Cooper Instruments MGT 12 Installation Manual

Page 1

SERIES MGT

USER’S GUIDE

www.cooperinstruments.com
PH: 540-349-4746 • FAX: 540-347-4755
Page 2

Controls

All series MGT torque gauges have three keys for controlling all functions of the instrument.
POWER Turns the instrument on and off. Also used in the calibration mode. Used to select Clockwise
Peak, Counterclockwise Peak or Normal (real time) display mode. The actual peak readings are always captured and can be displayed at any time.
ZERO Zeros any tare value (up to the full capacity of the gauge) and clears the peak readings stored in
memory.

Display

The display consists of a 4 ½-digit section and several indicators. Their functions are listed below.
LO BAT Low battery voltage indicator CW Clockwise torque indicator CCW Counterclockwise torque indicator CW PEAK Peak clockwise indicator CCW PEAK Peak counterclockwise indicator OZIN, LBIN, Units of measurement NCM, KGMM (Model dependent)
------ (dashes) Overload (>110% of range)
C W
P E A
W a
1
r
­r
8
e
0
n
0
t
-
o
3
n
4
,
4
V
-
3
A
9 2
U
1
S A
K

Operation

The default mode of operation is the normal (real time) mode. If the peak readings are to be observed as they occur, then the mode of operation can be changed by pressing PEAK until the desired mode (CW PEAK or CCW PEAK) appears on the display. Please note that this action affects only the display. The peak readings are captured automatically and can be cleared from the memory by either pressing ZERO or shutting off the gauge.
The gauge is equipped with an automatic shutdown feature and it will shut off after approximately 30 minutes of inactivity (readings do not change by more than ±10 counts and no keys are pressed). The entire display will flash for 5-7 seconds as a warning of the imminent shutdown. To change the default setting of 30 minutes, hold PEAK while turning on the gauge. Press PEAK repeatedly until ‘AOFF’ appears. Press POWER to select this function. The current setting will flash on the display. Use PEAK to scroll through the displayed choices and POWER to select. Press POWER again at the ‘donE’ prompt in order to save the setting.
The displayed units of measurement and the default mode of operation (peak or normal) can be changed by entering the setup mode as described above, pressing POWER at the ‘init’ prompt and selecting the desired settings using the PEAK and POWER keys.

Power

The instrument may be powered by the internal gV battery, or by the included AC adapter. The need for battery replacement is indicated by a 3-step sequence: 1- a steady LO BAT appears on the display indicating the last 10% of the battery life, 2- LO BAT begins to flash indicating the need for an immediate battery replacement, 3- the entire display except LO BAT flashes for several seconds and then the instrument shuts off.
CF37 2 07/02/01 (V-9/96)
Page 3

Calibration

To properly calibrate this instrument, application of a precise torque value equal to the full capacity of the gauge in inin or ozin (regardless of the displayed units) will be required. While holding PEAK, turn the gauge on. When ‘CAL’ appears on the display press POWER three times to select the calibration mode. At the ‘null’ prompt press ZERO. At the ‘SPAn’ prompt apply the calibration torque and press POWER. The display will show “uuuu” or “nnnn” if the calibration torque is insufficient or excessive accordingly. If this happens, the only way to terminate the calibration mode is by momentarily disconnecting the battery or connecting the AC adapter without plugging it into a wall outlet. This will stop the calibration procedure without making changes to the previous calibration data. Successful calibration is indicated by ‘donE’ on the display. Press POWER to save the changes and resume normal operation.

Specifications

ACCURACY ±0.5% of full scale ±1 digit SAMPLING RATE 30 readings/second DISPLAY RATE 2.5/s in normal mode, 30/s in peak mode SAFE OVERLOAD 150% of gauge capacity. Display shows ---(dashes) above 110% POWER 9V battery or AC adapter BATTERY LIFE 30 hours of continuous operation SIZE See drawing WEIGHT 1.8 lb (0.8kg)

WARRANTY REPAIR POLICY

Limited Warranty On Products

Any Cooper Instruments product which, under normal operating conditions, proves defective in material or in workmanship within one year of the date of shipment by Cooper will be repaired or replaced free of charge provided that a return material authorization is obtained from Cooper and the defective product is sent, transportation charges prepaid, with notice of the defect, and it is established that the product has been properly installed, maintained, and operated within the limits of rated and normal usage. Replacement or repaired product will be shipped F.O.B. from our plant. The terms of this warranty do not extend to any product or part thereof which, under normal usage, has an inherently shorter useful life than one year. The replacement warranty detailed here is the buyer’s exclusive remedy, and will satisfy all obligations of Cooper whether based on contract, negligence, or otherwise. Cooper is not responsible for any incidental or consequential loss or damage which might result from a failure of any and all other warranties, express or implied, including implied warranty of merchantability or fitness for particular purpose. Any unauthorized disassembly or attempt to repair voids this warranty.

Obtaining Service Under Warranty

Advance authorization is required prior to the return to Cooper Instruments. Before returning the item, contact the Repair Department c/o Cooper Instruments at (540) 349-4746 for a Return Material Authorization number. Shipment to Cooper shall be at buyer’s expense and repaired or replacement items will be shipped F.O.B. from our plant in Warrenton, Virginia. Non-verified problems or defects may be subject to a $100 evaluation charge. Please return the original calibration data with the unit.

Repair Warranty

All repairs of Cooper products are warranted for a period of 90 days from date of shipment. This warranty applies only to those items that were found defective and repaired; it does not apply to products in which no defect was found and returned as is or merely recalibrated. It may be possible for out-of-warranty products to be returned to the exact original specifications or dimensions.
* Technical description of the defect: In order to properly repair a product, it is absolutely necessary for Cooper to receive information specifying the reason the product is being returned. Specific test data, written observations on the failure and the specific corrective action you require are needed.
CF37 3 07/02/01 (V-9/96)