INSTRUCTION MANUAL
CLCI-100 BULK CURRENT INJECTION PROBE
with FCLC-100 CALIBRATION FIXTURE
List of Figures
FIGURE 1 - CLCI-100 Bulk Current Injection Probe Features 8
FIGURE 2 - FCLC-100 Calibration Fixture Features 9
FIGURE 3 - Dimensional Diagram for CLCI-100 Bulk Current Injection Probe 11
FIGURE 4 - Dimensional Diagram for FCLC-100 Calibration Fixture 12
FIGURE 5 - Installation of CLCI-100 Probe into FCLC Calibration Fixture 13
FIGURE 6 - Setup for Insertion/Coupling Loss in a 50Ω System (100Ω Loop) 14
FIGURE 7 - Setup for Insertion/Coupling Loss in a 150Ω System (300Ω Loop) 14
FIGURE 8 - Equivalent Schematic of Calibration Circuit 15
FIGURE 9 - Examples of Test Generator Systems 16
FIGURE 10 - Setup for Calibration Accessories for 150Ω System 20
FIGURE 11 - IEC 61000-4-6 (Method A) Test Level Calibration Setup 20
FIGURE 12 - IEC 61000-4-6 (Method B) Test Level Calibration Setup 21
FIGURE 13 - IEC 61000-4-6 (Method A) Test Setup 24
FIGURE 14 - IEC 61000-4-6 (Method B) Test Setup 25
FIGURE 15 - Test Level Calibration into 50 ohm System 26
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459-9600 ● com-power.com
REV051917
Page 4 of 28
1.0 Introduction
This manual includes descriptions of the CLCI-100 Bulk Current Injection Probe
connections, product specifications, safety precautions, operational instructions,
calibration instructions, measurement guidelines and warranty information.
Information contained in this manual is the property of Com-Power Corporation. It is
issued with the understanding that the material may not be reproduced or copied
without the express written permission of Com-Power.
INSTRUCTION MANUAL
CLCI-100 BULK CURRENT INJECTION PROBE
with FCLC-100 CALIBRATION FIXTURE
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459-9600 ● com-power.com
REV051917
Page 5 of 28
d
S
CLCI-100 BULK CURRENT INJECTION PROBE
with FCLC-100 CALIBRATION FIXTURE
2.0 Products Available from Com-Power
Antennas
AntennaKitsAbsorbingClamps
INSTRUCTION MANUAL
Coupling/Decoupling
Networks(CDN)
CombGenerators
ImpedanceStabilization
Networks(ISN)
Preamplifiers
TransientLimiters
CurrentProbes
LineImpedanceStabilization
Networks(LISN)
PowerAmplifiers
Turntables
EmissionsTest
Systems
AntennaMasts
pectrumAnalyzers
AntennaTripods
ConductedImmunity
TestSystems
Near‐FielProbeSets
ProductSafetyTest
Equipment
TelecomTest
Systems
www.com-power.com
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459-9600 ● com-power.com
REV051917
Page 6 of 28
3.0 Product Information
3.1 Incoming Inspection
If there is any evidence of shipping damage. If shipping damage to the product
or any of its accessories is suspected, or if the package contents are not
complete, contact Com-Power or your Com-Power distributor.
Please check the contents of the shipment against the package inventory in
section 3.2 to ensure that you have received all applicable items.
3.2 Package Inventory
STANDARD ITEMS:
9
CLCI-100 Bulk Current Injection Probe
9 Calibration Certificate and Associated Data
OTIONAL ITEMS:
9
FCLC-100 Calibration Fixture
9 ADA-515-2 150Ω to 50Ω Adapter Set
9 ATTN-06-100W Power Attenuator (6 dB, 100W)
9 ATTN-30-100W Power Attenuator (30 dB, 100W)
9 TERM-100W Termination (50Ω)
9 DCD-1000-100W Dual Directional Coupler
INSTRUCTION MANUAL
CLCI-100 BULK CURRENT INJECTION PROBE
with FCLC-100 CALIBRATION FIXTURE
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459-9600 ● com-power.com
REV051917
Page 7 of 28
INSTRUCTION MANUAL
CLCI-100 BULK CURRENT INJECTION PROBE
with FCLC-100 CALIBRATION FIXTURE
3.3 Product Safety Information
3.3.1 Product Hazard Symbols Definitions
The hazard symbols appearing on the product exterior are defined below.
The yellow triangle with an exclamation mark indicates the presence of
important operating and/or maintenance (servicing) instructions in the
literature accompanying the product.
3.3.2 Product Warning/Caution Statements
CAUTION: Hazardous voltages present during operation. Do not handle probe while test is in
progress.
3.3.3 General Safety Instructions
The following safety instructions have been included in compliance with safety standard
regulations. Please read them carefully.
• READ AND RETAIN INSTRUCTIONS - Read all safety and operating instructions before
operating the instrument. Retain all instructions for future reference.
• HEED WARNINGS - Adhere to all warnings on the instrument and operating instructions.
• FOLLOW INSTRUCTIONS - Follow all operating and use instructions.
• WATER AND MOISTURE - Do not use the instrument near water.
• WALL OR CEILING MOUNTING - Do not mount the instrument on a wall or ceiling.
• HEAT - The instrument should be situated away from heat sources such as heat registers or
other instruments which produce heat.
• POWER SOURCES - Connect the instrument only to the type of power source described in
the operating instructions or as marked on the instrument.
• POWER CORD PROTECTION - Place power supply cords so that they are not likely to be
walked on or pinched by items placed on them or against them.
• CLEANING – Clean the instrument only as recommended by the manufacturer.
• DEFECTS AND ABNORMAL STRESS - Whenever it is likely that the normal operation has been
impaired, make the equipment inoperable and secure it against further operation.
• DAMAGE REQUIRING SERVICE - Instrument should be serviced by qualified personnel when:
9 The power supply cord or the plug has been damaged.
9 Objects have fallen or liquid has been spilled into the instrument.
9 The instrument has been exposed to rain.
9 The instrument does not appear to operate normally.
9 The instrument has been dropped, or the enclosure has been damaged.
• SITTING OR CLIMBING - Do not sit or climb upon the instrument or use it as a step or ladder.
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459-9600 ● com-power.com
REV051917
Page 8 of 28
INSTRUCTION MANUAL
CLCI-100 BULK CURRENT INJECTION PROBE
with FCLC-100 CALIBRATION FIXTURE
3.4 Product Features
3.4.1 Description of CLCI-100 Bulk Current Injection Probe Features
1
(clamp closed)
SIDE VIEW
CLCI-100
Bulk Current Injection Probe
10 kHz to 100 MHz
Max Input: 100 Watts (CW)
www.com-power.com
CAUTION
Hazardous voltages present
during operation. Do not handle
probe while test is in progress.
TOP VIEW
2
(clamp open)
3
1
4
1
2
3
3
4
3
3
4
FIGURE 1 - CLCI-100 Bulk Current Injection Probe Features
Input/Output Port
1
When used for bulk current injection, this is the input port of the probe. When used as a
measuring device, this is the probe’s output port. It is fitted with an N-type coaxial
connector.
Spring-loaded Clasps
2
These two (2) clasps, when secured over their respective brackets, lock the clamp into its
closed position
Clasp Brackets
3
These two (2) brackets anchor the respective clasp brackets in order to lock the clamp into
its closed position.
Clamp Window
4
This clamp window is aperture in the center of the clamp through which the wire(s),
cable(s) or cable bundle(s) to be tested are passed through.
2
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459-9600 ● com-power.com
REV051917
Page 9 of 28
INSTRUCTION MANUAL
CLCI-100 BULK CURRENT INJECTION PROBE
with FCLC-100 CALIBRATION FIXTURE
3.4.2 Description of FCLC-100 Calibration Fixture Features
4
5
4
3
2
1
3
3
4
FCLC-100
4
www.com-power.com
4
4
FIGURE 2 - FCLC-100 Calibration Fixture Features
Fixture Opening
1
The CLCI-100 is installed within this opening so that the center conductor rod of the fixture
passes through the approximate center of the probe aperture.
Center Conductor Rod
2
This is the fixture’s center conductor.
Calibration Fixture Coaxial Ports
3
These are female N-type coaxial connectors providing input/output connections to the
fixture.
Thumb Screws for Removable Top Plate
4
These (6) screws secure the fixture’s top plate to the fixture assembly.
Removable Top Plate of Fixture
5
In order to install/remove the clamp into/from the calibration fixture, the top plate must be
removed by removing the six (6) thumb screws. The top cover and thumb screws must be
replaced prior to the performance of tests.
3
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459-9600 ● com-power.com
REV051917