INDICATIONS
The LM-900 is a cryosurgery system intended to be used in the process of controlled
destruction of defined areas of benign or pre-malignant lesions.
CONTRAINDICATIONS
The LM-900 is not indicated for therapeutic treatments to the central circulatory or central
nervous systems.
CAUTION
U.S. Federal law restricts this device to sale by or on the order of a licensed practitioner or
licensed medical facility.
WARNINGS
• Exposure to nitrous oxide may result in impairment of performance, cognition or dexterity.
Ventilate well. Refer to instructions for venting nitrous oxide gas.
• Do not sterilize. Device must only be cleaned and disinfected. See cleaning and disinfection
instructions below.
PRECAUTIONS
• This device utilizes high pressure gas and must be used with care to avoid unexpected
release of pressure. Maintenance and care must be taken to avoid injury to user
and patient.
• Use only medical grade gas under pressure range 600-800 psi (42-56 Kg/cm
2
) for
optimal freeze.
• Do not store gas cylinders above normal room temperature.
• It is the operator’s responsibility to fit the exhaust hose between cryosurgical unit exhaust and
disposal point to minimize gas exposure.
• Do not drag, slide or roll cylinders. Cylinders should be stored in upright position and
firmly secured to prevent falling or tipping.
• Do not crimp or tightly bend the gas hose. The hose should be coiled no tighter than an
8 in. (20.3 cm) circle. A damaged hose must be replaced.
• Do not use tools to tighten connections. Hand-tightening is all that is required.
• Before disconnecting the console from the gas cylinder, turn gas cylinder valve OFF.
• Before changing tips, make sure gas is off at the console.
• If cryotips are dropped, they should be returned to the factory for inspection.
GENERAL INSTRUCTIONS
Check the gas cylinder to be certain that it contains adequate gas. The pressure gauge should
read between 600 and 800 psi (42-56 Kg/cm2) for proper freezing to occur. A pressure below
600 psi (42 Kg/cm
2
) may not provide enough pressure for sufficient freeze or defrost, and is not
recommended. The gas cylinder should be securely seated on the mobile cart, and fastened
with the provided safety chain.
The tubing leading from the console to the cryogun should be examined by running the tube
between the thumb and forefinger. Check for any kinks or abrupt bends. A kink in the line can
cause a weakness, which can erupt under pressure.
Located between the tip and the gun barrel is an “O” ring seal, which should be inspected prior
to each use.
Before each use, test for proper operation by following the instructions below, except point the
tip towards the floor while squeezing the trigger for several seconds.
SCAVENGING HOSE
A twenty-foot (6.1 m) scavenging hose (P/N 3012-0314) is provided with the unit to divert gas
away from the operator and patient to a remote, well-ventilated area. One end of the scavenging
hose should be pushed over the exhaust muffler on the console of the cryo unit (mild soapy
water will ease this operation), and the other end should lead to a remote, well-ventilated area.
LM-900 Cryosurgery System • Directions for Use
2