Contec CMS-08A User Manual

Contec CMS-08A
Blood Pressure Monitor
Electronic Sphygmomanometer
After reading, please validly keeping to refer and consult at any moment.
Medical Test Supply www.medicaltestsupply.com
I
Contents
Chapter1 Safety Precautions: ....................................................................................................... 1
Chapter2 Main Unit ..................................................................................................................... 3
Chapter3 Button Functions .......................................................................................................... 4
Chapter4 External Interfaces ........................................................................................................ 4
Chapter5 Dry Battery/AC Adapter Installation ............................................................................ 5
5.1 Dry Battery Installation ..................................................................................................... 5
5.2 Using The AC Adapter ...................................................................................................... 5
Chapter6 Setting The Date And Time .......................................................................................... 6
Chapter7 Unit ............................................................................................................................... 7
Chapter8 User Switch .................................................................................................................. 7
Chapter9 Applying the Arm Cuff ................................................................................................. 7
Chapter10 BP Measurement ......................................................................................................... 9
10.1 Accurate Measurement Way ............................................................................................ 9
10.2 BP Measurement ............................................................................................................. 9
Chapter11 Memory Function ..................................................................................................... 10
11.1 Review The Memory Value ........................................................................................... 10
11.2 Delete Memory Values .................................................................................................. 11
Chapter12 Alarm Function ......................................................................................................... 12
Chapter13 SpO2 Measurement Function(Separate Sale) ........................................................... 12
Chapter14 Monitoring Procedure ............................................................................................... 14
Chapter15 Maintenance and Cleaning ....................................................................................... 15
Chapter16 Installation Of The Software .................................................................................... 16
16.1 Demand Of Editor ......................................................................................................... 16
16.2 Installation Of Software ................................................................................................ 16
Chapter17 Keys And Symbols ................................................................................................... 17
Chapter18 Error Message ........................................................................................................... 18
Chapter19 TROUBLESHOOTING ........................................................................................... 19
Chapter20 Clean And Maintenance ........................................................................................... 21
Chapter21 NIBP Specification ................................................................................................... 23
Chapter22 SpO2 Specification ................................................................................................... 25
Medical Test Supply www.medicaltestsupply.com
1
Chapter1 Safety Precautions:
Before use, carefully read "Safety Precautions" for a correct use. To prevent users suffered hurt or dandification due to improper use, see "Safety
Precautions", and use this product properly.
For safety reasons, be sure to comply with safety precautions.
Note
If not to use correctly, it exists that a potentially hazardous situation which may result in injury to
the user or patient or damage to the equipment or other property.
Note
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician.
Contact your physician for specific information about your blood pressure.
Please hand measurement results to the doctor who know your health to accept diagnosis.
For severe blood circulation disorder or arrhythmia patients, please use the device under the
guidance of a doctor.
Otherwise it may lead to acute hemorrhage, or measurement error as a result of squeezed arm.
This device is intended for using in measuring blood pressure and pulse rate.
Do not use for any other purpose.
Otherwise it may cause accident or holdback.
Please use special cuff.
Otherwise it is possible that measurement result is incorrect.
Do not disassemble or attempt to repair the unit or components without permission.
Otherwise it cannot measure correctly.
Operation for AC Adapter (Optional from MTS)
Note
Please use optional dedicated AC adapter sold by MTS
Otherwise it may cause trouble.
Optional dedicated AC adapter should be used with a separate power socket.
Otherwise it may cause electric shock or injury.
When there is breakage of the optional dedicated AC adapter plug or wire ,
please immediately pulled the plug from the power socket.
Otherwise it may cause electric shock or injury.
Do not plug or unplug the adapter power cord with wet hands.
Otherwise it may cause electric shock or injury
Medical Test Supply www.medicaltestsupply.com
2
.
Operation for battery
Note
Please use 4 "AA" size manganese or alkaline batteries, do not use batteries of other types.
Otherwise it may cause fire.
New and old batteries, different kinds batteries cannot be confusion.
Otherwise it may cause battery leakage, heat, rupture, and damage to Electronic
Sphygmomanometer.
+ and - polarities of the batteries must match the polarities of the battery compartment as
indicated. When the batteries power exhausts, replace with four new batteries at the same time.
Please take out the batteries when you do not use the device for a long time.
Otherwise it may cause battery leakage, heat, rupture, and damage to Electronic
Sphygmomanometer.
If battery fluid should get in your eyes, immediately rinse with plenty of clean water.
Contact a physician immediately.
Otherwise it will cause blindness or other hazards.
If electrolyte of the batteries immodestly glues on the skin or the clothes,
immediately rinse with plenty of clean water.
Otherwise it may hurt the skin.
Advice
Do not subject the device to strong shocks, such as dropping the unit on the floor.
Do not inflate before the cuff wraps around the arm
Do not inflect the cuff and the air tube forcibly
This device is intended for using in measuring blood pressure and pulse rate in the adult, pediatric and neonatal population.
Warning
Please use the device on the adult object who can read the user manual and the error message shown on the screen. Read the user manual before use the device in order to take actions according to the manual when something wrong with the device. For pediatric and neonatal population, measurement should be only performed by qualified personnel. And please make sure to select the right user mode and cuff before use.
Medical Test Supply www.medicaltestsupply.com
3
Chapter2 Main Unit
MEMORY Button
AC Adapter Jack
The product is in the package. Open the package and confirm whether the product is whole.
Cuff Plug
Cuff Air Plug
Display
UP/DOWN Button
MENU Button
USER SWITCH Button
Accessories
Adult Cuff Dry Battery
(Specification: limb circumference 22-32cm (middle part of upper arm), please choose suited cuff when measuring pediatric or other).
USB Data Line Software CD User Manual
ON/OFF
START/STOP Button
Button
USB Plug Jack
Separate Sale:
AC Adapter InputAC 100-240V 50-60Hz AC 500mA Output DC 6.0V 1.0A
SpO2 Probe Y10UCH150 SpO2 Measurement Range 35%~100 % Measurement Accuracy 70%~100% ±2% Pulse Measurement Range 30~250bpm
Medical Test Supply www.medicaltestsupply.com
4
Chapter3 Button Functions
All the operations to the Electronic Sphygmomanometer are through the buttons. The names of the buttons are above them. They are:
ON/OFF】】】】 Hold the button to start or close the device.
【【
START/STOP】】】】 Press to inflate the cuff and start a blood pressure measurement.
【【
When measuring, press to cancel the measurement and deflate the cuff.
The three buttons correspond with the hint in the LCD screen downside,
pressing any button will carry on corresponding function, eg:【MENU】【ENTER】【LIST】 etc.
Up and down buttons respectively carry on the functions of moving the cursor up and
down, changing the parameters and switching the status.
Chapter4 External Interfaces
Note
Please hold the air plug to remove the NIBP cuff.
① Cuff Socket
left side
The right side of the instrument is USB socket and AC adapter socket.
①USB socket
②AC adapter socket
right side
Medical Test Supply www.medicaltestsupply.com
5
Chapter5 Dry Battery/AC Adapter Installation
The product can use dry battery or AC adapter as power source.
5.1 Dry Battery Installation
Press theindicator on the battery cover and slide the cover off in the direction of the arrow. Install 4 "AA" size dry batteries so the +(positive) and (negative) polarities match the polarities of the battery compartment as indicated. Replace the battery cover.
Icon “ ”: the batteries power will exhaust. Replace with four new batteries (the same sort) at the same time.
Turn the unit off before replacing the batteries.
Note
Dispose of the batteries according to applicable local regulations about environmental.
5.2 Using the Optional AC Adapter
Connect device and the AC Adapter. Insert the AC Adapter Plug into the AC Adapter Jack on the right side of the device. Plug the AC Adapter into a AC outlet.
Note
Hold and pull the Housing to remove the AC Adapter from the electrical outlet. Do not
①①
remove by pulling on the cord.
Remove the AC Adapter plug from the unit.②②②
Please be sure to use dedicated AC adapter.
Note
Medical Test Supply www.medicaltestsupply.com
6
You'd better take off batteries when use the AC adapter as power source. If there is any damage to the AC adapter, you should use batteries to run the device. When adapter and batteries are both used at the same time, the battery power will not be consumed. Switch adapter and battery as power supply when the device is off, otherwise, the device may shutdown due to power failure.
Chapter6 Setting The Date And Time
It is necessary to set date and time after turning on the device. The Electronic Sphygmomanometer automatically stores measurement results with date and
time.
If dry battery power exhausts or removed, then, after the device turned on, the date resumes
from the last setting value and the time from 00:00:00, set the date and time again.
The Electronic Sphygmomanometer stores the measure results of three users automatically, and up to 100 items for every user. The results can be uploaded to PC via USB and processed with the PC software. If the date and time are set correctly, the date and time when measuring will be correct in the memory, otherwise it may not be correct.
Note
Correctly use data upload function:
1. Turn on the device to enter the main interface shown as the follow:
2.Press MENU】 button to enter SYSTEM MENU】 and select【SYSTEM TIME】 item in system menu. The current time will be displayed:
3.Press UP or DOWN buttons to set date and time.
4.After setting, select CONFIRM】 item and press ENTER】 button to confirm the setting value. If you do not want to change the time, select 【EXIT】 item and press ENTER】 button to return the previous menu.
Medical Test Supply www.medicaltestsupply.com
7
Note
Please choose the computer which should be ensured compliance with the requirements of IEC60950, or else it may damage the device.
Chapter7 Unit
There are two units: "mmHg" and "kPa". The default is: "mmHg". To switch "mmHg" and "kPa" units, enter the SYSTEM SETUP】 submenu in SYSTEM MENU, and complete switching in 【UNIT item.
Chapter8 User Switch
The Electronic Sphygmomanometer stores the measure results of three users automatically, and up to 100 items for every user. Press USER】 button in main interface to switch users. Or press USER PURVIEW】 item in SYSTEM SETUP menu to switch users.
Note
When the USER PURVIEW】 is set to be ALL】, current user can be switched under main interface; when set to a certain user, it will not be able to switch.
Chapter9 Applying the Arm Cuff
The measurement can be carried out by applying the cuff on left or right arm.
Remove tight-fitting clothing from your upper arms.
Medical Test Supply www.medicaltestsupply.com