Cooper Instruments DCM480 Specifications

DCM 480 LVDT SIGNAL CONDITIONING MODULE
LVDT Signal Conditioning Module
he DCM 480 offers low cost and simple conditioning for LVDT. The DCM 480 provides the necessary circuitry for
the transducer excitation, zero and scale adjustments. The housing is constructed of a robust die-cast aluminium enclosure, sealed to IP65 and fitted with IP65 rated cable glands which provide total protection against the ingress of water or dust.
GENERAL
• CHANNELS:
• TEMPERATURE
• MATERIAL:
• POWER REQUIRED
1 32º to 185º F Aluminum Housing-IP 65 12 to 36 Vdc Universal Polarity
TRANSDUCER
• EXCITATION:
• INPUTS ACCEPTED:
0.5V, 5mA rms 5kHz Sinusoidal 30 mv to 4V rms
AMPLIFIER
• ACCURACY:
• ANALOG OUTPUT:
• BANDWIDTH:
±0.05% F.S. 2-Wire, 4-20mA 250Hz, with Filter 25Hz (Flat)
PHYSICAL
• DIMENSIONS:
4.92" L x 3.15" W x 2.2" D
• SELECTABLE FILTERING
• FOR USE WITH LVDT TRANSDUCERS
• ALUMINUM HOUSING SEALED TO IP65
DCM 482
he DCM 482 low cost and simple conditioning for LVDT. The DCM 482 provides the necessary circuitry for the
transducer excitation, zero and scale adjustments. The housing is constructed of a robust die-cast aluminium inclosure, sealed to IP65 and fitted with IP65 rated cable glands which provide total protection against the ingress of water or dust.
LVDT SIGNAL CONDITIONING MODULE
GENERAL
CHANNELS:
TEMPERATURE:
MATERIAL:
POWER REQUIRED:
1 14º to 140º F Aluminum Housing-IP 65 ±6 to ±18 Vdc or 12 to 36 Vdc
TRANSDUCER
EXCITATION:
INPUTS ACCEPTED:
CALIBRATION METHOD:
BRIDGE RESISTANCE:
1V rms 5kHz, 25 mA ±250 mV Max Potentiometer 120 ohm min.
AMPLIFIER
ACCURACY:
ANALOG OUTPUT:
BANDWIDTH:
±0.04% ±4 or ±10 Vdc depending on load or
4-20 (3 wire) into 550 Ohms Max Flat to 500Hz
COOPER INSTRUMENTS & SYSTEMS
±5 VDC OR ±10 VDC OUTPUT
• CONDITIONS LVDT AND INDUCTIVE SENSORS
• EASY TO ACCESS ZERO AND GAIN
• BANDWIDTH FLAT TO 500Hz
PHYSICAL
DIMENSIONS:
www.cooperinstruments.com sales@cooperinstruments.com
6" L x 2.5" W x 1.34" D
1-800-344-3921