This document contains proprietary information protected by copyright law. No part of this
document may be photocopied, reproduced or translated to another language without prior written
consent of CONTEC.
CONTEC assumes no responsibility for any errors that may appear in this document, or for
incidental or consequential damage in connection with the furnishing, performance or use of this
material. The information contained in this document is subject to change without notice.
Responsibility of the Manufacturer
CONTEC only considers itself responsible for any effects on safety, reliability and performance of
the equipment if:
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by
persons authorized by CONTEC, and the electrical installation of the relevant room complies with
national standards, and the instrument is used in accordance with the instructions for use.
Note: This device is not intended for home use.
WARNING: This device is not intended for treatment.
If there is doubt as to fetal well-being after using the unit, further investigations should be
undertaken immediately using alternative techniques.
The accuracy of FHR is controlled by the equipment and can not be adjusted by user.
If the FHR result is distrustful, please use other method such as stethoscope to verify
or contact the local distributor or manufacture to get help.
Using This Label Guide
This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING label advises against certain actions or situations that could result in personal injury
or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
Note: A NOTE provides useful information regarding a function or procedure.
1 Safety Guidance
1.1 Introduction For the Safe Operation
The CMS800G Ultrasonic Fetal Doppler Monitor (Monitor) is Class I equipment and
designed to comply with IEC 60601-1.
Switching within 1 minute, at ambient temperatures between 5℃ and 40℃ . Ambient
temperatures that exceed these limits could affect the accuracy of the instrument and cause
damage to the modules and circuits. Allow at least 2 inches (5cm) clearance around the
instrument for proper air circulation.
The user must check the equipment, cables and transducers do not have visible evidence of
damage that may affect patient safety or monitoring capability before use. The recommended
inspection interval is once per week or less. If damage is evidence, replacement is
recommended before use.
The user must be serviced only by authorized and qualified personnel, The manufacturer can
not accept responsibility for safety compliance, reliability and performance if modifications
or repairs are carried out by unauthorized personnel. Identical replacement parts must be
used.
Perform period safety testing to insure proper patient safety. This should include leakage
current measurement and insulation testing. The recommended testing interval is once per
year.
The protection categories against electric shock of the patient connections are:
①FHR1 ②FHR2 ③TOCO MARK④
This symbol indicates that the instrument is IEC 60601-1 Type B equipment. Type B protection
means that these patient connections will comply with permitted leakage currents, dielectric
strengths and protective earthing limits of IEC 60601-1.
The monitor described in this user manual is not protected against:
A)The effect of defibrillator shocks
B)The effects of defibrillator discharge
C)The interference of high frequency currents
D)The interference of electrosurgery equipment
E)The interference of mobile phone
1.2 Ultrasound Safety Guide
Fetal Use
The Monitor is designed for continuous fetal heart rate monitoring during pregnancy and
labor . Clinical interpretation of fetal heart rate patterns can diagnose fetal and/or maternal
problems and complications.
Instructions for Use in Minimizing Patient Exposure
The acoustic output of the Monitor is internally controlled and can not be varied by the
operator in the course of the examination. The duration of exposure is, however, fully under
the control of the operator. Mastery of the examination techniques described in the User
Manual will facilitate obtaining the maximum amount of diagnostic information with the
minimum amount of exposure.
1.3 Safety Precautions
WARNING and CAUTION messages must be observed. To avoid the possibility of injury,
observe the following precautions during the operation of the instrument.
WARNING: EXPLOSION HAZARD-Do not use the in a flammable atmosphere where
concentrations of flammable anesthetics or other materials may occur.
WARNING: SHOCK HAZARD-the power receptacle must be a three-wire grounded
outlet. A hospital grade outlet is required. Never adapt the three-prong
plug from the monitor to fit a two-slot outlet. If the outlet has only two
slots, make sure that it is replaced with a three-slot grounded outlet before
attempting to operate the monitor.
WARNING: SHOCK HAZARD-Do not attempt to connect or disconnect a power cord
with wet hands. Make certain that your hands are clean and dry before
touching a power cord.
WARNING: The monitor should be installed by an authorized and qualified service
engineer.
WARNING: SHOCK HAZARD-Do not remove the top panel covers during operation or
while power is connected.
WARNING: Only connect the device to Contec supplied or recommended accessories, to
avoid the injury of the doctors and patient.
WARNING: Do not switch on device power until all cables have been properly connected
and verified.
WARNING: Don’t touch signal input or output connector and the patient simultaneously.
WARNING: Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards (e.g. IEC 950 for data
processing equipment and IEC 60601-1 for medical equipment). Furthermore
all configurations shall comply with the valid version of the system standard
IEC 60601-1-1. Everybody who connects additional equipment to the signal
input connector or signal output connector configures a medical system, and is
therefore responsible that the system complies with the requirements of the
valid version of the system standard IEC 60601-1-1. If in doubt, consult our
technical service department or your local distributor.
CAUTION: The device is designed for continuous and is “ordinary” (i.e. not drip or
splash-proof).
CAUTION: Keep the environment clean. Avoid vibration. Keep it far from corrosive
medicine, dust area, high-temperature and humid environment.
CAUTION: Do not operate the unit if it is damp or wet because of condensation or spills.
Avoid using the equipment immediately after moving it from a cold
environment to a warm, humid location.
CAUTION: Do not immerse transducers in liquid. When using solutions, use sterile wipes
to avoid pouring fluids directly on the transducers.
CAUTION
CAUTION
CAUTION: The temperature should not exceed 6
CAUTION: Electromagnetic Interference-Ensure that the environment in which the
CAUTION: The monitor must serviced by proper training and knowledge, practical
Do not use high temperature heating or gas to disinfect the monitor and its
④
accessories.
Turn off the power supply before clean the machine.
④
when clean the belt.
0℃
fetal monitor is installed is not subject to any sources of strong
electromagnetic interference, such as radio transmitters, mobile telephones,
etc.
personnel. The recommended testing interval is once twice year or under the
leakage current measurement and insulation testing.
CAUTION: The device and reusable accessories could be sent back to the manufacturer
for recycling or proper disposal after their useful lives.
2 Introduction
The Fetal Monitor can provide different configurations according to different user requirements:
The multi-crystal, broad beam ultrasound transducer is used for monitoring fetal heart
rate(FHR1).The ultrasound transducer operates at a frequency of 1.0MHz. Put the ultrasound
transducer on maternal abdomen to transmit lower energy ultrasound wave to fetal heart, then
receive the echo signal from it.
TOCO Transducer
④2④
This transducer is a toco tonometer whose central section is depressed by the forward
displacement of the abdominal muscles during a contraction. It is used for assessment of
frequency and duration of uterine contractions. It gives a subjective indication of contractions
pressure.
Remote Marker
④3④
The remote marker is a hand-held switch operated by patient. The mother is normally instructed
to push down the switch when feeling fetal movement.
Ultrasound TransducerⅠ④TOCO Transducer④Remote Marker are three in one transducers, their
sockets are marked FHR1/TOCO/MARK on the monitor panel.
Ultrasound TransducerⅡ is the transducer for FHR 2(Twins Configuration), it’s socket is
④4④
marked FHR2 on the monitor panel.
3.1.2 Left Side Sockets
Figure 3.2 Left Side Sockets
3.1.3 Interfaces and Symbols
FHR1/TOCO/MARK socket: Socket for FHR1/TOCO transducer and remote Marker
FHR2 socket: Socket for FHR2 Transducer
: Socket for Grounding Cable
NET.: Socket for network
: Warning Symbol
Push: LCD Screen rotation lock
3.1.4Main Interface
Figure 3.3 Twins Monitoring interface
The main monitoring interface(Twins Monitoring) is divided into 5 parts according to display
content, they are status bar ④data section④parameter section④indicate bar and wave display section.
It displays in status bar that sound channel and volume④connection status of ultrasound Transducer
I④ultrasound Transducer II(twins monitoring) and TOCO transducer, Freeze status④print
status④alarm on /off; It displays FHR 1 from Ultrasound Transducer I④FHR 2 from Ultrasound
Transducer II (Twins monitoring) and TOCO④Fetal Movement data; parameter section displays the
important parameter of current settings; time④bed number④time of pregnancy and age of pregnant
woman are displayed in indicate bar; waves from ultrasound transducer I channel④ultrasound
transducer II channel(Twins Monitoring) and TOCO transducer are displayed in wave display
section. Detail instruction as followed.
Status Bar
(1(
Sound channel and volume
④A④
Icon
④Sound Icon
④Sound off Icon
I④The No. of FHR sound channel, it is I under single fetal monitoring mode which is
default; I④II selectable under twins monitoring mode, it can be changed through
the main menu.
3④Volume level, ranging from 0-7, 0 stands for sound off. It can be changed through the
buttons on the panel or set in the main menu.
Connection status of ultrasound transducer
④B④
IconⅠ④: : Ⅱ
I④Channel No. of ultrasound transducer, there is only I under single monitoring mode,
there are I and II under twins monitoring mode
④Normal connection of ultrasound transducer
④Error connection of ultrasound transducer
④Ⅰ
:3 Ⅰ:0
Connection Status of TOCO transducer
④C④
④Normal connection of TOCO transducer
④Error connection of TOCO transducer
Freeze status
④D④
④Shows current screen is frozen; icon will disappear when unfrozen.
Recorder status
④E④
④Printing
④out of paper
④Failed to print
Alarm on/off status
④F④
④Alarm on
④Alarm off
Data Section
(2(
FHR 1 Data of Ultrasound Transducer I④3-digit data, it is in green color under normal
status, it will be in red when alarm occurs; it displays “―――” when there is no data.
FHR 2 Data of Ultrasound Transducer II: this data will show in twins monitoring mode, the
display format is the same with the FHR 1.
TOCO data④Display the relative contraction data, ranging from 0-100, it will be 10 after
Auto Zero.
Fetal Movement Data ④Display Fetal movement numbers, it will be “―――” after Auto
Zero.
Parameter Bar
(3(
This section displays important setting parameters: it contains alarm on/off status④alarm
upper limit④lower limit④alarm postpone time④print speed and print time.
Indicate Bar
(4(
In this item, it includes system time④bed No.④gestational age and patient age.
Wavefom Display Section
(5(
This section also be divided into 2 sections, FHR trend graph is displayed in the upper
section, TOCO waveform is displayed in lower section. FHR1 Trend is in green④FHR2 trend
is in Yellow(only displayed in twins monitoring)④the normal range of the fetal heart rate is
120-160bpm④which be showed in green on the screen.
Fetal movement mark “↑”④alarm mark ④event mark “↓” will also be showed in this section.
3.1.5 Buttons
There are several buttons of different functions on the front panel of fetal monitor. The diagram is
showed as Figure 3.4.
Figure 3.4 Buttons
Menu Button
(1(
Push Menu Button to enter setup menu, push it again to return monitor screen. When operating in
other menu, push this button to return this menu. Only turning knob button can exit wave review
mode.
Detailed operation please refer Figure 3.2
(2(
Function: Enable/Stop audio alarm when FHR is in alarm range.
When symbol appears, the alarm indicator status is shut off.
Press the button to enable audio alarm, the alarm indicator becomes , whenFHR is in alarm
situation, the alarm sound will be given out.
(3(
Function: Clear the screen, TOCO value back to 10 unit,
Press this button to clear the screen and adjust the present TOCO contractions trace/value to
reference point 10 when in the status of monitoring, after pressing the AUTO ZERO button, the
symbol “” will be recorded at the trace.
Alarm Button
Auto Zero Button
(4(
Function: Enable/Disable printing.
Press PRINT button in normal situation, if it not works, it begins real-time printing
Press PRINT button in frozen situation, it prints the waveform on the screen.
Print Button
(5(
: Volume down :Volume up
Function: adjust the audio volumeofthe Fetal heart Sound.
④6④
Function: Press this button to print an event symbol on the screen trend figure at the corresponding
time. If user want to mark an event on the trend figure, he/she could achieve this by pushing this
button.
④7④
Function: Freezing the screen. Press the button to stop drawing and the screen becomes in frozen
status, press the button again to continue drawing. This operation will clear the screen.
④8④
Volume Control Button
Event Button
Freeze Button
Knob Key
1 Press the button to activate the selected button, press it again to accept the configuration.
2 To choose and adjust the parameters by revolving the knob key.
Paper Cabinet Open Button
(9 (
Push this button for opening the paper cabinet.
(10(“
Function: Push this button to unlock the screen for rotation purpose.
Note: Please lock the LCD Screen during transportation to avoid any damage of the monitor.
3.2 Setup
Under Main monitoring interface, Press the Menu button or knob key to enter setup mode, the
diagram is showed as Figure 3.5
Push” – LCD Screen Rotation Lock
Figure 3.5 Setup
Revolving knob key to select different function. The Corresponding function and the adjustable
ranges are showed in table3.1.
Table 3.1 The setup function and adjustable ranges
NoFunctionAdjustable Ranges
1ALM SET(Alarm Setup)Enter Alarm Setup
2PAT SET(Patient Setup)Enter Patient Setup
3SYSTEM SET(System Setup)Enter System Setup
4REVIEW(Waveform Review)Enter Waveform Review
5PRINT SET(Print Setup)Enter Print Setup
6MONI TYPE(Monitor Mode)Optional: single, twins the default is
single fetus
7LANGUAGE(Language Selection) Optional: Chinese (CH), English (EN),
the default is EN.
8CHANNEL I(Audio Channel I
Setup)
9CHANNEL II(Audio Channel II
Setup)
10CHANNEL(Audio Channel)Optional: I, II, fetal heart audio come
11EXITExit main menu, back to main interface
Adjustable:1④7 and mute,
the default is 3
Adjustable:1④7 and mute,
the default is 3
from the selected channel.
Alarm Setup
(1(
Revolving the knob key to enter alarm setup, the diagram is showed as Figure 3.6:
Figure 3.6 Alarm Setup
Revolving the knob key to setup alarm function. The Corresponding function and the adjustable
ranges are showed in the table3.2:
Table 3.2 The alarm setup function and adjustable ranges
NoFunctionAdjustable Range
1FHR ALM(FHR Alarm) Optional: Turn on, shut off The default situation
is alarm on.
Note:
1
2
Patient Setup
(2(
2ALM HIGH(FHR Upper Limit of
Alarm)
3ALM LOW(FHR Lower Limit of
Alarm)
4ALM DELAY(FHR Alarm Delay)
When FHR is in alarm situation, alarm indicator becomes red
When FHR exceeds the alarm limit and time exceed the set alarm delay time
continuously, alarm will occur and an alarm symbol will appear on the screen.
Optional: lower limit of alarm-310, the unit is
BPM, and the default is 190
Optional: high limit of alarm 0 ④FHR, the unit is
BPM, and the default is 110
Optional:0④60,the unit is second, and the default
is 30 seconds
Revolving the knob key to enter patient setup, the diagram is showed as Figure 3.7:
Figure 3.7 Patient Setup
Revolving the knob key to setup patient function. The Corresponding function and the adjustable
ranges are showed in the table3.3:
Table 3.3 the patient setup function and adjustable ranges
NoFunctionAdjustable Ranges
1NAME(Name)Optional: 12 letter or numeral. The default is blank
2AGE(Age)Optional: numeral from 1-100. The default is 25
3BED NO.(Bed No.)Optional: numeral from 1-100. The default is 1
4PAT NO.(Case History No.)Optional: 12 letter or numeral. The default is blank
5ROOM(Ward No.)Optional: 5 letter or numeral. The default is blank
6LENGTH(Time of Pregnancy)Optional: 1-100.The unit is week. The default is 0
7BLOOD(Blood Type)Optional: A, B, AB, O, and N(unknown). The default is N.
8PARTUS NUM(Times of Giving BirthOptional: 0-20, the default is 0
9FETAL NUM.(Quantity of Fetus)Optional: 0-20, the default is 1
10PREGNANTNUM(Timesof
Pregnancy)
11WEIGHT(Patient’s Weight)
12HEIGHT(Patient’s Height)
13DELETE(Delete Information)Delete related patient information
14SAVE(Save Information) Save related patient information, return to previous menu
15ExitReturn to previous menu
Note: ① When adjustable range is letter or number, numeral and letter key will turn on
automatically after entering the setup, in which:
SP: Space bar
CAPS: Capital letters lock
OK: Setup finished, exit keyboard output mode
DEL: Delete, delete one selected letter or number after each push.
The main interface prompt box will show the patient’s data renewal after save the
②
patient’s data.
Optional: 0-20, the default is 1
2-250, interval④0.5Kg,unit: Kg or Pound; the default value
is 65Kg
20-300, interval ④ 0.5cm(inch) ④ unit: cm or inch; the
default value is 165cm
System Setup
(3(
Revolving the knob key to enter system setup, the diagram is showed as Figure 3.8:
Figure 3.8 System Setup
Time Setup
①
Revolving the knob key to enter time setup, the diagram is showed as Figure 3.9:
Figure 3.9 Time Setup
Revolving the knob key to enter time setup function. The Corresponding function and the
adjustable ranges are showed in the table3.4:
Table 3.4 The time setup function
NoFunctionAdjustable Ranges
1YEAR
2MONTH
3DAY
4HOUR
5MINUTE
6SECOND
7SAVE SET(Save)Save setup and return to previous menu
8EXIT(Exit)Exit to previous menu
NOTE: The main interface prompt box will show the time renewal after save the time setting.
NET NO.(
Be used for connecting with
System Update(
This device supports system update service. In system setup menu, revolving the knob key to enter
System update.
Note: Please enter password under the item “USR KEY” before click “CONFIRM”. This
password is provided by manufacturer or distributor when Contec add new function to upgrade the
system.
④Version
Revolving the knob key to enter System setup, choose version item and push the knob key to see
the equipment version
Wave review
(4(
Choose the WAVE REVIEW in the setup menu to enter wave review, and press WAVE REVIEW
in this item to review the history wave, which is showed as Figure 3.10
Display Content: bed No. , pregnancy age, age, single/twins type, paper speed, date,
time , volume, alarm status, transducer connection status, recorder status, FHR data
and wave, Contraction data and wave, Fetal move times and mark etc.
Print: Record Paper two-double type Z
Print Width: 112mm
Valid Print Width: 104mm
Paper output speed: 1cm/min④2cm/min④3cm/min(optional)
Data Precision: 5④④X Roll④④±1④④Y Roll
④
Record Content: bed No. ,name, pregnancy age, single/twins type, case No., paper speed,
date, time , FHR data and wave, Contraction data and wave, Fetal move times and mark
etc.
Signal Interface: RJ 45
Ultrasound probe: Nominal Frequency: 1.0MHz
Work Frequency: 1.0MHz±10
④
Negative peak sound pressure : P_④1MPa
2④
Output beam intensity : Iob④20mW/cm
The peak time space peak intensity: Ispta④100mW/cm
The average time space peak intensity: Ispta④10mW/cm
④
2④
④
2④
④
FHR Rang: 65BPM④210BPM④Allow error: within ±2bpm
Resolution: 1BPM
Accuracy : ±2
④
TOCO
TOCO range: 0④100
Resolution: 1
Nonlinear error: ④±10
④
④
④
RZ way: Manually
Fetal Marking
For the manual button (the operation of pregnant women), there will be a mark display
in the bottom area of FHR wave display section.
FHR Alarm:
Alarm for high and low FHR, which exceeds appointed limit.
A 1.2 Transducers
(1) Ultrasonic Transducer
System: Pulsed Doppler
Dimension: 90mm × 65mm
(2) TOCO Transducer
System: Passive Straingauge
Dimension: 102mm × 50mm
(3) Remote Marker
Length: 3.2m
Attachment 2 Troubleshooting
Note: If trouble occurs during operation, examine the product by the following ways. If it
not works, please contact the local distributor or manufacturer; do not open the machine by
the user.
1 The screen not display
Shut off the power; pull out the power cord, to check the electrical current goes through the socket,
and the power cord connects with the equipment properly. To check the fuse, if it is melt down,
change the fuse.
2 Noises
SymptomPossible CauseSolution
NoiseToo high volume setsAdjust the volume down
Interfered by handset or other
interfering source
3 Recorder Errors
SymptomPossible CauseSolution
Paper jamWrong feeding paper or paper is
affected with damp
Recorder does not workPRINT button is disabledPress the PRINT button again
Out of paperFeed paper
Just push print button, the printing of
last line not finished.
4 Ultrasound Monitoring of FHR
Keep the handset or other
interfering source far away
Feed paper correctly and keep
paper from moist
Waiting until it is finished.
SymptomPossible CauseSolution
Inconstant trace
Inconstant display
Doubtful FHRRecord Fetal heart rate wronglyChange the position of
Feint trace or no traceImproper paperUse the paper recommended
Wrong FHRNo
The pregnant woman is too fat No
Improper ultrasound transducer
position
Loose abdomen beltTighten abdomen belt
Superfluous coupling gelWipe off superfluous coupling
Fetal movementWait for a moment then monitor
Maternal movementRelax patient’s spirit
Inadequate coupling gelUse recommended coupling gel
The ultrasound transducer is not
placed well on the abdomen, and
the mixed noise has been
recorded
Change the position of
ultrasound transducer
gel
quantity
ultrasound transducer
Change the position of
ultrasound transducer
by manufacturer
5 Monitoring Contractions (External)
SymptomPossible CauseSolution
Worse trace quality or
fluctuant TOCO baseline
Too high TOCO sensitivity
(higher than 100 unit)
Attachment 3 Monitoring Figure
Too tight or too loose
abdomen belt or no
elasticity
Maternal MovementRelax patient’s spirit
Fetal MovementWait for a moment then monitor
The body pressure from
uterus to TOCO
transducer is far higher
than the average value.
Ensure the abdomen belt has been
used accurately and neither too
tight, nor too loose
Insure favorable contact for patient
skin with TOCO transducer. Change
the position of TOCO transducer, if
necessary.