MANUALP/4000_SERVICE_0712 July2012
21
FACTORY IR CO2CALIBRATION
(For use only if correct result cannot be obtained from USER Calibration)
These adjustments are normally carried out by a qualified service person and
require a DC Voltmeter.
There are two modes of calibration for the infra-red CO2system.
1. Zero Offset Calibration (This should never require adjustment)
2. Span calibration. (This may require a slight adjustment if the CO2 calibration
is incorrect)
NB:These adjustments do not calibrate the accuracy of the CO2concentration
measurement, but only adjust the output to match the measuring device.
Before carrying out these procedures insure that the cabinet has been allowed
to stabilise for at least 6 hours.
Before attempting these settings the CALIBRATION CONSTANT should be
centered to read 1.6 to allow for maximum user adjustment from the front
panel in the future.
E.G If the user calibration constant is 0.9 the easiest way to get it to read 1.6
is to follow the user calibration routine but when adjusting the calibration value
press the ‘UP’ button seven times (0.9 + 7 = 1.6). Wait for the calibration
constant to be displayed, if it doesn’t read 1.6 repeat above (either UP or
DOWN) until it does.
CO
2
ZERO OFFSET CALIBRATION
To adjust the ZERO of the CO2display, it is necessary to proceed as follows:
1 Connect a voltmeter to the CO2sensor board output terminals.
2 With JP5 OPEN, connect the shorting jumper onto pins of JP4, use the
‘UP’ or ‘DOWN’ buttons on the CO2PCB to adjust the voltmeter to read
0.005 volts DC. (The cabinet display will read about 0.2% CO2)
3 REMOVE the shorting jumper from pins of JP4.
CO2SPAN CALIBRATION
To adjust the SPAN of the CO2display, it is necessary to proceed as follows:
1 Connect a voltmeter to the CO2sensor board output terminals (V0 &
COM).
2 With JP5 CLOSED, connect the shorting jumper onto pins of JP4, use
the ‘UP’ or ‘DOWN’ buttons on the DCS-300 INFRA-RED CO2PCB to
adjust the voltmeter to read 0.650 volts DC. (The cabinet display will read