequency is 10 MHz, input level is 2 dBm ± 2 dB, input impedance is 50 Ohm. Output reference signal level is 3 dBm ± 2 dB into 50 Ohm impedance. Connector type is SMA female.
24
PREPARATION FOR USE
2.9 External Trigger Signal Input/Output Connector (R60 and R180
model only)
External Trigger Signal Input allows the user to connect an external trigger source. Connector type is SMA female. 3.3v CMOS TTL compatible inputs magnitude have at least 1 μs pulse width. Input impedance is at least 10 kOhm. The External Trigger Signal Output port can be used to provide trigger to an external device. The port outputs various waveforms depending on the setting of the Output Trigger Function: before frequency setup pulse, before sampling pulse, after sampling pulse, ready for external trigger, end of sweep pulse, measurement sweep.
25

3. GETTING STARTED

This section represents a sample session of the Analyzer. It describes the main techniques of measurement of reflection coefficient parameters of the DUT. SWR and reflection coefficient phase of the DUT will be analyzed.
The instrument sends the stimulus to the input of the DUT and then receives the reflected wave. Generally in the process of this measurement the output of the DUT should be terminated with a LOAD standard. The results of these measurements can be represented in various formats. The given example represents the measurement of SWR and reflection coefficient phase.
Typical circuit of DUT reflection coefficient measurement is shown in Figure 3.1.
Figure 3.1.
To measure SWR and reflection coefficient phases of the DUT in the given example you should go through the following steps:
Prepare the Analyzer for reflection measurement; Set stimulus parameters (frequency range, number of sweep points); Set IF bandwidth; Set the number of traces to 2, assign measured parameters and display
format to the traces;
Set the scale of the traces; Perform calibration of the Analyzer for reflection coefficient measurement; Analyze SWR and reflection coefficient phase using markers.
26
GETTING STARTED
Ready state features
The bottom line of the screen displays the instrument status bar. It should read Ready.
Note
You can operate either by the mouse or using a touch screen.

3.1 Analyzer Preparation for Reflection Measurement

Turn on the Analyzer and warm it up for the period of time stated in the specifications.
Connect the DUT to the test port of the Analyzer. Use the appropriate adapters for connection of the DUT input to the Analyzer test port. If the DUT input is type-N (female), you can connect the DUT directly to the port.

3.2 Analyzer Presetting

Before you start the measurement session, it is recommended to reset the Analyzer into the initial state. The initial condition setting is described in Appendix 1.
To restore the initial state of the Analyzer use the following softkeys in the right menu bar System > Preset.
Close the dialog by Ok.
27
GETTING STARTED

3.3 Stimulus Setting

After you have restored the preset state of the Analyzer, the stimulus parameters will be as follows: full frequency range of the instrument, sweep type is linear, number of sweep points is 201, power level is high, and IF is 10 kHz.
For the current example, set the frequency range from 100 MHz to 1 GHz.
To set the start frequency of the frequency range to 100 MHz use the following softkey in the right menu bar Stimulus .
Then select the Start field and enter 100 using the on-screen keypad. Complete the setting by Ok.
To set the stop frequency of the frequency range to 1 GHz select the Stop field and enter 1000 using the on-screen keypad. Complete the setting Ok. Close the Stimulus dialog by Ok.
28
GETTING STARTED
Note
You can also select the IF bandwidth by double clicking on the required value in the IFBW. The dialog will close automatically.

3.4 IF Bandwidth Setting

For the current example, set the IF bandwidth to 3 kHz.
To set the IF bandwidth to 3 kHz use the following softkey in the left menu bar Average.
Then select the IFBW field in the Average dialog.
To set the IF bandwidth in the IFBW dialog use the following softkeys 3 kHz > Ok.
29
GETTING STARTED
3.5 Number of Traces, Measured Parameter and Display Format
Setting
In the current example, two traces are used for simultaneous display of the two parameters (SWR and reflection coefficient phase).
To add the second trace use the following softkeys in the right menu bar Trace >
Add trace.
The added trace automatically becomes active. The active trace is highlighted in the list and on the graph.
To select the trace display format click on Format.
Set the Phase format by Phase > Ok.
To scroll up and down the formats list click on the list field and drag the mouse up or down accordingly.
30
GETTING STARTED
Note
To activate a trace use the softkey Active Trace.
To select the first trace display format click on Active Trace and on Format. In the Format dialog use the following softkeys SWR > Ok.
Close the dialogs by Ok.

3.6 Trace Scale Setting

For a convenience in operation, change the trace scale using automatic scaling function.
To set the scale of the active trace by the autoscaling function use the following softkeys in the right menu bar Scale > Auto Scale > Ok.
The program will automatically set the scale for the best display of the active trace.
If you use the softkeys Scale > Auto Scale All > Ok, the program will automatically set the scale for all traces.

3.7 Analyzer Calibration for Reflection Coefficient Measurement

Calibration of the whole measurement setup, which includes the Analyzer and other devices, supporting connection to the DUT, allows to enhance considerably the accuracy of the measurement.
31
GETTING STARTED
To perform full 1-port calibration, you need to prepare the kit of calibration standards: OPEN, SHORT and LOAD. Every kit has its description and specifications of the standards.
To perform proper calibration, you need to select the correct kit type in the program. In the process of full 1-port calibration, connect calibration standards to the test port one after another, as shown in Figure 3.2.
Figure 3.2. Full 1-port calibration circuit
In the current example Agilent 85032B/E calibration kit is used.
To select the calibration kit use the following softkeys in the left menu bar Calibration > Calibration Kit.
32
GETTING STARTED
Then select the required kit from the Calibration Kits list and complete the setting by Ok.
To perform full 1-port calibration you should execute measurements of the three standards. After that the table of calibration coefficients will be calculated and saved into the memory of the Analyzer. Before you start calibration, disconnect the DUT from the Analyzer.
To perform full 1-port calibration use the following softkey in the left menu bar Calibration.
33
GETTING STARTED
Connect an OPEN standard and click Open.
Connect a SHORT standard and click Short.
Connect a LOAD standard and click Load.
After clicking any of the Open, Short, or Load softkeys, wait until the calibration procedure is completed.
To complete the calibration and calculate the table of calibration coefficients click Apply. Then re-connect the DUT to the Analyzer test port.

3.8 SWR and Reflection Coefficient Phase Analysis Using Markers

This section describes how to determine the measurement values at three frequency points using markers. The Analyzer screen view is shown in Figure 3.3. In the current example, a reflection standard of SWR = 1.2 is used as a DUT.
34
GETTING STARTED
Figure 3.3 SWR and reflection coefficient phase measurement example
To enable a new marker use the following softkeys in the left menu bar Marker >
Add Marker.
Double click on the marker in the Marker List to activate the on-screen keypad and enter the marker frequency value.
Complete the setting by Ok.
35

4. MEASUREMENT CONDITIONS SETTING

4.1 Screen Layout and Functions

The screen layout is represented in Figure 4.1. In this section you will find detailed description of the softkey menu bars and instrument status bar. The channel windows will be described in the next section.
Figure 4.1 Analyzer screen layout

4.1.1 Left and Right Softkey Menu Bars

The softkey menu bars in the left and right parts of the screen are the main menu of the program. Each softkey represents one of the submenus. The menu system is multilevel and allows to access to all the functions of the Analyzer.
You can manipulate the menu softkeys by the mouse or using a touch screen.
36
MEASUREMENT CONDITIONS SETTING
Note
Type of saving is set by the user in the dialog form Save type (see section 8.1).
On-screen alphanumeric keypads also support data entering from external PC keyboard. Besides, you can navigate the menu by «Up Arrow», «Down Arrow»,«Enter», «Esc» keys on the external keyboard.
To expand the menu bar click on it and drag the cursor to the right or to the left accordingly. To collapse the menu bar click on it and drag the cursor to the right or to the left accordingly.
You can also click the softkey Menu Size to expand or to collaps the menu bar.

4.1.2 Top Menu Bar

The menu bar contains the functions of the most frequently used softkeys.
The softkey Recall State allows to recall the state from a file of Analyzer state (see section 8.1.2).
The softkey Save State allows to save the Analyzer state (see section 8.1.1).
37
MEASUREMENT CONDITIONS SETTING
The softkey Save Data allows to save the trace data in CSV format (see section
8.3.1).
The softkeys Add Marker and Delete Marker add and delete markers on the trace respectively.
The softkey Reference Marker allows to add the reference marker on the trace. To delete the reference marker reclick this key.
The softkeys Add Trace and Delete Trace add and delete traces respectively.
The softkey Memory trace enables trace saving into memory (see section 6.2).
38
MEASUREMENT CONDITIONS SETTING
The softkey Data Math pops up the corresponding dialog form for choosing the math operation type between data traces and memory traces (see section 6.2.4).
The softkey Auto Scale allows to define the trace scale automatically so that the trace of the measured value could fit into the graph entirely (see section 3.6).
The softkey Auto Ref Value executes the automatic selection of the reference level (see section 4.7.6).
The softkey Auto Scale All allows to define the trace scale automatically for all traces (see section 3.6).
39
MEASUREMENT CONDITIONS SETTING
Field
Description
Message
Instrument Status
DSP status
Not Ready
No communication between DSP and PC.
Loading
DSP program is loading.
Ready
DSP is running normally.
Standby
DSP is in energy saving standby mode.
Sweep status
Measure
Continuous sweep.
Hold
A sweep is on hold.
Factory calibration error
System Cal
Failure
ROM error of system calibration.
The softkey Inverse Color allows to change the interface color.

4.1.3 Instrument Status Bar

Figure 4.2 Instrument status bar
The instrument status bar is located at the bottom of the screen. It can contain the following messages (see Table 4.1).
Table 4.1 Messages in the instrument status bar
40
MEASUREMENT CONDITIONS SETTING
Field
Description
Message
Instrument Status
Error correction status
Correction Off
Error correction disabled by the user1.
System correction status
System
Correction Off
System correction disabled by the user.
1 Disabling of error correction does not affect factory calibration.
41
MEASUREMENT CONDITIONS SETTING
Note
The calibration parameters are applied to the whole Analyzer and affect all the channel windows.

4.2 Channel Window Layout and Functions

The channel windows display the measurement results in the form of traces and numerical values. The screen can display up to 4 channel windows simultaneously. Each window has the following parameters:
Frequency range; Sweep type; Number of points; IF bandwidth.
Physical analyzer processes the logical channels in succession.
In turn each channel window can display up to 4 traces of the measured parameters. General view of the channel window is represented in Figure 4.3.
Figure 4.3 Channel window
42
MEASUREMENT CONDITIONS SETTING
Note
To edit the channel title click on the softkey Edit to recall the on­screen keypad.

4.2.1 Channel Title Bar

The channel title feature allows you to enter your comment for each channel window.
To show/hide the channel title bar use the softkey Display.
Click on Caption field in the opened dialog.

4.2.2 Trace Status Field

Figure 4.4 Trace status field
43
MEASUREMENT CONDITIONS SETTING
Note
Using the trace status field you can easily modify the trace parameters by the mouse.
The trace status field displays the name and parameters of a trace. The number of lines in the field depends on the number of traces in the channel.
Each line contains the data on one trace of the channel:
Trace name from Tr1 to Tr4. The active trace name is highlighted in
inverted color;
Display format, e.g. Return Loss; Trace scale in measurement units per division, e.g. 0.5 dB/; Reference level value, e.g. -20.0 dB; Trace status is indicated as symbols in square brackets (see Table 4.2).
44
Table 4.2 Trace status symbols definition
Status
Symbols
Definition
Error Correction
RO
OPEN response calibration
RS
SHORT response calibration
F1
Full 1-port calibration
Data Analysis
Z0
Port impedance conversion
Dmb
De-embedding
Emb
Embedding
Pxt
Port extension
Math Operations
D+M
Data + Memory
D-M
Data - Memory
D*M
Data * Memory
D/M
Data / Memory
Maximum Hold
Max
Hold of the trace maximum between repeated measurements
Electrical Delay
Del
Electrical delay other than zero
Phase Offset
PhO
Phase offset value other then zero
Smoothing
Smo
Trace smoothing
Gating
Gat
Time domain gating
Conversion
Zr
Reflection impedance
Yr
Reflection admittance
1/S
S-parameter inversion
Conj
Conjugation
MEASUREMENT CONDITIONS SETTING
45
MEASUREMENT CONDITIONS SETTING
Status
Symbols
Definition
Trace display
Dat
Data trace
Mem
Memory trace
D&M
Data and memory traces
Off
Data and memory traces - off

4.2.3 Graph Area

The graph area displays the traces and numeric data (see Figure 4.5).
Figure 4.5 Graph area
46
MEASUREMENT CONDITIONS SETTING
Note
Using the graticule labels, you can easily control all the trace parameters by the mouse.
The graph area contains the following elements:
Vertical graticule label displays the vertical axis numeric data for the active
trace;
Horizontal graticule label displays stimulus axis numeric data (frequency,
time, or distance);
Reference level position indicates the reference level position of the trace; Markers indicate the measured values in different points on the active
trace. You can enable display of the markers for all the traces simultaneously;
Marker functions: statistics, bandwidth, flatness, RF filter; Trace number allows trace identification in the channel window; Current stimulus position indication appears when sweep duration exceeds
1 sec.

4.2.4 Markers

The markers indicate the stimulus values and the measured values in selected points of the trace (see Figure 4.6).
Figure 4.6 Markers
47
MEASUREMENT CONDITIONS SETTING
Symbol
Definition
--
No calibration data. No calibration was performed.
Cor
Error correction is enabled. The stimulus settings are the same for the measurement and the calibration.
C?
Error correction is enabled. The stimulus settings are not the same for the measurement and the calibration. Interpolation is applied.
The markers are numbered from 1 to 16. The reference marker is indicated with R symbol. The active marker is indicated in the following manner: its number is highlighted in inverse color, the stimulus indicator is fully colored.

4.2.5 Channel Status Bar

The channel status bar is located in the bottom part of the channel window (see Figure 4.7)
Figure 4.7 Channel status bar
The channel status bar contains the following elements:
Stimulus start field allows to display and enter the start frequency. This
field can be switched to indication of stimulus center frequency, in this case the word Start will change to Center;
Sweep points field allows to display and enter the number of sweep points.
The number of sweep points can have the following values: 2 - 100001;
IF bandwidth field allows to display and set the IF bandwidth. The values
can be set from 10 Hz to 30 kHz (100 kHz);
Power level field allows to display and enter the port output power; Stimulus stop field allows to display and enter the stop frequency . This
field can be switched to indication of stimulus span, in this case the word Stop will change to Span;
Error correction field displays the integrated status of error correction for
S-parameter traces. The values of this field are represented in Table 4.3.
Table 4.3 Error correction field
48
MEASUREMENT CONDITIONS SETTING
Symbol
Definition
C!
Error correction is enabled. The stimulus settings are not the same for the measurement and the calibration. Extrapolation is applied.
Off
Error correction is turned off.
Note
The manipulations described in this section will help you to perform the most frequently used settings only. All the channel functions can be accessed via the softkey menu.

4.3 Quick Channel Setting Using Mouse

This section describes the manipulations, which will enable you to set the channel parameters of R140 fast and easy. When you move a mouse pointer in the channel window field where a channel parameter can be changed, the mouse pointer will change its form and a prompt field will appear.

4.3.1 Active Channel Selection

You can select the active channel window when two or more channel windows are open. The border line of the active window will be highlighted (see Figure
4.8). To activate a channel click in its window.
Figure 4.8 Active channel window display
49
MEASUREMENT CONDITIONS SETTING

4.3.2 Active Trace Selection

You can select the active trace if the active channel window contains two or more traces.
The active trace name will be highlighted in inverted color. In the example given it is Tr2. To activate a trace click on the required trace or its status line.

4.3.3 Display Format Setting

To select the trace display format click on the format name in the trace status line.
Select the required format in the Format dialog and complete the setting by Ok.

4.3.4 Trace Scale Setting

50
MEASUREMENT CONDITIONS SETTING
To select the trace scale click in the trace scale field of the trace status line.
Enter the required numerical value using the on-screen keypad and complete the setting by Ok.

4.3.5 Reference Level Setting

To set the value of the reference level click on the reference level field in the trace status line.
Enter the required numerical value using the on-screen keypad and complete the setting by Ok.
51
MEASUREMENT CONDITIONS SETTING

4.3.6 Marker Stimulus Value Setting

The marker stimulus value can be set by dragging the marker or by entering the value from the on-screen keypad.
To drag the marker, move the mouse pointer to one of the marker indicators.
The marker will become active, and a pop-up hint with its name will appear near the marker. The marker can be moved either by dragging its indicator or its hint area.
To enter the numerical value of the stimulus in the marker data click on the stimulus value. Then enter the required value using the on-screen keypad.

4.3.7 Switching between Start/Center and Stop/Span Modes

To switch between the modes Start/Center and Stop/Span click in the respective field of the channel status bar.
Label Start will be replaced by Center, and label Stop will be replaced by Span.
52
MEASUREMENT CONDITIONS SETTING

4.3.8 Start/Center Value Setting

To enter the Start/Center numerical values click on the respective field in the channel status bar.
Then enter the required value using the on-screen keypad.

4.3.9 Stop/Span Value Setting

To enter the Stop/Span numerical values click on the respective field in the channel status bar.
Then enter the required value using the on-screen keypad.
53
MEASUREMENT CONDITIONS SETTING

4.3.10 Sweep Points Number Setting

To enter the number of sweep points click in the respective field of the channel status bar.
Select the required value in the Points dialog and complete the setting by Ok.

4.3.11 IF Bandwidth Setting

To set the IF bandwidth click in the respective field of the channel status bar.
Select the required value in the IFBW dialog and complete the setting by Ok.
54
MEASUREMENT CONDITIONS SETTING

4.3.12 Power Level Setting

To set the output power level click in the respective field of the channel status bar.
This way you can switch between high and low power settings.

4.4 Channel and Trace Display Setting

The Analyzer supports 4 channels, which allows measurements with different stimulus parameter settings. The parameters related to a logical channel are listed in Table 4.4

4.4.1 Setting the Number of Channel Windows

A channel is represented on the screen as an individual channel window. The screen can display from 1 to 4 channel windows simultaneously. By default one channel window is opened.
The program supports three options of the channel window layout: one channel, two channels, and four channels. The channels are allocated on the screen according to their numbers from left to right and from top to bottom. If there are more than one channel window on the screen, one of them is selected as active. The border line of the active window will be highlighted in inverted color.
55
MEASUREMENT CONDITIONS SETTING
Note
For each open channel window, you should set the stimulus parameters and make other settings.
Before you start channel parameter setting or calibration, you need to select this channel as active.
To set the number of channel windows displayed on the screen use the following softkey in the right menu bar Channels. Then select the softkey with the required number and layout of the channel windows.
In the Active Channel field, you can select the active channel. The repeated clicking will switch the numbers of all channels.
The measurements are executed for open channel windows in succession.

4.4.2 Channel Activating

Before setting channel parameters, you need to activate the channel.
56
MEASUREMENT CONDITIONS SETTING
To activate the channel use the following softkeys in the right menu bar Channels
> Active Channel.
Active Channel field allows viewing the numbers of all channels from 1 to 4.
Select the required number of the active channel.
To activate a channel, you can also click on its channel window.

4.4.3 Active Channel Window Maximizing

When there are several channel windows displayed, you can temporarily maximize the active channel window to full screen size.
The other channel windows will be hidden, and this will interrupt the measurements in those channels.
57
MEASUREMENT CONDITIONS SETTING
Note
Channel maximizing function can be controlled by a double mouse click on the channel.
To enable/disable active channel maximizing function use the following softkeys Channel > Maximize Channel.

4.4.4 Number of Traces Setting

Each channel window can contain up to 4 different traces. Each trace is assigned the display format, scale and other parameters. The parameters related to a trace are listed in Table 4.5.
The traces can be displayed in one graph, overlapping each other, or in separate graphs of a channel window. The trace settings are made in two steps: trace number setting and trace layout setting in the channel window. By default a channel window contains one trace. If you need to enable two or more traces, set the number of traces as described below.
58
MEASUREMENT CONDITIONS SETTING
To add a trace use the following softkeys in the right menu bar Trace > Add Trace.
To delete a trace use the following softkeys in the right menu bar Trace > Delete Trace.
All the traces are assigned their individual names, which cannot be changed. The trace name contains its number. The trace names are as follows: Tr1, Tr2 ... Tr4.
Each trace is assigned some initial settings: measured parameter, format, scale and color, which can be modified by the user.
By default the display format for all the traces is set to Return loss (dB).
By default the scale is set to 10 dB, reference level value is set to 0 dB, reference level position is in the middle of the graph.
The trace color is determined by its number.

4.4.5 Active Trace Selection

Trace parameters can be entered for the active trace. Active trace belongs to the active channel, and its name is highlighted in inverted color. You have to select the active trace before setting the trace parameters.
59
MEASUREMENT CONDITIONS SETTING
Note
A trace can be activated by clicking on the trace status bar in the graphical area of the program
To select the active trace use the softkeys in the right menu bar Trace.
Click the Active Trace to select the trace you want to assign the active.
60
Table 4.4 Channel parameters
N
Parameter Description
1
Sweep Range
2
Number of Sweep Points
3
IF Bandwidth
N
Parameter Description
1
Display Format
2
Reference Level Scale, Value and Position
3
Electrical Delay, Phase Offset
4
Memory Trace
5
Markers
6
Parameter Transformation
Table 4.5 Trace parameters
MEASUREMENT CONDITIONS SETTING
61
MEASUREMENT CONDITIONS SETTING
n Port at waveincident
m Port at wavedtransmitte
=S
mn

4.5 Measurement Parameters Setting

4.5.1 S-Parameters

For high-frequency network analysis the following terms are applied: incident, reflected and transmitted waves, transferred in the circuits of the setup (see Figure 4.9).
Figure 4.9
Measurement of magnitude and phase of incident, reflected and transmitted signals allow to determin the S-parameters (scattered parameters) of the DUT. An S-parameter is a relation between the complex magnitudes of the two waves:
R140 Analyzer has one measurement port which operates as a signal source and as a reflected signal receiver. That is why the Analyzer allows measuring only S11 parameter.

4.5.2 Trace Format

The Analyzer offers the display of the measured S-parameters on the screen in three formats:
rectangular format; polar format; Smith chart format.
62
MEASUREMENT CONDITIONS SETTING
bj+a=S
Format Type
Description
Label
Data Type (Y-axis)
Measurement Unit
(Y-axis)
Logarithmic
Magnitude
Log
Mag
S-parameter logarithmic
magnitude:
,
22
baS
Decibel (dB)

4.5.3 Rectangular Format

In this format, stimulus values are plotted along X-axis and the measured data are plotted along Y-axis (see Figure 4.10).
Figure 4.10 Rectangular format
To display S-parameter complex value along Y-axis, it should be transformed into a real number. Rectangular formats involve various types of transformation of an S-parameter
, where:
a – real part of S-parameter complex value;
b – imaginary part of S-parameter complex value.
There are nine types of rectangular formats depending on the measured value plotted along Y-axis (see Table 4.6).
Table 4.6 Rectangular formats
63
MEASUREMENT CONDITIONS SETTING
Format Type
Description
Label
Data Type (Y-axis)
Measurement Unit
(Y-axis)
Voltage
Standing
Wave Ratio
SWR
Abstract number
Phase
Phase
S-parameter phase from
–180to +180
a
b
arctg
180
Degree (°)
Expanded
Phase
Expand
Phase
S-parameter phase,
measurement range expanded
to from below –180 to over
+180
Degree (°)
Group
Delay
Group
Delay
Signal propagation delay within
the DUT:
d
d
,
a
b
arctg
,
f2
Second (sec.)
Linear
Magnitude
Lin Mag
S-parameter linear magnitude: Abstract number
Real Part
Real
S-parameter real part:
)(Srea
Abstract number
Imaginary Part
Imag
S-parameter imaginary part:
)(Simb
Abstract number
Cable Loss
Cable
Loss
Decibel (dB)
64
MEASUREMENT CONDITIONS SETTING
Format Type
Description
Label
Data Displayed by Marker
Measurement Unit
(Y-axis)
Linear
Magnitude
and Phase
Polar (Lin)
S-parameter linear magnitude
Abstrac number
S-parameter phase
Degree
Logarithmic
Magnitude
and Phase
Polar (Log)
S-parameter logarithmic
magnitude
Decibel (dB)
S-parameter phase
Degree
Real and
Imaginary
Parts
Polar
(Re/Im)
S-parameter real part
Abstract number
S-parameter imaginary part
Abstract number

4.5.4 Polar Format

Polar format represents the measurement results on the pie chart (see Figure
4.11). The distance to a measured point from the graph center corresponds to the magnitude of its value. The counterclockwise angle from the positive horizontal axis corresponds to the phase of the measured value.
Figure 4.11 Polar format
The polar graph does not have a frequency axis, so frequency will be indicated by the markers. There are three types of polar formats depending on the data displayed by the marker. The traces will remain the same on all the graphs.
Table 4.7 Polar formats
65
MEASUREMENT CONDITIONS SETTING
Format Type
Description
Label
Data Displayed by Marker
Measurement Unit
(Y-axis)
Linear
Magnitude and
Phase
Smith (Lin)
S-parameter linear
magnitude
Abstract number
S-parameter phase
Degree
Logarithmic
Magnitude and
Phase
Smith (Log)
S-parameter logarithmic
magnitude
Decibel (dB)
S-parameter phase
Degree
Real and
Imaginary
Parts
Smith
(Re/Im)
S-parameter real part
Abstract number
S-parameter imaginary part
Abstract number

4.5.5 Smith Chart Format

Smith chart format is used for representation of impedance values for DUT reflection measurements. In this format, the trace has the same points as in polar format.
Figure 4.12 Smith chart format
Smith chart format does not have a frequency axis, so frequency will be indicated by the markers. There are five types of Smith chart formats depending on the data displayed by the marker. The traces will remain the same on all the graphs.
Table 4.8 Smith chart format
66
MEASUREMENT CONDITIONS SETTING
Format Type
Description
Label
Data Displayed by Marker
Measurement Unit
(Y-axis)
Complex
Impedance
(at Input)
Smith
(R + jX)
Resistance at input:
)(
inp
ZreR
,
S
S
ZZ
inp
1
1
0
Ohm (Ω)
Reactance at input:
)(
inp
ZimX
Ohm (Ω)
Equivalent capacitance or
inductance:
0,
1
X
X
C
0, X
X
L
Farad (F)
Henry (H)
Complex
admittance
(at Input)
Smith
(G + jB)
Conductance at input:
)(
inp
YreG
,
S
S
Z
Y
inp
1
11
0
Siemens (S)
Susceptance at input:
)(
inp
YimB
Siemens (S)
Equivalent capacitance or
inductance:
0, B
B
C
0,
1
B
B
L
Farad (F)
Henry (H)
Z0 – test port impedance. Z0 setting is described in section 5.2.6
67
MEASUREMENT CONDITIONS SETTING

4.5.6 Data Format Setting

You can select the format for each trace of the channel individually. Before you set the format, first activate the trace.
To set the trace display format use the following softkey in the right menu bar Trace.
In the Trace dialog select the required trace from Active Trace and click on Format.
Then select the required format in the Format dialog. Complete the setting by Ok.
68
MEASUREMENT CONDITIONS SETTING

4.6 Trigger Setting

The trigger mode determines the sweep actuation of the channel at a trigger signal detection. A channel can operate in one of the following three trigger modes:
Continuous – a sweep actuation occurs every time a trigger signal is
detected;
Single – one sweep actuation occurs with trigger signal detection after the
mode has been enabled; after the sweep is complete the channel modes changes to hold;
Hold – sweep actuation is off in the channel, trigger signals do not affect
the channel.
The trigger signal applies to the whole Analyzer and controls the trigging of all the channels in the following manner. If more than one channel window are open, the trigger activates successive measurements of all the channels which are not in hold mode. Before measurement of all channels is complete, all additional triggers are ignored. When measurement of all the channels is complete, if there is as least one channel in continuous trigger mode, the Analyzer will enter waiting for a trigger state.
The trigger source can be selected by the user from the following four available options:
Internal – the next trigger signal is generated by the Analyzer on
completion of each sweep;
External – the external trigger input is used as a trigger signal source
(except R54);
Bus – the trigger signal is generated by a command communicated from an
external computer from a program controlling the Analyzer via COM/DCOM.
To set the trigger mode, use the following softkeys Trigger > Trigger Mode.
69
Then select the required trigger mode:
Hold Single Continuous
MEASUREMENT CONDITIONS SETTING
To set the trigger source, use the following softkeys Trigger > Trigger Source.
Then select the required trigger source:
Internal External Bus
70
MEASUREMENT CONDITIONS SETTING

4.6.1 External Trigger (except R54)

4.6.1.1 Point Feature
By default the external trigger initiates a sweep measurement upon every trigger event (See Figure 4.13 a, b). For the external trigger source, the point trigger feature instead initiates a point measurement upon each trigger event (See Figure
4.13 c, d).
To enable the point trigger feature for external trigger source, use the following softkeys Trigger > Trigger Input > Event { On Sweep | On Point }.
4.6.1.2 External Trigger Polarity
71
MEASUREMENT CONDITIONS SETTING
To select the external trigger polarity, use the following softkeys Trigger > Trigger
Input > Polarity { NegativeEdge | Positive Edge }.
4.6.1.3 External Trigger Position
The external trigger position selects the position when Analyzer expects the external trigger signal:
Before sampling, when the frequency of the stimulus port have been set.
The frequency change of the stimulus port begins after sampling (See Figure 4.13 a, c).
Before the frequency setup and subsequent measurement. The frequency
change of the stimulus port begins when the external trigger arrives (See Figure 4.13 b, d).
Depending on the Point Feature settings the external trigger is expected before each point or before the first point of the full sweep cycle.
72
MEASUREMENT CONDITIONS SETTING
To select external trigger polarity, use the following softkeys: Trigger > Trigger Input > Position { Before Sampling | Before Setup }.
4.6.1.4 External Trigger Delay
The external trigger delay sets the response delay with respect to the external trigger signal (see Figure 4.13). The delay value has range from 0 to 100 sec with resolution 0.1 μsec.
To set the external trigger delay, use the following softkeys: Trigger > Trigger
Input > Delay.
73
MEASUREMENT CONDITIONS SETTING
74
MEASUREMENT CONDITIONS SETTING
Figure 4.13 External Trigger
75
MEASUREMENT CONDITIONS SETTING

4.6.2 Trigger Output (except R54/R140)

The trigger output outputs various waveforms depending on the setting of the Output Trigger Function:
Before frequency setup pulse; Before sampling pulse; After sampling pulse; Ready for external trigger; End of sweep pulse; Measurement sweep.
Figure 4.14 Trigger Output (except Ready for Trigger)
Figure 4.15 Trigger Output (Ready for Trigger)
76
MEASUREMENT CONDITIONS SETTING
Note
When the Ready for Trigger function of the trigger output is selected the trigger source must be set to external to enable the output trigger.
4.6.2.1 Switching ON/OFF Trigger Output
To enable/disable the trigger output, use the following softkeys Trigger > Trigger
Output > Enable Out.
77
MEASUREMENT CONDITIONS SETTING
4.6.2.2 Trigger Output Polarity
To select the polarity of the trigger output, use the following softkeys Trigger >
Trigger Output > Polarity { NegativeEdge | Positive Edge }.
78
MEASUREMENT CONDITIONS SETTING
4.6.2.3 Trigger Output Function
To select the function of the trigger output (See Figure 4.14, Figure 4.15), use the following softkeys Trigger > Trigger Output > Position { Before Setup |
BeforeSampling | After Sampling | Ready for Trigger | Sweep End | Measurement }.
79
MEASUREMENT CONDITIONS SETTING

4.7 Scale Setting

4.7.1 Rectangular Scale

For rectangular format you can set the following parameters (see Figure 4.16):
Trace scale; Reference level value; Reference level position; Number of scale divisions.
Figure 4.16 Rectangular scale

4.7.2 Rectangular Scale Setting

You can set the scale for each trace of a channel. Before you set the scale, first activate the trace.
To set the scale of a trace use the following softkey in the right menu bar Scale.
80
MEASUREMENT CONDITIONS SETTING
Then select the Scale field and enter the required value using the on-screen keypad.
To set the reference level select the Ref. Value field and enter the required value using the on-screen keypad.
To set the position of the reference level select the Ref. Position field and enter the required value using the on-screen keypad.
To set the number of trace scale divisions1 select the Divisions field and enter the required value using the on-screen keypad.

4.7.3 Circular Scale

For polar and Smith chart format, you can set the outer circle value (see Figure
4.17).
Figure 4.17 Circular scale

4.7.4 Circular Scale Setting

1 The number of scale divisions affects all channel traces.
81
MEASUREMENT CONDITIONS SETTING
To set the scale of the circular graph use the following softkey in the right menu bar Scale.
Then select the Scale field and enter the required value using the on-screen keypad.

4.7.5 Automatic Scaling

The automatic scaling function allows the user to define the trace scale automatically so that the trace of the measured value could fit into the graph entirely.
In rectangular format, two parameters are adjustable: scale and reference level position. In circular format, the outer circle value will be adjusted.
To execute the automatic scaling use the following softkeys in the right menu bar Scale > Auto Scale.
82
MEASUREMENT CONDITIONS SETTING
tfj
eS=S
π2

4.7.6 Reference Level Automatic Selection

This function executes automatic selection of the reference level in rectangular coordinates.
After the function has been executed, the trace of the measured value makes the vertical shift so that the reference level crosses the graph in the middle. The scale will remain the same.
To execute the automatic selection of the reference level use the following softkeys in the right menu bar Scale > Auto Ref. Value.

4.7.7 Electrical Delay Setting

The electrical delay function allows the user to define the compensation value for the electrical delay of a device. This value is used as compensation for the electrical delay during non-linear phase measurements. The electrical delay is set in seconds.
If the electrical delay setting is other than zero, S-parameter value will vary in accordance with the following formula:
, where
f – frequency, Hz,
t – electrical delay, sec.
83
MEASUREMENT CONDITIONS SETTING
The electrical delay is set for each trace individually. Before you set the electrical delay, first activate the trace.
To set the electrical delay use the following softkey in the right menu bar Scale.
Then select the Electrical Delay field and enter the required value using the on­screen keypad.

4.7.8 Phase Offset Setting

The phase offset function allows the user to define the constant phase offset of a trace. The value of the phase offset is set in degrees for each trace individually. To set the phase offset, first activate the trace.
To set the phase offset use the following softkey in the right menu bar Scale.
84
MEASUREMENT CONDITIONS SETTING
Then select the Phase Offset field and enter the required value using the on- screen keypad.
85
MEASUREMENT CONDITIONS SETTING
Note
If you select segment frequency sweep, the Segment Table softkey will be become available in Stimulus dialog. For segment tables details see section 4.8.5.

4.8 Stimulus Setting

The stimulus parameters are set for each channel. Before you set the stimulus parameters of a channel, make this channel active.

4.8.1 Sweep Type Setting

To set the sweep type use the following softkey in the right menu bar Stimulus.

4.8.2 Sweep Span Setting

86
MEASUREMENT CONDITIONS SETTING
To enter the start and stop values of the sweep range use the following softkey in the right menu bar Stimulus.
Then select the Start Frequency or Stop Frequency field and enter the required values using the on-screen keypad.
If necessary, you can select the measurement units. The current measurement units are shown to the right from the value entry field.

4.8.3 Sweep Points Setting

To enter the number of sweep points use the following softkey in the right menu bar Stimulus.
87
MEASUREMENT CONDITIONS SETTING
Then click on Points field, select the required value from the list and complete the setting by Ok.

4.8.4 Stimulus Power Setting

The stimulus power level can take two possible values. High output power corresponds to the source signal power of -10 dB/m. Low output power corresponds to -30 dBm.
To enter the power level value use the following softkeys in the right menu bar Stimulus > Power.
Then select the Output power field to switch between the high and low settings of the power level.
The set power level can also be seen in the channel status bar.

4.8.5 Segment Table Editing

Frequency sweep span can be divided into segments. Each segment has start and stop values of the sweep range, number of points and measurement delay. IF filter and measurement delay can be enabled/disabled by the user.
88
MEASUREMENT CONDITIONS SETTING
The types of segment tables are shown below.
Each table line determines one segment. The table can contain one or several lines. The number of lines is limited by the aggregate number of all segment points, i.e. 100001
To edit the segment table use the following softkeys in the right menu bar Stimulus > Segment Table.
Select the segment frequency sweep to make the Segment Table softkey available (see section 4.8.1).
To add a segment to the segment table use Add.
To delete a segment from the table use Delete.
89
MEASUREMENT CONDITIONS SETTING
Note
The adjacent segments cannot overlap in the frequency domain.
To enter the segment parameters, move the mouse to the respective box and enter the numerical value. You can navigate the segment table using the «Up Arrow», «Down Arrow », «Left Arrow », «Right Arrow» keys.
To edit any parameter in the table, double click on the its value field and enter the required value using the on-screen keypad.
To enable/disable the IFBW filter column click on the List IFBW field.
To enable/disable the measurement delay column click on the List Delay field.
The segment table can be saved into *.seg file to a hard disk and later recalled.
To save the segment table use Save.
To recall the segment table use Recall.
90
MEASUREMENT CONDITIONS SETTING

4.9 Trigger Setting

The Analyzer can operate in one of three sweep trigger modes. The trigger mode determines the sweep actuation. The trigger can have the following modes:
Continuous – a sweep actuation occurs every time after sweep cycle is
complete in each channel;
Single – sweep actuation occurs once, and after the sweep is complete, the
trigger turns to hold mode;
Hold – sweep is stopped, the actuation does not occur.
If more than one channel window is displayed on the screen, a sweep will be actuated in them in succession.
Trigger source can be internal or bus (transferred through COM/DCOM).
To set the trigger mode use the following softkey in the right menu bar Trigger.
Then click on Trigger Mode field select the required mode from the list and complete the setting by Ok.
Close the Trigger dialog by Ok.
If you select Single Trigger Mode you can actuate sweep by clicking on the Trigger Event softkey in the right menu bar.
91
MEASUREMENT CONDITIONS SETTING

4.10 Measurement Optimizing

4.10.1 IF Bandwidth Setting

The IF bandwidth function allows the user to define the bandwidth of the test receiver. The IF bandwidth can be selected by user from the following values: 100 Hz, 300 Hz, 1 kHz, 3 kHz, 10 kHz and 30 kHz.
The IF bandwidth narrowing allows you to reduce self-noise and widen the dynamic range of the Analyzer. Also the sweep time will increase. Narrowing of the IF bandwidth to 10 will reduce the receiver noise to 10 dB.
The IF bandwidth should be set for each channel individually. Before you set the IF bandwidth, first activate the channel.
To set the IF bandwidth use the following softkey in the left menu bar Average.
To set the IF bandwidth click on IFBW field and select the required value from the list. Complete the setting by Ok.

4.10.2 Averaging Setting

The averaging function is similar to IF bandwidth narrowing, it allows reducing self-noise and widening the dynamic range of the Analyzer.
The averaging in each measurement point is made over several sweeps according to the exponential window method.
The averaging should be set for each channel individually. Before you set the averaging, first activate the channel.
92
MEASUREMENT CONDITIONS SETTING
To set the averaging use the following softkey in the left menu bar Average.
To toggle the averaging function on/off click on Average field.
To set the averaging factor click on Averaging Factor field and enter the required value using the on-screen keypad.

4.10.3 Smoothing Setting

The smoothing of the sweep results is made by averaging the measurement results of adjacent points of the trace determined by the moving aperture. The aperture is set by the user in percent from the total number of the trace points.
The smoothing does not increase the dynamic range of the Analyzer. It preserves the average level of the trace and reduces the noise bursts.
The smoothing should be set for each trace individually. To set the smoothing, first activate the trace.
93
MEASUREMENT CONDITIONS SETTING
To set the smoothing use the following softkey in the left menu bar Average.
To toggle the smoothing function on/off click on Smoothing field.
To set the smoothing aperture click on Smoothing Aperture field and enter the required value using the on-screen keypad.
94
MEASUREMENT CONDITIONS SETTING

4.10.4 Trace Hold Function

The Trace Hold function displays the maximum or the minimum of any given active measurement instead the real-time data. The held data is displayed as an active trace.
To toggle the Trace Hold function on/off use the following softkeys in the right menu bar Trace > Trace Hold.
Then select the required type (Maximum | Minimum) from the Hold Type list and complete the setting by Ok.
95
MEASUREMENT CONDITIONS SETTING

4.11 Cable Specifications

By default, the program does NOT compensate DTF measurements to account for the inherent loss of a cable. However, to make more accurate DTF measurements, the cable loss and velocity factor can be entered using one of the following methods:
Select a cable type from a list which contains the Cable loss in dB/meter
and Velocity factor;
Manually enter Cable loss and Velocity factor for the measurement.
Velocity factor is a property of the physical material of a cable. A VF of 1.0 corresponds to the speed of light in a vacuum, or the fastest VF possible. A polyethylene dielectric cable has VF = 0.66 and a cable with Teflon dielectric has VF = 0.7.
Cable Loss is specified in dB/meter. In addition to the length of the cable, loss is also directly proportional to the frequency of the signal that passes through the cable.

4.11.1 Selecting the type of cable

To select the type of cable use the following softkeys in the left menu bar
Analysis > Time Domain > Cable Correction > Cable type.
96
MEASUREMENT CONDITIONS SETTING
Select the required item from the Cable List and complete the setting by Ok

4.11.2 Manually specify Velocity Factor and Cable Loss

To set the parameters of cable, press the following softkeys in the left menu bar
Analysis > Time Domain > Cable Correction.
97
MEASUREMENT CONDITIONS SETTING
Click on Velocity Factor field to enter the value of velocity factor using the on­screen keypad.
Click on Loss field to enter the value of cable loss using the on-screen keypad.

4.11.3 Editing table of cables

To edit the table of cables, press the following softkeys in the left menu bar
Analysis > Time Domain > Cable Correction.
98
MEASUREMENT CONDITIONS SETTING
Click the left button of the mouse on the field Cable Type.
To add/delete rows in the table click Add/Delete.
Then select the required parameter in the table and double click on the corresponding cell.
Enter the required value Cable Name, Velocity, Cable Loss etc using the on-screen keypad.
To save the table of cables on the drive click the Save Cable List button.
To restore the table cables from the drive, press the Restore Cable List softkey.
99

5. CALIBRATION AND CALIBRATION KIT

5.1 General Information

5.1.1 Measurement Errors

S-parameter measurements are influenced by various measurement errors, which can be devided into two categories:
systematic errors, random errors.
Random errors comprise such errors as noise fluctuations and thermal drift in electronic components, changes in the mechanical dimensions of connectors subject to temperature drift, repeatability of connections. Random errors are unpredictable and hence cannot be estimated and eliminated in calibration. Random errors can be reduced by correct setting of the source power, IF bandwidth narrowing, maintaining constant environment temperature, observance of the Analyzer warm-up time, careful connector handling, avoidance of cable bending after calibration, and use of the calibrated torque wrench for connection of the Male-Female coaxial RF connectors.
Random errors and related methods of correction are not mentioned further in this section.
Systematic errors are the errors caused by imperfections in the components of the measurement system. Such errors occur repeatedly and their characteristics do not change with time. Systematic errors can be determined and then reduced by performing mathematical correction of the measurement results.
The process of measurement of precision devices with predefined parameters with the purpose of determination of measurement systematic errors is called calibration, and such precision devices are called calibration standards. The most commonly used calibration standards are SHORT, OPEN, and LOAD.
The process of mathematical compensation (numerical reduction) for measurement systematic errors is called an error correction.

5.1.2 Systematic Errors

The systematic measurement errors of vector network analyzers are subdivided into the following categories according to their source:
Directivity; Source match; Reflection tracking.
100
Loading...