Thank you for purchasing a Series 4 digital force gauge, designed for tension and compression force testing
applications from 0.12 lb to 500 lb (0.5 N to 2,500 N) full scale. The Series 4 is an essential component of a
force testing system, typically also comprising a test stand, grips, and data collection software.
With proper usage, we are confident that you will get many years of great service with this product. Cooper
force gauges are ruggedly built for many years of service in laboratory and industrial environments. This
User’s Guide provides setup, safety, and operation instructions.
Dimensions and specifications are also provided. For additional information or answers to your question s,
please do not hesitate to contact us. Our technical support and engineering teams are eager to assist you.
Before use, each person who is to use the Series 4 force gauge should be fully trained in appropriate
operation and safety procedures.
1 OVERVIEW
1.1 List of included items
Part No.
Qty.
1 12-1049 12-1049 12-1049 Carrying Case
1 AC1030 /
Lite software, MESURgauge DEMO
software, User’s Guide)
1.2 Safety / Proper Usage
Caution!
Note the force gauge’s capacity before use and ensure that the capaci ty is not exceeded.
Producing a force greater than 150% of the gauge’s capacity can damage the internal load cell. An
overload can occur whether the gauge is powered on or off.
Before use, each person who is to use the Series 4 force gauge should be fully trained in appropriate
operation and safety procedures.
Typical materials able to be tested include many manufactured items, such as springs, electronic
components, fasteners, caps, films, mechanical assemblies, and many others. Items that should not be used
with the gauge include potentially flammable substances or products, items that can shatter in an unsafe
manner, and any other components that can present an exceedingly hazardous situation when acted upon
by a force.
CF 182 1 32-1117 1010
The following safety checks and procedures should be performed before and during o peration:
1. Never operate the gauge if there is any visible damage to the AC adapter or the gauge itself.
2. Ensure that the gauge is kept away from water or any other electrically conductive liquids at all
times.
3. The gauge should be serviced by a trained technician only. AC power must be disconnected and
the gauge must be powered off before the housing is opened.
4. Always consider the characteristics of the sample being tested before initiating a test. A risk
assessment should be carried out beforehand to ensure that all safety measures have been
addressed and implemented.
5. Wear eye and face protection when testing, especially when testing brittle samples that have the
potential to shatter under force. Be aware of the dangers posed by potential energy that can
accumulate in the sample during testing. Extra bodily protection should be worn if a destructive
failure of a test sample is possible.
6. In certain applications, such as the testing of brittle samples that can shatter, or other applications
that could lead to a hazardous situation, it is strongly recommended that a machine guarding system
be employed to protect the operator and others in the vicinity from shards or debris.
7. When the gauge is not in use, ensure that the power is turned off.
2 POWER
The gauge is powered either by an 8.4V NiMH rechargeable battery or by an AC adapter. Since these
batteries are subject to self discharge, it may be necessary to recharge the unit after a prolonged period of
storage. Plug the accompanying charger into the AC outlet and insert the charger plug into the receptacle on
the gauge (refer to the illustration below). The battery will fully charge in approximately 8 hours.
Caution!
Do not use chargers or batteries other than specified or instrument damage may occur.
If the AC adapter is plugged in, a corresponding icon appears in the lower left corner of the display, as
follows:
If the AC adapter is not plugged in, battery power drainage is denoted in a three step process:
The gauge can be configured to automatically power off following a period of inactivity. Refer to the Other Settings section for details.
If battery replacement is necessary, the battery may be accessed by separating the two halves of the gauge.
Refer to the Mechanical Setup section for details.
CF 182 2 32-1117 1010
1. When battery life is greater than 30%, no indicator is present.
2. When battery life drops to approximately 30%, an indicator appears as follows:
3. When battery life drops to approximately 10%, the indicator will appear as follows:
4. When battery life drops to approximately 2%, the indicator from step 2 will be flashing. Several
minutes after (timing depends on usage and whether the backlight is turned on or off), a message
will appear, “BATTERY VOLTAGE TOO LOW. POWERING OFF”. A 4-tone audio indicator will
sound and the gauge will power off.
43 MECHANICAL SETUP
3.1 Loading shaft orientation
In order to accommodate a variety of testing requirements, the orientation of the loading shaft may be set up
in either of the two positions shown below. In order to change the loading shaft orientation, loosen the two
captive screws on the back side of the housing, separate the two housing halves, rotate one half 180
degrees, and reassemble. Contact between the two halves is made by the spring pins and contact pads on
the printed circuit boards.
3.2 Mounting to a plate
Although the gauge may be used by hand, proper mounting is important if attached to a fixture or test stand.
The round steel insert with a hole in the back of the housing is provided to withstand the load during a test. A
mating dowel pin should be used (see illustration below). Mounting plates on Cooper test stands inclu de a
dowel pin and clearance holes for the four threaded holes located near the corn ers of the housing. These
holes are designed to accommodate screws in order to hold the gauge in place (Cooper test stands include a
set of thumb screws for gauge mounting). The screws must not be used for load bearing purposes. Failure
to use a dowel pin properly can result in a hazardous situation.
3.3 Mounting attachments to the gauge
The force gauge’s threaded loading shaft is designed to accommodate common grips and attachments with
female mounting holes. To mount a grip, gently thread it onto the shaft. Other mounting adapters are also
available to prevent rotation. Ensure that the grip or fixture is positioned to ensure axial load with respect to
the loading shaft of the force gauge. When using a grip, ensure that it secures the sample in such a way that
it is prevented from slipping out during a test, preventing a potential safety risk to the5operator and others in
the vicinity. If using a grip or fixture from a supplier other than Cooper, ensure that it is constructed of suitably
rugged materials and components.
Do not use jam nuts or tools to tighten grips or attachments onto the shaft. Finger-tighten only. Antirotation
mounting adapters are available.
CF 182 3 32-1117 1010
4 HOME SCREEN AND CONTROLS
4.1 Home Screen
No. Name Description
1 Measurement direction
2 Peaks
3 Primary reading
4 Load bar
5 Units
6 Mode
7 Number of stored data
points
8 Battery / AC adapter
indicator
9 High / low limit indicators
CF 182 4 32-1117 1010
- indicates a compression (push) direction
- indicates a tension (pull) direction
These indicators are used throughout the display and menu.
The maximum measured compression and tension readings. These
readings are reset by pressing ZERO or by powering the gauge off and
on.
The current displayed force reading. See Operating Modes section for
details.
Analog indicator to help identify when an overload condition is
imminent. The bar increases either to the right or to the left from the
midpoint of the graph. Increasing to the right indicates compression
load, increasing to the left indicates tension load. If set points are
enabled, triangular markers are displayed for visual convenience. This
indicator reflects the actual load, which may not correspond to the
primary reading (depends on operating mode). The ZERO key does not
reset the load bar. See Operating Modes section for details.
The current measurement unit. Abbreviations are as follows:
lbF – Pound-force
ozF – Ounce-force
kgF – Kilogram-force
gF – Gram-force
N – Newton
kN – Kilonewton
mN – Millinewton
Note: not all gauge capacities measure in all the above units. Refer to
the capacity / resolution table in the Specifications section for details.
The current measurement mode. Abbreviations are as follows:
RT – Real Time
PC – Peak Compression
PT – Peak Tension
See Operating Modes section for details about each of these modes.
The number of stored data points in memory, up to 50. Displayed only
if Memory Storage is enabled for the DATA key.
Either the AC adapter icon, battery icon, or no icon will be shown,
depending on power conditions. Refer to the Power section for details.
Correspond to the programmed set points. Indicator definitions are as
follows:
– the displayed value is greater than th e upper force limit
– the displayed value is between the limits
– the displayed value is less than the lower force limit
10 Set points
The programmed force limits. Typically used for pass/fail type testing.
1, 2, or no indicators may be present, depending on the configuration
shown in the Set Points menu item.
4.2 Controls
Primary Label Primary Function
Powers the gauge on
ZERO
MENU
MODE
DATA
UNITS
and off. Press briefly to
power on, press and hold
to power off. Active only
when the home screen is
displayed.
Zeroes the primary
reading and peaks.
Enters the main menu.
Toggles between
measurement modes.
Stores a value to
memory and/or transmits
the current reading to an
external device,
depending on setup.
Toggles between
measurement units.
Turns the LCD backlight
on and off. Backlight
turns on automatically
upon powering on the
gauge.
Secondary Label Secondary Function
ENTER
(UP)
ESCAPE
(DOWN)
DELETE
DIRECTION
N/A
Various uses, as described in
the following sections.
Navigates up through the
menu and sub-menus.
Reverts one step backwards
through the menu hierarchy.
Navigates down through the
menu and sub-menus.
Enables and disables Delete
mode while viewing stored
data.
Reverses the display during
calibration, and toggles
between tension and
compression directions while
configuring set points and
other menu items.
N/A
4.3 Menu navigation basics
Most of the gauge’s various functions and parameters are configured through the main menu. To access the
menu press MENU. Use the UP and DOWN keys to scroll through the items. The current selection i s
denoted with clear text over a dark background. Press ENTER to select a menu item, then use UP and DOWN again to scroll through the sub-menus. Press ENTER again to select the sub-menu item.
For parameters that may be either selected or deselected, press ENTER to toggle between selecting and
deselecting. An asterisk (*) to the left of the parameter label is used to indicate when the parameter has been
selected.
For parameters requiring the input of a numerical value, use the UP and DOWN keys to increment or
decrement the value. Press and hold either key to auto-increment at a gradually increasing rate. When the
desired value has been reached, press ENTER to save the change and revert back to the sub-menu item, or
press ESCAPE to revert back to the sub-menu item without saving. Press ESCAPE to revert one step back
in the menu hierarchy until back into normal operating mode.
CF 182 5 32-1117 1010
Refer to the following sections for details about setting up particular functions an d parameters.
5 OPERATING MODES
Caution!
In any operating mode, if the capacity of the instrument has been exceeded by more than 110%, the
display will show “OVER” to indicate an overload. A continuous audible tone will be sounded until
the MENU key has been pressed or the load has been reduced to a safe level.
Three operating modes are possible with Series 4 gauges. To cycle between the modes, press MODE while
in the home screen.
5.1 Real time (RT)
The primary reading corresponds to the live measured reading.
5.2 Peak Compression (PC)
The primary reading corresponds to the peak compression reading observed. If the actual force decreases
from the peak value, the peak will still be retained in the primary reading area of the display.
Pressing ZERO will reset the value.
5.3 Peak Tension (PT)
Same as Peak Compression, but for tension readings.
6 DIGITAL FILTERS
Digital filters are provided to help smooth out the readings in situations where there is mechanical
interference in the work area or test sample. These filters utilize the moving average technique in which
consecutive readings are pushed through a buffer and the displayed reading i s the average of the buffer
contents. By varying the length of the buffer, a variable smoothing effect can be achieved. The selection of 1
will disable the filter since the average of a single value is the value itself.
To access digital filter settings, select Filters from the menu. The display will appear as follows:
8
Two filters are available:
Current Reading – Applies to the peak capture rate of the instrument.
Displayed Reading – Applies to the primary reading on the display.
Available settings: 1,2,4,8,16,32,64,128,256,512,1024. It is recommended to keep the current reading filter at
its lowest value for best performance, and the displayed reading filter at its highest value for best stability.
7 SET POINTS
7.1 General Information
Set points are useful for tolerance checking (pass/fail),
triggering an external device such as a motorized test stand,
CF 1826 32-1117 1010
or alarm indication in process control applications. Two limits, high and low, are specified and stored in the
non-volatile memory of the instrument and the primary reading is compared to these limits. The results of the
comparisons are indicated through the three outputs provided on the 15-pin connector, thus providing
“under”, “in range”, and “over” signaling. These outputs can be connected to indicators, buzzers, or relays as
required for the application.
7.2 Configuration
To configure set points, select Set Points from the menu. The screen will appear as follows:
Either one, two, or none of the set points may be enabled. To toggle between the tension and compression
directions, press the DIRECTION key.
If two set points have been enabled, they are displayed in the upper left corner of the display. If only one set
point has been enabled, the word “OFF” will appear in place of the value. If no set points have been enabled,
the upper left corner of the display will be blank.
9When set points are enabled, the following indicators are shown to the left of the primary reading:
Note: Set point indicators and outputs reference the displayed reading, not necessarily the current live load.
8 DATA MEMORY AND STATISTICS
Series 4 gauges have storage capacity of 50 data points. Readings may be stored, viewed, and output to an
external device. Individual, or all, data points may be deleted. Statistics are calculated for the data presently
in memory.
To enable memory storage, select DATA Key from the menu, then scroll to Memory Storage and press
ENTER. Then exit the menu. In the home screen, the data record number 00 will appear below the primary
reading. Press DATA at any time to save the displayed reading. The record number will increment each time
DATA is pressed.
To view, edit, and output stored readings and statistics, select Memory from the menu. The screen appears
as follows:
CF 182 7 32-1117 1010
8.1 View Data
All the saved data points may be viewed. The record number is displayed, along with the corre sponding
value and presently set unit of measurement. Any readings may be deleted individually. To do so, scroll to
the desired reading and press DELETE. The letter “D” will appear to the left of the record number, indicating
that the gauge is in Delete mode, as follows:
View Data
10
Press ENTER to delete the value. To exit Delete mode, press DELETE again. Any number of readings may
be individually deleted; however, all readings may also be cleared simultaneously. Refer to the Clear All Data section for details.
8.2 Statistics
Statistical calculations are performed for the saved values. Calculations include number of readings,
minimum, maximum, mean, and standard deviation.
8.3 Output Data
Press ENTER to output data to an external device. The display will show, “SENDING DATA…”, then
“DATA SENT”. If there was a problem with communication, the display will show, “DATA NOT SENT”.
Saved data can be downloaded by some Cooper data collection programs. Refer to their respective user’s
guides for details.
8.4 Output Statistics
Press ENTER to output statistics to an external device. The display will show, “SENDING STATS…”, then
“STATS SENT”. If there was a problem with communication, the display will show, “STATS NOT SENT”.
8.5 Output Data & Stats
Press ENTER to output data and statistics to an external device. The display will show, “SENDING
DATA”, then “SENDING STATS…”, then “DATA SENT”, then “STATS SENT”. If there was a problem with
communication, the display will show, “DATA NOT SENT” and/or “STATS NOT SENT”.
8.6 Clear All Data
Press ENTER to clear all data from the memory. A prompt will be shown, “CLEAR ALL DATA?”. Select
Yes to clear all the data, or No to return to the sub-menu.
For output of data and/or statistics, RS-232 or USB output must be enabled. Data formatting is
<CR><LF> following each value. Units can be either included or excluded. Output of data via the Mitutoyo
output is possible; however, output of statistics is not. Refer to the Communications section for details.
Note: Data is not retained while the gauge is powered off.
9 COMMUNICATIONS
Communication with Series 4 force gauges is achieved through the micro USB or 15-pin serial ports lo cated
at the bottom of the instrument, as shown in the illustration in the Power section. Communication is possible
only when the gauge is in the main operating screen (i.e. not in a menu or configuration area).
CF 182 8 32-1117 1010
9.1 Installing the USB driver
It is recommended that the USB driver be installed before physically connecting the gauge to the PC
with a USB cable.
1. Insert the Resource CD supplied with the gauge into the CD/DVD drive in the computer. Then, navigate in
Windows Explorer or My Computer to one of the following folders on the CD:
Windows 2000 through Vista - “Win_2K_XP_S2K3_Vista”
Windows 7 - “Windows_7”
2. Execute the installer application “Mark10USBInstaller.exe” by double-clicking it. When the program
launches, one of the following windows will appear, depending on the operating system:
11Click “Install”.
3. The next screen appears as follows:
Click “Continue Anyway”.
4. After installation completes the following screen may appear in non-Windows 7 operating systems.
Restart the computer before connecting a USB device.
5. After Windows as restarted, plug in the device. The following will occur:
CF 182 9 32-1117 1010
Windows 7 Operating Systems – When the USB device has been plugged into a USB port, the
driver will automatically be found. When the driver installation is complete, a message will ap pear as
follows: “The MARK-10 USB DEVICE driver is now installed and ready to use”.
Non-Windows 7 Operating Systems – When the USB device has been pl ugged into a USB port,
the following screen appears:
12
Select “No, not this time”, then click “Next”.
6. The next screen appears as follows:
Select “Install the software automatically (Recommended)”, then click “Next”.
7. The next screen appears as follows:
CF 182 10 32-1117 1010
Click “Continue Anyway”.
8. The next, and final, screen appears as follows:
Click “Finish”. The USB device is now installed and ready to use. The COM port number a ssigned by
Windows may be identified in Device Manager, or in the communication application being used, such as
MESURgauge or HyperTerminal.
9.2 Serial / USB
To set up RS-232 and USB communication, select Serial/USB Settings from the menu. The screen appears
as follows:
Select either RS-232 or USB input (output is always simultaneous through both the USB and RS-232 ports).
RS-232 must be selected when communicating through a test stand controller. When communicating from
the gauge directly to a PC or data collector, either RS-232 or USB can be selected as required. Configure
the baud rate and data format as required for the application. Default values are as follows:
CF 182 11 32-1117 1010
Baud Rate: 9,600
Data Format: Numeric + units
Other communication settings are permanently set to the following:
Data Bits: 8
Stop Bits: 1
Parity: None
Individual data points may be transmitted by pressing DATA.
The currently displayed reading may be requested from an external device by sending ASCII character ‘?’
followed by a Carriage Return character or with a Carriage Return/Line Feed combination. The gauge
responses are always terminated with a Carriage Return/Line Feed.
14
Any detected errors are reported back by means of the following error codes:
*10 Illegal command
*11 Not applicable
*12 Invalid specifier
*22 Value too large
9.3 Mitutoyo BCD settings
This output is useful for connection to data collectors, printers, multiplexers, or any other device capable
of accepting Mitutoyo BCD data. Individual data points may be transmitted by pressing DATA or by
requesting it from the Mitutoyo communication device (if available). To enable Mitutoyo output, select the
desired format – either with polarity or without polarity. The screen appears as follows:
9.4 Analog Output
This output can be used for chart recorders, oscilloscopes, data acquisition systems, or any other compatible
devices with analog inputs. The output produces ±1 volt at full scale of the instrument. The polarity of the
signal is positive for compression and negative for tension.
9.5 DATA Key Functions
The DATA key can be configured to perform several functions. To configure the DATA key, select DATA
Key from the menu. The display will appear as follows:
Three options are available:
Selection Function when pressing DATA
RS232/USB Output
Mitutoyo Output
Memory Storage
CF 182 12 32-1117 1010
Outputs data via the serial and USB ports
Outputs data via Mitutoyo (Digimatic) through the serial port
Stores a reading to memory (refer to the Memory section for details)
9.6 I/O Connector Pin Diagram (female)
Pin No. Description Input / Output
1 Signal Ground --2 Tension Overload Output
3 RS-232 Receive Input
4 RS-232 Transmit Output
5 +12V DC Output
6 Analog Output Output
7 Compression Overload Output
8 Mitutoyo Clock
Output Bit 2
9 Mitutoyo Data
Output Bit 0
10 Mitutoyo Request
Input Bit 3
11 “Under” Set Point Output
12 “Over” Set Point Output
13 “Within” Set Point Output
14 External Trigger Input
15 Mitutoyo Ready
Output Bit 1
Output
Output
Input
Output
10 CALIBRATION
10.1 Initial Physical Setup
The gauge should be mounted vertically to a test stand or fixture rugged enough to withstand a load equal to
the full capacity of the instrument. Certified deadweights or master load cells should be used, along with
appropriate mounting brackets and fixtures. Caution should be taken while handling such equipment.
10.2 Calibration Procedure
1. Select Calibration from the menu. The display will appear as follows:
16
2. Press DIRECTION to invert the display, if desired. ENTER to continue. The display will appear as follows:
CF 182 13 32-1117 1010
The gauge can be calibrated at up to 10 points in each direction. Enter the number of calibration points for
each direction (compression and tension). At least one point must be selected for each dire ction.
Note: To achieve the accuracy specification of ±0.1%, it is recommended to calibrate the gauge at 5 or more
even increments in both the tension and compression directions. For example, a gauge with capacity of 10
lbF should be calibrated at 2, 4, 6, 8, and 10 lb loads in each direction.
3. To escape the Calibration menu at any time, press ESCAPE. The display will appear as follows:
Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will return to the
menu without saving changes.
4. After the number of calibration points has been entered, press ENTER. The display will appear as follows:
5. Place the force gauge horizontally on a level surface free from vibration, then press ZERO. The gauge will
calculate offsets, and the display will appear as follows:
17
6. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply any
calibration loads. Then press ENTER.
7. The display will appear as follows:
CF 182 14 32-1117 1010
Optionally exercise the load cell shaft several times (at full scale, if possible), then press ENTER.
8. The display will appear as follows:
Apply a weight equal to the full scale of the instrument, then press ENTER.
9. After displaying “PLEASE WAIT…” the display will appear as follows:
Remove the load applied in Step 8, leave the fixtures in place, then press ZERO.
10. The display will appear as follows:
Use the UP and DOWN keys to adjust the load value as required. The load values default to even
increments, as indicated by the previously entered number of data points (even increments are
recommended for best results). For example, if a 50 lbF capacity gauge is calibrated, and 5 data points were
selected, the load values will default to 10, 20, 30, 40, and 50 lb. Apply the calibration load. Then press
ENTER.
Repeat the above step for the number of data points selected.
11. After all the compression calibration points have been completed, the display will appear as follows:
Press ENTER.
12. The display will appear as follows:
CF 182 15 32-1117 1010
Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees. Press DIRECTION to invert
the display. Then attach weight fixtures. The following screens will step through the same procedure as with
the compression direction. Proceed in the same manner.
13. At the completion of the tension calibration, the display will appear as follows:
To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.
14. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
The calibration weight does not match the set value.
The entered calibration point is too close to the previous point.
CF 182 16 32-1117 1010
11 OTHER SETTINGS
11.1 Automatic Shutoff
The gauge may be configured to automatically power off following a period of inactivity. Inactivity is defined
as the absence of any key presses or load changes of 100 counts or less. To access these settings, select
Automatic Shutoff from the menu. The display will appear as follows:
Select Disabled to disable automatic shutoff. Select Enabled to enable it. The length of time of inactivity is
programmed in minutes via the Set Minutes parameter. Available settings: 5-30, in 5 minute increments.
11.2 LCD Contrast
The contrast of the display may be adjusted. Select LCD Contrast from the menu. The screen will appear as
follows:
Press ENTER to modify the contrast. Select a value from 0 to 25, 25 producing the most contrast. Higher
contrast ratios will deplete battery power more quickly.
11.3 Beeps
Audible tones can be enabled for all key presses and alerts, such as overload, set point value reached, etc.
The Set Point alert can be configured to be either a momentary tone or a continuous tone (until the load is
restored to a value between the set points). To configure the functions for which audible tones will apply,
select Beeps from the menu. The screen will appear as follows:
11.4 Initial settings
This section is used to configure the initial settings upon powering on the gauge. The initial units of
measurement and the primary reading measurement mode may be configured. To access these settings,
select Initial Settings from the menu. The screen will appear as follows:
The default values are LBF and Real Time.
CF 182 17 32-1117 1010
11.5 Information / Welcome Screen
The following screen is displayed at power up and can be accessed at any time by selecting Information
from the menu:
12 SPECIFICATIONS
12.1 General
Accuracy:
Sampling rate:
Power:
Battery life:
Measurement units:
Outputs:
Configurable settings:
Safe overload:
Weight (gauge only):
Included accessories:
Warranty:
±0.2% of full scale ±1 digit
3,000 Hz
AC or rechargeable battery. Low battery indicator appears when battery level is low, and
gauge powers off automatically when power reaches critical stage.
Backlight on: up to 7 hours of continuous use
Backlight off: up to 24 hours of continuous use
lbF, ozF, gF, kgF, N, kN, mN (depending on model)
USB / RS-232: Fully configurable up to 115,200 baud.
Mitutoyo (Digimatic): Serial BCD suitable for all Mitutoyo SPC-compatible devices.
Analog: ±1 VCD, ±0.25% of full scale at capacity,
General purpose: Three open drain outputs, one input.
Set points: Three open drain lines.
Carrying case, chisel, cone, V-groove, hook, flat, extension rod, AC adapter, battery,
USB cable, resource CD (USB driver, MESUR Lite software, MESURgauge DEMO
software, and user’s guide), NIST-traceable certificate of calibration with data
3 years (see individual statement for further details)
12.2 Factory Settings
Parameter Setting
Set points
Upper
Lower
Filters
Current
Displayed
DATA Key
RS-232/USB
Mitutoyo
Memory Storage
Serial/USB
RS-232 Selected
USB Selected
Baud Rate
Data Format
Mitutoyo BCD Disabled
CF 182 18 32-1117 1010
Disabled (defaults to 80% of full scale when enabled)
Disabled (defaults to 40% of full scale when enabled)
Any Cooper Instruments product which, under normal operating conditions, proves defective in material or in
workmanship within one year of the date of shipment by Cooper will be repaired or replaced free of charge
provided that a return material authorization is obtained from Cooper and the defective product is sent,
transportation charges prepaid, with notice of the defect, and it is established that the product has been
properly installed, maintained, and operated within the limits of rated and normal usage. Replacement or
repaired product will be shipped F.O.B. from our plant. The terms of this warranty do not extend to any
product or part thereof which, under normal usage, has an inherently shorter useful life than one year. The
replacement warranty detailed here is the buyer’s exclusive remedy, and will satisfy all obligations of Cooper
whether based on contract, negligence, or otherwise. Cooper is n ot responsible for any incidental or
consequential loss or damage which might result from a failure of any and all other warranties, express or
implied, including implied warranty of merchantability or fitness for particular purpose. Any unauthorized
disassembly or attempt to repair voids this warranty.
Obtaining Service under Warranty
Advance authorization is required prior to the return to Cooper Instruments. Before returning the item,
contact the Repair Department c/o Cooper Instruments at (540) 349-4746 for a Return Material Authorization
number. Shipment to Cooper shall be at buyer’s expense and repaired or replacement items will be shipped
F.O.B. from our plant in Warrenton, Virginia. Non-verified problems or defects may be subject to a $100
evaluation charge. Please return the original calibration data with the unit.
Repair Warranty
All repairs of Cooper products are warranted for a period of 90 days from date of shipment. This warranty
applies only to those items that were found defective and repaired; it does not apply to products in which no
defect was found and returned as is or merely recalibrated. It may be possible for out-of-warranty products
to be returned to the exact original specifications or dimensions.
*Technical description of the defect: In order to properly repair a product, it is absolutely necessary for
Cooper to receive information specifying the reason the product is being returned. Specific test data, written
observations on the failure and the specific corrective action you require are needed.
CF 182 20 32-1117 1010
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.