TRACEABLE® MANOMETER/PRESSURE/VACUUM INSTRUCTIONS
SPECIFICATIONS
RANGE AND RESOLUTION CHART
Cat. No. 3460 3461 3462 3463
Maximum Allowed PSI 20 30 60 150
PSI Range –5.000 to 5.000 –15.00 to 15.00 –30.00 to 30.00 –100.0 to 100.0
PSI Resolution 0.003 0.01 0.02 0.1
mbar Range –345.0 to 345.0 –1034 to 1034 –2068 to 2068 –6895 to 6895
mbar Resolution 0.2 1 2 4
bar Range –0.3450 to 0.3450 –1.034 to 1.034 –2.068 to 2.068 –6.895 to 6.895
bar Resolution 0.001 0.001 0.002 0.004
cm/H2O Range –351.0 to 351.0 –1055 to 1055 –2109 to 2109 –7031 to 7031
cm/H2O Resolution 0.2 1 2 4
inch/H2O Range –138.0 to 138.0 –415.0 to 415.0 –830.0 to 830.0 –2768 to 2768
inch/H2O Resolution 0.1 0.3 0.5 2
feet/H2O Range –12.00 to 12.00 –35.00 to 35.00 –69.00 to 69.00 –231.0 to 231.0
feet/H2O Resolution 0.01 0.02 0.04 0.2
mm/Hg Range –259.0 to 259.0 –776.0 to 776.0 –1551 to 1551 –5171 to 5171
mm/Hg Resolution 0.2 0.5 1 3
inch/Hg Range –10.00 to 10.00 –31.00 to 31.00 –61.00 to 61.00 –204.00 to 204.00
inch/Hg Resolution 0.01 0.03 0.1 1.0
kg/cm2 Range –0.350 to 0.350 –1.060 to 1.060 –2.110 to 2.110 –7.030 to 7.030
kg/cm2 Resolution 0.001 0.001 0.002 0.004
oz/inch2 Range –80.00 to 80.00 –240.0 to 240.0 –480.0 to 480.0 –1600 to 1600
oz/inch2 Resolution 0.05 0.2 0.3 1
kPa Range –34.50 to 34.50 –103.0 to 103.0 –207.0 to 207.0 –690.0 to 690.0
kPa Resolution 0.02 0.1 0.2 0.4
Display:
Accuracy: ±0.3% of full scale at ±25oC
Size/Weight: 23/4 x 71/8 x 11/8 / 5 oz
Response time: 0.5 seconds
Operating Temp Range:
Features: Memory feature,
CONTROL BUTTONS
Powers unit off and on
REC Record mode
HOLD "Freeze" the pressure reading on the display
DIF - Differential mode
UNIT Cycle meter through pressure units of measurement
Backlight function
OPERATION
On / Off / Auto Power Off
Press the button to turn the meter on. The meter will shut off
automatically approximately 20 minutes after every power on.
The “auto power off” function can be disabled for recording or
operating over longer periods of time by pressing POWER and
HOLD buttons simultaneously before power on. An “n” will
appear in the middle of the screen at which time the HOLD button
can be released, and then the power button. The “auto power
off” mode has now been disabled.
1
/2", 4- digit LCD.
32-122oF (0-50oC)
Backlight display,
RS-232 capability.
Backlight Feature
Press the button and the backlight illuminates for
approximately 20 seconds then turns off automatically.
Measurement Modes
The meter can be set to measure pressure in the following
modes: bar, mmHg, ozin2, kgcm2, psi, inH20, kPa, ftH20, inHg,
cmH20 and mbar. Each press of the UNIT button will cycle
the meter through each of the measurement modes. An
arrow at the bottom of the LCD will indicate the mode
selected. The meter will default to the mode last selected
before meter was turned off.
Clearing Display to Zero
To clear the display to zero disconnect from all pressure and
press the HOLD button for 3 seconds. “o o o o” will appear
on the LCD and the meter value will be cleared to zero.
Hold Mode
Pressing the HOLD button once will “freeze” the measurement reading. “HOLD” will appear at the upper left of the LCD
indicating the meter is in the hold mode. Press the HOLD
button again to return the meter to pressure measurement
mode.
(Continued on next page )
TRACEABLE® MANOMETER/PRESSURE/VACUUM INSTRUCTIONS
(Continued from previous page)
Differential Mode
Differential pressure is the measurement of the difference
between two pressures.
EXAMPLE: Use differential pressure sensor to measure
gauge pressure by leaving one process connection open
at atmosphere and connecting the second sensor port to
your system.
Press the DIF button, “DIF” appears at the top of the LCD.
The display indicates the relative zero – only the amount
of pressure change will be indicated. Press DIF button
again and the meter returns to normal pressure measurement mode.
Recording
1. Turn the meter on and press the HOLD button for 3
seconds. “o o o o” will appear on the display and the
pressure reading will be cleared to zero.
2. Press the REC button to begin recording a reading.
“REC” will appear in the lower left of the LCD, the
timer will begin counting up and the pressure value
will display.
3. To turn the record feature off press and hold the REC
button for 3 seconds.
NOTE: Only the REC, power and backlight buttons will
be operable in the Record mode.
Minimum/Maximum Readings
While in the recording mode the minimum and maximum
readings can be recalled.
1. Press the REC button and “MAX” will display in the
lower left of the LCD. The display will show the
maximum pressure value recorded and the time of
recording.
2. Pressing the REC button again displays the “MIN”
icon on the screen and the time and value of the
minimum pressure recording.
3. Another press of the REC button returns the display to
the current timer and pressure readings.
Error Messages Displayed
“Err.1” - Pressure value exceeds maximum range
“Err.2” - Pressure value is less than minimum range
“Err.3” - When operating in DIF mode, differential pressure value
is larger than maximum range
“Err.4” - Tubing was left connected while setting zero (try reset-
ting zero), if this does not clear the error the sensor or
meter is damaged.
“E10L” or “E2UL” (when operating RS232) - pressure source is
less or over range
ATTACHMENTS
Hose/tubing
RS-232 serial computer output
ABS plastic carrying case
ALL OPERATIONAL DIFFICULTIES
If this unit does not function properly for any reason, replace the
battery with a new high-quality battery (see “Battery Replacement” section). Low battery power can occasionally cause any
number of “apparent” operational difficulties. Replacing the
battery with a new fresh battery will solve most difficulties.
BATTERY REPLACEMENT
Erratic readings, a faint display, no display, flashing display or
“BAT” appearing on the display are all indicators that the battery
must be replaced. To replace the battery remove the tubing from
the unit. Lay the unit face down on a clean, flat surface. Remove
the screw from the battery cover and slide the cover off in the
direction indicated. Replace the exhausted battery with a new 9
volt alkaline battery. Replace the battery cover and screw into
place. Replacement battery Cat. No. 1112.
ACCESSORIES
Cat. No. 8045 Data Acquisition System Accessory
Captures, displays, and stores readings on any PC computer.
Information can be imported into spreadsheets, word processors,
databases, and statistical program. Supplied complete with serial
cable and CD.
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, TX 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E-mail sales@control3.com
www.control3.com
Control Company is ISO 9001 Quality-Certified by DNV and ISO 17025
accredited as a Calibration Laboratory by A2LA.
Cat. No. 3460/3461/3462/3463
Traceable® Manometer/Pressure/Vacuum Instructions
92-3460-00
Rev. 3 010505
Traceable® is a registered trademark of Control Company
© 2005 Control Company