Cooper Instruments DFI-INFINITY CS User Manual

DFI INFINITY CS Quick Start Manual
Use this Quick Start Manual to set up your Strain Meter and begin operation. Information is provided on how to:
• Connect ac power • Set basic options for operation Connect the sensor • Scale the meter.
Features with are for the “B” version which has three­color programmable “Big” LED display - All segment characters shown are for the “B” version.
NOTE: This Quick Start Manual includes specific configuration parameters for bridge sensors with an output range of 0–100 mV and 10 V excitation. Other sensor types may require different parameters or additional ones. When this is the case, we refer you to the Operator's Manual for detailed instructions.
Safety Consideration
This device is marked with the international Caution symbol.
The instrument is a panel mount device protected in accordance with Class I of EN61010-1 (AC units) and Class III (DC units). Remember that the unit has no power-on switch. Building installation should include a switch or circuit-breaker that must be compliant to IEC 947-1 and 947-3.
SAFETY:
• Do not exceed voltage rating on the label located on the top of the instrument housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
EMC:
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wire close to the instrument if EMC problems persist.
Mount the Unit
1. Cut a panel opening using the dimensions shown to the right.
2. Position the unit in the opening, making sure the front bezel is flush with the panel.
3. Install retaining clips on both sides of the meter and tighten against the panel.
Wiring
Warning: Do not connect AC power to your device until you have completed all input and output connections. This device must only be installed by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!
1. Remove the panel at the back of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each terminal as shown in the following figure and tighten the lockdown screws.
4. Tug gently on the wires to verify the connections.
2
External Fuse Required:
Time-delay, UL 248-14 listed Time-lag, IEC 127-3 recognized
175 mA (115 Vac line) 125 mA (115 Vac line) 80 mA (230 Vac line) 63 mA (230 Vac line)
AC Powered Unit Connections
10-32 Vdc, 7.5 W
When using DC power, do not use internal excitation or Isolated Analog Output for high color brightness. For low or medium brightness, internal excitation is limited to 24 V @ 25 mA; 5 V, 10 V, 12 V @ 35 mA.
In order to maintain the same degree of protection as the AC units, always use a Safety Agency Approval DC source with the same Overvoltage Category and Pollution Degree.
DC Powered Unit Connections
Connect the Sensor
1. Locate the TB2 connector on the rear of the unit.
2. Attach the sensor wires and tighten the lockdown screws. The diagram below shows the wiring for bridge sensors with internal excitation.
NOTE: Refer to the Operator’s Manual for setup requirements for other sensor types.
Bridge Sensor with Internal Excitation
3. Tug gently on the wires to verify the connections.
4. Replace the panel at the back of the unit.
34
Using the Configuration Menu
To configure the meter, you use the buttons on the front panel.
To: Take This Action:
Display the Press the MENU button. The first function Configuration Menu on the menu, INPT, displays. Select a submenu 1. Press MENU until the function you
function want is shown.
2. Press /TARE.
The information you can change flashes.
Select a value 1. Press /NT/GRS to display the option for that submenu you want.
function 2. Press MENU to store it.
STRD quickly flashes, indicating that the selection has been stored in memory. Then the next menu function displays.
Go back to previous Press RESET once. menu function
Exit the Press RESET twice. The unit displays Configuration RST as it reinitializes. When a numeric
Menu value displays, the unit is in run mode.
(Optionally, you can press MENU to move through all the menu functions until the unit reinitializes.)
MENU SUBMENU /TARE DESCRIPTION
INPT 100M , ±50M,!1 0V , !± 5V , 0-20*Input DEC.P FFFF*, F.FFF, FF.FF, FFF.F Decimal Point RD.S.O IN!1, RD!1, IN!2 , RD!2 Scale and Offset RD.CF R.1=T*, R.1=N
R.2=0, R.2=1 , R.2=2, R.2=3, R.2=4 R.3=F*, R.3=U
R.4=P, R.4=G COLR GRN , RED, AMBR Display Color S1.CF S.1=A*, S.1=B
S.2=U*, S.2=L
S.3=N*, S.3=G S2.CF S.1=A*, S.1=B
S.2=U*, S.2=L
S.3=N*, S.3=G S1.DB 0003 S2.DB 0003 OT.CF O.1=E*, 0.1=D Analog Output
OT.S.O RD!1, OUT1, RD!2 , OUT2 LK.CF RS=E*, RS=D
BRIT M.BrT , L.BrT, H.BrT
* Factory Default Settings
*
*
O.2=C*, 0.2=V
O.3=A*, 0.3=P
SP=E*, SP=D
L3=0*, L3=1
shown if R.4=G
shown if R.4=G
Reading Configuration
*
Setpoint 1 Configuration
Setpoint 2 Configuration
Setpoint 1, Deadband
Setpoint 2, Deadband
Configuration
Output Scale & Offset
Lockout Configuration
Display Brightness
To Set the Input Type
1. Press MENU until the unit displays:
2. Press /TARE. The unit displays:
3. For this application you want 100M. If 100M is not
displayed, press /NT/GRS until it appears. Other choices are 50M, 10V, 5V and 0-20.
NOTE: Refer to the Operator's Manual for more information
on changing ranges.
4. Press MENU to select the sensor shown. The meter displays the next menu item. If you changed input type, the meter displays:
To Set the Decimal Point
1. If it's not already shown, press MENU until the unit displays:
2. Press /TARE. The unit displays:
Press /NT/GRS to move the decimal
3. point to the desired location. The
factory settings is FFFF. The other choices are F.FFF, FF.FF, and FFF.F.
4. Press MENU to select the decimal point position shown. The unit displays:
5678
To Scale the Meter
You can scale the meter in one of two ways:
1. With a known load — This method uses input (load) information sent from another device such as a scale or a simulator for voltage or current.
2. Without a known load — This involves calculating the load based on transducer specifications and manually entering it to the meter.
For both methods, you must first identify the minimum input load IN!1 and the corresponding display reading you want RD!1 . Then you identify the maximum input load IN!2 and its corresponding display reading RD!2 .
The decimal point is for display purposes only — you set it where you want it to display for your application. When entering IN!1 and IN!2 values, ignore any decimal point on the display. However, you must enter RD!1 and RD!2 values with the decimal point in the desired position.
Scaling With Known Loads
To identify the minimum known load (IN!1 and RD!1
1. If it's not already shown, press MENU until the unit displays:
2. Apply the minimum known load (0%).
3. Press /TARE. The unit displays:
4. Press /TARE again. The unit displays the last setting for
IN!1
.
):
Scaling With Known Loads (Continued)
6. Press MENU to store IN!1 . The unit
displays:
7. Press /TARE. The unit displays the
last setting for RD!1 .
8. Change RD!1 as necessary:
• Press 䊳/TARE to scroll to the digit(s) you want to change (it flashes on the display).
• Press
/NT/GRS to change the value of the flashing digit. Values can range from 0 to 9. For the first digit, you can also enter a minus sign (–) or –1.
9. Press MENU to store the value shown
for RD!1 . The unit displays:
To identify the maximum known load (IN!2 and RD!2
1. Apply the maximum known load (100%).
2. Repeat steps 4–9 for IN!2 and RD!2 .
Once you've completed all steps, the unit displays:
To begin operation:
Reinitialize the unit (press RESET twice or press MENU until RST
flashes on the display). When a numeric reading appears, the unit is operational.
):
Scaling Without Known Loads
For 0–100 mV sensors, the values for the minimum and maximum input loads are always as follows:
• Minimum load (IN!1)— 0
• Maximum load (IN!2)— 9999. If your installation uses a different sensor type, you must calculate the values for IN!1 and IN!2 before proceeding
with the steps below. Use the formula provided in the Operator's Manual.
To define the minimum load
1. If it's not already shown, press MENU until the unit displays:
2. Press /TARE. The unit displays:
3. Press /TARE again. The unit displays the last setting for IN!1 .
(The first digit flashes.)
4. Change IN!1 as necessary:
Press
/NT/GRS to set or change the digit's current value. Continue to press desired value for the flashing digit. Values can range from 0 to 9. For the first digit, you can also enter a minus sign (–) or –1.
• Press 䊳/TARE to scroll to the digit(s) you want to change.
5. Press MENU to store IN!1 .
The unit displays:
6. Press /TARE. The unit displays
the last setting for RD!1 . (The first digit flashes.)
7. Change RD!1 as necessary:
• Press 䊱/NT/GRS until the meter displays the desired value for the flashing digit. Values can range from 0 to 9. For the first digit, you can also enter a minus sign (–) or –1.
• Press
/TARE to scroll to the digit(s) you want to change.
(
IN!1 and RD!1
/NT/GRS until the meter displays the
):
Scaling Without Known Loads (Continued)
To define the maximum load (IN!2 and RD!2
1. Repeat steps 3–8 above, entering the values for IN!2 and RD!2 .
2. Once you've completed all steps, the unit displays:
To begin operation: Reinitialize the unit (press RESET twice or press MENU until RST flashes on the
display). When a numeric reading appears, the unit is operational.
):
Determining Reading Offset
The run mode reading for meters scaled without known loads
RD!1
may reflect an offset. For example, say you set
RD!2
to 100, but when the minimum load is applied, a
negative value of –1.5 displays on the front panel. To correct the reading offset:
1. With zero load applied, note the reading on the display.
2. Subtract that amount from the RD!1 and RD!2 values you originally entered.
In the example, the offset would be –1.5. If RD!1 is to read 0 in Run Mode, it must be reentered as 1.5. RD!2 must
likewise be reentered as 101.5 if the meter is to read 100 when the maximum load is applied.
Repeat the steps for "Scaling Without Known Loads," but
3. when the values for IN!1 and IN!2 display, do not change
them. Instead, press MENU to move to the prompts for RD!1 and RD!2 and make the necessary changes.
4. Reinitialize the unit and resume operation.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as it contains important information relating to safety and EMC.
to 0 and
5. Press /TARE again. The unit
displays the actual reading being received from the sending device.
8. Press MENU to store the value
shown for RD!1 . The unit displays: