Control Company TOTAL-RANGE THERMOMETER User Manual

TRACEABLE
®
TOTAL-RANGE THERMOMETER INSTRUCTIONS
SPECIFICATIONS
Range: –328°F ~ 2498°F or –200°C ~ +1370°C Resolution: 0.1° between –199.9° ~ 1369.9° Accuracy*: ±1°C between –50°C to 740°C, otherwise ±2°C Probe: Type K Battery: Cat. No. 1105
*Accuracy is specified for unit operating temperatures between 64°F
and 82°F (18°C and 28°C). The accuracy specification does not include thermocouple error.
OPERATION
Insert the Type K probe plug into the unit’s keyed receptacle. Press the round, yellow ON/OFF button, the timer begins to count and the tem­perature displays.
Te m perature Measurement
Press °F/°C to select desired unit.
Hold/Test Function
Press HOLD - HOLD sign flashes on display to retain temperature tested. Press HOLD again and unit will return to testing mode and HOLD sign will disappear.
Maximum and Minimum Function
Press MAX/MIN, MIN appears with the minimum temperature achieved. Press again MAX appears with the maximum temperature achieved.
Te m perature Alarm Setting
1. High Alarm Setting Press HI AL - HI sign appears. –199°C or –327°F and AL
are displayed.
Press SET/TIMER - the hundred digit flashes. The number can be
chosen from –1 to 13 for °C or –3 to 24 for °F. Press HI AL to increase.
Press LOW AL - the unit digit flashes, numbers 00 to 99 can be
chosen. Press HI AL to increase.
Press SET/TIMER to set High Alarm. The maximum temperature is 2497°F (or 1369°C). When set tem-
perature is reached, the alarm beeps. Press any key to stop
the alarm.
2. Low Alarm Setting Press LOW AL - LOW sign appears. –199°C or –327°F and AL
are displayed.
Press SET/TIMER - the hundred digit flashes. The number can be
chosen from –1 to 13 for °C or –3 to 24 for °F. Press HI AL to increase.
Press LOW AL - the unit digit flashes, numbers 00 to 99 can be
chosen. Press HI AL to increase.
Press SET/TIMER to set Low Alarm. The minimum temperature is –327°F (or –199°C). When the set
temperature is reached, the alarm beeps. Press any key to stop
the alarm.
Backlight Function
Press the light icon - the back light is on. Press the light icon again ­light is off.
PVC Case and Flip-Out Stand
The PVC case protects the thermometer from moisture and outdoor elements. A flip-out stand is attached to the back of the case for bench or table use.
ALL OPERATIONAL DIFFICULTIES
If this thermometer does not function properly for any reason, please replace the battery with a new high quality battery (see “Battery Replacement” section). Low battery power can occasionally cause any number of “apparent” operational difficulties. Replacing the battery with a new fresh battery will solve most difficulties.
BATTERY REPLACEMENT
Erratic readings or faint/no display are all indications that the battery must be replaced. To remove the thermometer from the PVC case pull the sides of the case away from the front of the thermometer and pull off the case. Using a coin, turn the two screws of the battery cover on the back of the unit. Remove the exhausted batteries and replace with 3 new Cat. No. 1105 batteries. Replace battery cover and case.
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281-482-1714
Fax 281-482-9448
E-mail sales@control3.com
www.control3.com
Control Company is
ISO 9001 Quality-Certified by DNV
and ISO 17025 accredited
as a Calibration Laboratory by A2LA.
Timer
1. Alarm Setting for Count Up Timing Press SET/TIMER. AL and ((( appears and the timer displays 0:00:00
(representing hour: minute : second). Press SET/TIMER again. The hour digit flashes. Press HI AL to increase. Press LOW AL to move to setting minutes. The minute digit flashes. Press the HI AL to increase. Press SET/TIMER to return to the normal mode. When the alarm is on, ((( appears on the display. When the set time is reached the unit makes audible beeps. Press any key to stop the alarm or alarm will turn off automatically.
2. Alarm Cancellation The alarm can be cancelled by turning the unit off.
Cat. No. 4015
Traceable® Total-Range Thermometer Instructions
92-4015-00
Rev. 2 051205
Traceable® is a registered trademark
© 2005
of Control Company.
CONTROL COMPANY