Features
LINE
LINE
EQUIPMENT
UNDER
TEST
POWER
SOURCE
EARTH GROUND
LISN
NEUTRAL
NEUTRAL
GROUND
GROUND
LINE 1
NEUTRAL
LINE 1
NEUTRAL
EQUIPMENT
UNDER
TEST
POWER
SOURCE
EARTH GROUND
LISN
LISN
LINE 2
LINE 3
LINE 2
LINE 3
GROUND
GROUND
GROUND
GROUND
Frequency range of 9 kHz to 30 MHz
Fully compliant with CISPR 16-1-2 / ANSI C63.4
“Air-core” inductors to prevent saturation
Individual Calibration Included
Three-Year Warranty
Description
Line Impedance Stabilization Network
LI-215A
The LI-215A Line Impedance Stabilization Network
(LISN) provides the necessary measurement
platform for performing power line conducted
emissions compliance testing as required by most
worldwide standards for commercial products.
The LI-215A is compliant with both CISPR 16-1-2 and
ANSI C63.4.
The LISN provides dened stable impedance and
isolates the EUT from power source inuences,
thereby providing accurate and repeatable results.
The LI-215A includes one pair of single-conductor
networks, housed together, to be installed in series
with each current-carrying conductor in a single-
phase, dual-phase or DC power system. A second
LI-215A pair can be used to accommodate 3-phase
power systems (Wye or Delta congurations).
The LI-215A is equipped with IEC connector at
the mains (power input) and NEMA 5-15R at EUT
(power output) ports.
protection for your measurement instrument is
highly recommended and available from ComPower.
All Com-Power LISNs are individually calibrated
in compliance with the relevant requirements of
CISPR 16-1-2 and ANSI C63.4. Impedance, Phase,
Isolation, and Insertion Loss data is supplied with
each unit, along with the calibration certicate.
Typical Connection Diagrams
Single Phase connection with one set of LISN
Three Phase connection with two sets of LISNs
This LISN uses air-core inductors to prevent
saturation and permeability variation. The
mounting plate of the LI-215A is left unpainted in
order to facilitate connection to earth ground in its
installation, which is essential due to high leakage
current.
Use of a Transient Limiter for impedance matching,
reduction of out-of-band emissions and transient
Com-Power Corporation 19121 El Toro Rd., Silverado, California 92676 (949) 459 - 9600 www.com-power.com
Rev. D04.14
Line Impedance Stabilization Network
LI-215A
Application
Product Name Line Impedance Stabilization Network (LISN)
Specication CISPR 16-1-2 / ANSI C63.4
Application
Frequency Range
RF Connector
Current Rating
Voltage Rating
Inductors
Mains & EUT Connections
Dimensions (each network)
Power line conducted emissions tests
9 kHz to 30 MHz
50Ω N-type (female)
15 Amperes
270 VAC (Line to Ground), 380 VDC
50 µH (air-core)
IEC 60320 @Mains and NEMA 5-15R @EUT
17 x 16 x 7 inches / 43 x 40 x 18 cm
Weight (each network) 21 lbs / 9.5 kg
Insertion Loss
Isolation
All specications are subject to change without notice.
All values are typical, unless specied.
70
60
50
40
30
Impedance (ohm)
20
10
0
0.009 0.09 0.9 9 90
< 6 dB (150 kHz to 30 MHz)
> 20 dB (150 kHz to 30 MHz)
Frequency (MHz)
, 10 Amperes
(AC)
(DC)
LISN Model No. LI-215A
CISPR 16-1-2, Ed.1.2, Tbls
3/4 (50/250uH)
Impedance Limits (±20%)
ANSI C63.4-2009, Tbl B.1
(50/250uH) Impedance
Limits (±20%)
80
70
60
50
40
30
20
10
Phase Angel (degree)
0
-10
-20
0.009 0.09 0.9 9 90
Com-Power Corporation 19121 El Toro Rd., Silverado, California 92676 (949) 459 - 9600 www.com-power.com
Frequency (MHz)
LISN Model No. LI-215A
CISPR 16-1-2 (Phase), Tbls
3/4 (50/250uH)
Rev. D04.14