SPECIFICATIONS
Conductivity—
Range: 0 to 1999 µS (micromhos)
Resolution: 1
Range: 0.00 to 19.99 mS (mhos)
Resolution: 0.01
Automatic Temperature Compensation—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Normalization: 20 °C or 25 °C (user selectable)
Coeffi cient: 0.0 to 4.0% per °C, 0.1 increments
Temperature Measurement—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Resolution: 0.1°
OPERATION
This section contains basic operating instructions, to
insure accurate results and for more advanced functions,
refer to the additional sections that follow.
1. Remove the protective cover from the probe head.
2. Turn the meter on by pressing the
ON/OFF
SET
button.
3. Place the probe head into the solution to be measured.
4. To insure Automatic Temperature Compensation,
immerse the probe head and temperature sensor
completely. For accurate Automatic Temperature
Compensation, the temperature sensor must reach
equilibrium with the liquid being measured, this could
take several minutes.
5. Insure that there is adequate movement of the liquid
around the tip of the probe head (0.2 - 0.3 meters per
second minimum). This can be achieved by stirring the
meter in the liquid.
The meter will automatically select the range that is the
most appropriate for the value being measured. For
example, when measuring a 500 µS solution, the meter
will select the µS range and display 500 µS as opposed
to displaying 0.50 mS. To manually select the range,
press and hold the
Ç
button for approximately three
(3) seconds, if the value being measured is able to be
displayed in the alternate range, the meter will switch
ranges (see the “Range Mode Selection” section).
6. Take the desired measurements.
7. Press the
ON/OFF
SET
button to turn the meter off.
To preserve battery life, always turn the meter off when
not in use.
8. Rinse the probe head with distilled/deionized water and
replace the protective cover. Always replace the
protective cover to prevent damage to the probe head.
If no button is pressed for twenty (20) minutes, the meter
will turn off automatically to preserve battery life (see the
“Automatic Shutoff” section).
POWER ON DISPLAY SEQUENCE
When the meter is turned on (by pressing the
ON/OFF
SET
button), the display will cycle through showing the
following items:
All Segments - Display check
tnr Normalization temperature value (see the
“Selecting Temperature Normalization Value”
section)
tco Temperature Compensation Coeffi cient value
(see the “Setting Temperature Compensation
Coeffi cient” section).
ran Range display mode (see the “Range Mode
Selection” section).
CONDUCTIVITY CALIBRATION
One calibration point may be entered for each of the
measurement ranges (0 to 1999 µS and 0.00 to 19.99 mS).
To insure accuracy:
–Always use Traceable® Conductivity Standards
–Select a Traceable® Conductivity Standard with a value
as close as possible to the value of the unknown sample.
–If possible, the Traceable® Conductivity Standard and
unknown sample should be at the same temperature.
The ideal temperature is 25 °C.
Previous calibration points are cleared when the meter is
calibrated to a new value.
1. Remove the protective cover from the probe head.
2. Turn the meter on by pressing the
ON/OFF
SET
button.
3. Place the probe head into the Traceable® Conductivity
Standard.
4. To insure Automatic Temperature Compensation,
immerse the probe head and temperature sensor
completely. For accurate Automatic Temperature
Compensation, the temperature sensor must reach
equilibrium with the liquid being measured, this could
take several minutes.
5. Insure that there is adequate movement of the liquid
around the tip of the probe head (0.2 - 0.3 meters per
second minimum). This can be achieved by stirring the
meter in the liquid.
6. Press and hold the
È
button for approximately
HLD/CAL
three (3) seconds, CAL will appear at the top of the
display and the conductivity measurement will fl ash.
Ç
7. Press the
and/or
È
buttons to adjust the
HLD/CAL
display to the know value of the Traceable® Conductivity
Standard. The meter may only be adjusted by ±30%
of the detected value.
8. With the the know value of the Traceable® Conductivity
Standard appearing and CAL no longer fl ashing,
press the
ON/OFF
SET
button, the meter will return to the
measurement display.
VIEWING CALIBRATION DATA
1. Turn the meter on by pressing the
2. Press and hold the
ON/OFF
SET
ON/OFF
SET
button.
button for approximately
three (3) seconds, P1.0 will appear at the bottom of
the display.
3. Press the
appear at the bottom of the display.
4. Press the
Ç
button two (2) times, P4.0 will
ON/OFF
SET
button, P4.1 will appear at the
bottom of the display. The top of the display will show
the value that the meter was calibrated to for the
range of 0 to 1999 µS. (If no user calibration has been
performed, “- - - ” will appear.)
Ç
5. Press the
button, P4.2 will appear at the
bottom of the display. The top of the display will
show the value that the meter was calibrated to for the
range of 0 to 19.99 mS. (If no user calibration has
been performed, “- - - ” will appear.)
6. Press the
bottom of the display.
7. Press and hold the
ON/OFF
SET
button, P4.0 will appear at the
ON/OFF
SET
button for approximately
three (3) seconds, the meter will return to the
measurement display.
SELECTING °F OR °C
1. Turn the meter on by pressing the
2. Press and hold the
ON/OFF
SET
ON/OFF
SET
button.
button for approximately
three (3) seconds, P1.0 will appear at the bottom of
the display.
3. Press the
ON/OFF
SET
button, the temperature unit of
measure will fl ash on the display.
4. Press the
Ç
button, the temperature reading will
switch to the opposite unit of measure. If the meter
had been displaying °F, it will now display °C, and
Ç
vice-versa. Each press of the
button will switch
the display to the opposite unit of measure.
5. With the desired unit of measure appearing, press the
ON/OFF
SET
button three (3) times, P1.0 will appear at the
bottom of the display.
6. Press and hold the
ON/OFF
SET
button for approximately
three (3) seconds, the meter will return to the
measurement display.
The unit of measure selected will be the default unit of
measure until changed.
RANGE MODE SELECTION
The meter will automatically select the range that is the
most appropriate for the value being measured. For
example, when measuring a 500 µS solution, the meter
will select the µS range and display 500 µS as opposed
to displaying 0.50 mS.
To manually select the display range, press and hold
the
Ç
button for approximately three (3) seconds, if
the value being measured is able to be displayed in the
alternate range, the meter will switch ranges.
Each time the
Ç
button for pressed and held, the range
mode will cycle through the available modes:
ATO Automatic range selection
1 0 to 1999 µS
2 0.00 to 19.99 mS
HOLD FUNCTION
The hold function allows the readings on the display to be
“frozen” so that they may be recorded.
È
1. Press and release the
button to “freeze”
HLD/CAL
the display at the current reading. “HOLD” will
appear at the top of the display.
2. Once the reading has been recorded, press and
release the
È
button to return the display to
HLD/CAL
the current reading. “HOLD” will no longer
appear on the display.
AUTOMATIC SHUTOFF (Disable)
If no button has been pressed for twenty (20) minutes,
the Automatic Shutoff feature will turn the meter off to
preserve battery life.
To disable the Automatic Shutoff feature:
1. With the meter turned off, press and hold the
button, then press and hold the
“n ” appears on the display.
2. Release the
È
HLD/CAL
and
ON/OFF
SET
ON/OFF
SET
buttons. The meter
È
HLD/CAL
button until
will power on and the Automatic Shutoff feature will
be disabled.
The Automatic Shutoff feature is automatically enabled
when the meter is powered off and turned back on.
SELECTING TEMPERATURE NORMALIZATION
VALUE
Conductivity is greatly infl uenced by temperature. Most
fl uids increase in conductivity as temperature increases.
Depending on the value selected here, Automatic
Temperature Compensation (ATC) will automatically
make corrections to the reading and display a value as
if the sample was 25 °C or 20 °C, no matter what the
actual sample temperature is. The international standard
temperature normalization value is 25 °C.
The following procedure is used to change the
temperature normalization value from 25 °C to 20 °C, or
from 20 °C to 25 °C.
ON/OFF
SET
1. Turn the meter on by pressing the
2. Press and hold the
ON/OFF
SET
button for approximately
button.
three (3) seconds, P1.0 will appear at the bottom of
the display.
3. Press the
ON/OFF
SET
button two (2) times, the temperature
normalization value will fl ash on the display.
4. Press the
value will change. Each press of the
Ç
button, the temperature normalization
Ç
button will
switch between 25.0 °C (77.0 °F) and 20.0°C (68.0°F).
5. With the desired temperature normalization value
appearing, press the
P1.0 will appear at the bottom of the display.
6. Press and hold the
ON/OFF
SET
button two (2) times,
ON/OFF
SET
button for approximately
three (3) seconds, the meter will return to the
measurement display.
The temperature normalization value selected will be the
default value until changed.
SETTING TEMPERATURE COMPENSATION
COEFFICIENT
Conductivity is greatly infl uenced by temperature. Most
fl uids increase in conductivity as temperature increases.
The value set here will determine how much the
Automatic Temperature Compensation (ATC) feature will
adjust the value for every 1 °C in order to automatically
make corrections to the reading and display a value as
if the sample was 25 °C or 20 °C (see the “Selecting
Temperature Normalization Value” section), no matter
what the actual sample temperature is.
Although every fl uid will have a different temperature
coeffi cient, a temperature compensation coeffi cient of
2.0% is a generally accepted standard compensation
value. If the temperature coeffi cient of the solution being
measured is known, that value should be entered in
order to obtain the most accurate results. To take nontemperature compensated measurements, a temperature
compensation coeffi cient of 0.0% should be entered.
The following procedure is used to set the temperature
compensation coeffi cient.
ON/OFF
SET
1. Turn the meter on by pressing the
2. Press and hold the
ON/OFF
SET
button for approximately
button.
three (3) seconds, P1.0 will appear at the bottom of
the display.
3. Press the
temperature coeffi cient value will fl ash on the display.
4. Press the
ON/OFF
SET
button three (3) times, the
Ç
and/or
È
buttons to adjust the
HLD/CAL
display to the desired temperature coeffi cient value.
5. With the desired temperature coeffi cient value
appearing, press the
P1.0 will appear at the bottom of the display.
6. Press and hold the
ON/OFF
SET
button one (1) time,
ON/OFF
SET
button for approximately
three (3) seconds, the meter will return to the
measurement display.
The temperature coeffi cient value selected will be the
default value until changed.
RESET TO FACTORY DEFAULTS
Resetting the meter will clear all calibration data and
reset the user selectable values to the factory defaults.
Factory defaults are as follows:
Temperature display: °C
Normalization Temperature: 25 °C
Temperature Coeffi cient: 2.1%
The following procedure is used to reset the meter to the
factory default values.
ON/OFF
1. Turn the meter on by pressing the
2. Press and hold the
ON/OFF
SET
SET
button.
button for approximately
three (3) seconds, P1.0 will appear at the bottom of
the display.
3. Press the
appear at the bottom of the display.
4. Press the
5. Press the
6. With
Ç
button one (1) time, P3.0 will
ON/OFF
SET
button, the
Ç
button, they will fl ash on the display.
y
appearing, press the
n
will fl ash on the display.
ON/OFF
SET
button, P3.0
will appear at the bottom of the display and the values
will have been reset to the factory defaults.
7. Press and hold the
ON/OFF
SET
button for approximately
three (3) seconds, the meter will return to the
measurement display.
DISPLAY MESSAGES
Upper Display Messages:
--- Indicates that the meter is in the manual
range selection mode and the value being
measured is higher than 1999 µS.
E03 Indicates that the value of the solution being
measured exceeds the measurement range of
the unit.
E04 Indicates that the value of the solution being
measured exceeds the measurement range of
the unit based on a temperature measurement
error (see the “Lower Display Messages” section).
Lower Display Messages:
E0
1
Indicates that the temperature sensor is damaged.
E0
2
Indicates that the temperature of the solution
being measured is below the measurement
range of the unit.
E0
3
Indicates that the temperature of the solution
being measured is above the measurement
range of the unit.
BATTERY REPLACEMENT
An erratic display, faint display, no display, or a battery
symbol appearing on the display are all indicators that
the batteries need replacement. The battery cover is
located at the top of the unit. Unscrew the battery cover
by turning it counter-clockwise. Remove the exhausted
batteries and replace them with four (4) new #357/LR44
size silver oxide batteries. Make certain to insert the
new batteries with the proper polarity as indicated by the
“+” and “–” symbols in the battery compartment. The
“+” side of the silver oxide battery is the fl at side with
the engraving. Replace the battery cover and tighten
securely.
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281 482-1714 • Fax 281 482-9448
E-mail sales@control3.com • www.control3.com
Control Company is ISO 9001 Quality-Certifi ed by DNV
and ISO 17025 accredited as a Calibration Laboratory
by A2LA.
TRACEABLE
CONDUCTIVITY
METER
INSTRUCTIONS
®
ALL OPERATION DIFFICULTIES
If this meter does not function properly for any reason,
replace the batteries with a new, high quality batteries
(see the “Battery Replacement” section). Low battery
power can occasionally cause an number of “apparent”
operational diffi culties. Replacing the batteries with new
fresh batteries will solve most diffi culties.
Control Company Cat. No. 4365
Traceable® registered trademarks of Control Company.
2009 Control Company. 92-4365-00 Rev. 0 112409
©