IMPORTANT! Read and save these instructions. This manual to be left with the equipment.
2588787_A_EN_1811_US_OM
OPERATION AND
MAINTENANCE MANUAL
Ultrasonic Humidifier
US-Series
Humidification and Evaporative Cooling
Page 2
Thank you for choosing Condair
Installation date (DD/ MM / YYYY):
Commissioning date (DD / MM / YYYY):
Site:
Model:
Serial number:
Proprietary Notice
This document and the information disclosed herein are proprietary data of Condair Ltd. Neither this
document, nor the information contained herein shall be reproduced, used, or disclosed to others
without the written authorization of Condair Ltd., except to the extent required for installation, operation
or maintenance of the customer's equipment.
Liability Notice
Condair Ltd. does not accept any liability due to incorrect installation, maintenance or operation of the
equipment, or due to the use of parts/components/equipment that are not authorized by Condair Ltd..
Copyright Notice
Copyright 2018, Condair Ltd., All rights reserved.
Technical modification rights reserved.
Page 3
CONTENTS
1Introduction7
1.1Before You Start7
1.2General7
1.2.1Limitations of the Manual7
1.2.2Symbols Used in This Manual7
1.2.3Storage of Manual7
1.2.4Available Manuals and Languages8
1.3Safety Precautions8
1.4Compliance8
2For Your Safety9
3Hygiene12
3.1Important Notes on Hygiene12
3.2Hygiene Risk Management12
3.2.1Guidelines to Keep a Clean System and Prevent Micro-organism Growth13
3.3Risk Monitoring and Assessment13
4Product Description15
4.1General Description15
4.2Functional Description17
5Product Specifications19
5.1Technical Data19
5.1.1Capacity19
5.1.2Water Supply19
5.1.3Fill Rate21
5.1.4Drain Rate21
5.1.5Weights22
5.1.6Dimensions22
5.2Operating Data23
5.2.1Control Signal23
5.2.2Ambient Conditions23
6User Interface24
6.1User Interface Controls24
6.2Control Software24
6.2.1Home Screen24
6.2.2Navigation Elements25
6.2.3Operating and Fault Status26
6.2.4Maintenance and Service Status27
6.2.5Main Menu27
6.2.6Configuration Menu28
6.2.6.1 Features Menu28
6.2.6.2 Control Settings29
6.2.6.3 General31
6.2.7Service32
6.2.8Changing Your Settings33
US
2588787_A_EN_1811_US_OM
Contents|3
Page 4
6.2.8.1 Change the period between flushing cycles.33
6.2.8.2 Activate the Ultrasonic Blower Pack.33
6.2.8.3 Change how the large humidifier reacts to demand.33
6.2.8.4 Use the humidifier at X% capacity. To set up and configure the manual capacity control type.33
6.2.8.5 Humidify the space at X% setpoint, based on a control signal.33
6.2.8.6 Put the humidifier on standby.33
6.2.8.7 Change the fan speed of the nebulizing module.34
6.2.8.8 Change the LCD display settings.34
6.2.8.9 Change the system language.34
6.2.8.10 Restore factory defaults.34
6.2.8.11 Reset the service reminder (remove the service warning).34
7Operation35
7.1General35
7.2Inspecting the Unit36
7.3Starting Up36
7.3.1Restarting the Unit36
7.4Shutting Down the Unit37
7.5Remote Monitoring37
7.6Draining the Unit37
8Maintenance38
8.1General38
8.2Cleaning and Disinfecting38
8.3Maintenance List39
8.4Maintenance Procedures42
8.4.1Removing and Installing the Door Panels43
8.4.2Cleaning, Removing, and Installing the Water Reservoir44
DAppendix:WiringDiagram Ultrasonic Humidifier - Large (A-unit)68
EAppendix:WiringDiagram Ultrasonic Humidifier - Large (B-unit)70
US
2588787_A_EN_1811_US_OM
Contents|5
Page 6
This page intentionally left blank.
Page 7
1Introduction
1.1Before You Start
Thank you for purchasing the Ultrasonic Humidifier.
The Ultrasonic Humidifier incorporates the latest technical advances and meets recognized safety
standards. Improper use of the Ultrasonic Humidifier may result in danger to the user or third parties,
and/or damage to property.
To ensure safe, proper, and efficient operation of the Ultrasonic Humidifier, please observe and comply
with all information and safety instructions contained in this manual, as well as all relevant documentation of components of the installed humidification system.
If you have additional questions, please contact your Condair representative. They will be glad to assist
you.
1.2General
1.2.1Limitations of the Manual
The subject of this manual is the Ultrasonic Humidifier in its different versions. The various options and
accessories that are described in this manual pertain to the proper operation of the equipment. Additional information and instructions are supplied with the options and accessories.
This manual is restricted to the operation and maintenance of the Ultrasonic Humidifier. This manual is
intended for well trained personnel who are suitably qualified for their respective tasks.
1.2.2Symbols Used in This Manual
CAUTION!
The catchword "CAUTION" in conjunction with the general caution symbol is used to provide safety
instructions that, if neglected, may cause damage and/or malfunction of the unit or damage to property.
WARNING!
The catchword "WARNING" in conjunction with the general warning symbol is used to provide safety
instructions that, if neglected, may cause injury to personnel. Other specific warning symbols may also
be used in place of the general symbol.
Danger!
The catchword "DANGER" in conjunction with the general danger symbol is used to provide safety
instructions that, if neglected, may cause severe injury to personnel or even death. Other specific
danger symbols may also be used in place of the general symbol.
1.2.3Storage of Manual
Keep this manual in a place where it is safe and readily accessible. If the equipment is moved to
another location, please ensure that it remains with the equipment.
If the manual is lost or misplaced, please contact your Condair representative for a replacement.
US
2588787_A_EN_1811_US_OM
Introduction|7
Page 8
1.2.4Available Manuals and Languages
Table 1: Document (Part) Numbers and Languages of Available Manuals
Installatio n
Manual
Operation and
Maintenan ce
Manual
Installatio n,
Operation, and
Maintenan ce
Manual
Spare Parts
List
Engineering
Manual
Ultrasonic Humidifier US
Ultrasonic
Blower PackUB
Reverse
Osmosis System
RO-U
2588786_A_EN_
1811_US_IM
1.3Safety Precautions
The Ultrasonic Humidifier and Ultrasonic Blower Pack is NOT SUITABLE for humidification in healthcare premises. Due to the nature of ultrasonic humidification, it is NOT recommended to use these units
in environments where the occupants or materials in the space are adversely sensitive to airborne bacteria, viruses, dust, and debris. Special care must be observed to properly install, maintain, clean, and
disinfect the Ultrasonic Humidifier and Ultrasonic Blower Pack to preserve acceptable levels of hygiene.
Regard all safety precautions described in this manual.
For more information, refer to "For Your Safety" on page9.
1.4Compliance
Follow all local guidelines and regulations regarding the installation and operation of the unit; this
includes (and not limited to) mounting, plumbing, water, gas, ventilation, and electrical connections.
Please consult your local Condair representative to learn more.
2588787_A_EN_
1811_US_OM
——
——
—
2588705_A_EN_
1811_UB_IOM
2594385_A_EN_
1811_RO_IOM
2588788_A_EN_
1811_US_SPL
——
——
—
8|Introduction
2588787_A_EN_1811_US_OM
US
Page 9
2For Your Safety
General
Every person who is tasked with the installation, operation and/or maintenance of the Ultrasonic Humidifier must read and understand this manual, as well as any respective installation, operation and/or
maintenance manuals before performing any work. Knowing and understanding the contents of these
manuals is a basic requirement for protecting personnel against any kind of danger, preventing faulty
operation, and operating the unit safely and correctly.
All labels, signs and markings applied to the Ultrasonic Humidifier must be observed and kept in a readable state.
Personnel Qualifications
All procedures described in this document must only be performed by personnel who are adequately
qualified, familiar with the humidifier unit, well trained, and are authorized by the customer.
For safety and warranty reasons, any activity beyond the scope of this manual must only be performed
by qualified personnel authorized by Condair.
All personnel working with the Ultrasonic Humidifier unit must be familiar with, and comply with the
appropriate regulations on workplace safety and prevention of accidents.
Intended Use
The Ultrasonic Humidifier is intended exclusively for air humidification using a Condair-approved Ultrasonic Blower Pack within specified operating conditions. Refer to "Technical Data" on page19 for
details. Any other type of application, without the express written consent of Condair, is considered to
be NOT conforming to its intended purpose, and may lead to dangerous operation and will void the warranty.
In order to operate the equipment in the intended manner all information contained in this manual, in
particular the safety instructions, must be observed closely.
Not Intended Use
l NOT recommended to for humidification in healthcare premises.
l NOT recommended to use these units in environments where the occupants or materials in the
space are adversely sensitive to airborne bacteria, viruses, dust, and debris.
Special care must be observed to properly install, maintain, clean, and disinfect the Ultrasonic Humidifier and Ultrasonic Blower Pack to preserve acceptable levels of hygiene.
Safety Precautions that Must be Observed
DANGER!
Risk of electric shock!
The unit is mains powered. Using the On/Off switch on the humidifier to shut down the unit will
NOT shut off power to the control cabinet. Live parts may be exposed when the access panels or
doors are removed. Touching live parts may cause severe injury or even death.
Prevention: Before performing any work inside the humidifier, shut down the humidifier properly, as
described in "Shutting Down the Unit" on page37. The unit must be connected to the mains only after
all mounting, installation, and maintenance work has been completed, checked for correct workmanship, and the access panels/doors are closed and fastened securely. Secure the unit against accidental power-up.
Only qualified personnel should perform maintenance or repairs on the unit.
US
2588787_A_EN_1811_US_OM
For Your Safety|9
Page 10
CAUTION!
Electrostatic discharge (ESD)!
The electronic components inside the control cabinet in the humidifier are sensitive to electrostatic discharge (ESD).
Prevention: Take appropriate measures to protect the electronic components inside the unit against
damage caused by electrostatic discharge (ESD). Refer to ANSI/ESD-S20.20.
WARNING!
Heavy object. Risk of injury!
The humidifier is heavy. Injury, damage, or even death may occur if the humidifier falls or tips
over.
Prevention: Ensure that the humidifier is secured properly. Use an appropriate lifting device, or
request assistance to raise the humidifier into position.
WARNING!
Risk to Hygiene and Health!
Improper and infrequent maintenance of the unit presents the risk of spreading pathogens (such
as bacteria , viruses, and parasites) originating from contaminated water, air, and equipment. Failure to reduce the risks increases the likelihood of causing injuries (requiring medical care), infection, and death.
Prevention: Use, regularly clean, and replace the air and water inlet filters. Regularly flush and dis-
infect the humidifier, the blower pack, and their components (tanks, hoses, reservoirs, etc.). Use potable, treated water (potable, R.O., or D.I.).
Follow regular maintenance routines as described in "Maintenance List" on page39 Follow local regulations.
WARNING!
Do NOT leave the unit without power. Risk of contamination.
Do NOT leave the unit unpowered for more than 48 hours, or else a total system disinfect will be
required. Without power, the system will not flush properly, resulting in standing water and conditions that may promote the growth of micro-organisms.
Prevention: Leave your unit powered on. If you do not need humidification during the 48 hours, set
the humidifier to standby or set the demand to 0.
Preventing Unsafe Operation
All personnel working with the Ultrasonic Humidifier must immediately report to the customer any alterations or changes to the unit that may affect safety, and secure the unit against accidental power-up.
If it is suspected that the unit cannot be operated safely for any of the reasons listed below, shut it down
immediately.
Alterations and changes that may affect safe operation include:
l damage to the unit
l loose or damaged electrical connections
l improper operation
l leaks in the distribution system
10|For Your Safety
2588787_A_EN_1811_US_OM
US
Page 11
Modifications to the Unit Prohibited
Modifications are NOT permitted on the Ultrasonic Humidifier without the express written consent of
Condair.
Safety Labels
The safety label (hygiene) is found on the right wall inside the plumbing cabinet (on the panel separating the electrical cabinet).
US
2588787_A_EN_1811_US_OM
For Your Safety|11
Page 12
3Hygiene
3.1Important Notes on Hygiene
Observe the local health and safety executive's technical guidance on the control of Legionella in water
systems.
The user is responsible to ensure that the water system complies with local regulations, bylaws, and
guidelines (such as the HSE ACoP L8, VDI 6022, ISO 22000, HACCP, or equivalent). If inadequately
maintained, water systems, of which any humidifier is a part, can support the growth of microorganisms,
including the bacterium responsible for Legionnaires' disease.
Any risks or hazards relating to the system, including during installation and maintenance, should be
identified by a competent health and safety representative who is responsible for introducing effective
control measures.
3.2Hygiene Risk Management
Cold water humidification presents the risk of spreading pathogenic microorganisms into the air. The
Ultrasonic Humidifier and Ultrasonic Blower Pack distribute tiny water droplets (mist) that are blown into
the space before being absorbed into the air. If the mist was contaminated before leaving the unit and
then evaporates, these contaminants are deposited onto surfaces in the space, or are otherwise
inhaled or redistributed into the ventilation system.
Pathogens include, and are not limited to, Legionella, E. coli, and viruses, and may originate from contaminated water, air, and equipment.
Regular cleaning and disinfecting is required to reduce health risks. Refer to "Hygiene Risk Man-
agement" above for more information on managing hygiene hazards.
Follow all safety precautions and local regulations. Refer to "For Your Safety" on page9.
There are multiple points in the humidification system where a contamination can occur. It is the
responsibility of the customer and of Condair to ensure that health hazards due to contamination are
reduced to acceptable levels. The use of a hygiene risk assessor is recommended.
Regularly clean and maintain the humidifier and its components, and do NOT miss or ignore the warnings in this manual and service messages displayed on the humidifier.
Hygiene|12
Supply Air and Water
Pathogens can originate from the water supply, such as a tank or water line. It is important that standing
water in the tank or in the water line is flushed frequently to prevent the buildup of contaminants, and
the water is filtered and treated. Refer to "Water Supply Requirements" for source water recommendations.
Pathogens may also enter the humidification system through the ventilation system or from within the
humidifier space. A fan blows air out of the water reservoir into flexible tubes, and further into a blower
pack and into the atmosphere. Contaminated air can join the mist, so it is recommended to use air filters
in the ventilation system, in the humidifier, and in the Ultrasonic Blower Pack. Ensure to use the supplied air filters at all times and replace them regularly (as recommended in the Ultrasonic Humidifier
Maintenance and Operation Manual).
Inside the Humidifier
A small risk of contamination may occur during manufacture, testing, delivery, and commissioning. Any
water that was present during these processes is considered stagnant. Ensure that all applicable components (i.e. wet surfaces, such as flexible tubes, reservoirs, and hoses) of the humidification system
has been cleaned and disinfected before installation. Refer to the maintenance routines as described in
"Maintenance List" on page39., and "Maintenance Procedures" on page42.
2588787_A_EN_1811_US_OM
US
Page 13
The humidifier possess a UV assembly that uses ultraviolet radiation to kill microorganisms that enter
the unit from the water supply. The UV assembly is powered by the humidifier and must be inspected
and replaced regularly.
To prevent standing water inside the unit, the humidifier software activates a flushing cycle that empties
the water from hoses and reservoirs into the drain. The humidifier is replenished with new water.
During normal operation and use, standing water may be present in cracks, scratches, and joints within
the reservoir, flexible tubes, mist channels, and hoses. To prevent stagnant water from accumulating,
flexible tubes and channels, water reservoirs, and hoses must be disassembled regularly and cleaned.
Reduce the frequency of microscopic scratches and cracks by using nonabrasive cleaning materialsand agents when cleaning the humidifier (i.e. do NOT use a hard or metallic brush). Refer to "Cleaning
and Disinfecting" on page38. Follow regular maintenance routines as described in "Maintenance List"
on page39.
Note: Stagnant water may also be formed after shutting down the humidifier (and after inactivity).
Ensure that the humidifier is air dried for inactivity, and cleaned before restarting.
3.2.1Guidelines to Keep a Clean System and Prevent Micro-organism Growth
l Perform a risk assessment of the water system using a competent person, and implement an
appropriate monitoring and control program.
l Connect a RO system to a clean, potable main water supply.
l Ensure that the system is cleaned and/or flushed regularly.
l Ensure that the buffer tank is clean and sterile before use.
l Enter a service contract that suits the needs of the company.
l Stop the system if polluted drinking water is found in the area.
l Avoid water temperatures above 20 °C (68 °F). Temperatures between 20-60 °C (68-140 °F)
favor the growth of Legionella.
l Do not stop the system unless it is faulty or leaking (to avoid water stagnation).
l If the power to the system is interrupted (such that the unit fails to flush or drain) for more than 48
hours, complete a total system disinfection.
l Refrain from closing sections of the system, unless there is a leakage or fault (to avoid water
stagnation).
l Disinfect at least once per year, and after every maintenance or repair.
l Analyze the water and test for harmful bacteria at least once per year.
l Conduct follow-up measurements until the system is clean if bacteria has been detected in the
system.
The Condair service team can help with expert technicians who can provide:
l Bacteriological troubleshooting on-site*.
l Cleaning and disinfecting.
l Preventative maintenance.
l Repair and fault finding.
l Training and guidance.
Note: * Condair uses a quick method for measuring bacterial activity in the water with the use of
BactiQuant (BQ). Once the water sample has been taken, the bacteriological quality of the water is read
within 30 minutes, and the Condair service team will disinfect the system if necessary.
Please contact your local Condair representative for further information about our services.
3.3Risk Monitoring and Assessment
Water hygiene assessments must be performed at the site to determine the risk and severity of hygiene
hazards that affect the health and safety of the public. Performing a risk assessment with regular
US
2588787_A_EN_1811_US_OM
Hygiene|13
Page 14
monitoring will allow procedures to be adjusted.
Ensure that the person conducting the assessment is qualified.
A typical assessment inspects the following:
l Supply water quality. Refer to "Water Supply" on page19.
l Environment around the humidifier.
l Maintenance procedures. Refer to "Maintenance Procedures" on page42 .
l Water pre-treatment options (such as using UV light, using reverse osmosis or deionized water,
etc.).
Proper maintenance procedures along with recommended site controls will help ensure compliance to
local regulations and protect the public.
It is recommended to monitor the humidification system weekly, and then monthly once satisfactory
levels are maintained.
Ensure that the results of every assessment is recorded.
IMPORTANT! Failure to make adequate provision against the outbreak of disease, such as Legionnaires' disease, may cause fatalities, illness, and liabilities.
14|Hygiene
2588787_A_EN_1811_US_OM
US
Page 15
4Product Description
4.1General Description
The Ultrasonic Humidifier is a steam-less humidifier. To humidify the space, the Ultrasonic Humidifier
creates mist using piezoelectric transducers. Mist is expelled from the unit with internal fans, and an
Ultrasonic Blower Pack (if installed), into the space. Mist is then absorbed into the air.
The Ultrasonic Blower Pack comes in 4 different output capacities; 3 kg/h (6.6 lb/h); 6 kg/h (13.2 lb/h); 9
kg/h (19.8 lb/h); and 18 kg/h (39.6 lb/h).
There are two housing sizes; medium and large. The medium and large units can supply mist to other
rooms or directly into the space using any installed Ultrasonic Blower Packs. The large units require two
Ultrasonic Blower Packs. Refer to "Technical Data" on page19 and "Operating Data" on page23 for
more information about capacity, power and water consumption.
US
2588787_A_EN_1811_US_OM
Product Description|15
Page 16
1
1
1
1
1
2
1
2
4
5
3
6
7
16|Product Description
2588787_A_EN_1811_US_OM
US
Page 17
Figure 1 - Typical Installations*
1. Installation with the Ultrasonic Blower Pack
2. Flexible tubes
3. Electrical disconnect, high voltage supply
4. Drain line to air gap funnel
5. Trap
6. Expansion tank
7. R.O. system
Note: * Water, electrical, and drain connections not shown for the large model.
4.2Functional Description
The Ultrasonic Humidifier is an adiabatic unit that uses transducers in the humidifier's water reservoir to
create a fine mist using high-frequency vibrations (1.7 MHz).
Treated water enters the humidifier's reservoir from an inlet located at the bottom of the unit. The water
that enters the humidifier's reservoir must be treated to remove minerals, bacteria and fungi, and should
otherwise be pure to prolong the service cycle of the unit and, above all, safeguard the public. An accumulation of minerals will result in scale, and an accumulation of bacteria and fungi will result in
unhygienic conditions in the space. Refer to "Hygiene" on page12.
Transducers in the reservoir receive a signal for humidification and will activate. The transducers will
vibrate in the water to create a mist of fine water particles. The mist rate is controlled through the humidifier's control board, and may be modulated on demand or on humidity signal.
The mist is blown from the reservoir into flexible tubes that exit from the top of the humidifier. The mist is
channeled directly into the Ultrasonic Blower Pack. The water particles in the mist are then absorbed
into the air and increase the relative humidity of the space.
Condensate formed in the flexible tubes between the blower pack and the humidifier re-enters the reservoir where it can be reused.
Water is drained from the reservoir through an outlet located at the bottom of the unit.
Water Supply
Water enters the humidifier through the fill valve into the fill cup with integral air gap, and then flows
down through the UV light into the water reservoir
Water Level Monitoring
The water level in the reservoir is continuously monitored with a level sensor. If the water level drops
below the sensor switching point, from normal operation, the sensor supplies a signal to the controller.
This opens the inlet valve and the water reservoir is filled. When the water level reaches the operating
level the sensor supplies a signal to the controller to close the inlet valve.
UV Pre-Treatment
Water enters the UV reactor, from the inlet, and is disinfected with UV radiation in the UV reactor in the
UV assembly. The water exits the UV assembly into the fill cup which monitors the water level.
Mist Generation
When the humidifier receives a signal to supply mist, the unit fan and the piezoelectric transducers start
operating. The fine mist is blown out of the humidifier outlets by the unit fan.
US
2588787_A_EN_1811_US_OM
Product Description|17
Page 18
Air Supply
Air supplied to the humidifier comes from the space surrounding the humidifier and through the humidifier. The air should be filtered and clean to prevent contaminating the mist with microorganisms and
dust.
The use of the supplied air filters (recommended) is advised for the Ultrasonic Humidifier and Ultrasonic
Blower Pack.
Flushing
To prevent the growth of bacteria in the unit, a flush cycle is implemented to replace the content of the
water reservoir with fresh water. The flushing is carried out at an adjustable time interval.
The production of mist is interrupted during the flushing process. The drain valve will open and the
reservoir will be completely discharged though the water discharge (drain) connection. The inlet valve
will open and flush out any stagnant water in the water supply line. The drain valve is closed and the
inlet valve is closed after the water level inside the reservoir reaches the normal operation level. After a
flush cycle the unit resumes normal operation.
Control
Mist production can be controlled using the continuous proportional (P) or the proportional-integral (PI)
controller internal to the humidifier, or an external modulation controller, or with an On/Off control signal
from an external humidistat.
18|Product Description
2588787_A_EN_1811_US_OM
US
Page 19
5Product Specifications
1
2
D.I. (optional - not supplied)
R.O.
5.1Technical Data
5.1.1Capacity
Table 2: Ultrasonic Humidifier Capacity
ModelHousing Size
US 7Medium3 kg/h (6.6 lb/h)250 W
US 136 kg/h (13.2 lb/h)250 W
US 209 kg/h (19.8 lb/h)650 W
US 40*Large18 kg/h (39.6 lb/h)1250 W
Note: *Equipped with dual stainless steel reservoirs – each with its own dedicated driver board.
5.1.2Water Supply
Capacity
at 120V/208/240V/1
Mist ProductionPower Rating
Figure 2 - Water Supply Requirements
1. Water from the source entering the RO-U system
2. Water entering the US
US
2588787_A_EN_1811_US_OM
Product Specifications|19
Page 20
1
Mains Water Supply to RO-U
ConditionsValue
QualityPotable, drinkable
Pressure2.0-5.0 bar (30-75 psi)
Temper ature *10-20°C ( 50-68°F)
Turbidity(max)1.0 NTU
Max pH without Softener*8.0
pH
pH 6.5-7.0pH 7.1-7.5pH 7.6-8.0
Max Hardness *
Max US Hardness *
Max Conductivity (at 20 °C / 68°F)*,**
Max Total Dissolved Solids (TDS) *
SiltIndex (max)3
KMnO4Potassium permanganate
(max)
Fe Iron (max)0.2 mg/L
Mn Manganese (max)0.05 mg/L
Ca Calcium (max)300 mg/L
Cu Copper (max)1 mg/L
SiO2Silica (max)1 mg/L
Ammonium (max)0.5 mg/L
30 °dH17 °dH10 °dH
31 gpg18 gpg10 gpg
1300 µS/cm750 µS/cm600 µS/cm
800 mg/l500 mg/l375 mg/l
pH of water supply
10 mg/L
Chloride (max)300 mg/L
Sulphates (max)250 mg/L
Colony-forming Unit (CFU) (max)200 CFU/ml ; no pathogens allowed
* If the water entering the RO-U system isgreater than the values indicated in this table, install a water softener
between the water supply and the Reverse Osmosis System.
** Contact your Condair representative ifthe conductivity of the supplywater is greater than 350 µS/cm.
20|Product Specifications
2588787_A_EN_1811_US_OM
US
Page 21
2
Water Supply Requirements to the US (after treatment with RO and/or DI)
ConditionsValue
Pressure2-5.5 bar (30-80 psi)
spike pressure limited to 8 bar (120psi)
Temper ature1-25 °C (34-77 °F)
Min-Max Conductivity * , Max Total
Dissolved Solids (TDS)
QualityRO or DI water
Colony-forming Unit (CFU)< 200 CFU/ml for industrial applications**
* if the conductivityof the water entering the RO-U system isgreater than 350 µS/cm, contact your Condair representative .
** CFU leaving the unit via mist should be <1000 CFU/ml for industrialapplications. Keep the CF U entering the humidifier as low as possible, such that the CFU leaving the unit iswithin acceptable guidelines.
5.1.3Fill Rate
Fill Rate
US 7-20Minimum: 3.3 L/min (0.9 gpm)
US 40*Minimum: 6.6 L/min (1.8 gpm)
2-15 µS/cm , TDS: 9 mg/L
refer to the water supply requirements for the RO-U
Refer to Table 8 below for the maximum external fusing (single-phase) for the Ultrasonic
Humidifier.Refer to 5.2 above for the maximum external fusing (single-phase) for the Ultrasonic Blower
Pack.
Table 8: Ultrasonic Humidifier, Maximum External Fusing — Single-Phase
mADC
15-24 VDC On
Mod-el110-120V/1~208V/1~220-240V/1~
Maximum
Output
US 7 744 W6.2 A15 A744 W3.6 A15 A744 W3.1 A15 A
US13744 W6.2 A15 A744 W3.6 A15 A744 W3.1 A15 A
US20744 W6.2 A15 A744 W3.6 A15 A744 W3.1 A15 A
US401440 W12.0 A15 A1440 W7.0 A15 A1440 W6.0 A15 A
Maximum
Current
Maximum
External
Disconnect
Maximum
Output
Maximum
Current
Maximum
External
Disconnect
Maximum
Output
Maximum
Current
Maximum
External
Disconnect
US
2588787_A_EN_1811_US_OM
Product Specifications|23
Page 24
6User Interface
1
2
3
6.1User Interface Controls
The user interface of the Ultrasonic Humidifier is located at the front of the unit. It consists of a display
screen with four buttons and an On/Off switch. Refer to "Navigation Elements" on the next page.
Ensure that all safety precautions are observed. Refer to "For Your Safety" on page9.
Figure 3 - Ultrasonic Blower Pack Location of Operator Interface
1. Display
2. Navigation buttons (up, down, enter, menu)
3. On/Off button
6.2Control Software
The control software runs in the background and controls all the functions of the Ultrasonic Humidifier.
The user interacts with the control software through the screens on the display.
Note: The screens shown in this manual are for representation purposes only, and may differ from
what is actually displayed on the humidifier.
6.2.1Home Screen
When the Ultrasonic Humidifier is turned on, it will perform system checks to ensure proper functioning
of the unit. When initialization and system checks are complete, the unit will enter normal operating
mode. The home screen will appear on the display.
The home screen displays the following information:
l Control signal type
l Demand requested
l Operating/Service/Warning/Fault status message
l Graphic indicator of the mist output
Note: Status information for dual reservoirs are displayed simultaneously on large models, US40.
User Interface|24
2588787_A_EN_1811_US_OM
US
Page 25
1
2
3
Figure 4 - Ultrasonic Humidifier Home Screen - Single Reservoir
1
2
4
5
3
1. Humidity control information - shows the control signal type and requested demand
2. Operating/Service/Warning/Fault status message
3. Visual indication of the current mist output level
Figure 5 - Ultrasonic Humidifier Home Screen - Dual Reservoir
1. Humidity control information - shows the control signal type and requested demand
2. Humidifier A visual indication of the current mist output level
3. Humidifier A Operating/Service/Warning/Fault status message
4. Humidifier B Operating/Service/Warning/Fault status message
5. Humidifier B Visual indication of the current mist out put level
6.2.2Navigation Elements
View humidifier settings by pressing the menu button.
Navigate through the menus using the up and down buttons, and make selections by pressing enter.
Note: Settings changed are only saved after pressing enter. If a timeout occurs (when no user input is
detected), or if the menu button is pressed before pressing enter, the setting will not save.
Refer to Figure 6 - on the next page.
US
2588787_A_EN_1811_US_OM
User Interface|25
Page 26
1
234
Figure 6 - Navigation Elements
1. Menu button - press this button to display and hide the menu
2. Up button
3. Down button
4. Enter button - press this button to make a selection in a menu or save a setting
6.2.3Operating and Fault Status
The operating and fault status is shown at the bottom of the home screen (refer to "Home Screen" on
page24). Refer to Table 9 on the facing page to see the descriptions of each operating and fault status.
26|User Interface
2588787_A_EN_1811_US_OM
US
Page 27
Table 9: Operating and Fault Status Descriptions
MessageDescription
StartingThe humidifier isstarting up.
IdleThe humidifier isidle and waiting for demand.
Idle DrainThe humidifier isidle and the reservoir isempty.
FillingT he reservoir is filling with water.
MistingThe humidifier is producing mist.
FlushingThe humidifier isflushing the water reservoir.
Ext. DisabledT he humidifier has been stopped via an external enable-contact (remote enable/disable).
Safety Loop Open The safety loop isinterrupted.
Blower PackFault A fault has occurred in one of the blower packfans. The humidifier stops producing mist.
Service StopThe humidifier has stopped operation and needs service.
Filling TimeoutThe humidifier has exceeded the maximum allowed filling time. It will continue filling after a wait
period.
Drain FaultThe humidifier detected a problem with the draining of the reservoir and stopsoperation.
UV FaultThe humidifier detected a fault with the UV assembly. The humidifier stops operation.
High Temperature
Fan FailureThe interval fan hasa malfunction. The humidifier stops operation.
The humidifier has detected a high temper ature in the electronics.The humidifier stops operation.
6.2.4Maintenance and Service Status
Service reminders are displayed on the home screen. Generic service reminders are described in .
Table 10: Service Status Descriptions
Message Description
Service
Warning
Service
Required
UV Service
Required
This warning message appears if the control software has detected that the humidifier has been running
for more than 2000 hours. The humidifier willcontinue to operate for another 500 hours before shutting
down for service. The message will then change to "Service Required".
This message appears when the control softwar e has detected that the humidifier has been running for
more than 2500 hours without service. The humidifier stops oper ation.
This message appears when the UV assemblyneeds to be serviced.
6.2.5Main Menu
Access the Main Menu by pressing the menu button on the humidifier. Refer to "Navigation Elements"
on page25.
US
2588787_A_EN_1811_US_OM
User Interface|27
Page 28
Figure 7 - Main Menu Screen
The following menus may be accessed from the Main Menu:
l Configuration
Features Menu - Set the flush cycle for each reservoir and activate the Ultrasonic Blower
Pack. Refer to "Features Menu" below.
Control Settings - Change settings such as, fan speed, manual capacity, set-point, control
mode, signal type, bandwidth, and integral time. Refer to "Control Settings" on the facing
page.
General - Change the language settings, LCD settings, restore the settings to the factory
default, and view the software version. Refer to "General" on page31.
l Service - View and reset general service and UV service times.
6.2.6Configuration Menu
Configure the operation of the Ultrasonic Humidifier through the configuration menu.
28|User Interface
Figure 8 - Configuration Menu
6.2.6.1 Features Menu
Configure the flush cycle of the Ultrasonic Humidifier and connect any installed Ultrasonic Blower Pack
units.
2588787_A_EN_1811_US_OM
US
Page 29
Figure 9 - Features Menu
Flush Cycle A/B
Set the desired interval time (hours) to flush
the reservoir. For humidifiers with dual-reservoirs, this setting will allow each reservoir to
cycle independently of each other.
Range: 1-24 hours
Factory Setting: 1 hour
6.2.6.2 Control Settings
Figure 10 - Control Settings Menu.
Blower Pack A/B
Activates the Ultrasonic Blower Pack attached
to the humidifier. Each Ultrasonic Blower Pack
should connected to its own reservoir. Selecting
Installed will allow the Ultrasonic Blower Pack
to start up.
Options: Installed or Not Installed
Factory Setting: Not Installed
US
2588787_A_EN_1811_US_OM
User Interface|29
Page 30
Dual Humidifier Mode
Change this setting to adjust how the dual
nebulizing modules (water reservoir assembly
with transducers, drivers, and fans) react to
humidifier demand. This setting only appears
on units possessing dual nebulizing modules.
Options:
l Parallel - The modules operate in par-
allel. This setting lowers the turn-on
ratio and creates even wear on both
modules.
l Series - The modules operate in
sequence to meet humidifier demand.
Module A operates in a range
between 0-50 % demand. Module B
operates in a range between 51-100%
demand.
l Independent - The modules operate
independently to meet the demand
and are based on control signals from
two separate zones.
Factory Setting: Series
Control Configuration
Select a control signal channel for the humidifier.
Changing the Control Configuration to Single or
Dual Channel will change the Control Settings
menu to display and allow access to Channel
settings.
Changing the Control Configuration to Manual
Capacity will change the Control Settings menu
to display and allow access to the Manual Capacity menu.
Options:
l Single Channel - Single control
l Dual Channel - Two control signals (typ-
ically one control signal and one highlimit signal)
l Manual Capacity - The demand of the
humidifier is set manually
Factory Setting: Manual Capacity
Manual Capacity
Set the demand capacity for the humidifier.
Range: 0-100 %
Factory Setting: 0 %
Fan Speed
Set the blower fan speed inside the unit. Reducing this value will decrease the mist production
capacity. This setting does not affect the Ultrasonic Blower Pack fan speed.
Range: 0-100 %
Factory Setting: 100 %
Control Channel 1/2
Configure options for each control channel. Refer to Table 11 on the facing page.
30|User Interface
2588787_A_EN_1811_US_OM
US
Page 31
Table 11: Control Channel Options
SettingDescriptionOptions/RangeFactory
Setting
SetpointSet the humidifier to operate at a fixed humidity
setpoint value.
Con trol
Mode
Signal
Type
Band width Adjust the proportional range of the internal P/PI
I TimeSet the integral time for error corrections by the
Select the general control mode for the incoming
control signal.
Select the signal type of the incoming control signal.
controller as a percentage of relative humidity
that the humidifier willmodulate between 0-100
% output.
internal PI controller.
0-100 %50 %
Demand - Control signal isfr om an
external controller
RH P - Control signal usesinternal
proportional controller to interpret
sensed humidityinto demand
RH PI - Control signaluses the
internal proportional-integral controller to interpret sensed humidity
into demand
On/Off - Contr ol signal is from an
external On/Offhumidistat
Set the length of time the LCD screen remains
brightened after inactivity from the user.
Range: 0-60 seconds
Factory Setting: 30 seconds
Software Version
View the software version of the controller and installed driver boards.
6.2.7Service
View and reset the status reminders.
Figure 12 - Service Menu
Factory Defaults
Restore the humidifier settings to the factory settings.
General Service Info
View the current run time hours and remaining hours until the unit must be serviced.
UV A/B Service Info
View the current run time and remaining
hours of the UV assembly until the assembly
must be serviced.
General Service Reset
Reset the general service reminder timer. Perform this task after servicing the unit.
UV A/B Service Reset
Reset the UV service reminder after a UV service has been performed. Separate UV Service
Reset options are displayed for units with dual
reservoirs.
32|User Interface
2588787_A_EN_1811_US_OM
US
Page 33
6.2.8Changing Your Settings
This section shows a brief overview of available settings and where to go to change them.
Do not forget to press enter to save your change.
I want to...
6.2.8.1 Change the period between flushing cycles.
(refer to "Flush Cycle A/B" on page29)
Main Menu > Configuration > Features > Flush Cycle A/B
6.2.8.2 Activate the Ultrasonic Blower Pack.
(refer to "Blower Pack A/B" on page29)
Main Menu > Configuration > Features > Blower Pack A/B
6.2.8.3 Change how the large humidifier reacts to demand.
In parallel, series, or indepen dently (refer to "Dual Humidifier Mode" on page30).
Main Menu > Configuration > Control Settings > Dual Humidifier Mode
Note: For dual reservoir units only.
6.2.8.4 Use the humidifier at X% capacity. To set up and configure the
manual capacity control type.
(refer to "Control Configuration" on page30 and "Manual Capacity" on page30)
1) Go to Main Menu > Configuration > Control Settings > Humidifier A/B* >
Control Configuration > Manual Capacity
2) Return to previous menu then, Control Settings > Humidifier A/B* > Manual
Capacity
Note: *For dual reservoir unitsonly. Single units: Configuration > Control Settings > Control Configuration.
US
6.2.8.5 Humidify the space at X% setpoint, based on a control signal.
To set up and conf igure the humidity co nt rol (refer to "Control Channel 1/2" on page30).
1) Go to Main Menu > Configuration > Control Settings > Humidifier A/B* >
Control Configuration > Single Channel (or) Dual Channel
2) Return to previous menu then select a channel, Control Settings > Humidifier
A/B* > Channel 1 (or) Channel 2
3) Input the Setpoint Control Mode, Signal Type, Bandwidth, and I-Time.
Note: *For dual reservoir unitsonly. Single units: Configuration > Control Settings > Control Configuration.
6.2.8.6 Put the humidifier on standby.
(refer to "Control Settings" on page29)
Set the demand to 0%.
2588787_A_EN_1811_US_OM
User Interface|33
Page 34
6.2.8.7 Change the fan speed of the nebulizing module.
(refer to "Fan Speed" on page30).
Main Menu > Configuration > Control Settings > Fan Speed
6.2.8.8 Change the LCD display settings.
(refer to "LCD Contrast" on page31and "LCD Backlight" on page32).
Main Menu > Configuration > General
6.2.8.9 Change the system language.
(refer to "Language" on page31).
Main Menu > Configuration > General
6.2.8.10 Restore factory defaults.
(refer to "Factory Defaults" on page32).
Main Menu > Configuration > General
6.2.8.11 Reset the service reminder (remove the service warning).
Perform this on ly after servicin g th e humidifier (refer to "Service" on page32 ). General service and UV
assembly service.
Main Menu > Service
34|User Interface
2588787_A_EN_1811_US_OM
US
Page 35
7Operation
7.1General
Personnel Qualifications
The Ultrasonic Humidifier must only be operated by personnel who are adequately qualified, welltrained, and are authorized by the customer.
Safety
Observe the following safety precautions.
DANGER!
Risk of electric shock!
The unit is mains powered. Using the On/Off switch on the humidifier to shut down the unit will
NOT shut off power to the control cabinet. Live parts may be exposed when the access panels or
doors are removed. Touching live parts may cause severe injury or even death.
Prevention: Before performing any work inside the humidifier, shut down the humidifier properly, as
described in "Shutting Down the Unit" on page37. The unit must be connected to the mains only after
all mounting, installation, and maintenance work has been completed, checked for correct workmanship, and the access panels/doors are closed and fastened securely. Secure the unit against accidental power-up.
Only qualified personnel should perform maintenance or repairs on the unit.
CAUTION!
Electrostatic discharge (ESD)!
The electronic components inside the control cabinet in the humidifier are sensitive to electrostatic discharge (ESD).
Prevention: Take appropriate measures to protect the electronic components inside the unit against
damage caused by electrostatic discharge (ESD). Refer to ANSI/ESD-S20.20.
WARNING!
Heavy object. Risk of injury!
The humidifier is heavy. Injury, damage, or even death may occur if the humidifier falls or tips
over.
Prevention: Ensure that the humidifier is secured properly. Use an appropriate lifting device, or
request assistance to raise the humidifier into position.
WARNING!
Risk to Hygiene and Health!
Improper and infrequent maintenance of the unit presents the risk of spreading pathogens (such
as bacteria , viruses, and parasites) originating from contaminated water, air, and equipment. Failure to reduce the risks increases the likelihood of causing injuries (requiring medical care), infection, and death.
Prevention: Use, regularly clean, and replace the air and water inlet filters. Regularly flush and dis-
infect the humidifier, the blower pack, and their components (tanks, hoses, reservoirs, etc.). Use potable, treated water (potable, R.O., or D.I.).
Follow regular maintenance routines as described in "Maintenance List" on page39 Follow local regulations.
US
2588787_A_EN_1811_US_OM
Operation|35
Page 36
WARNING!
Do NOT leave the unit without power. Risk of contamination.
Do NOT leave the unit unpowered for more than 48 hours, or else a total system disinfect will be
required. Without power, the system will not flush properly, resulting in standing water and conditions that may promote the growth of micro-organisms.
Prevention: Leave your unit powered on. If you do not need humidification during the 48 hours, set
the humidifier to standby or set the demand to 0.
7.2Inspecting the Unit
Perform regulation inspections of the Ultrasonic Humidifier and the humidification system when it is in
operation.
Regularly inspect the following when the unit is in operation:
□
Inspect the water and mist connections
□
Ensure that the drain lines are not blocked or kinked, and empty into the air gap funnel.
□
Check that the unit and other system components have not shifted from their mountings, and are
fastened securely.
□
Visually check the electrical wiring for any damage.
□
Check the Home screen on the unit for messages.
□
Check that the doors on the unit are closed and fastened securely.
If any irregularities, such as leakages, fault messages, or damaged components, are found, shut down
the unit before servicing. Refer to "Shutting Down the Unit" on the facing page. If assistance is required,
please contact your Condair representative.
7.3Starting Up
IMPORTANT! Clean and disinfect the unit if the unit has been unpowered for an extended period of
time (over 48 hours). Refer to "Cleaning and Disinfecting" on page38.
To start the Ultrasonic Humidifier, press the On/Off button at the front of the humidifier. Refer to "User
Interface Controls" on page24.
Once the humidifier is turned on, the control software will perform a system check. When the system
check is completed, the Home screen is displayed on the LCD screen. Refer to "Navigation Elements"
on page25 and the humidifier enters normal operating mode.
Before the humidifier begins to create mist, the unit will flush itself to remove water from the water lines
and reservoir(s). Once flushed, the humidifier will fill the reservoir(s) (this process can take several
minutes, depending on water pressure). After filling the reservoir, the humidifier will remain idle until a
demand for humidification is received.
Note: The humidifier will check the operation of the water level sensor on start-up, and after a flushing
cycle.
7.3.1Restarting the Unit
The Ultrasonic Humidifier can be restarted after performing maintenance on the unit. Ensure the unit is
secure and undamaged before restarting the unit.
IMPORTANT! Clean and disinfect the unit if the unit has been unpowered for an extended period of
time (over 48 hours). Refer to "Cleaning and Disinfecting" on page38.
36|Operation
2588787_A_EN_1811_US_OM
US
Page 37
To restart the unit:
1. Examine the humidifier and other system components for damage. Check for loose or damaged
electrical connections, and ensure that the humidifier and its components have not shifted from
their mountings, and are securely fastened.
2. Turn on the water supply to the humidifier and check for leaks.
3. Reinstall and secure any removed door panels.
4. Remove the lock and tag from the external disconnect switch, and restore power to the humidifier.
5. Press the On/Off button to the On position.
The humidifier will power up and remain idle until a humidity demand signal is received.
7.4Shutting Down the Unit
Strictly observe all safety precautions and perform only the tasks described in this manual. Refer to "For
Your Safety" on page9.
IMPORTANT! Avoid unpowering your unit. To stop humidification, set your demand to 0. This will leave
your unit powered and in 'standby' mode to allow your unit to flush regularly and prevent microbial
growth.
To shut down the Ultrasonic Humidifier:
1. Shut off the water supply to the humidifier.
2. Make a note of the fault code, if any, displayed on the Home screen.
3. Set the On/Off button to the Off position.
4. Wait for the humidifier to drain the reservoir and shut down. The drain valve will remain open for
70 seconds to drain the unit. After this period, the drain valve will close.
5. Disconnect the power supply to the humidifier at the external disconnect switch. To prevent accidental power-up, lock out and tag out.
7.5Remote Monitoring
The Ultrasonic Humidifier can be monitored remotely by connecting the fault relay to a remote monitoring system.
7.6Draining the Unit
To drain the reservoir:
1. Turn off the humidifier by pressing the On/Off button at the front of the unit. Refer to "User Inter-
face Controls" on page24. The drain valve will open and remain energized for 70 seconds to
fully drain the reservoir (independent of the current water level in the unit).
During a flushing cycle, the unit will drain from full. If the water level sensor indicates a full tank after 70
seconds has passed, the unit will enter a Drain Fault and stop operating.
Note: There is no manual drain valve. The drain valve is operated with electricity and is normally
closed. In addition to its main power source, it is also powered by the back-up battery (when the mains
power is not available). Ensure that the back-up battery is installed and fully functional or the unit
will not drain during an outage, and will create stagnant water.
US
2588787_A_EN_1811_US_OM
Operation|37
Page 38
8Maintenance
8.1General
Strictly observe all safety precautions, and perform only those maintenance tasks described in this
manual. Refer to "For Your Safety" on page9.
Use only original Condair replacement parts.
All maintenance work must be performed only by licensed personnel authorized by the customer. It is
the customer's responsibility to verify qualifications of the personnel.
The maintenance work may require removal of the cover on the unit. Refer to "Removing and Installing
the Door Panels" on page43.
The Ultrasonic Humidifier may need to be unplugged to safely maintain and operate on the unit. Refrain
from unpowering the unit (for more than 48 hours) unless it is deemed necessary, as unpowered units
will not flush.
8.2Cleaning and Disinfecting
Cleaning and disinfecting should be performed by authorized persons. Contact your Condair representative for more details.
Refer to "Maintenance List" on the next page for the general procedure and frequency of maintenance.
Disinfectants are mandatory, and must NOT produce toxic residue. Observe all safety precautions
labeled on the disinfectants and cleaner.
CAUTION!
Risk of damage to the humidifier components!
Do NOT use solvents, aromatized, or halogenized hydrocarbons, or other harsh chemicals for
cleaning.
Do NOT use cleaning materials that could leave scratches on wet, plastic and metal surfaces of
the unit.
Do NOT use hard or metallic brushes to clean the unit.
Condair recommends the following:
l Avoid recontaminating the components. Use clean gloves (that are safe to use with the agents
listed below) when handling components and cleaning agents.
l For cleaning: dish washing soap
l For descaling: citric acid solution
Note: pH 2, or 20g citric acid in 1 L water, or 0.18 lbs for every 1 gallon
l For disinfecting: 0.1% hydrogen peroxide, H
l At the end of each cleaning, allow the humidifier to fill and create mist for 2-3 minutes, then per-
2O2
form a full drain cycle. This should be performed manually or through changing the demand signal (full demand to mist; zero stop mist and allow to drain).
Condair provides a cleaning and disinfecting kit for the Ultrasonic Humidifier (P/N: 2594996)
IMPORTANT! Avoid cleaning with hard or rough materials, such as steel wool, or metal brushes. Abrasions on the surface of any components may prevent micro-organisms from being cleaned easily.
Instead use a clean cloth or soft brush.
Avoid materials that will leave particles behind.
Avoid using a hard, or metal (or otherwise abrasive) brush.
Maintenance|38
2588787_A_EN_1811_US_OM
US
Page 39
To disinfect a component:
1. Review the maintenance instructions for the component that will be cleaned and disinfected.
Specific considerations must be taken into account for some components.
2. Ensure UV assembly is shut down before disinfection.
3. Clean the component with hot water and soap, and rinse with water. Remove biofilms and any
surface impurities. Avoid using hard or metallic brushes (to avoid surface abrasions) to clean
and disinfect.
4. Descale components using citric acid* (pH 2), then rinse with water.
l *Let the component soak for 20 minutes and avoid scrubbing. Do NOT agitate the piezo-
electric transducers because they are very fragile.
5. Apply disinfecting agent (hydrogen peroxide) to the component and allow to air dry (this allows
the agent to penetrate and kill pathogens).
6. Rinse thoroughly with clean water (RO or DI water) to remove residue.
To disinfect hoses:
1. Clean and disinfect hose connections, as described above.
2. Remove any residue or particles from the hoses.
3. Flush the hoses with hot water.
4. Flush the hoses with disinfecting liquid and allow to air dry.
5. Rinse the hoses with clean water (RO or DI water) to remove residue from disinfectant.
8.3Maintenance List
Regular maintenance should be performed on the Ultrasonic Humidifier. Refer to Table 12 below. For
replacement procedures, refer to "Maintenance Procedures" on page42.
Following cleaning and disinfecting procedures. Refer to "Cleaning and Disinfecting" on the previous
page.
Ensure that the humidifier and its components are shut down. Refer to "Shutting Down the Unit" on
page37.
Equipment that may be needed for cleaning:
l Screwdriver (cleaned and disinfected).
l Hot water and room-temperature or warm water.
l Cleaning agent (ex. dish soap).
l Descaling agent (ex. citric acid: part number 2589797). Avoid bleach.
l Disinfecting agent (ex. hydrogen peroxide).
l Compressed air.
l Cleaning material (ex. soft bristle brush, cloth).
l Rubber mallet.
Table 12: Maintenance and Cleaning List
ComponentFrequencyDescription
Drain valveAnnuallyInspect and clean the drain valve of any scale and debris. Refer to "Removing
and Installing the Drain Valve" on page49.
US
2588787_A_EN_1811_US_OM
Maintenance|39
Page 40
ComponentFrequencyDescription
Drain cupAnnuallyInspect and clean the drain valve of any scale and debris. Wash with soap and
warm water. Rinse thoroughly. Refer to F igure 13 - on the facing page.
Fill valveAnnuallyClean the fillvalve strainer with a soft-bristled brush to r emove any scale and
debris. Rinse thoroughly. Refer to F igure 13 - on the facing page.
Fill cupMonthlyRemove the cover from the fillcup and clean the inside of the fill cup to remove
any scale and debris. Disinfect and rinse the fillcup before reattaching the
cover. Refer to Figure 13 - on the facing page.
FlexibletubesWeekly -
Monthly
Flexibletubes
and outlets
Water reservoirWeekly -
Electrical wiresAnnuallyInspect allcablesfor damage and insulation breakdown. Replace any dam-
FusesAnnuallyInspect fuseslocated on the driver board (315 mA), the battery backup boar d
Piezoelectric
transducers
WeeklyMonthly
Monthly
AnnuallyInspect the piezoelectric transducers on the reservoir. Replace annually, or as
Rinse thoroughly with hot water . In some cases scale buildup forms inside the
hoses. Use a rubber mallet to knock out the scale. Refer to Figure 13 - on the
facing page.
Clean and disinfect weekly. If the resultsfrom inspectionsare satisfactory (as
per your risk assessor), flexible tubes and outletsmay be cleaned monthly.
Clean and disinfect weekly. Refer to F igure 14 - on page42. If resultsfrom
inspections are satisfactory (as per your riskassessor), reservoir and flexible
tubes may be cleaned monthly. Refer to "Cleaning, Removing, and Installing
the Water Reservoir" on page44.
Remove the fan assembly and mist outlet flange from the reservoir to access
the inside of the reservoir, the heat sink, mist outlet flange, and fan. Clean the
mist outlet flange and inside the reservoir with hot water and soap. Remove
any scale and then disinfect. Rinse the mist outlet flange and reservoir with
clean water.
Clean the fan by using an object (ex. screwdriver) to secure the fan in place
and blow compressed air into the fan assemblyto remove dust. Clean the heat
sinkby blowing compressed air through the ridges.
aged cablesand ensure that allcables are secured properly.
(4 A), and on the piezo-boar ds (3.15 A). Replace the fuses ifnecessary. Refer
to "Replacing Fuses" on page51.
necessary before 10,000 hours of use. Refer to "Replacing the Transducers
(Piezoelectric)" on page47.
40|Maintenance
Transducer
driver board
Backup BatteryAnnuallyInspect the battery and ensure ithas a charge. Check the cables from the bat-
UV assemblyWeekly -
Ultrasonic
Blower Pack
AnnuallyInspect the cables, wires, transistors, and the thermal paste on each tr ans-
ducer driver board. Replace if necessary. Refer to "Replacing the Transducer
Driver Board" on page55.
tery. Replace the cables or battery as necessary. Refer to "Replacing the Bat-
tery" on page56. Do NOT sho rt circuit t he battery.
Inspect for any damage. Replace annually, or as necessary before 10,000
Monthly
WeeklyMonthly
hours of use. Refer to "Removing and Installing the UV Assembly" on page46.
Clean and disinfect the flexible tubes and mist channelweekly. If results from
inspections are satisfactory (as per your riskassessor), flexibletubes and channels may be cleaned monthly. Refer to the UltrasonicBlower Pack Installation,
Operation, and Maintenance Manual.
2588787_A_EN_1811_US_OM
US
Page 41
ComponentFrequencyDescription
2
3
1
4
5
6
7
Reverse
Osmosis System
Air filtersWeekly-
Monthly Annually
Monthly
Replace membranes and filters as necessary. Refer to the Reverse Osmosis
System Installation, Operation, and Maintenance Manual.
Replace allair filters for air leading to the humidifier and UltrasonicBlower
Pack. If the air quality is bad, replace the air filters weekly, and if the air qualityis
good, replace the air filter monthly. Depending on per formance and air quality,
this may be extended — contact your Condair representative.
Figure 13 - Components to clean in the Ultrasonic Humidifier
1. Hoses
2. Fill valve
3. Strainer
4. Fill cup
5. Fill cup cover
6. Drain valve
7. Drain cup
US
2588787_A_EN_1811_US_OM
Maintenance|41
Page 42
Figure 14 - Cleaning the Water Reservoir
8.4Maintenance Procedures
Ensure that the humidifier is shut down properly and observe all safety precautions described in this
manual . Refer to "Shutting Down the Unit" on page37 and "For Your Safety" on page9.
42|Maintenance
2588787_A_EN_1811_US_OM
US
Page 43
8.4.1Removing and Installing the Door Panels
1
2
3
4
5
7
6
Removing the Door Panels
Remove the two door panels in the humidifier as follows. Refer to Figure 15 - below.
1. Turn the captive screw in each door panel counter-clockwise with a screwdriver.
2. Lift up the door panel slightly so that the tabs clear the notches in the cabinet, then carefully
remove the panel.
Installing the Door Panels
Install the two door panels in the humidifier as follows. Refer to Figure 15 - below.
1. Position each door panel so that the two tabs in the top rear of the panel align with the notches
in the cabinet.
2. Lower the door panel to seat the tabs in the notches, while ensuring that the long tab on the side
of the panel slides behind the stationary panel.
3. Make sure that the door panel fits flush on the front and the side. Turn the captive screw in the
door panel clockwise with a screwdriver to secure the door panel. Check that it is fastened
securely.
US
Figure 15 - Door Panel Removal and Installation
1. Stationary panel
2. Door panel, control cabinet
3. Captive screw
4. Door panel, plumbing cabinet
5. Top tab, door panel
6. Long tab, door panel
7. Notch, cabinet
2588787_A_EN_1811_US_OM
Maintenance|43
Page 44
8.4.2Cleaning, Removing, and Installing the Water Reservoir
Cleaning and Disinfecting Considerations: Water Reservoir
The water reservoir houses very delicate piezoelectric transducers that may break during cleaning/removal/installation. To avoid damaging the transducers, avoid removing the water reservoir from
the unit unless it is necessary. This may impact cleaning/disinfecting procedures, so care must be
taken:
l Gently spray the reservoir with clean water to rinse away residue, or allow the reservoir to fill
and drain several times to rinse it out.
l Clean with a soft cloth.
l After cleaning with soap, instead of scrubbing off scale, soak the water reservoir in citric acid
(pH2) for 15-20 min. Approximately 3 L (0.8 gal) is necessary to fill the reservoir without it overflowing through the fan opening.
l To disinfect, spray the water reservoir (drained) with hydrogen peroxide (0.1% solution) and
allow to air dry. The unit may be wiped with disinfectant, however, care must be taken that the
sensitive piezoelectric transducers are not damaged.
If the reservoir should be removed from the humidifier, refer to the steps in "Removal" below.
Citric acid can be ordered from Condair (PN: 2589797).
The following instructions can only be performed while the water reservoir is removed from the unit. It is
suggested to also inspect and perform maintenance on the UV assembly and transducers before the
water reservoir is reinstalled into the humidifier.
l "Removing and Installing the UV Assembly" on page46
l "Replacing the Transducers (Piezoelectric)" on page47
l "Removing and Installing the Drain Valve" on page49
Removal
Remove the water reservoir as follows. Refer to Figure 16 - on the facing page.
1. Shut down the humidifier as described in "Shutting Down the Unit" on page37.
2. Remove the door panels. Refer to "Removing and Installing the Door Panels" on the previous
page.
3. Remove the flexible tubes on top of the water reservoir.
4. Disconnect the two electrical connectors on the water reservoir. Refer to Figure 17 - on
page46.
5. Disconnect the earth connector from the earth tab. Refer to Figure 17 - on page46.
6. Unscrew fastening bolt on the top of the water reservoir. Refer to Figure 16 - on the facing page.
7. Lift the water reservoir out of the drain valve by wiggling the reservoir.
8. Discard the O-ring in the drain valve. The O-ring must be replaced when reinstalling a reservoir.
Installation
Install the water reservoir as follows. Refer to Figure 17 - on page46.
1. Lubricate a new O-ring with water – do NOT use grease, oil, or any other lubricant. For safety
reasons, always install a new O-ring in the drain valve.
2. Make sure the new reservoir is of the same model as the one removed. This can be determined
by the size of the reservoir or the number of piezoelectric transducers.
3. Position the water reservoir above the drain valve. Carefully slide the drain connection of the
reservoir into the drain valve.
44|Maintenance
2588787_A_EN_1811_US_OM
US
Page 45
4. Insert and tighten the two fastening bolts on top of the reservoir to secure the reservoir in place.
1
2
Refer to Figure 16 - below.
5. Connect the earth connection on the earth tab. Refer to Figure 17 - on the next page.
6. Connect the two electrical connectors on the water reservoir. Refer to Figure 17 - on the next
page.
7. Install the flexible tubes on top of the water reservoir.
8. Restart the humidifier as described in "Restarting the Unit" on page36.
Figure 16 - Water Reservoir Mounting
1. Screw (x2)
2. Water reservoir
US
2588787_A_EN_1811_US_OM
Maintenance|45
Page 46
1
2
3
4
Figure 17 - Water Reservoir Connections
1. Flexible tubes
2. Water reservoir
3. Electrical connections
4. Ground connection
8.4.3Removing and Installing the UV Assembly
Removal
Remove the UV assembly as follows. Refer to Figure 18 - on the facing page.
1. Remove the water reservoir from the humidifier . Refer to "Cleaning, Removing, and Installing
the Water Reservoir" on page44.
2. Disconnect both water hoses.
3. Remove the two screws that secure the UV assembly to the cabinet.
4. Disconnect the electrical connectors from the ballast if necessary.
Installation
Install the UV assembly as follows. Refer to Figure 18 - on the facing page.
1. Connect the electrical connectors to the ballast.
2. Install the UV assembly and secure it to the cabinet with the two screws.
3. Connect both water hoses.
4. Install the water reservoir. Refer to "Cleaning, Removing, and Installing the Water Reservoir" on
page44.
46|Maintenance
2588787_A_EN_1811_US_OM
US
Page 47
5
4
6
3
1
2
Figure 18 - Removal and Installation of the UV Assembly
1. Water connection to fill cup
2. Water connection from fill valve
3. Screws (x2)
4. UV assembly
5. Electronic ballast
6. UV reactor
8.4.4Replacing the Transducers (Piezoelectric)
Removal
1. Remove the water reservoir from the humidifier as described in "Replacing the Transducers
(Piezoelectric)" above.
2. Position the reservoir upside down on a flat surface to access the piezo transducers.
3. Remove the two screws that secure the fan assembly to the water reservoir. Refer to Figure 19 -
on the next page.
4. Remove the two nuts that secure the transducer holder to the water reservoir. Lift the transducer
holder upward to remove it from the reservoir. Refer to Figure 20 - on page49.
5. Disconnect the two transducer wires from the transducer driver board.
Installation
1. Position the new transducer over the threaded studs and slide it downward. Make sure the transducer is mounted in the right direction. The contacts on the transducer holder should point to the
center of the water tank. Refer to Figure 20 - on page49.
2. Secure the transducer holder with the two nuts to the water reservoir.
3. Connect the two wires of the transducer to the transducer driver board.
4. Install the fan assembly and secure it to the water reservoir with the two screws. Refer to Figure
19 - on the next page.
US
2588787_A_EN_1811_US_OM
Maintenance|47
Page 48
5. Install the water reservoir into the humidifier as described in "Cleaning, Removing, and Installing
1
2
3
the Water Reservoir" on page44.
Figure 19 - Removal of Fan Assembly
1. Screw (x2)
2. Fan assembly
3. Water reservoir
48|Maintenance
2588787_A_EN_1811_US_OM
US
Page 49
3
1
2
4
5
6
7
8
9
Figure 20 - Replacing Piezoelectric Transducers
1. Rubber seal
2. Notch position
3. Empty holder
4. Piezoelectric transducer
5. Water reservoir
6. Threaded studs
7. Transducer spacer
8. Transducer with holder
9. Nut (x2)
8.4.5Removing and Installing the Drain Valve
Removal
Remove the drain valve as follows. Refer to Figure 21 - on the next page.
1. Remove the water reservoir. Refer to "Cleaning, Removing, and Installing the Water Reservoir"
on page44.
2. Release the hose clamp, and disconnect the fill hose (fill cup to cylinder) from the drain valve.
3. Disconnect the electrical connectors and the ground wire from the drain valve.
4. Remove the two screws that secure the drain valve to the cabinet.
5. Remove and discard the O-ring in the drain valve. This will be replaced with a new O-ring when
reinstalling.
6. To disassemble the valve, loosen the brass nut and remove the solenoid assembly from the
valve body. Note the orientation of the spring.
7. Disassemble the solenoid assembly, and clean all components.
US
2588787_A_EN_1811_US_OM
Maintenance|49
Page 50
Installation
1
2
3
4
7
8
5
6
8
2
1
7
Install the drain valve as follows. Refer to Figure 21 - below.
1. Install a new O-ring in the drain valve.
2. Assemble the solenoid assembly. Orient the tapered end of the spring towards the nut and solenoid.
3. Assemble the solenoid assembly to the valve body, and hand-tighten. Then turn an additional
quarter turn.
4. Install the drain valve and secure it to the cabinet with the two screws.
5. Connect the fill hose to the drain valve and secure it with the hose clamp.
6. Connect the electrical connectors to the drain valve.
7. Install the water reservoir. Refer to "Cleaning, Removing, and Installing the Water Reservoir" on
page44.
50|Maintenance
Figure 21 - Drain Valve Removal and Installation
1. Fill hose (fill cup to cylinder)
2. Screw (x2)
3. O-ring
4. Drain valve
5. Plunger
6. Spring (note orientation)
7. Brass nut and solenoid
8. Solenoid
2588787_A_EN_1811_US_OM
US
Page 51
8.4.6Replacing the Air Intake Filter
1
2
Figure 22 - Air Intake Filter Replacement
1. Air intake port
2. Filter
The air intake filter should be replaced regularly to minimize the intake of airborne contaminants.
To replace the air intake filter:
1. (Recommended) Shut down the humidifier as described in "Shutting Down the Unit" on page37.
If the humidifier is not turned off, the air drawn into the humidifier will be unfiltered and it is possible that the humidifier will become contaminated.
2. Pull the air intake filter down from the humidifier, then replace the air intake filter.
8.4.7Replacing Fuses
Only a qualified service technician authorized by the customer should replace the internal fuse.
Replace the internal fuse in the control cabinet as follows. Refer to Figure 23 - on the next page.
1. Shut down the unit as described in "Shutting Down the Unit" on page37.
2. Remove the door panel on the control cabinet and swing the front panel open.
3. Replace the fuse with a 10A slow-acting fuse.
4. Close the front panel and install the door panel. Ensure the panels are fastened securely.
5. Remove the lock and tag from the external disconnect switch and restore power to the humidifier.
US
2588787_A_EN_1811_US_OM
Maintenance|51
Page 52
1
Figure 23 - Internal Fuse Replacement
1. Fuse, slow-acting (10A)
DANGER!
Risk of electric shock!
The unit is mains powered. Using the On/Off switch on the humidifier to shut down the unit will
NOT shut off power to the control cabinet. Live parts may be exposed when the access panels or
doors are removed. Touching live parts may cause severe injury or even death.
Prevention: Before performing any work inside the humidifier, shut down the humidifier properly, as
described in "Shutting Down the Unit" on page37. The unit must be connected to the mains only after
all mounting, installation, and maintenance work has been completed, checked for correct workmanship, and the access panels/doors are closed and fastened securely. Secure the unit against accidental power-up.
Only qualified personnel should perform maintenance or repairs on the unit.
CAUTION!
Electrostatic discharge (ESD)!
The electronic components inside the control cabinet in the humidifier are sensitive to electrostatic discharge (ESD).
Prevention: Take appropriate measures to protect the electronic components inside the unit against
damage caused by electrostatic discharge (ESD). Refer to ANSI/ESD-S20.20.
52|Maintenance
Only a qualified service technician authorized by the customer should replace the fuses in the humidifier.
Always replace defective fuses that match the specifications given in the procedure.
NEVER short-circuit and bypass the fuses.
There are three locations where fuses should be replaced:
l "Fuses on the Backup Battery Board and Driver Board" on the facing page
l "Fuses on the Transducer Driver Board" on the facing page
2588787_A_EN_1811_US_OM
US
Page 53
Fuses on the Backup Battery Board and Driver Board
Replace the backup battery fuse and the driver board fuse as follows. Refer to Figure 24 - on the next
page.
1. Remove the door panel on the control cabinet. Remove the locking screw and swing the front
panel open.
2. Replace the fuse “F1”, located on the driver board, with a new slow-acting 315 mA fuse.
3. Replace the fuse “F1”, located on the battery backup board with a new slow-acting 4 A fuse
4. Close the front panel and install the locking screw to secure it.
5. Install the door panel, making sure that it is fastened securely.
Fuses on the Transducer Driver Board
Replace the fuse on the transducer driver board as follows. Refer to Figure 25 - on page55.
1. Remove the door panel on the plumbing cabinet. Remove the locking screw and swing the front
panel open.
2. Replace the fuse “F1”, located on the Piezo driver board, with a new 3.15 A slow-acting fuse.
3. Close the front panel and install the locking screw to secure it.
4. Install the door panel, making sure that it is fastened securely.
US
2588787_A_EN_1811_US_OM
Maintenance|53
Page 54
2
1
3
4
5
Figure 24 - Driver Board and Backup Battery Fuse Replacement
1. Fuse, 4 A slow-acting
2. Fuse, 315 mA slow-acting
3. Driver board
4. Battery backup board
5. Battery
54|Maintenance
2588787_A_EN_1811_US_OM
US
Page 55
1
1
Figure 25 - Fuse Replacement on Transducer Driver Board
1. Fuse, 3.15 A slow-acting
8.4.8Replacing the Transducer Driver Board
Removal
Remove the transducer driver board as follows. Refer to Figure 26 - on the next page.
1. Unplug the power connector
2. Unplug the transducer connections
3. Remove the 3 screws, that secure the driver board to the water reservoir
4. Remove the driver board
Installation
Install the transducer driver board as follows. Refer to Figure 26 - on the next page and Figure 27 - on
the next page.
1. Apply thermal paste on the transistors. Start with a dot of thermal paste and evenly spread the
paste on the transistors. Refer to Figure 27 - on the next page.
2. Secure the transistors to the water reservoir with two screws.
3. Secure the board to the water reservoir with one screw.
US
2588787_A_EN_1811_US_OM
Maintenance|55
Page 56
4. Connect the transducer wires (2 orange and 2 yellow) to the driver board.
3
2
1
4
5
3
2
1
5. Connect the power wire to the board.
Figure 26 - Replacing the Transducer Driver Board
1. Transducer driver board
2. Screw
3. Power connector
4. Transducer wires
5. Screw (x2)
Figure 27 - Applying Thermal Paste to the Driver Board
1. Dot of thermal paste
2. Transistor
3. Thin layer of thermal paste
8.4.9Replacing the Battery
56|Maintenance
The battery on the Ultrasonic Humidifier helps prevent stagnant water by energizing the drain valve to
allow water to flow out of the unit when power is shut off.
2588787_A_EN_1811_US_OM
US
Page 57
DANGER!
Risk of electric shock!
The unit is mains powered. Using the On/Off switch on the humidifier to shut down the unit will
NOT shut off power to the control cabinet. Live parts may be exposed when the access panels or
doors are removed. Touching live parts may cause severe injury or even death.
Prevention: Before performing any work inside the humidifier, shut down the humidifier properly, as
described in "Shutting Down the Unit" on page37. The unit must be connected to the mains only after
all mounting, installation, and maintenance work has been completed, checked for correct workmanship, and the access panels/doors are closed and fastened securely. Secure the unit against accidental power-up.
Only qualified personnel should perform maintenance or repairs on the unit.
CAUTION!
Electrostatic discharge (ESD)!
The electronic components inside the control cabinet in the humidifier are sensitive to electrostatic discharge (ESD).
Prevention: Take appropriate measures to protect the electronic components inside the unit against
damage caused by electrostatic discharge (ESD). Refer to ANSI/ESD-S20.20.
Only a qualified service technician authorized by the customer should replace the battery in the humidifier.
Always replace defective batteries that match the specifications given in the procedure.
IMPORTANT! NEVER short-circuit and bypass the battery.
Replacement
1. Remove the door panel on the control cabinet. Remove the locking screw and swing the front
panel open.
2. Disconnect the battery board cables from the battery.
3. Replace the battery located on the backup board with a new one. Refer to Figure 24 - on
page54.
4. Connect the cables to the new battery.
5. Close the front panel and install the locking screw to secure it.
6. Install the door panel, making sure that it is fastened securely.
US
2588787_A_EN_1811_US_OM
Maintenance|57
Page 58
9Troubleshooting
9.1General
Troubleshooting of the Ultrasonic Humidifier must only be performed by well qualified and well trained
personnel. Electrical repairs must only be performed by a licensed electrician authorized by the customer, or by Condair service technicians.
Troubleshooting the Ultrasonic Humidifier may require the user to access the inside of the humidifier to
the control and plumbing cabinets, which may expose the user and equipment to hazards described in
"For Your Safety" on page9.
Warning and Fault Messages:
l "Filling timeout" below
l "Drain fault" on the next page
l "Blower pack fault" on the next page
l "High temperature" on the next page
9.2Fault Indication
When a condition that is not normally expected occurs, it is detected by the control software and a warning or fault message is displayed on the Home screen. Refer to "Home Screen" on page24.
9.3Warning and Fault List
Warnings and faults are displayed on the Home screen of the humidifier. The warning and fault messages can be reset, however, if the cause of the warning or fault persists, the message will return. Refer
to "Resetting Fault Status" on page60. Refer to "Maintenance List" on page39 for cleaning instructions
and "Maintenance Procedures" on page42.
Filling timeout
Probable CauseCorrective Action
Permissible filling time exceeded
Water supply line or water filter is
clogged, shutoff valve closed
l "Fan failure" on the next page
l "UV fault" on page60
l "Service stop" on page60
l "Driver board (A/B) disconnected" on page60
nWait 3 minutes and the unit will attempt to fill the unit again.
nCheck the supply line, shutoff valve, filter, and water pressure.
Troubleshooting|58
Leak in the water supply
Pressure in the buffer tank is too high
2588787_A_EN_1811_US_OM
nClean the strainer in the inlet valve. Replace the inlet valve if
necessary.
nHigh buffer tank pr essure may cause leaksin the water line.
Lower the air pressure in the buffer tank. Refer to "Adjusting
the Buffer Tank Pressure".
US
Page 59
Drain fault
Probable CauseCorrective Action
"Drain fault" appears when the permissible draining time has been exceeded and the sensors still
detect a water in the reservoir.
Drain line clogged
Clogged or defective drain valve
Water level sensor stuck
nClean the drain line.
nClean the drain valve. Replace the drain valve if necessary.
nClean the water levelsensor inside the reservoir.
Blower pack fault
Probable CauseCorrective Action
Blower pack not connected or installed
Blower pack defective
nDisable the Ultrasonic Blower Pack. Refer to "Blower Pack
A/B" on page29.
nConsult the Ultrasonic Blower Pack manual.
High temperature
Probable CauseCorrective Action
Ambient temperature too high
nCheck if the ambient temperature is within operating range.
Refer to "Operating Data" on page23.
Fan failure
Probable CauseCorrective Action
Blower fan blocked
Blower fan defective
nClean the fan. Refer to Figure 14 - on page42.
nReplace the blower fan inside the reservoir.
US
2588787_A_EN_1811_US_OM
Troubleshooting|59
Page 60
UV fault
Probable CauseCorrective Action
UV tube (reactor) defective
UV ballast defective
nCheck connections on the UV tube and replace if necessary.
nCheck connections on the UV ballast. Replace the UV ballast
or assemblyif necessary.
Service stop
Probable CauseCorrective Action
A service interval is overdue
nPerform the needed service on the humidifier and reset the
serviceinterval. Refer to "Service" on page32.
Driver board (A/B) disconnected
Probable CauseCorrective Action
This message appears when there is no communication between the controller and the driver
board.
Communication cable between the
driver board and the control board is
not connected properly
S2 switch on the driver board incorrectly set
nReconnect the RS485 cable is connected to the RS485 port
on the driver board, and isconnected to the SAB on the con-
trol board.
nSet the S2 switch to "0" on driver board A and "1" on dr iver
board B. Power cycle the humidifier.
Unexpected driver board fault
9.4Resetting Fault Status
If the cause of the fault is not resolved, the fault message will return on the Home screen.
Reset a condition in the humidifier with a fault status as follows:
1. Press the On/Off button at the front of the humidifier to Off and wait for 5 seconds.
2. Press the On/Off button to On.
nContact your Condair representative.
60|Troubleshooting
2588787_A_EN_1811_US_OM
US
Page 61
10Decommissioning
10.1General
If the Ultrasonic Humidifier needs to be replaced or removed from service for disposal, strictly follow the
instructions in this section.
Personnel Qualifications
All decommissioning work must be performed only by a qualified service technician authorized by the
customer. It is the customer’s responsibility to verify proper qualifications of the personnel.
Safety
The decommissioning work will require removal of the covers/door panels on the Ultrasonic Humidifier.
Refer to "For Your Safety" on page9.
10.2Removing Unit from Service
Remove the Ultrasonic Humidifier from service for disposal, or long-term storage, as follows:
1. Drain the reservoir in the humidifier, and shut down the humidifier. Follow all safety precautions.
2. Disconnect the humidifier from the power source.
3. Disconnect the control signal inputs (including any power and security loop connections) from
the terminal strips in the humidifier.
4. Disconnect the water and drain connections to the humidifier. Empty out all fluids.
5. Remove the Ultrasonic Humidifier from its mounting surface.
6. If the unit is to be put in long-term storage, store the unit in its original packaging inside a protected area. Refer to "Receiving and Storage".
7. If the unit is to be disposed, refer to "Disposing or Recycling the Unit" below.
10.3Disposing or Recycling the Unit
The Ultrasonic Humidifier and its components must NOT be disposed in domestic waste, and should
only be disposed in accordance with local regulations at authorized collection facilities.
If you have any questions, please contact the appropriate local authorities, or your local Condair representative.
US
2588787_A_EN_1811_US_OM
Decommissioning|61
Page 62
APPENDIX
Appendix
2588787_A_EN_1811_US_OM
US
Page 63
A Appendix:
Maintenance Checklist
* Period may be extended based on water and air quality assessments.
Comments:
Follow all safety precautions. Refer to "For Your Safety" on page9. Shut down the unit before performing maintenance procedures. Refer to "Shutting
Down the Unit" on page37. Do NOT ignore service, warning and fault messages that appear on the Home screen. Refer to "Troubleshooting" on
page58.
For information on cleaning procedures and cleaning agents, refer to "Cleaning and Disinfecting" on page38. Refer to "Maintenance List" on page39
and "Maintenance Procedures" on page42 for descriptions on maintenance procedures.
Table 13: Maintenance Log
ComponentFrequencyDates Last Performed (DD/MM/YY) and Initials
Drain valve12 months
Drain cup12 months
RO-U/DI systemsee Condair representative
Fill valve12 months
Fill cup12 months
Flexibletubes and outlets Weekly-Monthly*
Water reservoirWeekly-Monthly*
Electrical wires12 months
Fuses12 months
Backup battery12 months
Piezoelectric transducers 12 months
Transducer driver board12 months
UV assembly12 months
Ultrasonic Blower Pack6 months
Air filters (replace)Weekly-Monthly
see Condair representative
US
2588787_A_EN_1811_US_OM
Maintenance Checklist|63
Page 64
LEVEL
SWITCH
NTC
5K
MAIN
FAN
BK
RD
YE
OR
OR
WH
WH
GY
RD
RD
GY
GY
RD
GY
RD
GY
RD
GY
RD
C
N
1
C
N
1
C
N
1
C
N
1
C
N
1
C
N
1
+
4
8
V
4
8
V
-
+
4
8
V
-
+
-
+
4
8
V
4
8
V
+
-
4
8
V
+
-
F
1
3
,
1
5
m
A
s
l
o
w
F
1
3
,
1
5
m
A
s
l
o
w
F
1
3
,
1
5
m
A
s
l
o
w
F
1
3
,
1
5
m
A
s
l
o
w
F
1
3
,
1
5
m
A
s
l
o
w
F
1
3
,
1
5
m
A
s
l
o
w
C
N
8
C
N
7
C
N
1
0
C
N
9
C
N
8
C
N
7
C
N
1
0
C
N
9
C
N
8
C
N
7
C
N
1
0
C
N
9
C
N
9
C
N
1
0
C
N
7
C
N
8
C
N
9
C
N
1
0
C
N
7
C
N
8
C
N
9
C
N
1
0
C
N
7
C
N
8
T1T7
T8
T2
T9T3
T10
T4
T11
T5
T12
T6
1 2
3
4
5
6 7 8
1 2
3
4
MU4-6 MU1-3
X1X2
MU-1
MU-2
MU-3
MU-4
MU-5
MU-6
YEL
ORG
YEL
ORG
YEL
ORG
ORG
YEL
YEL
ORG
YEL
ORG
YEL
ORG
ORG
YEL
YEL
ORG
YEL
ORG
YEL
ORG
ORG
YEL
B Appendix:
WiringDiagram Ultrasonic Humidifier - Reservoir
Figure 28 - Wiring Diagram of the Ultrasonic Humidifier Reservoir
Jumper, removed on driver board A, installed on driver board B
Power supply 24 VDC 50 VA
Power supply 48 VDC 600 VA
On/Off switch
Rotary switch module identification: 0
Level, temperature and fan connection
Missing unitspower connection
Supply voltage terminal block
Control ter minal block
US
2588787_A_EN_1811_US_OM
WiringDiagram Ultrasonic Humidifier -
Large (B-unit)|71
Page 72
Notes:
Page 73
Warranty
Condair Inc. and/or Condair Ltd. (hereinafter collectively referred to as THE COMPANY), warrant for a
period of two years after installation or 30 months from manufacturer’s ship date, whichever date is
earlier, that THE COMPANY’s manufactured and assembled products, not otherwise expressly warranted, are free from defects in material and workmanship. No warranty is made against corrosion,
deterioration, or suitability of substituted materials used as a result of compliance with government regulations.
THE COMPANY’s obligations and liabilities under this warranty are limited to furnishing replacement
parts to the customer, F.O.B. THE COMPANY’s factory, providing the defective part(s) is returned freight
prepaid by the customer. Parts used for repairs are warranted for the balance of the term of the warranty
on the original humidifier or 90 days, whichever is longer.
The warranties set forth herein are in lieu of all other warranties expressed or implied by law. No liability
whatsoever shall be attached to THE COMPANY until said products have been paid for in full and then
said liability shall be limited to the original purchase price for the product. Any further warranty must be
in writing, signed by an officer of THE COMPANY.
THE COMPANY’s parts or materials that are considered consumables, including but not limited to: cylinders, filters, nozzles, membranes, media, gaskets, O-rings, etc. are NOT covered by the warranty.
THE COMPANY makes no warranty and assumes no liability unless the equipment is installed in strict
accordance with a copy of the catalog and installation manual in effect at the date of purchase and by a
contractor approved by THE COMPANY to install such equipment.
THE COMPANY makes no warranty and assumes no liability whatsoever for consequential damage or
damage resulting directly from misapplication, incorrect sizing or lack of proper maintenance of the
equipment.
THE COMPANY makes no warranty and assumes no liability whatsoever for damage resulting from
freezing of the humidifier, supply lines, drain lines, or quality of the water used.
THE COMPANY retains the right to change the design, specification and performance criteria of its
products without notice or obligation.
THE COMPANY’s limited warranty on accessories, not of the companies manufacture, such as controls,
humidistats, pumps, etc. is limited to the warranty of the original equipment manufacturer from date of
original shipment of humidifier.
Extended Warranty
Extended warranties are available to purchase under the conditions listed above.
Page 74
CONSULTING, SALES, AND SERVICE:
U.S.A.
826 Proctor Avenue
Ogdensburg, NY 13669
CANADA
2740 Fenton Road
Ottawa, ON, K1T 3T7
Tel: 1.866.667.8321
Fax: 613.822.7964
Email: na.info@condair.com
Website: www.condair.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.