Please read the User Manual carefully before using this product. The User Manual which
describes the operating procedures should be followed strictly. This manual detailed
introduce the steps must be noted when using the product, operation which may result in
abnormal, the risk may cause personal injury and product damage and other contents, refer to the
chapters for details. Any anomalies or personal injury and device damage arising from use,
maintain, store do not follow requirements of the User Manual, Our company is not responsible
for the safety, reliability and performance guarantees! The manufacturer’s warranty service does
not cover such faults!
Note: Please read the User Manual carefully before using this product.
Described in this User Manual is in accordance with practical situation of the product. In case of
modifications and software upgrades, the information contained in this document is subject to
change without notice.
The warning items
Before using this product, you should consider the safety and efficacy of the following
described:
The safety classification of this product isclass Ⅱ equipment, type BF.
Described each measurement results combined with clinical symptoms by qualified doctors.
The reliability and operation of using this product whether meets the operation of this
manual relate to the maintenance instructions.
Warning: Replace accessories which not provided by our company may lead to the
occurrence of errors.Without our company or other approved maintenance organizations
trained service personnel should not try to maintain the product.
Responsibility of operator
The operator must carefully read the User Manual before use this product, and strictly
follow the operating procedure of the User Manual
Fully consider the security requirements during product design, but the operator should not
ignore the observation for the patient and the state of machine.
The operator has the responsibility to provide the use condition of the product to our
company.
Responsibility for our company
Our company have the responsibility to provide qualified product which conform to
company standard of this product.
Our company have the responsibility to complete product installation, debugging and
technical training according to the contract.
Our company have the responsibility to complete product maintenance according to the
contract.
Our company have the responsibility to respond the requirements of user in time.
The User Manual is writen by our company. All rights reserved.
I
Contents
Chapter1 Functions and Purpose ............................................................................................. 1
Description of functions:
The Sphygmomanometer apply to measure the non-invasive blood pressure of human (adult,
pediatrics, neonate), each user can store 100 items records of measurement results at most. Each
record includes detailed measuring time, systolic pressure, diastolic pressure, average pressure,
pulse rate and record number, etc. With 2.8 inch color LCD screen, clear interface, the function
of data review is complete. User can implement ON/OFF, manual measuring, system setup,
parameters change and other operations with seven buttons which are located on the front panel
of the device. The device uses audible and visual alarm, when the battery power is low, the
buzzer will intermittent buzzing and LCD screen displays "Low Power" to prompt user replace
batteries. When the measurement data exceeds the set alarm limit, the font color of measurement
results will change to red and the audible alarm will occur, user can open and close the alarm
sound according to needs. Users can send measurement results which be automatically saved by
device to mobile phone with bluetooth transmission function. Refer to the help or explication of
the related software for specific operation.
Purpose:
The device apply to measure the non-invasive blood pressure and SpO
(optional) of human.
2
Record parameter value of blood pressure to provide the reference for the health care
professional
1
q
p
Chapter2 Safety Precautions
●Before use, carefully read the "Safety Precautions" for a correct use.
●To prevent users suffered hurt or damnification due to improper use,see "Safety
Precautions",and use this product properly.
For safety reasons, be sure to comply with safety precautions.
Note
If not to use correctly, it exists that a potentially hazardous situation which may result in
injury to the user or patient or damage to the equipment or other property.
Note
Human and the surrounding environment are in the range of electromagnetic radiation
which generated by this device,there is no energy hazard to human or the surrounding
environment, also not affected by the electromagnetic interference of the external device.
This device has slight energy radiation,should be avoided too close to the high precision
re
uirements equi
ment.
2
Note
Self-diagnosis and treatment using measured results may be dangerous.Follow
the instructions of your physician.
Please hand measurement results to the doctor who know your health to accept
diagnosis.
For severe blood circulation disorder or arrhythmia patients, please use the
device under the guidance of a doctor.
Otherwise it may lead to acute hemorrhage, or measurement error as a result of squeezed
arm.
Infant and the person who can't express oneself can't use the device
Otherwise it may cause accident or dissension
Please do not use for any other purpose.
Otherwise it may cause accident or holdback.
Please use special cuff.
Otherwise it is possible that measurement result is incorrect.
Please do not keep the cuff in the over-inflated state for a long time
Otherwise it may cause risk
Do not disassemble,repair and change the device
Otherwise it can not measure correctly
2.1 Operation about AC Adapter
Note
Please use the dedicated AC adapter of this device.
Otherwise it may cause trouble.
Dedicated AC adapter must use AC 100V~220V.
Otherwise it may cause fire or electric shock.
When there is breakage of dedicated AC adapter plug or wire,please do not use it.
Otherwise it may cause fire or electric shock.
Please do not plug or unplug the adapter on the socket with wet hands
Otherwise it may cause electric shock or injury.
3
y
2.2 Operation about Batteries
Note
Please use 4 "AA" manganese or alkaline batteries, do not use batteries of other
types.
Otherwise it may cause fire.
New and old batteries, different kinds batteries can not be confusion.
Otherwise it may cause battery leakage, heat, rupture, and damage to Electronic
Sphygmomanometer.
Please don't put wrong the positive and negative of battery. When the batteries
power exhausts, replace with four new batteries at the same time.
Please take out the dry batteries when you do not use the device for a long time.
Otherwise it may cause battery leakage, heat, rupture, and damage to Electronic
Sphygmomanometer.
If electrolyte of the batteries immodestly get in your eyes, immediately rinse with
plenty of clean water.
It will cause blindness or other hazards,Should immediately go to the nearest hospital
for treatment
If electrolyte of the batteries immodestly glues on the skin or the clothes,
immediately rinse with plenty of clean water.
Otherwise it ma
hurt the skin.
Advice
Do not subject the device to strong shocks, such as dropping the unit on the floor;
Do not inflate before the cuff wrapps around the arm;
Do not inflect the cuff and the air tube forcibly.
4
g
y
Cuff Plug
Cuff Air Plu
Chapter3 Main Unit
Displa
Up/Downbutton
Memory button
AC adapter socket
Menu button
On /Off button
Start /stop button
USB socket
Accessories:
Adult Cuff Dry Battery
(Cuff specification: limb circumference 22-32cm (middle part of upper arm))
User Manual
Optional Accessories:
AC adapter
Input: voltage: AC 100V~220V
frequency: 50Hz/60Hz
Rated current:AC 150mA
Output: DC 6.0V±0.2V 1.0A
or DC5.0V±0.2V 1.0A
5
SpO2 Probe: Y10UCH150
A. SpO2 measurement
Range:0%~100%
Error: 70~100%:±2%;Below 70%:unspecified
B. Pulse rate measurement
Range:30bpm~250bpm
Error: ±2bpm or ±2%(select the larger)
C. Optical sensor: red light(wavelength: 660nm,
6.65mW) infrared light(wavelength: 880nm, 6.75mW)
6
Chapter4 External Interfaces
Note
Please hold the air plug to remove the NIBP cuff.
①Cuff Socket
Left side
The right side of the device is USB socket
and AC adapter socket
USB socket
AC adapter socket
Right side
Note
The cuff socket and AC adapter socket of the sphygmomanometer only connect to the
stated equipment which accord with the requirements of IEC 60601-1: 1988 “Medical
electrical equipment-Part 1: General requirements for safety”.
7
①③②
Chapter5 Dry Battery/AC Adapter Installation
This product can use dry batteries as well as AC adapter.
5.1 Dry Battery Installation
① Demount the battery cover in the direction of the arrow.
② Install "AA" dry batteries according to
③ Slide to close the battery cover.
“” :the batteries power will exhaust. Replace with four new batteries (the same
Icon
sort) at the same time. Test while low power may cause data deviation and other problems.
Turn the unit off before replacing the batteries.
polarities.
Note
Dispose of the batteries according to applicable local regulations about environmental.
5.2 Using the AC adapter
① Connect sphygmomanometer and the AC adapter. Insert the AC Adapter Plug into the
AC Adapter socket on the right side of the device.
② Please insert the power plug of AC adapter into the AC 100V~240V socket.
Note
When cut off the power, first cut off the connection of power socket and the regulated
power supply, then cut off the connection of regulated power supply and the
sphygmomanometer.
Please be sure to use dedicated AC adapter.
Note
When regulated power supply and dry batteries are both used at the same time, the dry
8