DIGITAL TORQUE TESTER/WRENCH DTW 200/200N
USER’S GUIDE
LIST OF EQUIPMENT
1. Torque Tester
2. Stepless CW one-way clutch (included for DTW 200, available for DTW 750)
3. AC adapter/charger
4. Carrying case
Read First: Safety Information
For safety, and for damage avoidance, be sure to read this manual thoroughly. The warranty is only valid when the
product is used following the instructions provided within this manual.
• Do not use tester in high temperature, high humidity, or in damp or wet areas.
• Recommended operating temperature is between 0-42ºC (32-100ºF).
• Do not apply torque exceeding the rated capacity (180 lbf-in), regardless of whether the unit is On or Off. Avoid
shock load. Do not use with impact wrenches.
• When charging the battery, be sure to use the provided AC adapter/charger exclusively.
• Do not use lacquer thinner or any solvent to clean the unit.
• Do not disassemble or modify the unit.
1. ON/OFF Switch Press once to turn on, press again to turn off. After 10 minutes of non-use, unit will shut off
automatically.
2. MODE Select GO (Real Time), PP (Peak), PD (Peak Down), C (Continuous Output Mode)
3. CLEAR Reset display to zero and send data to memory
4. LCD Display Displays torque value, battery icon, mode and units
5. GO/NG Indicator Green for values between LO and HI setpoints, flashing red for values above HI, solid green
for data output.
6. DOWN Change values or numeric places
7. UP Change values or options
8. PRG Enter programming mode or enter values
9. MEM Display memory locations and data
10. STAT Display statistics; number of records, Max, Min, and Ave.
USB Port Send data to computer via USB cable
Charge light Lights when charging, off when complete
DC IN AC Charger /adapter receptacle
Reset System reset button
CF 139 1 DTW 200/200N (12/10)

General Operation
1. Press and hold MODE for one second to select from the following measuring modes:
Real Time – displays torque transients (no output or indicator)
PP Mode – Peak, captures peak torque (peak data output, PP appears on the display)
PD Mode – First Peak, capture first peak value (peak data output, PD appears on the display)
C Mode – Continuous RS-232 output, Display and output torque transients (12 or 180 d ata/sec., C appears
on the display.
2. Place the appropriate size socket onto the drive, and if desired, the one-way clutch also.
3. Rotate the wrench head so the display may be viewed while measuring.
4. Press the CLEAR button on the display unit to zero the display. Insert the socket onto the bolthead, hold
perpendicularly (not at an angle) and turn to measure.
5. After measuring, press CLEAR to zero display for the next test.
Programming
Click ON/OFF to turn on. Press PRG for one second. The display shows HI and then
the High setpoint value. This confirms the tester is ready for the following
programming steps.
1. High Setpoint (HI)
After HI is displayed, and the high setpoint value, press DOWN to move the
numeric place and press UP to select values, then press PRG to enter.
2. Low Setpoint (LO)
After high value is entered, LO is displayed, then the low setpoint value. Press
DOWN to move the numeric place and press UP to select values, then press
PRG to enter.
3. Peak Down Minimum (PdLO)
After low value is entered, PdLO is displayed, then the PdLO value. Press down
to move the numeric value and press up to select values, and press PRG to enter.
PdLO sets a minimum torque value for Peak Down mode. For example, if PdLO value is set at 5.0 lbf-in, only a
reading over that will be measured in Peak Down mode.
4. Continuous Data Output Minimum
After Peak Down value is entered, CLO is displayed, then the CLO value. . Press DOWN to move the numeric
place and press UP to select values, then press PRG to enter.
5. Auto Zero Reset (AC)
After Continuous Data Output Minimum is entered, AC is displayed and then the Auto Zero Reset duration
value. Press Up or Down to select 0.0C-0.1C-0.5C-1.0C-1.5C-2.0C-2.5C-3.0C, and press PRG to enter (0.1C
for 0.1 second and 0.0C for Manual Reset)
6. Batch Counter (CO)
After the Zero Value is entered, CO is displayed then the CO value. Use the UP or DOWN keys to select a
value and PRG to enter (0-99, clockwise only)
7. Interval (In)
After the Batch Counter value is entered, In is displayed then the In value. the UP or DOWN keys to select a
value and PRG to enter (0-24 in 22-second increments)
CF 139 2 DTW 200/200N (12/10)