Cooper Surgical Leisegang LM-900 Directions For Use Manual

Models 50501 and 50502
LM-900 Cryosurgery System
Directions for Use
®
INTRODUCTION
Model 50501 requires Medical Grade N2O (Nitrous Oxide) gas. Model 50502 requires Medical
Grade CO2(Carbon Dioxide) gas.
Each LM-900 has been tested prior to shipment. Proper care and maintenance should ensure a long useful life.
This Directions for Use document includes general and specific directions for use, and information on warnings and precautions, storage, cleaning, and disinfection.
Please contact our Customer Service department with any comments or questions.
Send correspondence to:
CooperSurgical Customer Service Dept. 95 Corporate Drive Trumbull, CT 06611 USA
Phone:
800-243-2974 Toll Free 203-601-5200
THE LM-900 CRYOSURGERY SYSTEM
No tools are required for assembly and operation of the LM-900 Cryosurgery System.
The LM-900 is delivered in an attache carrying case along with interchangeable cryotips, a package of replacement “O” rings, cleaning plugs and a gas scavenging hose. On the end of the cryogun barrel is a tip protector, which is removed when a cryotip is used, and replaced when the cryogun is not in use.
An empty 20 lb N2O (Nitrous Oxide) or CO2(Carbon Dioxide) gas cylinder is available for the LM-900. The N2O or CO2gas is used as the refrigerant for the cryogenic process. The LM-900 should be used only in a well-ventilated area. Consult your approved local compressed gas supplier. A twenty-foot (6.1 m) scavenging hose is provided to divert used gas away from the operator and patient. The mobile hand cart should be used for cylinder movement. Compressed gas cylinders should be refilled only by qualified suppliers. Do not allow the temperature where cylinders are stored to exceed normal room temperature.
READ THIS ENTIRE DOCUMENT BEFORE OPERATING
THE LM-900 CRYOSURGERY INSTRUMENT
20 POUND CYLINDER
CARRYING CASE
CLEANING PLUGS & “O” RINGS
QUICK-CONNECT
CONSOLE TIPS CRYOGUN
MOBILE CART
LM-900 Cryosurgery System
Directions for Use
1
INDICATIONS
The LM-900 is a cryosurgery system intended to be used in the process of controlled destruction of defined areas of benign or pre-malignant lesions.
CONTRAINDICATIONS
The LM-900 is not indicated for therapeutic treatments to the central circulatory or central nervous systems.
CAUTION
U.S. Federal law restricts this device to sale by or on the order of a licensed practitioner or licensed medical facility.
WARNINGS
• Exposure to nitrous oxide may result in impairment of performance, cognition or dexterity.
Ventilate well. Refer to instructions for venting nitrous oxide gas.
Do not sterilize. Device must only be cleaned and disinfected. See cleaning and disinfection
instructions below.
PRECAUTIONS
• This device utilizes high pressure gas and must be used with care to avoid unexpected release of pressure. Maintenance and care must be taken to avoid injury to user and patient.
• Use only medical grade gas under pressure range 600-800 psi (42-56 Kg/cm
2
) for
optimal freeze.
• Do not store gas cylinders above normal room temperature.
• It is the operator’s responsibility to fit the exhaust hose between cryosurgical unit exhaust and disposal point to minimize gas exposure.
• Do not drag, slide or roll cylinders. Cylinders should be stored in upright position and firmly secured to prevent falling or tipping.
• Do not crimp or tightly bend the gas hose. The hose should be coiled no tighter than an 8 in. (20.3 cm) circle. A damaged hose must be replaced.
• Do not use tools to tighten connections. Hand-tightening is all that is required.
• Before disconnecting the console from the gas cylinder, turn gas cylinder valve OFF.
• Before changing tips, make sure gas is off at the console.
• If cryotips are dropped, they should be returned to the factory for inspection.
GENERAL INSTRUCTIONS
Check the gas cylinder to be certain that it contains adequate gas. The pressure gauge should read between 600 and 800 psi (42-56 Kg/cm2) for proper freezing to occur. A pressure below 600 psi (42 Kg/cm
2
) may not provide enough pressure for sufficient freeze or defrost, and is not recommended. The gas cylinder should be securely seated on the mobile cart, and fastened with the provided safety chain.
The tubing leading from the console to the cryogun should be examined by running the tube between the thumb and forefinger. Check for any kinks or abrupt bends. A kink in the line can cause a weakness, which can erupt under pressure.
Located between the tip and the gun barrel is an “O” ring seal, which should be inspected prior to each use.
Before each use, test for proper operation by following the instructions below, except point the tip towards the floor while squeezing the trigger for several seconds.
SCAVENGING HOSE
A twenty-foot (6.1 m) scavenging hose (P/N 3012-0314) is provided with the unit to divert gas away from the operator and patient to a remote, well-ventilated area. One end of the scavenging hose should be pushed over the exhaust muffler on the console of the cryo unit (mild soapy water will ease this operation), and the other end should lead to a remote, well-ventilated area.
LM-900 Cryosurgery System • Directions for Use
2
Loading...
+ 5 hidden pages