TRACEABLE® DUAL-DISPLAY LIGHT METER INSTRUCTIONS
SPECIFICATIONS
Ranges:
Lux: 0 to 1999 Lux
2000 to 19,990 Lux
20,000 to 50,000 Lux
Foot Candle: 0 to 199.9 Fc
200 to 1999 Fc
2000 to 5000 Fc
Relativity: 0 to 1999% (relative to desired range and measured value)
Resolution: 1 Lux (0 to 1999 Lux)
10 Lux (2000 to 19,990 Lux)
100 Lux (20,000 to 50,000 Lux)
0.1 Foot Candle (0 to 199.9 Fc)
1 Foot Candle (200 to 1999 Fc)
10 Foot Candle (2000 to 5000 Fc)
Sampling time: Approximately 0.4 seconds
Case: ABS plastic
Fail safe: Low battery Indicator
Power: One (1) 9-Volt alkaline battery
Accessories
Supplied: photo sensor probe, battery, Traceable
Instructions
®
Certificate,
FIGURE 1:
DUAL-DISPLAY LIGHT METER OPERATION
1. Turn the meter on by pressing the POWER BUTTON (3-2, fig. 1)
2. Zero Adjustment Cover the LIGHT SENSOR (3-12, fig. 1) with the
SENSOR COVER (3-13, fig. 1) for a more accurate zero reading. Set
the RANGE SWITCH (3-1, fig. 1) to 2000 LUX position. This is the far
left setting. Press the ZERO BUTTON (3-9, fig. 1). The display should
read 0.0. Remove the sensor cover.
3. Select the desired measurement unit by pressing the LUX/FC BUTTON
(3-4, fig. 1). The display will alternate between “Lux” and “Ft-cd” with
each press of the button.
4. Select the desired lighting source by pressing the LIGHT SOURCE
BUTTON (3-8, fig. 1). The lower left portion of the display will alternate
between “L” (Tungsten), “F” (Fluorescent), “S” (Daylight), and “C”
(Mercury) with each press of the button.
5. Select the desired range by switching the RANGE SWITCH (3-1, fig. 1).
The left position (2000 Lux/ 200 Fc) measures 0 to 1999 Lux/ 0 to
199.9 Fc. The middle position (20,000 Lux/ 2000 Fc) measures 2000 to
19,990 Lux/ 200 to 1999 Fc. The right position (50000 Lux/ 5000 Fc)
measures 20,000 to 50,000 Lux/ 2000 to 5000 Fc.
NOTE: If the display shows “-------” then there is too much light for the
meter to read in this range. Select the next higher range. If the display
shows “____” then there is not enough light for the meter to read in
this range. Select the next lower range.
6.
Place the LIGHT SENSOR (3-12, fig. 1) directly under the light source
you wish to measure. Hold for at least 0.4 seconds. The meter will
measure the light and display the value. When the range is set to
20000 Lux, “0” will appear on the lower line of the display. This is the
last digit of the reading. When the range is set to 50000 Lux, “00” will
appear on the lower line of the display. These are the last two digit of
the reading for example, if the range is set to 2000 and the reading
displays the following:
7. To hold a measurement on the display, press the HOLD BUTTON (3-3,
fig. 1) while a measurement is being taken. The LCD will show OH
DESCRIPTION
3-1 LCD Display
3-2 Power Button
(Data Hold) in the upper left portion of the display to indicate that the
value is a “held” value. To cancel the data hold feature, simply press
the HOLD BUTTON a second time.
3-3 Hold Button: press to hold reading
3-4 LUX/FC (Foot Candle) Button
3-5 LCD Contrast Adjustment Knob
3-6 Record Button: press to record results.
3-7 Recall Button: press to show minimum and maximum readings.
3-8 Light Source Button
3-9 Zero Button
3-10 % Button (Relativity)
3-12 Light Sensor
3-13 Sensor Cover
3-14 Sensor Plug
3-15 Sensor Input Socket
3-16 Computer Output Socket
3-17 Battery Compartment
8. To show the relative % light measurement, press the % BUTTON (310, fig. 1) while a measurement is being taken. Subsequent
measurements will be shown as a percentage of the value of the
reading when the % BUTTON was pressed. To cancel the % feature,
simply press the % BUTTON a second time.
9. To record a measurement for both Memory Recall and RS—232 Data
Acquisition, press the RECORD BUTTON (3-4, fig. 1) while a
measurement is being taken. The LCD will show REC in the lower left
portion of the display to indicate that the value is being recorded. To
deactivate the record function, press the RECORD BUTTON again.
Traceable® is a registered trademark of Control Company ©2009 Control Company. 92-3252-00 Rev. 1 031909
The actual reading is 15620 Lux.
10. Memory Recall: Press the RECORD BUTTON. Use the RECALL
BUTTON (3-6, fig. 1) to recall the minimum and maximum readings.
With the REC symbol on the display, press the RECALL BUTTON (3-6,
fig. 1) once. The maximum recorded value will be displayed. The letters
“Max” will also appear, indicating that this is the maximum or highest
reading. A second press of the RECALL BUTTON will display the
minimum reading. The letters “Mm” will appear indicating that this is the
minimum or lowest reading.
NOTE: The Data Record function must be on to utilize the Memory recall
features. Once the RECORD BUTTON has been pressed a second time to
deactivate the data record function, the minimum and maximum values are
no longer stored.
COMPUTER INTERFACE
This unit features computer output. A COMPUTER OUTPUT SOCKET (3-11,
fig. 1) is located on the top of the unit. To utilize this feature, connect the unit
to a PC with an accessory Data Acquisition System. See ACCESSORIES.
BATTERY LIFE
If the letters ‘LBT’ appear on the left corner of the display, it indicates the
batteries are low and need to be replaced. To replace the battery, slide the
battery cover located on the back of the unit away from the unit. Remove the
old battery and replace it with a new 9-Volt alkaline battery. Use an alkaline
battery, NOT a regular or heavy duty battery. Properly connect the battery.
Replace the battery cover. Incorrectly installed batteries may damage
electronics.
The unit has an automatic shut off feature to prolong battery life. If no button
on the unit is pressed for ten minutes, the unit will automatically shut off. To
deactivate this feature, press the RECORD BUTTON while a measurement
is being taken.
LCD CONTRAST ADJUSTMENT
The Contrast of the LCD may need to be changed either because of light
conditions, battery voltage drop, or user preferences. The LCD contrast can
be adjusted using the LCD CONTRAST ADJUSTMENT KNOB (3-5, fig. 1).
Simply turn the knob to the left or right until the optimal LCD contrast is
obtained. If the display is filled with “88888” when no reading is being taken,
the contrast needs to be adjusted. Turn the contrast knob to the left until a
single “0” appears.
ALL OPERATIONAL DIFFICULTIES
If this Light Meter does not function properly for any reason, please replace
the battery with a new 9-Volt alkaline battery (see Low Battery section,
above). Low battery power can occasionally cause any number of “appare nt”
operational difficulties. Replacing the battery with a new fresh battery will
solve most difficulties.
ACCESSORIES
Cat. No. 4138 Easy-Use Accessory Adaptor 115 VAC
Bonus Free Disk This demonstration disk shows the data acquisition capabilities of
this digital product when connected to a computer. See Cat No. 4136 below to order
complete Data Acquisition System.
Cat. No. 4136 Data Acquisition System Accessory Powerful and easy to use
computer data capture/data logging program works with Traceable
computer output. Records interval readings from 1 to 10,000 seconds; displays
minimum/maximum readings; and utilizes an alarm mode that permits the user to be
notified visually, audibly, and by email when an alarm is triggered. Data is stored to a
file that can be printed in any report or spreadsheet format. Networking server/client
capability allows the captured data to be monitored on a remote workstation and/or by
email. It is designed to work with Windows
®
98/Me/NT/2000/XP/Vista. Includes a CD,
®
Instruments with
a 6-foot cable (supplied USB and serial connections) that plugs into the instrument and
computer. Accessory extension cables expand cable length to 300 feet.
Cat No. 4325 Data Logger Complete DAS–4™ System captures and stores up to
8000 bytes (over 1000 readings). Readings may be taken at intervals from 1 second to
8 hours. Stored readings may be downloaded to any PC and viewed. Can be read “asis” or imported to spreadsheets, databases, and statistical programs. Supplied
complete with 36-inch cable with D9F computer plug, CD, Windows
alkaline batteries.
Cat. No. 4326 Accessory Adaptor 115 VAC for Data Logger
®
, and four AA
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E-mail sales@control3.com
www.control3.com
Control Company is ISO 9001 Quality-
Certified by DNV and ISO 17025 accredited
as a Calibration Laboratory by A2LA.
Traceable® is a registered trademark of Control Company ©2009 Control Company. 92-3252-00 Rev. 1 031909