Cooper Instruments DPG1000DRBL Installation Manual

DPG 1000DR
Digital Pressure Gauge
USER’S GUIDE
www.cooperinstruments.com
PH: 540 -349-4746 FAX: 540-347-4755
Install or remove the gauge-using wrench on hex fitting only. Do not attempt to tighten by turning housing or any other part of the gauge. Use fittings appropriate for the pressure range of the gauge. Do not apply vacuum to gauges not designed for vacuum operation. Due to the hardness of 316 stainless steel, it is recommended that a thread sealant be used to ensure leak -free operation. NEVER insert objects into the gauge port or blow out with compressed air. Permanent damage not covered by warranty will result to the sensor.
Electrical Connection
NEVER connect the gauge wires directly to 115 VAC or permanent damage not covered by warranty will
result. The DPG1000DR and DPG1000DRBL can be powered by any 9 to 32 VDC or 8 to 24 VAC 50/60 Hz power
source. An inexpensive unregulated low voltag e source can be used. The magnitude of the supply voltage has negligible effect on the gauge calibration as long as it is within the stated voltage ranges. Do not allow the gauge supply voltage fall below 9 VDC or 8 VAC RMS. Operation below these values may cause erratic or erroneous readings or output. Models with 4-20 mA output power the current loop. Use a power source with sufficient voltage to operate the current loop.
Connection is made with the 4-conductor cable at the gauge rear. This cable accommodates both the gauge power supply and retransmission output. This cable has one RED and one BLACK lead. If using a 9 to 32 VDC power source, connect the (+) supply to the RED lead and the (–) supply to the BLACK lead. If using a 8 to 24 VAC 50/60 Hz power source, connect to the RED and BLACK leads. When using low voltage AC power, there is of course, no polarity consideration.
The (+) retransmission output appears on the WHITE lead, and the (–) retransmission output appears on the GREEN lead. Use of the shield (drain) wire of the retransmission output is optional. It is not generally needed for 4-20 mA current loops unless very long cable lengths are used in electrically noisy environments.
The output is a continuous analog signal based on the transducer output rather than the display. This output is filtered to improve noise immunity and has a response time of about 50 milliseconds.
The power supply (–) lead is tied to the retransmission output ground. Therefore, if a DC supply is used, the power supply (–) lead should be considered common with regard to the retransmission output (–) connection.
Using the Retransmission Output
NEVER connect retransmission output wires together
or to an external power source or permanent damage not covered by warranty will result.
For 4-20 mA output models, be sure to observe the output compliance (voltage drive) capabilities of the gauge. The compliance, and therefore the maximum loop resistance the output can drive, is a function of the supply voltage to the gauge. Consult the graph shown at right for maximum loop resistance vs. gauge supply voltage. Too large a loop resistance will cause the gauge output to “limit” or saturate before reaching its full 20 mA output.
When using the 0-2 volt retransmission output, do not allow the resistive load on the output to fall below 5K ohms. Also, avoid large capacitive loads (greater that 1000 pF) such as those caused by long runs of shielded cable. For long retransmission runs, use the 4-20 mA output model instead.
CF 144 1 DPG 1000DR
Loading...
+ 2 hidden pages