Control Company HYGROMETER User Manual

TRACEABLE® HYGROMETER INSTRUCTIONS
SPECIFICATIONS
Humidity Range: 10 to 95%
Resolution: 0.1oRH Accuracy: ±2%RH mid-range, otherwise ±4% Temperature Range: 0 to 199.9o F Resolution: 0.1
o
Accuracy: ±1oC between 0o and 40oC, otherwise
±2oC.
Computer\ Output:
Baud rate: 1200 bps Data bit: 7 Stop bit: 1 Parity: none Format: Txxx.xF:Hxx.x% cr 1f
1
2
Txxx.xC:Hxx.x% cr 1f
Storage Temp: –20oC to 50oC Operating Temp: 0oC to 60oC Power: One (1) 9-volt alkaline battery
OPERATION
MEASURING MODE
This is the basic function to measure temperature (oF or oC) and relative humid­ity. Press the ON key (1, Fig. 1) to start the measurement.
SLEEP MODE OPTION
1. The unit has a non-sleep mode and a 20 minute sleep mode.
2. To deactivate the sleep mode, turn off the meter (2, Fig. 1), and then press “ON” (1, Fig. 1) and “MN/MX” (5, Fig. 1) keys simultaneously, until “nSL” is displayed.
MN/MX MODE
Read minimum and maximum values by pressing MN/MX key (5, Fig. 1) sequentially. Press it again to return to normal mode. The meter will automati­cally return to normal mode in 10 seconds.
HOLD MODE
Press the HOLD key (6, Fig. 1) once to hold the current measurement. The screen will display “HOLD”. Press HOLD again to return to normal mode.
COMPUTER OUTPUT
1. Plug the earphone jack of the cable into the computer port socket on the meter and connect 9-pin D-sub to the computer’s COM1 or COM2.
2. Press “ON” key (1, Fig. 1) to start the measurement.
ERROR MESSAGES
1. If the display shows “OP” and the unit continually beeps, then the probe is not connected properly.
2. If display shows 0% or 99% for an extended period of time, then probe is damaged and needs to be replaced.
3. If the display shows Er 1, Er 3 or Er 4 with a beep, then there has been a circuit error.
4. If the display shows Er 2 with a beep, then there has been improper calibra­tion, the probe is damaged, or the humidity reference needs to be changed.
Figure 1
DESCRIPTION:
1. ON key Power on
7
2. OFF key Power off
3 RESET key Reset Min/Max memory
4. °C / °F key Switch between display of °C and °F
5. MN/MX key Minimum/Maximum selection
6. HOLD key Hold display
6
7. PROBE SOCKET For probe connection
4 3 5
ALL OPERATIONAL DIFFICULTIES
If this Hygrometer does not function properly for any reason, please replace the battery with a new Cat. No. 1112 battery. Low battery power can oc­casionally cause any number of “apparent” operational diffi culties. Replacing the battery with a new fresh battery will solve most diffi culties.
Accessories
Cat. No. 4394 33% RH Accessory Calibration Standard with calibration instructions (Not
4395 75% RH Accessory Calibration Standard with calibration instructions (Not
8045 Data Acquisition System Accessory Powerful and easy to use
1112 Replacement battery
Warranty, Service, or Recalibration
For warranty, service, or recalibration contact:
Traceable, but an ideal standard for checking the unit.)
Traceable, but an ideal standard for checking the unit.)
computer data capture/data logging program works with Traceable® Instruments with computer output. Records interval readings from 1 to 10,000 seconds; displays minimum/maximum readings; and utilizes an alarm mode that permits the user to be notifi ed visually, audibly, and by email when an alarm is triggered. Data is stored to a fi le that can be printed in any report or spreadsheet format. Networking server/client capability allows the captured data to be monitored on a remote workstation and/or by email. It is designed to work with Windows® 98/Me/NT/2000/XP/Vista. Includes a CD, a 6-foot cable (supplied USB and serial connections) that plugs into the instrument and computer. Accessory extension cables expand cable length to 300 feet.
Cat. No. 4394 / 4395
CONTROL COMPANY
4455 Rex Road Friendswood, Texas 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E-mail sales@control3.com www.control3.com
Control Company is ISO 9001 Quality-Certified by DNV and ISO 17025 accredited as a Calibratin Laboratory by A2LA
Cat. No. 4185 Traceable® Hygrometer 92-4185-00
Traceable® is a registered trademark of Control Company © 2009 Control Company
Rev. 2 040709