CONTEC MEDICAL SYSTEMS CO.,LTD
Pulse Oximeter Probe user manual
Product name:
Product model: ESA0008,ESA0015,ESC0029,ESA0004,ESA0014,ESA0005,ESA0016,ESA0061
Scope of application:
This product is used to match CONTEC Patient Monitor 、Pulse Oximeter and Electronic sphygmomanometer ,collect and transmit the
signal from patient with continuance and no trauma. It is inapplicable to monitor the weak perfusion moving state and monitor for
SpO
2
long, so check the measuring position or change for another position per 4 hours.
Taboo disease:
Don't fix the product on the position with tissue injury. It is inapplicable for the patient or users allergic to PVC、TPU、TPE、ABS plastic.
Product performance:
1) The range of SpO2 measurement: 70%~100%;
Accuracy: 70~100%:±2%; Below 70%: unspecified.
2) The range of pulse measurement: 30~250bpm;
Accuracy: ±2bpm or ±2%(select larger).
3) Optical Sensor:
Red light (wavelength is 650~670nm, 6.65mW)
Infrared (wavelength is 880~910nm, 6.75mW)
Main configuration: Consisting of plug, cable and probe.
Power supply requirement:
and Electronic sphygmomanometer which are applicable to the requirements of IEC60601-1.
Directions for use:
This product is type BF applied part;
Note:
Pulse Oximeter Probe
The special power is supplied by from the equipments of CONTEC Patient Monitor 、Pulse Oximeter
Sketch map Model explanation Applied crowds placement
Reusable adult
finger-clip SpO2 probe
(ESA0008,ESA0004,
ESA0005,ESA0014,
ESA0061)
Reusable wrap SpO2
Reusable child finger-clip
SpO2 probe
(ESA0015,ESA0016)
probe(type Y probe)
(ESC0029)
Weight>40Kg adult
Weight10~40kg
child
Weight3-10kg
neonate
Recommendatory
placement:forefinger
Recommendatory
placement:forefinger
Recommendatory
placement:sole of foot
Figure 1
1) As Figure 1, the pulse oximeter probe of different types is applied to different crowds.
2) Select proper probe and put recommendatory placement according to Figure 1.
3) Arrange the cable along the back of hand when place the pulse oximeter probe.
4) Connect Pulse oximeter probe with Pulse oximeter ,Patient Monitor or Electronic sphygmomanometer and check if the operating
procedure accords with the procedure introduced in user manual.
5) Pulse Oximeter probe ESC0029 needs the help of the FST0001 Pulse Oximeter probe extension cable to be connected in to the jack
of the Pulse Oximeter CMS60D,CMS70A or Patient Monitor CMS8000(old model),Pulse Oximeter probe ESC0029 needs the help of the
FST0004 Pulse Oximeter probe extension cable to be connected in to the jack of the Electronic Sphygmomanometer
CONTEC08A,Pulse Oximeter probe ESC0029 needs the help of the FST0002 Pulse Oximeter probe extension cable to be connected in
to the jack of the Patient Monitor PM50. Pulse Oximeter probe ESC0029 needs the help of the FST0014 Pulse Oximeter probe
extension cable to be connected in to the jack of the Patient Monitor CMS8000(new model)
1
CONTEC MEDICAL SYSTEMS CO.,LTD
Notice items:
1) pulse oximeter probe placement, the position without ductus arteriosus, BP cuff and vein input pipe is top-priority.
2) If the pulse oximeter probe can't monitor the state of pulsation, it shows that the position of probe is improper, or the position is too
thick, too thin or having too deep pigment to reach a proper translucidus effect. If above things has happened, place the probe again or
select probe of other type.
3) This pulse oximeter probe should be applied to the special medical equipment. Operator is responsible to check the compatibility
Incompatible fittings or device will influence the measuring result.
4) The disposal of scrap instrument and its accessories and packing (including battery, plastic bags, foams and paper boxes) should
follow the local laws and regulations.
Maintenance/cleaning/disinfection:
1) Check if the product is undamaged and clean before using.
2) This product is not allow to use disinfection liquid for disinfection, this probe belong to one-off products.
Note: Don't immerse the product in the liquid, and don't expose it under the strong ultra-violet radiation
Service life: Suggest this product use only once, don't use again.
Environment requirements:
Transport and storage
1) Temperature: -10℃~+40℃
2) Humidity: less than 80%
3) Pressure:86kPa~106kPa
Operating
1) Temperature: 10℃~+40℃
2) Humidity: 30% ~ 75%
3) Pressure: 700hPa~1060Pa
Statement:
1) pulse oximeter probe needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided in User Manual and test report.
2) Portable and mobile RF communications equipment can affect pulse oximeter probe.
Warning:
1) The use of cables other than those specified, with the exception of cables sold by CONTEC as replacement parts for internal
components, may result in increased emissions or decreased immunity of pulse oximeter probe.
2) pulse oximeter probe should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary,
the pulse oximeter probe should be observed to verify normal operation in the configuration in which it will be used.
3) Improper usage can result in inaccurate measurement.
4) Using it under too strong light will cause inaccurate measurement, in case of that, please set a opaque stuff around the probe to cut
light off.
5) You should move the probe to other position per 4 hours at least. Because the state of local skin can influence the ability of skin to
enduring probe, it is necessary to replace the position of probe according to the state of patient. Please do that when skin integrity
changes.
6) The dyestuff in blood vessel cab cause the inaccurate measurement.
7) The performance of pulse oximeter probe is influenced by movement easily, so it is not suitable for active patient to use it.
8) Don't fix the probe with belt or bundle it tightly, because the vein pulsation can cause inaccurate SpO
9) Same as other medical equipment, the cable should be set properly to avoid enlacing or asphyxiate patient.
10) Don't use it in the process of MRI scan, because the conductor current may burn the skin of patient, moreover, the probe will
influence MRI image and MRI set will also influence the accuracy of SpO
11) Don't change the product at will, otherwise the capability or accuracy of product will be influenced.
12) The probe is not intended for use during patient transport outside the healthcare facility.
13) DO NOT use the probe while the patient is being scanned by MRI or CT.
2
measurement.
2
measurement.
2