Cooper Instruments RANGER 5 Specifications

XAA 911 - LOAD CELL SUMMING BOX
he Cooper XAA 911 load cell summing box has four 5 position terminal strips for load cell input and one 7
T
position terminal strip with internally jumped sense terminals for output to the instrument. A total of 4 load cells can be summed. Trimming is accomplished with four 25 turn trim pots. Calibration is completed by turning each of the four Trim pots fully clockwise to obtain the highest possible output from each load cell.
ACCESSORIES
RANGER 5 - TRANSDUCER SIMULATOR
he model Ranger 5 is an inexpensive and easy to operate
instrument for use in testing, troubleshooting and calibrating
T
strain gage instruments and systems. It features a switchable vernier control for obtaining exact instrument readings while adjusting setpoints and overrange blanking. The Ranger 5 simulates load, force, pressure and torque.
SPECIFICATIONS
ACCURACY: MATERIAL: OUTPUT: VERNIER: BRIDGE RESISTANCE: SIZE:
±0.05% (with Vernier off) ABS Plastic
0.5 mv/v to 4.5 mv/v 10 turn with locking graduated dial 350 ohm 3 3/4 x 6 1/4 x 1 3/4
COOPER INSTRUMENTS & SYSTEMS
www.cooperinstruments.com sales@cooperinstruments.com
1-800-344-3921
4/9/04