Control Company FLOW METER User Manual

SPECIFICATIONS:
Flow Rate Measurement Unit of Measure Range Resolution
Milliliters per minute (ml/min) ........................ 4.00 to 30.00 ...............0.01
Liters per minute (l/min) ................................ 0.0040 to 0.0300 .........0.0001
Cubic Centimeters per minute (cm3/min) .....4.00 to 30.00 ...............0.01
Fluid Ounces per minute (oz/min) ................ 0.13 to 1.01 .................0.01
Gallons per minute (gal/min) ........................ 0.00105 to 0.00742 .....0.00001
Imperial Fluid Ounces per minute (imp oz)... 0.14 to 1.05 .................0.01
Imperial Gallons per minute (imp gal)........... 0.00088 to 0.00659 .....0.00001
Accuracy: ±2% full scale @ 20 to 30 °C liquid temperature, ±4% full scale otherwise Measurement Modes: Current Flow Total Flow Fluid Viscosity: 0 to 200 centipoises Fluid Temperature Range: 0 to 50 °C
Acceptable Tubing: All types Tubing Connector Size: Inside diameter of nominal 1/8” ID (3.2 mm) Wetted Parts: Nylon and PharMed® Tubing CAUTION: Prior to use, a chemical resistance chart should be used to insure
that wetted parts will not be affected by the chemical being pumped.
BUTTON FUNCTIONS
OPERATION
1. Plug the power supply into the unit.
2. Plug the power supply into an electrical outlet.
3. Remove red covers from the inlet and outlet barbed pump fi ttings.
4. Connect the incoming tubing to the Inlet barb.
5. Connect tubing to the Outlet barb.
6. Press the POWER button to turn the unit on.
7. Press the UNITS button to select the desired unit of measure to display.
8. Press the FLOW/TOAL to toggle between the Current Flow Rate and Total Flow display.
9. Press the POWER button to turn the unit off.
SETTING THE K-FACTOR
To insure the most accurate measurement, the K-factor should be set. The
meter is calibrated by fl owing know amounts of water at known temperatures through the fl ow meter. The measurement accuracy of the meter can be affected by a diffi rent liquids viscosity, liquid temperature, system pressure, tubing diameter, etc. Insert the fl ow meter into your system, and zero the Total
Flow. Pump a known volume of liquid through your system. Calculate the K-factor by dividing the known liquid volume by the Total Flow indicated by the Flow Meter and set the K-factor accordingly.
To set the K-Factor:
1. Press the FLOW/TOTAL button to select the Total Flow mode.
2. Press the UNITS button until the K-factor is diplayed, indicated by “K=”.
3. Press the ZERO button to increment the selected digit.
4. Press the POWER button to change the digit selected.
5. With the desired K-factor set, press the TOTAL/FLOW button exit from the
K-factor setting mode.
Toggles the LCD display between showing the Current Flow rate and
Total Flow.
Selects the unit of measure to be displayed. Also used when setting
the K-Factor.
Zeros the Total Flow. When the fl ow rate is less than 10 mL/min, zeros
the current fl ow rate. Also used when setting the K-Factor
Turns the unit On/Off. Also used when setting the K-Factor.
WARRANTY, SERVICE OR RECALIBRATION
For warranty, service or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281 482-1714 Fax 281 482-9448
E-mail sales@control3.com www.control3.com
Control Company is ISO 9001 Quality-Certifi ed by DNV and
ISO 17025 accredited as a Calibration Laboratory by A2LA.
TRACEABLE
®
FLOW METER
INSTRUCTIONS
Control Company Cat. No. 3500
Traceable® is a registered trademark of Control Company.
2010 Control Company. 92-3500-00 Rev. 0 01262011
©