1.1.3.1 Display Menu .........................................................................................................................................3
1.1.3.2 Program Menu .......................................................................................................................................3
1.1.3.3 Function Menu .......................................................................................................................................3
2.3.3.2 Resistance Temperature Detector (RTD) Inputs .................................................................................11
2.3.3.3 Linear Inputs ........................................................................................................................................11
2.3.4.2 Solid State Relay Option......................................................................................................................12
3.2.5.1 Bar Charts............................................................................................................................................16
3.2.5.2 Digital Windows ...................................................................................................................................17
3.2.5.3 Alarm/Events Data Window .................................................................................................................17
3.2.5.5 Transient Windows ..............................................................................................................................18
3.3 Browsing, Compressing and Searching Data ............................................................................18
3.10 Recording Data ...........................................................................................................................25
3.10.1 Using the Recorder to format a floppy/zip disk or PCMCIA card .................................... 26
3.10.2 Selecting the Record Mode .................................................................................................26
3.10.2.1 Fill to End ...........................................................................................................................................26
3.10.2.3 Average or Instantaneous Recording ................................................................................................ 26
3.10.3 Programming the unit for recording................................................................................... 27
3.10.3.1 Set the record mode ..........................................................................................................................27
3.10.3.3 Record Rate.......................................................................................................................................27
4.1.1.1 Memory Test (RAM).............................................................................................................................29
4.1.1.2 ROM Test.............................................................................................................................................29
4.1.3.2 Full Init..................................................................................................................................................30
4.2.1.4 Version .................................................................................................................................................32
4.2.1.5 Media Status ........................................................................................................................................33
4.2.2 PROG (Program) Menu ..........................................................................................................33
4.2.3 FUNC (Function) Menu ..........................................................................................................33
4.2.3.1 Record On/Off......................................................................................................................................33
4.2.3.2 Activate Point .......................................................................................................................................34
4.2.3.3 Bypass Point ........................................................................................................................................34
4.2.3.4 Reset Point ..........................................................................................................................................34
4.2.3.6 Record speed.......................................................................................................................................34
5.4.2 Time Format............................................................................................................................ 38
5.4.3 Power Up Display...................................................................................................................38
5.4.3.1 Unit Tag................................................................................................................................................38
5.5.2.1 Scale Type ...........................................................................................................................................41
5.6.2.1 Choosing a Point Number....................................................................................................................44
5.6.2.2 Setup an Unprogrammed Point ...........................................................................................................44
5.6.3 Point Options.......................................................................................................................... 44
5.6.3.1 Setup a Point by Copying ....................................................................................................................44
5.6.3.2 Setup a Point by Restoring ..................................................................................................................44
5.6.3.3 Modify an Existing Point.......................................................................................................................45
5.6.3.4 Delete an Existing Point....................................................................................................................... 45
5.6.4 Programming Point Types.....................................................................................................45
5.6.5.1 Point Tag..............................................................................................................................................46
5.6.5.9 Eng Units..............................................................................................................................................48
5.6.5.13 Reset Control .....................................................................................................................................51
5.6.5.14 Time Period........................................................................................................................................53
5.6.5.18 Set Equation ......................................................................................................................................54
5.6.5.19 Set Cndtionl .......................................................................................................................................54
5.6.12.2 Hi Peak ..............................................................................................................................................58
5.6.12.3 Lo Peak..............................................................................................................................................58
5.6.12.4 HiLo Difference ..................................................................................................................................59
5.6.12.5 Moving Average .................................................................................................................................59
5.6.12.6 Time Average.....................................................................................................................................59
5.7.3 Record Mode...........................................................................................................................61
CF 46 vi 2/06
5.7.3.1 Fill Mode...............................................................................................................................................61
5.7.6.3 All Rates...............................................................................................................................................63
5.7.7 Disk Full Alarm .......................................................................................................................64
5.7.7.1 Disk Full Setpoint .................................................................................................................................64
5.7.7.2 Disk Full Alarm Output .........................................................................................................................64
5.7.8 Format Disk.............................................................................................................................64
5.7.9 Save ConFiGuration File........................................................................................................64
5.9 Digital I/O .......................................................................................................................................66
5.9.2.3 Record on/off .......................................................................................................................................68
5.9.2.6 Record Rate.........................................................................................................................................68
5.10.1.2 Port Setup ..........................................................................................................................................70
5.10.1.3 Network ID .........................................................................................................................................70
5.11 System .........................................................................................................................................72
5.11.3.3 Save English ......................................................................................................................................73
5.11.3.4 Save Alt. Lang....................................................................................................................................73
5.11.4.3 Data Interval.......................................................................................................................................74
5.11.4.5 Data Lines..........................................................................................................................................74
7.3.1.1 Exported File Format ...........................................................................................................................78
APPENDIX B: CONFIGURING A 4-20 MA INPUT TO EQUAL 0 TO 150 GPM........................85
APPENDIX C: CONFIGURING A J TYPE T/C INPUT FOR A 100-750
O
C RANGE .................91
APPENDIX D: CONFIGURING THE TOTALIZER.....................................................................96
APPENDIX E: SETTING THE JUMPERS FOR THE 3 OR 6 CHANNEL MECHANICAL OR
SOLID STATE RELAY.............................................................................................................100
CF 46 viii 2/06
SAFETY NOTICE
This Safety Notice has been included to emphasize the DANGER OF HAZARDOUS VOLTAGES on the REAR
TERMINAL PANEL of your instrument. USE EXTREME CAUTION WHEN INSTALLING OR SERVICING your
instrument. Pleas read the entire contents of the Installation and Wiring Chapter of this manual before attempting
to install or service your instrument.
Use Extreme caution when servicing the rear terminal of your instrument.
CF 46 1 2/06
1.0 General Description
Figure 1-1 Solid State Data Recorder
This manual is a user reference guide for the Solid State Paperless Data Recorder (Figure 1-1). The manual
provides detailed instruction for installation, operation, programming, calibration and maintenance of the instrument.
The recorder is a sophisticated piece of equipment that requires some level of programming before use. The user
is advised to read this manual in its entirety before proceeding with the installation and programming. Refer to
Chapter 3, Getting Started for minimum configuration. Step by step instructions for a few sample applications are
included in Appendixes A-D.
1.1 Recorder Description
The Solid State Data Recorder is a Paperless Recording instrument. Data is stored in either an internal floppy disk,
a removable PCMCIA memory card or a Zip Drive.
All data is stored in MSDOS format and may be archived or analyzed on any IBM compatible PC running
Microsoft’s Windows 3.1 or Windows 95/98 using the available Companion Software. The instrument retains all the
features of a traditional Paper Chart Recorder by virtue of its large STN monochrome or TFT color Liquid Crystal
Display (LCD) which presents the data in the traditional chart mode as well as in bar graph or digital numeric form.
The unit has many features and functions, which are unique and cannot be performed on traditional paper
recorders such as data compression and historic data browsing. The recorder is programmed via a touch screen
keypad on the display.
The recorder will measure and process up to six direct inputs, calculated, conditional, or external points for logging,
trending, or data manipulation.
If direct inputs are not desired, the Data Recorder will accept up to fifteen points from a combination of calculated,
conditional, or external point types.
1.1.1 Inputs
Direct input sources may come from voltage, current, dry contacts, thermocouple, or RTD sources. The voltage and
current ranges accepted by the instrument include: ±150mV, 0 to ±1.25V, ±2.5V, ±12.5V and ±25V; 4 to 20mA, 0 to
20mA and 10 to 50mA current. Thermocouple inputs include J, K, T, E, R, S, B, C and N. RTD inputs accepted
include 10 ohm Cu, 100 ohm Platinum, 200 ohm Platinum, 120 ohm Nickel and 1000 ohm Nickel.
CF 46 2 2/06
1.1.2 Instrument Size
The instrument is sized to fit in a DIN standard panel cutout of 5.43 inches x 5.43 inches (138mm x 138mm) and
requires only 7.41 inches [188.2 mm] behind panel depth, not including power and input source cable space
needed. Actual dimensions of the instrument are shown in Figure 2-1 Recorder Dimensions in Chapter 2 of this
manual.
1.1.3 Menus
The instrument’s features are accessed through a series of menus. Press the MENU button displayed in the
bottom right hand corner of the LCD screen. The STATUS bar along the top of the screen displays the various
recorder parameters. (Refer to Section 3.1). The Command button bar contains three user programming option
buttons - DISPLay, PROGram, and FUNCtion (see to Figure 1-2 below). Each menu level features easy-to-follow
prompts that simplify operation.
Figure 1-2 The Command Button Bar
1.1.3.1 Display Menu
The Display Menu is accessed by pressing the Display button (DISPL) on the Command button bar. This menu
allows you to display any programmed point or series of points, or any current alarm or series of alarms on the
STATUS line along the top of the display. The version of operating software can also be shown in a pop up window.
1.1.3.2 Program Menu
The Program Menu is accessed by pressing the Program button (PROG) on the Command button bar. This menu
item may be passcode protected. The Program Menu allows you to define the system operating parameters. Menu
driven prompts, answered by yes, no or by entering the desired value, enable you to customize the Instrument to
meet your application requirements. Refer to Chapter 5 for detailed instructions on programming.
1.1.3.3 Function Menu
The Function Menu is accessed by pressing the Function Button (FUNC) on the Command button bar. This Menu
item may be passcode protected. The Function Menu allows you to Activate, Bypass, and/or Reset a point. This
menu also allows you to change between high and low display chart speed or record speed, turning Alarm Check
on or off, and choosing Scale Set 1 or 2.
1.1.3.4 Hidden Menu
The Hidden Menu can only be accessed by pressing a certain combination of buttons and may be passcode
protected. This menu allows the user to Initialize the recorder, set Passcodes, perform Calibrations and perform
Diagnostics on the recorder. Refer to Chapter 4 for detailed information.
1.1.4 Memory
All the Random Access Memory (RAM) in the Recorder is battery backed. This enables the unit to recover in the
event of a mains failure with minor data loss. Any programming will be protected in the event of power removal and
past browse data is maintained in the off state. The battery is a non-rechargeable lithium and will keep memory
intact for at least 12 months.
1.1.5 Clock
A real time clock keeps time and date in the event of a power loss. It uses the same battery as the memory.
1.1.6 Recorder Construction
The Data Recorder features modular construction. Power Supply and Analog conditioning modules are
conveniently accessible for fast and simple troubleshooting and/or removal. The floppy disk, PCMCIA or Zip drive
can be accessed through the front of the unit.
1.2 Recording Options
The Recorder can be ordered the with one of three storage mediums, a standard PC compatible floppy disk drive,
an industry-standard PCMCIA memory card drive, or a Zip drive. The Companion software provided with the unit
CF 46 3 2/06
supports all three media types.
1.2.1 Floppy Disk Drive
The Floppy disk drive uses PC compatible 3½-inch (89mm) floppy disks. These disks store 1.44 Megabytes of
data, which translates to approximately 700,000 data samples (each sample is 16 bits). The disks are magnetic
media and the drives are mechanical. This limits their use in harsh environments where vibration or temperature is
a problem, but they are convenient as they are very inexpensive and can simply be plugged into any IBM
compatible PC.
1.2.2 PCMCIA Memory Card
The PCMCIA card (Personal Computer Memory Card Interface Association) is a small solid state card containing
Flash memory. The PCMCIA drive supports memory cards up to 200 Megabytes in size, which is approximately
100,000,000 data samples. The cards do require a special drive in your PC which is readily available, inexpensive
and already standard on many laptops and desktop machines. Contact your dealer for details.
NOTE: It is recommended that a Sandisk 4-200 Meg Flash card be used.
1.2.3 Zip Drive
The Zip Drive stores 100 Megabytes of information, which is approximately 50,000,000 samples for a 100
Megabyte Disk. This Recorder does not support 250 Megabyte Zip drives.
1.3 Recorder Options
Additional functions and capabilities can be added to the Data Recorder as options and are described in the
following paragraphs.
1.3.1 Digital Input and Output
This option provides three or six mechanical form C (Normally Open, Common, Normally Closed contacts), relay
outputs capable of switching 250 VAC as well as three isolated digital control inputs, or three or six solid state open
collector type contacts able to switch 30 Vdc at 0.5 Amps. The relay outputs can be programmed to respond to
alarm events while the digital inputs can be used to trigger events such as changing recording speeds or stopping
and starting recording.
1.3.2 Communications Interface
There are two communications options available:
1. An isolated RS232/RS485 interface using the Modbus protocol with the recorder acting as a slave device in
a Modbus RTU or Modbus ASCII environment.
2. An 802.3 compatible Ethernet option with RJ45 four wire connection.
Full scale ranges: ±150mV, ±1.25V and ±2.5V Accuracy ±0.06% ±12.5V, and ±25V Accuracy ±0.1%
DC Current: 4 to 20mA, 0 to 20mA and 10 to 50mA. Accruacy ±0.15% using external 50
2,3, or 4 wire connection. Cable compensation to ±50 ohm
Open and short circuit detection.
10 ohm Cu
100 ohm Pt 385
100 ohm Pt 392
200 ohm Pt 385
200 ohm Pt 392
120 ohm Ni
Input Resolution0.0015% of full scale, 16 bit unless otherwise stated
Input Impedance>10 megohms on 150mV, 1.25V and 2.5V Ranges, >100K on 12.5V, and 25V ranges
Input Channels2,4, or 6 direct
Maximum Input 50 Vdc
Isolation 250 Vdc or peak AC channel to channel, 300 Vdc or peak AC to Chassis
Measurement Rate 8 times per second on all direct input channels
Common Mode Noise
Rejection
Normal Mode Noise
Rejection
Math Functions+, -, x, /, logarithms, totalization, powers, averages, and timers
RECORDING
Recording RatesUser programmable from 8 samples per second to 1 sample every 600 seconds (10 Minutes).
Data FormatProprietary binary format. User File Naming.
Data Storage
Capacity
Removable Medial
Types
Internal Media One Megabyte RAM (Non-volatile)
File typesData files, Alarm and Event file, Configuration files, Language Files. Multiple files of different
DISPLAY
Display Type MonoCCFL backlit STN Liquid Crystal Display (5.0 inch) with touchscreen control.
Resolution Mono 240 (H) X 128 (V) pixels
Display Type Color CCFL backlit Active Matrix TFT Liquid Crystal Display (5.6 inch) with touchscreen control.
Resolution Color 320 (H) X 240 (V) pixels.
Display ModesGraphics (Trending Vertical or Horizontal), Bar Graphs (vertical or horizontal), Digital Meter,
Virtual Chart SpeedProgrammable: 0.5in/hr to 600in/hr (10mm/hr to 15000mm/hr)
Virtual Chart Scales2 sets of 8 scales
Display WindowsTime/Date, Graphics (Bars, Large Digital, Trends), Disk Status, System Status, Menu Button
Power Requirements100 to 240 Vac, 50/60 Hz, 35VA Max. Optional 24 Vdc ±15%
Power fail protectionProgrammed parameters stored in non-volatile memory. Clock battery backed. Data retention
OTHER
Operating Range -10°C to 50°C, 10% to 80% RH (5° to 40°C for floppy media)
SafetyMeets the requirements of UL 3111-1 and EN61010- 1 when installed in accordance with
1000 ohm Ni
>100 dB, 50/60 Hz
>50 dB at 50/60 Hz
Channels independently programmed
Data stored in non-volatile RAM and recorded automatically to on board removable media
3½ inch (89mm) floppy disk - approximately 700,000 samples for a 1.44 Megabyte Disk
PCMCIA Flash cards - approximately 100,000,000 samples for a 200 Megabyte Card
Zip Disk – approximately 50,000,000 samples for a 100 Megabyte Disk
Two Megabyte RAM (Non-volatile) - optional
names on a single disk.
Alphanumeric Alarm and Event data, or combinations on a split screen. Review trended data.
Bar, Unit Identification, Alarms/Events.
time without power > 12 months.
instructions in this manual
UL and cUL approved – File No. E175096
EMCMeets the requirements of EN61326:1998 and CE directive 89/336/EEC
R
S
B
C
-270 to –100
-100 to 1300
-50 to 1768
-50 to 1768
0 to 1820
0 to 2400
±2.5
1.5
o
-70 to 170
-220 to 850 oC
-180 to 820
-220 to 400
-180 to 400 oC
-70 to 300
-60 to 209
C
o
o
o
C
o
C
±3
±3
±4
±3
o
C (0 oC to 50 oC)
C
C
-450 to –150
-150 to 2372
-58 to 3200
-58 to 3200
32 to 3300
32 to 4350
-94 to 338
-364 to 1560 oF
-292 to 1500
-364 to 750
-292 to 750 oF
-94 to 570
-76 to 408
±5
±3
±6
±6
±7
±6
o
F
o
F
o
F
o
F
o
F
Input Voltage100 to 240 Vac, 50/60 Hz or 125 Vdc @ 35VA max.
WeightApprox. 7 lbs (3.2 kg) – weight will vary depending on options installed.
SAFETY NOTICE
This Safety Notice has been included to emphasize the DANGER OF HAZARDOUS VOLTAGES on the REAR
CF 46 5 2/06
TERMINAL PANEL of your instrument. USE EXTREME CAUTION WHEN INSTALLING OR SERVICING your
instrument. Pleas read the entire contents of the Installation and Wiring Chapter of this manual before attempting
to install or service your instrument.
WARNING
ELECTRIC SHOCK HAZARD
MAY CAUSE INJURY OR DEATH.
USE EXTREME CAUTION
WHEN INSTALLING OR SERVICING
REAR TERMINAL PANEL.
FOLLOW INSTRUCTIONS BELOW.
POWER INPUTS WARNING
When connecting power to the Rear Terminal Panel of your instrument, it is important to ensure that the AC mains
cable has an effective ground and provide a low impedance earth ground connection (Safety Ground) to the screw
terminal on the rear panel labeled “
exposed on the Rear Terminal Panel and is exposed inside the instrument case. When wiring, use the supplied AC
mains cable or recommended plug, make sure the HOT wire, or Line 1 is connected to L/H. Make sure the
NEUTRAL wire, or Line 2, is connected to N, and make sure a low impedance SAFETY GROUND wire is
connected to “
Use extreme caution when wiring signal input connections. Hazardous potentials may exist on signal input
terminals, which are floating, with respect to instrument ground. These hazardous potentials may be exposed
inside the instrument case and on the Rear Terminal Panel of your instrument. Any voltage potential at the signal
source will exist on the instrument’s respective signal input terminal: e.g. power generator stator winding
temperature-monitoring thermocouples.
Use extreme caution when wiring contact output connections. Hazardous potentials may exist on contact output
terminals, which are floating, with respect to instrument ground. These hazardous potentials may be exposed
inside the instrument case and on the Rear Terminal Panel of your instrument. Any voltage potentials at the
contact circuit will exist on the instrument’s respective contact output terminals: e.g. line-powered circuits.
” or “GND”.
” or “GND” to prevent the possibility of electrical shock. Power may be
SIGNAL INPUTS WARNING
CONTACT OUTPUT TERMINALS WARNING
2.0 Installation and Wiring
This chapter provides information and procedures on installing and wiring the Recorder. Included are handling
procedures, installation and wiring specifications, and instructions for both standard and optional equipment.
2.1 Equipment Handling
2.1.1 Initial Inspection
Exercise care when unpacking the instrument from the shipping carton. The instrument is packed in a shockproof
foam retainer to prevent damage during normal transit. If damage to the shipping carton is evident, ask the carrier’s
representative to be present when the instrument is unpacked.
2.1.2 Unpacking Procedure
To unpack your Recorder, first remove the foam retainer and instrument from the shipping carton. Then, carefully
remove the instrument from the foam retainer.
CF 46 6 2/06
2.1.3 Detected Damage
If damage is detected after unpacking the instrument, re-pack the instrument and return it to the factory as
described in the following section.
2.1.4 Equipment Return
Before returning a damaged or malfunctioning instrument to the factory for repairs, contact the sales organization
from which you purchased the instrument. A Return Merchandise Authorization number must be obtained from the
factory before returning an instrument for any reason.
2.1.5 Storage
For prolonged storage before installation, re-pack the Recorder in the shipping container. Cushion the Recorder
with foam molding or an equivalent and store in a cool, dry area. We do not recommend storage of the Recorder for
more than one year. If longer storage time is required, contact the factory for additional storage information.
2.2 Installation
The instrument is intended to operate in the following environment:
Indoor Use Only
Installation Category II per IEC 664
Pollution Degree Level II per UL3111-1/IEC1010-1
Temperature 5°C to +40°C (41°F to 104°F) per UL3111-1/IEC1010-1 with disk drive -10 °C to 50°C
with PCMCIA drive or Zip drive
Humidity 10% to 80% RH per UL3111-1/IEC10101-1
AC Mains supply 100 - 240 VAC~ 50/60 Hz 35Vac
NOTE: The recorder is designed to be panel mounted and as such should be considered as permanently
connected. Disconnection from the supply must be possible via a customer supplied switch or circuit
breaker. This disconnection device must be included in the panel installation and should be clearly marked,
in close proximity to the Recorder and easily accessible to the operator.
The Recorder is sized to fit in a standard DIN panel cutout of 5.43 inches x 5.43 inches (138mm x 138mm) and
requires only 7.41” (188.2 mm) behind panel depth not including power and input source cable space needed.
Actual dimensions of the instrument are shown in Figure 2-1.
Figure 2-1 Recorder Dimensions in Inches [mm]
WARNING: Do not use this instrument in any manner inconsistent with these operating instructions or
under any conditions that exceed the environmental specifications stated.
2.2.1 Panel Mounting
The Recorder should be mounted in a vertical panel to ensure proper operation. Ensure you have the proper
clearances and proceed as follows:
2.2.1.1 Cut a panel opening 5.43 x 5.43 inches (138mm x 138mm).
2.2.1.2 Remove any packaging material from the recorder. Always handle the unit carefully to avoid damaging the
LCD display or scratching the display surface.
CF 46 7 2/06
Figure 2-2 Front Panel Insertion
2.2.1.3 Remove the locking bars from the Recorder enclosure and ensure the gasket is not twisted on the
enclosure.
2.2.1.4 Insert the Recorder, rear end first, into the panel opening from the front of the panel. Ensure the
gasket is between the front bezel of the Recorder and the front panel.
2.2.1.5 With the Recorder held firmly in place against the panel, install one of the locking bar assemblies
by sliding the locking bar notch into the slot on the side of the Data Chart enclosure.
Figure 2-3 Rear View Panel Installation
2.2.1.6Using a slotted screwdriver, tighten the screw until the locking bar is just pressing against the
panel.
CF 46 8 2/06
2.2.1.7 Install the other locking bar assembly into the slot on the opposite side of the Recorder enclosure
and tighten as before.
2.2.1.8Using the screwdriver, tighten both screws so that the Recorder is held firmly in place. Do not
over tighten.
2.3 Wiring Specifications and Procedures
2.3.1 Power Requirements
The Recorder operates on any voltage from 100 to 240 VAC ±10%, 50/60 Hz enabling it to be used in most
countries. The maximum apparent power required by the unit is 35 VA.
2.3.2 Power Connections
NOTE: The Recorder is designed to be panel mounted and as such should be considered as permanently
connected. Disconnection from the supply must be possible via a customer supplied switch or circuit
breaker. This disconnection device must be included in the panel installation and should be clearly marked,
in close proximity to the recorder and easily accessible to the operator.
All connections to the Recorder are made to the Rear Terminal Panel. Any wiring carrying hazardous voltages
must conform to all applicable local and national safety codes. AC Mains connection is via an internationally
accepted IEC 320 AC mains connector or screw terminal (Figure 2-4 and Figure 2-5).
WARNING
Figure 2-4 AC
Connector
Ensure all mains power is turned off before proceeding with installation. This unit is
provided with a mating connector for the ac power socket or with a compatible three wire
grounded cable that may be terminated with a plug. Always ensure the ground wire (green
or green and yellow) or ground pin of the plug, is connected to a low impedance safety
ground (earth) within the ac power distribution system you are using. Always use the
recommended mating connector and an approved three wire cable to connect this unit to
the ac mains. Always provide a low impedance safety ground wire to the ground lug on the
rear panel marked
Figure 2-4 shows the IEC 320 AC mains socket on the rear of the Recorder. The center pin
is the ground termination. If a mating plug is provided, it will be marked with the Ground,
LINE (L) or hot, and NEUTRAL (N) or return. In the United States, an approved cable with
integral plug (NEMA 5-15 P) is provided. In some instances, a cable with no plug may be
provided. In this instance, the user must connect an approved plug to the cable prior to
connecting to the AC source.
.
Figure 2-5 shows the screw terminal power connections on the rear of the Recorder. The
right terminal is ground, the center terminal is LINE (L) or hot and the left terminal is
NEUTRAL (N) or return. The wire color codes are as follows:
COUNTRY
USA WHITE BLACK GREEN
EEC BLUE BROWN GREEN/YELLOW
Figure 2-5 Screw
Terminal
Figure 2-5a
Ground Lug
This unit is equipped with an AC mains fuse internally. If this fuse should blow, it generally indicates a serious
problem with the Recorder. THE FUSE SHOULD NOT BE REPLACED BY AN OPERATOR. The fuse is a quick
acting 5 x 20mm type rated at 2.0 Amps 250 VAC (~).
CF 46 9 2/06
Figure 2-5a shows the ground lug on the rear panel. This screw terminal must be
connected to an earth wire which in turn is connected to the ground or earth of the AC
power distribution system.
NEUTRAL (RET) LINE (HOT)GROUND
An optional AC mains plug retention clip is available - contact the factory.
2.3.3 Signal Input Wiring
SIGNAL INPUT CONNECTIONS. HAZARDOUS POTENTIALS MAY EXIST ON SIGNAL INPUT
TERMINALS WHICH ARE FLOATING WITH RESPECT TO CASE GROUND. THESE
HAZARDOUS POTENTIALS MAY BE ON THE REAR TERMINAL PANEL OF YOUR
INSTRUMENT. ANY VOLTAGE POTENTIAL AT THE SIGNAL SOURCE WILL EXIST ON THE
INSTRUMENT’S RESPECTIVE SIGNAL INPUT TERMINAL. (I.E. POWER GENERATOR STATOR
WINDING).
The Recorder accepts up to six direct inputs. Input connection is via plug in screw terminal
connectors on the rear panel. Inputs can be mixed in any combination of thermocouple, RTD,
milliamps, millivolts, volts or contact inputs. There is a common ground lug, marked with a for
connection of signal cable shields or screens. Read the following procedures before connecting
inputs to the terminals.
Figure 2-6 DFR 2000 Rear Panel Connections
WARNING: Ensure The Power Is Off Before Connecting Signal Inputs To The Unit.
The plug in screw terminal connectors are of the clamping screw variety, putting even pressure on the signal wire. It
is therefore not necessary to terminate the wires with lugs, however you may do so if you wish. The maximum
gauge wire that can be accommodated is 14 AWG or 2.5mm
2
.
You will need a small screwdriver and a pair of wire cutters and strippers. The use of shielded twisted lead wire is
recommended to minimize electromagnetically induced noise.
WARNING: All unused inputs must have all contacts commoned together.
CAUTION: Never run signal and power or control wiring together in the same conduit. This is to prevent
possible recording errors due to induced signals between lines. Route signal wires away from
power wires at the rear panel.
CF 46 10 2/06
NOTE: Ground cable shields at one end only to eliminate the possibility of interference due to ground loop
currents. When grounded transducers are used, the shield should be grounded at the sensor end
only.
Figure 2-7 Thermocouple, RTD and Linear Inputs
2.3.3.1 Thermocouple Inputs
Thermocouple input connections are made as shown in Figure 2-7 T/Cs. Note that a link must be installed between
the “-“ terminal and “C” terminal.
2.3.3.2 Resistance Temperature Detector (RTD) Inputs
Two, three or four wire RTDs may be used for connection with cable compensation ±50 ohms. Refer to Figure 2-7
and manufacturing specifications.
2.3.3.3 Linear Inputs
Current inputs: 4 - 20 milliamps, 0 –20 milliamps or 10 to 50 milliamps, using an external 50 ohm shunt.
Voltage inputs: ±150 millivolts, ±1.25 volts, ±2.5 volts, ±12.5 volts, ±25 volts and normally open/closed contact
inputs. Note that a link must be installed between the “-“ terminal and “C” terminal. Refer to Figure 2-7 for details.
2.3.4 Relay Output, Contact Input
WARNING!!!
To prevent the possibility of electrical shock, use extreme caution when wiring contact
output connections. Hazardous potentials may exist on contact output terminals that are
floating with respect to instrument ground. These hazardous potentials may be exposed on
the rear terminal panel of your instrument. Any voltage potentials at the contact circuit will
exist on the instrument’s respective contact output terminals (i.e. Line-powered circuits).
2.3.4.1 Mechanical Relay Option
The Recorder may be equipped with an optional Digital Input Output Board which has three or six potential free
Form C relay contacts and three opto-isolated digital inputs. A terminal block as shown in Figure 2-8 below, is
provided for the three alarm output Potential Free Form C connections: normally open (NO), common (C), and
normally closed (NC), and the three digital inputs which share a common. The relay contacts are capable of
switching 125 VAC ~ at 0.5 Amp or 30 VDC at 2 Amp. The potential free relay contacts are protected internally with
300 volt Metal Oxide Varistors (MOVs) to prevent contact arcing.
The Recorder may be equipped with an optional Digital Input Output Board which has three or six solid state open
collector outputs and three opto-isolated digital inputs. A terminal block as shown in Figure 2-9 below, is provided
for the alarm output connections. The output connections are polarized and are intended for DC operation only.
Reverse polarity protection is provided and each output is fused at 1 Amp. The fuse is solid state and will reverse
once the load is removed. The outputs are rated at 30 VDC at 0.5 Amp and are optically isolated from the
recorder.
Figure 2-9 Solid State Relay Connections (3 Channel Shown)
2.3.4.3 Opto-isolated Inputs/Outputs
The opto-isolated inputs require an external potential of 5 to 12 volts DC @ 10 milliAmps. The three inputs are
isolated from the unit, but not from each other as they share a common. The positive voltage connects to the
terminals marked 1, 2 or 3 and the common connects to the terminal marked C. It is possible to use potential free
contacts to operate the digital inputs. This requires opening the unit and setting jumpers on the relay board. Refer
to Appendix A for details.
2.4 Serial Interface Option
Data can be accessed to download Configuration files or Data files using the RS232C option and a modem. The
RS485 option allows the Recorder to be installed into an existing Modbus network or it can be used to connect to
an IBM PC compatible computer using a null modem cable and the RS232 can support cable runs up to 50 feet
[16m]. The RS485 connection is via two wire (twisted pair) cable (a DB9 Female connector is required) and can
support cable runs up to 4000 feet [1300m].
The Serial Interface contains an isolated switching unit for RS232 and RS485 access with a standard DB9 Female
connector. When switch 2 is in the ON position, RS485 is enabled. When switch 2 is in the OFF position, RS232 is
enabled.
When more than one recorder are connected in a series, it is necessary to apply a termination resistor on the last
recorder. Switch 1 in the ON position applies this necessary termination resistor and should be positioned only on
the last recorder in series.
CF 46 12 2/06
Figure 2-10 RS232/RS485 Modbus
All Serial Interface connections are made through the DB9 female connector.
The RS232 Connection to the DB9 female connector are as follows:
DB9 PIN CONNECTION DIRECTION
2 RxD Received Data In
3 TxD Transmit Data Out
4 DTR Data Terminal Ready Out
5 Common N/A
7 RTS Request to Send Out
8 CTS Clear to Send In
The RS485 (Half Duplex Mode) Connection to the DB9 female connector are as follows:
The 10 BaseT Ethernet option allows direct connection to a LAN or WAN and the standard TCP/IP protocol enables
the transfer of data over the Internet.
Figure 2-11 Ethernet Option
The RJ45 Connection for the Ethernet are as follows:
Receive Differential Pair
3 RX6 RX+
Transmit Differential Pair
7 TX-
8 TX+
Two LEDs on the rear panel show Ethernet activity. The Link LED indicates a good connection and the RxTx LED
indicates communication activity.
2.6 Cleaning
The unit may be cleaned by wiping with a soft cloth. The front panel and display / keypad may be wiped with a
slightly damp soft cloth containing soapy solution or a mild detergent. Do not use any lemon based (citric acid)
product to clean the display / keypad.
2.7 Contrast Adjust (Monochrome Only)
The contrast or viewing angle of the monochrome LCD display can be adjusted from the side panel. If not adjusted
correctly, the display may look dim, dirty, completely black, washed out or blank. Insert a small star or Philips type
screwdriver in the hole in the side panel as shown in Figure 2-12 and turn to adjust the display contrast. The
CF 46 13 2/06
adjustment potentiometer has a 270° rotation. Do not apply excess pressure or attempt to turn it beyond the end stops.
Figure 2-12 Contrast Adjust
3.0 Getting Started
The Instrument is an extremely versatile solid state data recorder. It has a liquid crystal display capable of complex
graphical representation and either a floppy disk drive, Zip drive or PCMCIA Memory Card for data storage. The
unit is easily programmable and the average user will probably never need to use most of the features or functions
available in the recorder. This chapter will give the user a brief system overview and guide the first-time user into a
simplified setup that will enable you to begin recording with the least amount of effort.
3.1 Moving About the Screen
The Recorder has an LCD Graphics Screen that also acts as a touch keypad. Areas of the screen are active as
push buttons, the exact areas that are sensitive depends on what is currently displayed. The user has only to lightly
touch the screen area depicting the button to activate the function. If the buzzer is turned on, the unit will provide
audible feedback, as a short beep, each time a “button” press is registered. The default display is shown in Figure
3-1 below.
Figure 3-1 Recorder Screen
The screen is divided into three distinct areas, the BUTTON BAR, along the bottom of the screen (containing the
time/date stamp), the STATUS LINE, across the top of the screen, and the GRAPHICS AREA between them.
Under normal operating modes, when not in a menu, the BUTTON BAR area is active for “Button” pressing. The
MENU button on the bottom right of the screen, will bring up the command menu button bar, see below, which
allows the user to do a number of functions and select the option of programming the unit. Refer to Chapters 4 and
5 for details.
Figure 3-2 The Command Menu Button Bar
CF 46 14 2/06
The VIEW button is on the lower left side of the screen, and it enables the user to scroll through the selectable
display options in the graphics area of the screen. These views might be charts, bar graphs, digital or Alarm
information. Each time the “button” is pressed, the next view is presented. The Graphics display can be either a full
screen of information, or the screen can be split, allowing combinations of the primary screens to be displayed. The
user can also choose whether the screen is horizontally or vertically oriented.
The area along the top of the display is the Status Bar or Status Line, and it is used to display a number of user
programmed functions such as the unit tag (identification) or the digital values of the various channels or alarm
status. On the right hand side of the Status Line is the disk status information (disk info). This displays the current
condition of the disk that is currently being used to save data. When the unit is not recording, this area shows REC OFF. When recording it displays REC↑ when recording at the fast rate or REC• when recording at the slow rate,
as well as XX% used, where XX is the amount of disk space already recorded. The default display is set using the
DISPL option on the Command Menu button bar. The JOG button is used to switch between channel data.
To the left of the Status Line is the position of the ACK button (not shown). This is the Alarm ACKnowledge and is
only present when there is an alarm condition, at which time it blinks until the user presses it to acknowledge the
alarm condition. Pressing the Acknowledge button will also reset any Output relays if this option is installed and
programmed. Note that the ACK button will always be the top-most button, always rising to the surface when
covered by other items such as menus. The user can thus acknowledge an alarm at any time, even while in the
programming mode.
As the user moves through the menu options, more or fewer buttons will be shown. The key buttons are always
displayed on the button bar along the bottom of the display. ENTER and EXIT buttons are always in the same place
on the button bar allowing rapid movement through menus.
The BROWSe button is only displayed on screen views that can actually be browsed. These are the full Chart
Screen and the full Alarm Status Screen. The user can choose to browse RAM (Random Access Memory) which is
the screen trace data, or File, in which case a file menu will be presented so that the user can choose a file from
the disk. The BROWS button also allows the user to SEARCH historic data either by TIME or VALUE, as well as
COMPRESS data in time.
Pressing the BROWS button puts a freeze on real time display and allows the user to scroll back in time to browse
through historical data that has effectively dropped off the screen or has been recorded previously, even on another
machine. The actual Chart Screen browse buffer is dependent on the memory options installed, the number of
traces being displayed and the effective chart speed. With the standard buffer, a chart speed of one inch per hour
with four traces active, it is possible to browse back around 140 hours (5½ days). The user may choose to browse
files on the disk rather than the current memory buffer even though the unit is currently recording. The user can
then browse files that had been recorded earlier, or may even browse the file that is currently being recorded. Once
the data has been recorded to disk, it is possible to use the file browser or the PC and the companion software to
view data as far back as the start of recording, irrespective of how long that may be, within the constraints of disk
capacity.
3.2 What the Screens Mean
The basic screen layout is shown in Figure 3-3 below. Each area is used to present different information to the
user.
3.2.1 Status Line
The Status Line is used to show the Unit Tag, Point information, rear terminal
Junction Temperature or Alarm information. The choice as to what is transiently
displayed is set in the DISPL menu (Chapter 4.2.1) or the user can set the
default display for this line in the “PROGram - Display - Powerup disp” menu
(Chapter 5.4.3). At any time the user can press the JOG button to display point
data on the Status line. Each time JOG is pressed the next point data will be
shown. The data can be set to auto jog from the “PROGram – Display - Powerup
disp - Autojog” menu (Chapter 5.4.3.2).
Figure 3-3 Basic Screen Layout
CF 46 15 2/06
The Point information is displayed on the Status line as “ Point No. Value Units” as shown below:
1 0.496 VOLTS
The Alarm information is shown on the Status line as “Point No. Value Alarm Type/Number” - where H1 is
1 0.496 HI
will show –
The Unit Tag is shown on the Status Line and may be up to 20 characters long. It is entered from the “PROGram Display - Powerup disp - Unit Tag” menu (Chapter 5.4.3.1) and displays as entered –
High Alarm #1. There are 5 possible alarms per channel, H = High, L= Low R = Rate. If more than
one alarm is active they will cycle on the Status Line. If there are no alarms active the Status Line
NO ALARMS
This is a UNIT TAG
3.2.2 Disk Status
The area in the top right corner marked DISK STATUS is used to show Disk Status. The following messages may
appear in this area:
OFF 15% Unit is not recording to disk. Record Mode is OFF - Disk is 15% full.
REC ↑ 15% Unit is recording to Disk in HIGH speed mode - Disk is 15% full.
REC• 75% Unit is recording to Disk in LOW speed mode - Disk is 75% full.
FORMAT Disk is being Formatted.
SAVE CFG Saving Configuration to Disk.
LOAD CFG Loading Configuration from Disk.
TRIG 75% Unit is waiting for an Event or Alarm to turn the Record Mode on - Disk is
75% full.
3.2.3 Button Bar
The Button Bar is the area across the bottom of the screen where the main menu buttons appear. The function of
these buttons varies according to which mode the unit is in. Figure 3-3 shows the Normal mode button layout.
3.2.4 Date/Time Window
The Date /Time window sits at the bottom of the screen in the Button Bar and continuously shows the current date
and time. Date is shown numerically in either American Month/Day/Year or European Day/Month/Year format. The
format is selected in the “PROGram - Display - Time format” menu (Chapter 5.4.2)
3.2.5 Graphics Window
The graphics window is used to display various information in different formats. The VIEW button is used to switch
between the various graphic displays, which can be Trend Charts, Bar Graphs, Digital Windows or Alarm/Event
data, or combinations of these. Furthermore the graphics can tend to be horizontally or vertically oriented. The
orientation is selected in the “PROGram - Chart/Pens - Direction” menu (Chapter 5.5.4)
3.2.5.1 Bar Charts
Bar graphs can be displayed on their own or as part of a split screen. Up to twelve can be displayed
at a time. The user can select to display bar graphs either horizontally or vertically. The assignment
of points to bar graphs is made in the “PROGram - Display - Bar assign” menu (Chapter 5.4.4). The
format of the vertical bar graph is shown opposite in Figure 3-4. The very top of the bar has the
actual/real time digital value of the point, in this case 0.385. Immediately below this is the
Engineering Units shown as ENG UNITS. There can be a maximum of five characters. At the very
bottom of the bar graph is the POINT TAG. This is a ten-character description of the point.
NOTE: Not all characters can be displayed on the bar graph.
Figure 3-4
Bar Chart
CF 46 16 2/06
Between the ENG UNITS and POINT TAG is a bar that represents the actual value as a function of
the full scale value. To the right of the bar are scale values. The alarm setpoints, if any are set, are
indicated on the bar by a “◄” in the text area. If any alarm is active, the bar will be blinking. The
Engineering Units, Point Tag, alarm setpoints and scaling of the bar is done in the “PROGram Points” menu (Chapter 5.6)
3.2.5.2 Digital Windows
Digital Windows can be displayed on their own or as part of a split screen. Up to twelve
can be displayed at a time. The assignment of points to digital windows is made in the
“PROGram - Display - Digital assign” menu (Chapter 5.4.5). The format of the digital
window is shown opposite in Figure 3-5. The large numbers in the center of the window
are the real time point value. Above this point value is the Point Tag shown as POINT
Figure 3-5 Digital
Window
TAG. This is a ten character description of the point. Below the point value is the
Engineering Units shown as ENG UNITS. There can be a maximum of five characters.
The Engineering Units, Point Tag, and scaling of the digital value is done in the
“PROGram - Points” menu (Section 5.6).
3.2.5.3 Alarm/Events Data Window
Figure 3-6 Alarm/Event Window
The alarm/event window is used to display alarms, events, reset power
up and power down information. This data may also be recorded to disk
by enabling Alarm data in the “PROGram – Disk functions - Alarm
on/off” menu. (Chapter 5.7.2).
The second line of the display indicates the status of the alarm
checking. If alarm checking is enabled this will indicate ‘Alarm Check
On’ or it will indicate ‘Alarm Check Off’ in which case no alarms will
occur.
The state of alarm checking is changed in the “FUNCtion - Alarm
Check” menu (Chapter 4.2.3.7).
The format for any entry in the Alarm/Event file is – Date Time Point Status Value. The Point value is shown as
Pt X, where X is the point number. The Value shown is dependent on the Event. The Event can be an Alarm, an
input contact closure (event) or a reset, either automatic or manual.
For linear inputs, the alarm event description is HIGH1 or LOW2, where the number is the alarm count up to a
maximum of 5. For conditional inputs, the alarm event descriptions are OPEN, CLOSE, TRUE or FALSE. The value
is the actual value at the time the alarm was registered. An event message of “*” is an alarm return to normal
condition, the time the point came out of alarm.
Resettable points, such as totalizers, will show the event as RESET, with the actual value at the time of reset.
Some Resettable point types, such as HI PEAK will have two entries (Pt 6 above). The first entry is the date and
time that the high peak occurred, with the peak value, the second entry, above it, is the actual time the reset
occurred, with the value at that time. Note: Resettable points will not print unless the Reset Print option is turned on
in the “PROGram - Points” menu (Section 5.6).
For external events via the digital inputs, the user can define separate messages for each of the three inputs, one
for input activated (Close) and one for input deactivated (Open). These event messages can be entered, up to a
maximum of ten characters, in the “PROGram - Digital I/O - Event msgs” menu (Chapter 5.9.3). This message will
appear in the log as - date - time - Switch number (Sw X) – Event message. Default event messages are CLOSE
and OPEN as seen in the data log example above.
3.2.5.4 Trend Window
The trend window is the one that looks like a “paper” recorder. It has traces or pens and emulates the paper chart,
moving the “paper” across the screen, and is shown in the vertical mode in Figure 3-7 below. The direction of the
trending can be vertical, from top to bottom, or horizontal, from right to left.
The direction can be changed in the “PROGram - Chart/Pens - Direction” menu (Chapter 5.5.4). There may be as
many as twelve pens on the chart at a given time. Pens are assigned to the chart in the “PROGram - Chart/Pens Pens” menu (Chapter 5.5.3).
Consider the vertical chart below, the same features are found on the horizontal chart. Along the top of the chart
are the pen pointers, one for each pen that is on. These track the real time value of the points and identify the origin
CF 46 17 2/06
of the trace.
Figure 3-7 Vertical Trend Window
There are grids on the screen, both horizontal and vertical. In the above example, the vertical grid spacing is a
function of the divisions on the chart scales. The horizontal grid lines indicate the scales for the chart, and if there is
more than one scale set, they alternate. On the full screen trend view, the scales can be toggled through by
pressing the middle of the scale (top center of the display). Each scale has its end points marked along the top of
the chart with the scale value, the 0.00 and 1.00 indicating that the trace has a value of zero when it is hard to the
left, and 1.00 when it is hard to the right. There is a marker on top of the pen pointers to indicate which pens
reference the current scale. Thus when interpreting the data for pens 1 and 2 use a value of 0 to 1.00 full scale.
When the scale changes, so will the markers on the pen pointers to indicate the pens for the next scale. Also along
the top of the chart are the scale UNITS.
3.2.5.5 Transient Windows
Transient windows are those that appear momentarily. They pop up over any
existing window to inform the user of a problem or of a background task being
completed. They require a user response, normally pressing the “OK?” button,
to acknowledge the message.
Figure 3-8 Transient Window
3.3 Browsing, Compressing and Searching Data
Once at the full trend screen (using the VIEW button), the user can choose to browse, search or compress either
data from the disk, including the file currently being recorded to disk, or immediate past data which is buffered to
memory (RAM). The operation of browsing is the same whether the user is browsing memory or file data. When
BROWS is pressed a new button bar is presented with three buttons, FILE , RAM and EXIT as shown in Figure 3-9
below.
Figure 3-9 The Browse Source Button Bar
Select FILE to browse prerecorded files on the disk, RAM to browse Trend data from memory, or EXIT to return to
the real time view.
Browsing allows the user to view historic data of the pens currently trending on the display or saved on the disk,
without affecting any real time data acquisition. Alarm data can also be browsed from memory by selecting BROWS
on the full view Alarm window. See section 3.3.5 for the added step required to browse from disk.
On the Trend screen, pressing BROWS brings up the Browse Source button bar. Press RAM to browse memory or
FILE to browse from disk. This brings up the Browse Mode Button Bar shown below (Figure 3-10).
CF 46 18 2/06
Figure 3-10 The Browse Mode Button Bar
The FINDV and FINDT buttons initiate the SEARCH functions. FINDV is FIND by V
The “→←“ and “←→” are to COMPRESS and EXPAND data respectively. BROWS enters the interactive Browse
Screen and EXIT returns to the prior screen.
alue, FINDT is to FIND by Time.
3.3.1 Compressing Data
The data on the screen may be compressed up to 32 times the normal view, this enables long term trends to be
seen on a single screen. Each time the “→← “ (Compress) button is pressed the data is compressed by a factor of
2, and it may require a short time for the recorder to process the data. Five presses give the maximum 32 times
compression. The data may be expanded back by pressing the “←→” (Expand) button. The user can enter the
interactive browse mode with any level of compression by simply pressing the BROWS button.
3.3.2 Searching Data By Time
The historic data can be searched by time, by pressing the FINDT (FIND Time) button. This will bring up a transient
window showing the time to be searched. If the displayed time is correct press ENTER, if not press NO. This will
bring up a series of windows to allow the required time to be entered. The Date will then be displayed. Press NO to
change or ENTER to accept. The data will be searched and the sample corresponding to the required time and
date will be placed under the cursor near the center of the screen. The search may take some time, especially if
searching a large disk file. If no point is found the unit will display “No Point Found”. The user can then enter the
interactive browse mode by pressing the BROWS button.
3.3.3 Searching Data By Value
The historic data can be searched by value, by pressing the FINDV (FIND Value) button. This will bring up the
Search Value button bar shown in Figure 3-11 below.
Figure 3-11 The Search Value Button Bar
To set up the actual Value to search for, press the VALUE button. Use the numeric keypad to enter the value to
search for. Note that you cannot search for an exact match, you will be looking for a value immediately greater than
or less than the value you enter. Once you have entered the value you want, press the ENTER button.
Before you begin a search, you have to decide which PEN or point you are going to use as the search reference,
and which direction you wish to search in.
Use the PEN button to select the pen trace to search. Each time this button is pressed, the next pen is selected.
The current pen is shown in the status line at the top of the screen, together with the value of the trace under the
cursor and its actual time and date stamp.
The search direction is determined by the button to the left of the PEN button. This button toggles between BACK
and FWD. BACK will search BACKWARDS in time from the current cursor position (into older data), while FWD will
search FORWARDS in time from the current cursor position (into newer data).
To initiate a search press either the FIND> button to find the first point GREATER than the search value, or FIND<
to find the first point LESS than the current search value. Each time either of these buttons is pressed, the next
point that meets the search criteria is found. Note that once a point is found, the next point to be found will be the
one that is after the first point that does not meet the criteria. The search “hops” across all other values so that you
are not bogged down finding useless information. Therefore if you are searching for a point on a sine wave, for
example, you will find the same point on the same phase of the wave (360° apart). Otherwise you would find every
point less than or greater than the search point which could be every point in the trend.
The point that meets the search criteria will be placed under the cursor near the center of the screen. The search
may take some time, especially if searching a large disk file. If no point is found the unit will display “No Point
Found”. The user can continue to search for other points in the same direction or change direction and search
again. Once an acceptable point is found, the user can enter the interactive browse mode by first pressing the EXIT
button to return to the Browse mode button bar and then pressing the BROWS button.
CF 46 19 2/06
3.3.4 Interactive Browse
The interactive browse mode enables the user to uniquely identify points on the screen by time, date and value,
and move around in time. Once the Browse source has been selected, FILE or RAM, pressing the BROWS button
again, immediately or after a search or compress operation, will enter the Interactive Browse mode, and the Browse
Button Bar shown in Figure 3-12 below will appear.
Figure 3-12 The Browse Button Bar
The cursor appears near the center of the screen and it can be moved with the UP Arrow and Down Arrow keys in
the Vertical mode or with the Left Arrow and Right Arrow keys in the Horizontal mode. Pressing these keys a single
time, moves the cursor one sample at a time, holding these keys will cause the cursor to move in 10 sample
increments. As the cursor moves, the time and date information is updated in the status window at the top of the
screen. To move about more speedily, the PAGE- and PAGE+ buttons move the data forward or backward an
entire screen at a time.
When browsing graphics, the Status Bar at the top of the screen shows the Date and Time stamp and actual value
for the trace data directly under the cursor. Note that if you are browsing Memory there will be an “M” in the right
hand corner of this line or if browsing Files, there will be an “F” as follows:
07/27/96 17:03:06 point 1: 0.318 Volts
To see the actual pen data value press the PEN button to toggle the data through each trace if more than one trace
is displayed. As the data is toggled, the traces for all pens but the selected one, disappear. Press EXIT to return to
real time viewing from the RAM browser or to the directory from the FILE browser.
F
3.3.5 File Browsing
File Browsing can be done at any time, note however that if you are browsing from disk while recording, there may
be delays in retrieving data from the disk as writing to disk has priority over reading from disk. Press BROWS to
bring up the Browse Select Button Bar and choose FILE, to browse files from the disk.
Note: Ensure that a disk is present and that it has suitable files or you will get an error message.
Once FILE has been pressed the unit will check the disk then bring up a directory. If there is more than one file per
point, the files for that point will be displayed one under the other with an “<” indicating the current file selected as
shown below. The list will scroll if it is longer than what will fit on the screen.
File Directory
P o i n t N o : 1
B AT C H 1 . D T 1 <
B AT C H 2 . D T 1
B AT C H 3 . D T 1
If only one file is available per point, the directory selection will default to that file, if no files are available for that
point, the directory will indicate “No Files Found”. Use the POINT button to select the point you want.
Once you have the directory listing of files for the specific point, use the ↑ (UP Arrow) and ↓ (Down Arrow) keys to
select the file you wish to browse, indicated by the “<” mark. Press the ENTER button to browse the selected file.
Once the file is loaded, it is browsed, compressed or searched in the same manner as the memory browser
described above. Press EXIT to select a different file or EXIT again to return to real time viewing.
3.4 Getting to the Points
The Instrument is a 15 point recorder of which up to 6 channels may be direct or real world inputs. These are
typically voltages or currents. Any channel not being used to record or display live inputs may be used as a
computational channel. Inputs can be conditioned or scaled to display any range of engineering units. Refer to
CF 46 20 2/06
Figure 3-13 below.
Figure 3-13 Data Flow
There are four distinct levels of data handling, namely conversion, conditioning, scaling and display/record.
Conversion applies only to live inputs and is the process of converting real world analog signals into a 16-bit digital
value that can be used by the Recorder. There are five full-scale ranges for all conversions, 150 milliVolts, 1.25 V,
2.5 V, 12.5 V or 25 V, and DC. The converted values pass to a conditioning block which converts the binary value
which is effectively a percentage of full scale, into a value useful to the user. Conditioning includes converting the
binary value into a representative voltage, conditioning and linearizing this voltage to represent for example, a real
world temperature as might be input by a thermocouple or RTD, and applying any other computation as required.
The conditioning block handles 15 channels. The outputs from the conditioning block are referred to as base points and may be fed back to the inputs of the conditioning block to form the basis (base point) for other base
point computations. This includes tracking peak or valley values, doing moving averages, timed averages,
difference, totalization or any other user-entered equation.
The input to any channel in the conditioning block can be any one of the base points, or it can be the live inputs.
Live inputs can have input and output scaling applied in the conditioning block unless this is pre-defined by the
nature of the input, i.e., thermocouples. An example of input/output scaling is given in section 3.5.
The outputs of the conditioning block are also the values represented in the Digital Windows and are real-world
engineering values. The outputs of the conditioning block are also fed through to a scale block where a chart scale is applied to each point. The chart scales determine what part of the full scale range will be used by the
display/record block. There are two sets of 8 scales available to the user. The second set is an alternate set which
can be applied to the base points in place of the normal set when triggered by an external event. Each chart scale
may be applied to any one or more of the base points to provide an output scaled point. These scale points are
then applied to the display/record block which consists of the visual information or recorded information that the
user requires. Up to twelve scaled points are applied to the bar graph display. Up to twelve may be displayed on
the chart and up to fifteen may be recorded to disk. Note that points recorded to disk need not be the same as
points displayed on the bar graph or the chart, however, in the case of the chart, any point which is not recorded to
CF 46 21 2/06
disk will have limited browse capability. The user should also be aware that applying scaling other than full scale
output to recorded data limits the data to the range between scale endpoints. It is possible to store the base point
with different scaling than is used to display it on the chart using one of the computational channels.
3.5 Using Chart Scales
To better understand the use of input and output scaling, and how scaling base points affects the data, consider the
following example, highlighted by Figure 3-14.
Figure 3-14 Scaling Example
Assume the user wants to record the output from a pressure transducer that gives a 0 to 5 volt DC output signal.
This coincides to a pressure of 0 to 3000 pounds per square inch (PSI). The process being monitored typically runs
at 2200 PSI ±10%, this is the area of interest. The output of the transducer that peaks at 5 volts is connected to live
input channel 1 and the 12.5 volt full-scale range is selected to cover this range. Since the maximum input voltage
will be only 5 volts on a 12.5 volt range, use the input scale option to set the low end at 0.0, and the high end at
5.000 volts. To convert this directly to PSI, the output scale is set at low point 0.0 and high point 3000.0. The base point now becomes 0 to 3000 for an input of 0 to 5 volts, scaled linearly across the range. The engineering units
can be set to PSI and the point tag can be set to any label that identifies the process. All of this is accomplished in
the Points menu - Section 5.6. Apply scale A to the base point, this too is selected in the Points menu.
It is then necessary to define chart scale A to suit the requirement - Section 5.5.2. In order to maximize the display
resolution, the user is interested in pressure ranges from 2000 to 2500 only. Scale A is thus set for a low end of
2000, a mid-range of 2250 and a high end of 2500. This scale point is now assigned to the bar graph and the chart
graph. The chart will thus display from 2000 to 2500 as will the bars, maximizing the display resolution for the value
of interest. The user can choose to record this particular value, in other words, what you see on the chart is what is
recorded to disk, values from 2000 to 2500. Or, if so desired, the user can choose to record a second point that will
show pressures from 0 to 3000. To do this, use the channel 1 as a base point for a computational channel and
multiply this value by 1 which is then recorded to disk.
3.6 Programming
While the unit may be programmed remotely, this exercise deals with manually programming the unit from the front
panel using the display and the touch screen keypad. Various buttons and keys will be presented on the display.
The user simply makes a choice by pressing the appropriate key or button. In some instances, direct selection is
not possible but rather a series of choices is made by using an UP or DOWN arrow button, pressing ENTER or
EXIT. Once programming is complete, the unit automatically LEARNS the current setup, which is stored in
nonvolatile memory and remembered indefinitely.
When powering up the unit for the first time or if the data in the nonvolatile memory should become corrupt, the
user will be asked to perform a Smart Initialize, select No. The user will then be requested to perform a full
initialization. Select Yes at this stage to reset all values to the default.
There are many parameters that can be programmed, including time and date, the display, the virtual chart and
pens, the points, the recording capability, the alarms, the digital outputs and the communication port.
CF 46 22 2/06
Loading...
+ 81 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.