COOK Medical Bird's Nest BNF-45, Bird's Nest BNF-75 Suggested Instructions

ORDERING INFORMATION
Proven PE Protection
May be used in vena cava up to • 40 mm in diameter.
Con trolled insertion for use in varying anatomic conditions while creating a fi ne wire multiplane fi ltering system.
Secure fi xation prevents migration.
Introducer system fl exibility permits • options in access site selection.
Filter
Global Product Number
Femoral
G50687 BNF-45 40 11.0 45 12.0 .038
Jugular
G50686 BNF-75 40 11.0 75 12.0 .038
Set Order Number
Maximum
Diameter
mm
Filter
Catheter
French Size
Filter
Catheter
Length
cm
Introducer
Sheath
French Size
Accepts
Wire Guide
Diameter
inch
SUGGESTED INSTRUCTIONS FOR FEMORAL PLACEMENT
SUGGESTED INSTRUCTIONS FOR FEMORAL PLACEMENT
1. Introduce fi lter/sheath assembly
Insert the fi lter catheter into the introducer sheath and Luer lock it into position. The tip of the fi lter catheter will then extend approximately 1.5 cm beyond the tip of the sheath. While performing hand injections of contrast medium through the fi lter catheter sidearm, the fi lter catheter/introducer sheath assembly can be properly positioned to initi­ate fi l ter placement. When using a femoral
approach, position the fi lter catheter tip just below (caudad to) the renal veins.
2. Pull fi lter/sheath
Mark
Loosen the Tuohy-Borst valve and, while hold­ing the wire guide pusher stationary, withdraw the fi lter catheter/introducer sheath assembly to the mark on the wire guide pusher. This movement will expose the distal anchoring hooks. Fluoroscopically, this corresponds to the position of the junction point of the hook wires remaining just within the tip of the fi lter catheter. If this can be clearly visualized with fl uoroscopy, the mark can be ignored.
NOTE: At the end of the fi lter, the V-shaped hook wire struts are readily visible, but fi lter wires are too fi ne to be identifi ed fl uoroscopically.
3. Anchor hooks
Be certain junction point of distal pair of hooks is in catheter tip when positioning these hooks into caval wall. If there is uncertainty regarding hook fi xation, the catheter may be advanced up over the hook wires to the edge of the hooks. The catheter may then be repositioned, and the fi lter placement re-initiated in a slightly different caval orientation or level.
CAUTION: Overly forceful or multiple jabs with the fi lter catheter/introducer sheath assembly could result in perforation of the wall of the inferior vena cava by the exposed hooks and struts of the fi lter.
4. Pull fi lter/sheath
Again, hold the wire guide pusher stationary and withdraw the fi lter catheter/introducer sheath assembly 1-3 cm. This will facilitate subsequent passage of the fi lter wires through the tip of the fi lter catheter and provide room for fi lter formation within the vena cava.
CAUTION: Do not withdraw the fi lter catheter/ introducer sheath assembly into the iliac vein. Repositioning into the inferior vena cava may prove diffi cult.
5. Push fi lter wires
Holding the fi lter catheter/introducer sheath assembly stationary and with fl uoroscopic control, advance the wire guide pusher with a steady, smooth motion until the junction point of the proximal hook wires is seen fl uoroscopically at the fi lter catheter tip.
CAUTION: Do not advance too quickly or forcefully as wire kinking within fi lter catheter may occur, making advance ment diffi cult.
6. Push fi lter/sheath assembly
Advance the fi lter catheter/introducer sheath assembly so that the proximal junction point is in close proximity to or overlaps the junction point of previously positioned hook wires.
NOTE: The approximation or slight overlap of the junction points ensures adequate compression and formation of the fi lter wires and places the second pair of hooks in the inferior vena cava. Placement of a hook in a renal vein, or passage of fi lter wires distal to the fi rst hook position (wire prolapse), may occasionally occur and is of no consequence.
7. Pull sheath
While maintaining slight forward pressure on the wire guide pusher, slowly withdraw the fi lter catheter/introducer sheath assembly. This will permit the proximal hook wires to exit the catheter, springing open to engage with the caval wall. This is readily seen fl uoroscopically.
To be sure the hooks are completely free from the fi lter catheter tip, the fi lter catheter/ introducer sheath assembly should be withdrawn to within approximately 1 cm of the Bird’s Nest Filter handle attached to the wire guide pusher.
CAUTION: Do not advance the wire guide pusher after the proximal hook wires have exited the fi lter catheter tip. Forceful advancement of wire guide pusher may alter the orientation of the fi lter attachment site, making disengagement diffi cult.
8. Anchor hooks
Gently tug or use a slight to-and-fro motion on the wire guide pusher to secure the second pair of hooks to the vena cava wall before disengaging the fi lter.
9. Detach fi lter
Fig. 9a
Fig. 9b
Fig. 9c
To release fi lter:
a
. Maintaining wire guide pusher position,
remove red safety lock from handle by grasping both sides and pulling up and away from the handle.
b
. Holding the wire guide pusher steady,
use the handle to fully depress the thumb grip of the wire guide pusher forward into the cannula. Maintaining this position will release the fi lter. Separation can be seen fl uoroscopically.
c
. Release pressure on the thumb grip of
the handle.
d
. Pull the wire guide pusher’s distal end
back into the fi lter catheter.
Remove the wire guide pusher and fi lter catheter, leaving the introducer sheath in place. Reposition the sheath for a fi nal vena cavogram.
Fig. 9d
Gently advance the fi lter catheter/introducer sheath assembly one time, 1-3 mm, to secure the hooks to the vena cava wall.
AORTIC INTERVENTION CARDIOLOGY CRITICAL CARE ENDOSCOPY PERIPHERAL INTERVENTION SURGERY UROLOGY WOMEN’S HEALTH
www.cookmedical.com
Refer to Suggested Instructions For Use for indications, warnings and precautions.
COOK MEDICAL INCORPORATED
P.O. Box 4195, Bloomington, IN 47402-4195 U.S.A. Phone: 812.339.2235, Toll Free: 800.457.4500, Toll Free Fax: 800.554.8335
COOK (CANADA) INC.
111 Sandiford Drive, Stouffville, Ontario, L4A 7X5 CANADA Phone: 905.640.7110, Toll Free: 800.668.0300
WILLIAM A. COOK AUSTRALIA PTY. LTD.
Brisbane Technology Park, 12 Electronics Street, Eight Mile Plains Brisbane, QLD 4113 AUSTRALIA, Phone: +61 7 38 41 11 88
WILLIAM COOK EUROPE ApS
Sandet 6, DK-4632, Bjaeverskov, DENMARK, Phone: +45 56 86 86 86
© COOK 2008 PI-BM-BNFMP-EN-200807