Com-Power CDN-T2E User Manual

Features
Frequency Range 150 MHz to 230 MHz
Designed for IEC / EN 61000-4-6
One unscreened, balanced pair
Individual calibration included
Three Year Warranty
Coupling Decoupling Network
CDN-T2E
Description
Com-Power CDN-T2E is a part of the series of
for testing product for conducted immunity per IEC/EN 61000-4-6.
The CDN-T2E is designed for testing products that uses one unscreened, balanced pair for communication such as Ethernet. It has a RJ11 connectors for both EUT and AE connection. The CDN-T2E can handle up to 2 Amps of current.
The RF disturbance signal coupling port is female BNC. It can handle up to 40V of RF Input Voltage. The bottom surface of the CDN is not painted for
easy and eective grounding.
All Com-Power CDNs are individually calibrated. The Com-Power CDN-T2E fully complies with the requirements contained in the IEC 61000-4-6 and CISPR 16-1-2.
All Com-Power CDNs can be purchased separately or as part of the CIS series conducted immunity test system. This is a pre-packaged solution that includes
an ACS series power amplier and all accessories
required for the test.
Application
Before you begin testing with the CDN-T2E you will need to establish calibrated drive levels corresponding to your desired test levels. During drive level calibration, the RF signal level being injected to the CDN is adjusted incrementally until the
voltage level measured at the 150Ω to 50Ω adapter
(ADA-515-2) connected to the EUT port of the CDN is approximately equal to the Umr value given in the table below. The ADA-515-2 and accessories needed for this test are available from Com-Power.
Test Levels
Open Circuit Voltage
Umr
1 0.167
3 0.5
10 1.67
Umr= Voltage level measured at the output of the 150Ω to 50Ω
adapter (ADA-515-2)
Rev. D01.14
Com-Power Corporation 19121 El Toro Road, Silverado, California 92676 (949) 459 - 9600 www.com-power.com
Coupling Decoupling Network
Specications
Product Name Coupling Decoupling Network (CDN) Compliant Test Standards IEC / EN 61000-4-6 Application One unscreened, balanced pair Frequency Range 150 kHz to 230 MHz RF Input Voltage 40V (Max) RF Input Connector 50Ω BNC (Female) Voltage Rating 124 VAC / 125 VDC (Line to Ground) Current Rating 2 Amps (Max) AE and EUT Connections RJ 11 Common Mode Impedance 150 kHz - 26 MHz: 150Ω ± 20Ω
26 MHz - 80 MHz: 150Ω + 60Ω / – 45Ω 80 MHz - 230 MHz: 150Ω + 60Ω / – 60Ω
Voltage Division Factor 9.5 dB +4/ -1
CDN-T2E
Shorting Adapter Set ADA-T2
Decoupling of Common Mode Disturbance
150 KHz ≥ 45 dB (RF/AE) 500 KHz ≥ 50 dB (RF/AE) 230 MHz ≥ 50 dB (RF/AE)
Dimensions 8.5 x 4.5 x 3.5 inches
21.5 x 11.4 x 8.8 cm
Weight 2 lbs.
0.9 kg
Accessories Available from Com-Power for setting test levels and running the test
ADA-T2 shorting adapters
ADA-515-2 150Ω to 50Ω adapters TEP-050 50Ω Terminator 1, 3, 6, 10, 20, 30 dB Power Attenuators
Directional Coupler
ACS series Power Ampliers
All values are typical values unless otherwise specied. Specications are subject to change without notice.
Typical Data
Impedance
230
210
190
170
150
130
Impedance (Ω)
110
90
70
50
0 1 10 100 1000
Frequency (MHz)
ADA-515-2 Adapter Set
TEP-050 Terminator
Phase
25
20
15
10
5
0
-5
Angle (degrees)
-10
-15
-20
-25 0 1 10 100
Frequency (MHz)
Rev. D01.14
Com-Power Corporation 119121 El Toro Road, Silverado, California 92676 (949) 459 - 9600 www.com-power.com
Loading...