Commax JNS-70MN User Manual

User Manual
Thank you for purchasing COMMAX products.
Thank you for purchasing COMMAX products.
Please carefully read this User’s Guide (in particular, precautions for safety) before using a product and follow
Please carefully read this User’s Guide (in particular, precautions for safety) before using a product and follow
instructions to use a product exactly.
instructions to use a product exactly.
The company is not responsible for any safety accidents caused by abnormal operation of the product.
The company is not responsible for any safety accidents caused by abnormal operation of the product.
Contents
Chapter 1. Nurse call function .............................................................................................. 1
1. Main screen composition .................................................................................... 2
2. Conversation with an inpatient ......................................................................... 3
2.1. Using the speed dial (After setting the bed number) .............. 3
2.2. Using dial button ........................................................................................ 4
3. Receiving a call from an inpatient................................................................... 5
3.1. Request of an inpatient call ................................................................... 5
3.2. Seeing the stand-by call list (in case of offering bed call
function) ........................................................................................................................ 6
3.3. Outside call(IP-Phone) function(Option function) ....................... 7
4. STAFF CALL function .............................................................................................. 8
5. Emergency situation function ............................................................................ 9
5.1. Emergency situation notice ................................................................... 9
6. Simultaneous situation function (Emergency + Call) ........................... 10
6.1. Simultaneous situation notice ............................................................ 10
7. Broadcasting function ......................................................................................... 11
7.1. Whole broadcasting ................................................................................ 11
7.2. Group broadcasting ................................................................................. 12
8. Smart control .......................................................................................................... 13
8.1. Searching/saving the bed ..................................................................... 14
8.2. Designating the bed group ................................................................. 16
8.3. Deleting the bed ....................................................................................... 20
8.4. Recent calls .................................................................................................. 21
8.5. Bed designation in charge ................................................................... 22
9. Bed, group number setting .............................................................................. 33
9.1. To set the number .................................................................................... 33
9.2. Reset ............................................................................................................... 34
1
Chapter 2 . Configuration ..................................................................................................... 35
Users notice
A class equipment
(Broadcasting communication
equipment for business)
Sellers or users should be advised of this
equipment as the equipment suitable for
electromagnetic waves for business use (A class)
and it aims at usage in region except house.
Setting the environment ..................................................................................... 36
COMMAX user option .......................................................................................... 36
Password change .................................................................................................... 36
COMMAX connection ........................................................................................... 37
Network ....................................................................................................................... 38
Cable connection .................................................................................................... 38
Sound and display ................................................................................................. 38
Space to save ........................................................................................................... 39
Language and keyboard...................................................................................... 40
Date and time .......................................................................................................... 40
System Info. ............................................................................................................... 41
Chapter 3. Appendix ................................................................................................................ 42
How to use Nurse Call Log Viewer ......................................................................... 42
1. Connection to PC .......................................................................................... 42
2. Log In .................................................................................................................. 43
3. System Information ...................................................................................... 44
4. Network Information ................................................................................... 44
5. System setting ................................................................................................ 45
6. Nurse call log .................................................................................................. 46
* Diagram ............................................................................................................................ 47
2
Chapter 1. Nurse call function
1. Main screen composition
2. Conversation with an imp
3. R
eceiving a call from an impa
4. S
taff call function
5. Emergency situation func
6. Simultaneous situation function
7. Br
8. Smar
oadcasting func
t contro
8-
1. Searching the bed
8-2. Designat
3. Deleting the bed
8-
8-4. Recent calls
8-5. Bed designation in cha
l function
ing the bed group
tion
atient
tient
tion
rge
9. Bed, group number setting
3
1. Main screen composition
① , ② Page movementRoom numberBed numberWhole broadcasting
Group broadcastingSmart controlRemote control
③ ④ ① ② ⑤ ⑥ ⑦
Main screen
4
2. Conversation with an inpatient
2.1. Using the speed dial (After setting the bed number)
Pick the receiver or press the speaker button in front.After pressing the relevant room number, press the bed number.
Then, you can speak with by calling the relevant patient. (In case
of speaker phone mode, you can speak with if you pick the
handset.)
Screen using the speed dial
* , By pressing (, ②), you can check the bed
numbers of previous and next rooms. (If 12 rooms, 8 beds are inputted, max 24 rooms, 16 beds are set.)
5
2.2. Using dial button
Pick the handset or press the speaker button in front.After entering the relevant room number with the button in front,
press the asterisk(*). If you press the pound button (#) after
entering bed number, you can speak with the relevant patient by
calling him or her. (In case of speaker phone mode, you can
speak with if you pick the handset.)
Screen using number button
6
3. Receiving a call from an inpatient
3.1. Request of an inpatient call
The number of bed and room requesting call are displayed as
green and yellow wall paper is switched.
Call request screen
If you pick the handset or press the speaker phone button, the
number of bed and room requesting call are displayed.
Incoming progress screen
7
3.2. Seeing the stand-by call list (in case of offering bed call function)
If there’s call request from another bed during the conversation with a
patient, stand-by call list will be shown.
If you select the stand-by list while talking, previous call will be ended and
conversation with selected bed will be started. Stand-by list will disappear
when the conversation has done.
Seeing stand-by call list
8
3.3. Outside call(IP-Phone) function(Option function)
During the nurses absence in the nurse station, call request is
delivered by the information of nurse and schedule registered in the server(IP-PBX) to the nurses cell phone. Once the nurse get
the phone, a main receiver is displayed the number of bed on a
screen. In a main receiver, outside call should be ended in order to
use call function.
Mark of outside call
* This function is needed additional devices such as IP-PBX, Media Gateway
for its use.
9
4. STAFF CALL function
In case that the nurse judges the situation as emergency when the nurse presses the presence switch from patients call and visits the bed
of the patient, in a bed if you try call request on the screen of main
receiver in a nurse station the number of bed and emergency sound
will break out. (Same as call request screen)
Call request screen
10
5. Emergency situation function
5.1. Emergency situation notice
Can check emergency request occurred in a ward or outside
lavatory or shower room.
If you press the call button once more or pull the emergency
switch in the place occurred in the emergency situation,
emergency will be cancelled and stand-by screen will appear.
General emergency notice screen in a ward
Outside emergency notice screen
* In case of ward of general, lavatory and shower room are marked as same number.
In case of outside, room number represents designated number (9997) and lavatory
or shower room are marked differently in bed number
11
6. Simultaneous situation function (Emergency +
Call mark
Emergency situation notice
Call)
6.1. Simultaneous situation notice
In case of emergency during conversation or incoming call in the
middle of emergency, you can check them at the same time.
Simultaneous situation notice screen
12
7. Broadcasting function
7.1. Whole broadcasting
Able to broadcast for whole rooms in connection with receiver.
Pick the handset or deliver a message after pressing the whole
broadcasting in speaker mode.
Whole broadcasting implementation screen
13
7.2. Group broadcasting
Can broadcast with bed in a group. (Group setting in remote control
mode) Pick the handset or press the group broadcasting at the
bottom of menu in speaker phone mode.
With using the dial button in front or selecting the group to
broadcast, press the group number and deliver a message after
pressing group number and pound button (#).
Group broadcasting screen
14
8. Smart control
This function is provided in order to register and change the bed
phone in the main receiver. Because ID of first released bed phone is
identified, each bed phone can be used through remote control.
15
8.1. Searching/saving the bed
Max bed numbersWard exploration room number to startWard exploration room number to end
Start searching button End button
② ③ ④
It can be registered automatically by exploring the shortcut as
appointed numbers. As ID of bed telephone is identical, this function is
used after ID change.
After selecting max beds number in the blank to enter, enter the all
numbers of the installed bed inside the ward. (the most installed beds
number inside the ward or the highest bed number inside the ward)
After selecting the edit box of start room number, enter the lowest
ward number in the ward.
Select the edit box for end room number. Then, enter the highest
ward number in a ward.
16
If the three information input is completed, press the start button of
exploration and then exploration will be starting.
After completed the exploration, press the exit button. Then, in a
main screen, each beds hot key(shortcut) registered automatically can
be checked.
Exploration progress screen
ex) A certain ward in a general hospital has from second ward to fourth
ward, the ward which has the most beds is No. 1010 and No. 1011 and
they accommodate 8 beds.
Max bed number => Enter 8.
Ward exploration start => Enter 1001.
Ward exploration end => Enter 1020.
17
8.2. Designating the bed group
Bed group filterBed selection buttonBed ID (Room no.-Bed no.)Bed group ID
Bed group ID selectionBed group change buttonBed group deletion buttonQuit button
① ② ③ ④ ⑤ ⑥ ⑦
Able to set as identical group by selecting the bed to want number
group designation of registered bed.
18
Through Bed group filter, search the registered beds as all or each
group.
Group selection screen
After selecting the beds you want to change, press the Bed group ID
selection button and select the group number to change. (Group F is
default setting number)
Bed selection and Bed group ID selection button screen
19
Group Number changing screen
After selecting Bed ID and group number to change, press the
change button. Then, group number of Bed ID is changed.
Group number change progress screen
20
If you press the delete button after selecting Bed ID and group
number, group ID of default state is changed to F.
21
8.3. Deleting the bed
Bed deletion screen
After selecting the room number of the ward to delete, agreement
message of deletion pops up if you press the relevant bed.
If you press Yes, the relevant bed is deleted. In case of return to main screen, press the Home button of navigation
button or (Home) at the right bottom of the screen.
22
8.4. Recent calls
Incoming callsOutgoing callsBed number list
Dialing time listTalk time list
① ② ③ ④ ⑤
Recent calls between user and patient can be checked. It can check the
place to send and receive, date and time to start a call, Talk time.
23
8.5. Bed designation in charge
Telephone number box Name box Nurse select button Users list (Red part)
User employment status button User creation button User Info. Edit button User Info delete button
① ② ③ ④ ⑤ ⑥ ⑦
This is the function which is connected to smart phone of the nurse in charge during a nurses absence in nurse station. This setting is possible to
set with smart phone or main receiver.
Currently, this can be used the Database must be installed on a IP-PBX
server. (PSTN approach supported later).
8.5.1 User registration
24
After entering the name and phone number of a nurse, press the
create button. Then, the contents in a list will be shown
If you want to edit the information of a nurse, select a nurse in the
list. Then, the name and phone number of the selected nurse are
shown in a blank to enter at the top. After editing the contents by
selecting the blank to enter, press the edit button. Then, the contents
will be modified.
In case of deleting the information for the saved nurse, after
selecting the information of a nurse to delete in the list, press the
delete button. Then, the information of the nurse in the list will be
disappeared.
It can be set the employment status with the user employment
status. The nurse who is not set cannot receive a call.
25
8.5.2 Time organization
Time arrangement list select button
Time arrangement list (MAX 4)
Profile name boxStart time setting button
End time setting buttonCreation buttonDelete button
④ ⑤ ⑥
After entering a name in a blank to enter of profile name, enter the
star time by pressing the start time setting button. After entering the
end time with pressing the end time setting button, press the
creation button. Then, it is added in the list. (Max 4)
26
Screen of start time setting
In case of editing the information, if you select the profile to edit in
the list, the selected name is shown in the edit box of the profile
name. If you set the time with time setting button, the changed
contents in the list can be checked.
In case of deleting the information, if you press the delete button
after selecting the profile to delete in the list, it is deleted from the
list.
27
8.5.3 Time table
Previous month change buttonPresent monthNext month change buttonPresent month schedule
User select buttonTime arrangement list and buttonCreation buttonEdit button
① ② ③ ④ ⑤ ⑥ ⑦
To set the working schedule, select the relevant time zone in the time
schedule list() after selecting the saved user() to set the working
schedule of the user. If you press the creation button after designating
the month to work in the relevant time zone and date, it is completed.
In order to edit the working schedule of user, press the edit button
after selecting the user to change in the list() and then it is
completed.
Caution) Creation button is needed first once a month at least. Since then, the
schedule changes of the relevant month will be edited.
28
8.5.4 Majority designation
Bed select button Room list (Red part) Bed list (Red part)Setting button
Select all beds button User select buttonUser list
① ② ③ ④ ⑤ ⑥ ⑦
In case that users responsible beds are same, you can set it easily
through not individual designation but majority designation.
After selecting the bed and room to set, choose the relevant user in
the users list. Then press the setting button and responsible user for
the relevant bed will be saved.
29
8.5.5 Individual designation
User select buttonBed select button
Whole bed listEdit button
① ② ③
Able to set the responsible bed of a nurse individually.
After selecting the user to set, choose the bed in charge. Then press
the edit button and it will be saved.
30
8.5.6 Call type
Bed name Connection form Time intervalBed selection button
Interval set button Select all button Delete all buttonSetting button
① ② ③
⑥ ⑦ ⑧
Able to select call system in order and concurrently of call connection system with responsible nurses smart phone.
After checking the bed number to set(), select the concurrence or
not by pressing next setting button. In case of editing the time interval, press the time interval setting button() and set the interval.
It will be set at once with Select all(), Cancel all button()
31
8.5.7 Setting
Mark of main receiver confirmation
Send button
Bed list should be saved in the server so that the user selects the
relevant bed with main receiver. Designation in charge for bed function can be used if this process is preceded.
Set screen through main receiver
In case of setting in the main receiver, check mark(①) is shown.In order to saving all bed list connected with the main receiver to
the server, press the transmission button(②) and then it will be
saved in the server.
32
Set screen through smart phone
Enter the number of main
receiver.
Enter IP address
Main receiver confirmation mark
① ② ③
To user the designation function for the bed in the smart phone,
select the assigned number of the main receiver() and enter the
assigned phone number.
33
For the saved bed list in the server, the record of user, loading the
time organization history, press the relevant server IP and port
number.
During setting a smart phone, confirmation mark(③) of the main
receiver is shown as unchecked.
34
9. Bed, group number setting
① ② ③
④ ⑤ ⑥
9.1. To set the number
Press the remote control in the main screen while talking on the
phone and select the number to change. (Bed number, group
number)
Check the existing number in (Before) and correct the number
in (After) and press (OK). Then the value will be changed.
To change the bed number, enter the room number and after
pressing the asterisk(*) button, enter the bed number and OK
button.
To change the group number, enter the new group number and
press the OK button.
Press the remote control again and confirm the changed number.
Number setting screen
35
9.2. Reset
1. This is the function to reset the power of bed connected with
the main receiver.
2. Press the OK button and Yes button. Then, you can reset
the power of the bed.
36
Chapter 2 . Configuration
37
Setting the environment
In the mode of environment setting, you can move to stand-by
screen and previous screen by pressing (HOME) or
(BACK) button on the top of the screen.
COMMAX user option
Able to change password.
Password change
You can change the password to off emergency situation and go out mode.
1. Click the password-changing button.
2. If the entering password screen is displayed, click the password
box. Then you can use virtual keyboard mode.
3. Enter the current password and then click the next button.
4. Enter the new password and click the next button, enter the new
password again to verify whether the password you enter earlier is
correct.
38
5. After clicking the complete button, you can go back to the
password changing mode.
6. Press the OK button to complete new password changing.
Do not forget the password. Password is 4 digit number. In case of inconsistency with previous password, password is not
changed. In case that new password and new password confirmation are different
each other, password is not changed. In case that server is not connectable, icons are inactivated and cannot
set.
COMMAX connection
To COMMAX connection, configuration can be set. Do not operate the
setting except for installers. In case that right value is not entered,
COMMAX connection and management are not operated properly.
1. Server IP setting
Able to set local server, SIP server, update, CCTV server, device server, IP
address of home server
1. Press the icon to set
2. Enter the relevant IP address after pressing the screen to input if
the window to input is marked.
2. Call prix fixe setting
1. Press the call prix fixe to select in the menu.
2. If call prix fixe input screen is displayed, enter the relevant number
and press the OK button.
39
Network
Cable network to use the COMMAX connection and management can be
set.
Cable connection
Enter the IP address, gate way, net mask, DNS (Inquire to installation
manager)
Sound and display
1. Mute mode
Able to convert into Mute mode. (Except for media and alarm)
1. Press the mute mode.
2. If mute mode is checked, sound setting icon at the bottom is
inactivated. And is displayed the task bar on the top of the
screen. . If checking of the mute mode is cancelled, the icon of the
sound set at the bottom is activated and in the task bar of
the top is removed.
2. Bell sound/media volume
1. Press the bell sound volume or media volume.
2. Adjust the volume by moving the stick if the window for volume is
marked in a screen.
3. By pressing the Okay button, setting is saved.
3. Call/Alarm bell sound
1. Press the call bell sound or alarm bell sound.
2. In case of telephone bell sound, list screen of caller ’s location is
displayed. (In case of bell sound, select the list right)
3. After selecting the location to set, select the bell sound. (Can here
the bell sound in advance)
4. Press the Okay button to set the selected bell sound.
40
4. Screen brightness
1. Press the brightness.
2. Bright window is displayed on a screen.
3. Adjust the volume by moving the tool.
4. Setting is saved by pressing the Okay button.
5. Setting the automatic OFF time of the screen
1. Press the time restriction on the screen.
2. The list which can restrict the time on the screen is marked.
3. If you press the time to want, the setting is saved. If you press the
cancellation button, it is not set.
6. Touch screen correction
1. Press the correction of touch screen.
2. According to the guide of screen, press the cross icon in order.
3. If the correction is completed, completion of correction is marked.
Space to save
The space to save of the product can be checked. SD card, whole size of
internal space to save and size of the saving space available are marked.
41
Language and keyboard
The language marked in the equipment and simulated keyboard can be set.
1. Change of language
1. Press the language selection.
2. Select the language to set in the list.
2. Key board setting
When entering the text, decide the default keyboard to be marked. In
case that additional keyboard is supported except for Android keyboard
(English input), when entering the text, it can be used as the keyboard the
user want. If you check the keyboard to use as default keyboard, when
entering the text, the relevant keyboard is marked.
1. Check the keyboard to use as a basic keyboard.
2. When entering the text, checked keyboard is marked.
Date and time
Set the date and time.
1. Date and time setting automatically
1. Check the auto
2. Time server (NTP) and synchronization interval icon are activated.
3. After pressing the time server (NTP), enter the time server. If it is
connected with the entered time server rightly, it is changed to the
current time.
4. Synchronizing with server every time set at an interval of
synchronization.
2. Date and time setting manually
1. Cancel the check automatically.
2. Date and time setting icon are activated.
3. Press the date setting
4. If the window of date setting is displayed, set the time with +
42
button and – button and press the OK button.
5. By pressing the time setting, set screen is marked.
6. Set the time on reference to the date setting.
3. Designating time zone
Press the time zone.
4. In the list, select the time zone of the city that host lives currently.
5. Designating the 24hours form
1. Check or cancel the 24hours form
2. If checked, the time is marked as 24hours. If cancelled, the time is
marked as 12 hours.
3. Press the date form selection.
4. The date form selection window is marked on a screen.
5. Press the form to set.
System Info.
Able to check the component installed in system and product firmware info.
1. Checking component version
1. Press the component.
2. The version of installed component is marked.
2. Checking product firmware info.
1. The version of Boot loader Kernel and Build number are marked.
43
Chapter 3. Appendix
Computer IP
JNS-70MN IP
IP Example
10.10.1.12
10.10.1.11
Subnet Mask Example
255.0.0.0
255.0.0.0
Gateway Example
10.0.0.1
10.0.0.1
How to use Nurse Call Log Viewer
1. Connection to PC
1.1 Connect the LAN port of PC to the JNS-70MN with LAN cable.
1.2 Change IP band of computer regarding JNS-70MN IP band.
1.3 Progress Ping Test from PC to JNS-70MN IP.
1.4 If the network connection is normal provides the following
window.
44
2. Log In
2.1. Enter IP address of Nurse call main receiver into web browser.
(Ex: Nurse call Main receiver IP: 10.10.1.11)
.
2.2. Enter ID and Password.
(Initial ID/Password: root/root)
.
45
3. System Information
You can check product information.
3.1. Click the menu [System Info].
4. Network Information
You can change IP of Nurse Call main receiver.
4.1. Click the menu [Network Info].
4.2. Enter new IP information to change.
4.3. Click Update button.
4.4. If you want to apply changed information, reboot the power of
Nurse Call main receiver.
46
5. System setting
You can change the password.
5.1. [System Setting] Click the menu.
5.2. Enter the password that you will change and click Update button
5.3. ID cant be changed .
47
6. Nurse call log
Nurse call log is appeared on the screen.
6.1. [Nurse Log] Click the menu.
6.2. You can search by specifying the period
6.3. If you click excel icon, you can download excel file
Logs are saved for up to 3,000.
48
* Diagram
Product name : nurse calling device
Line item : JNS-70MN
Power source : DC24V
Max current consumption : 400mA
49
513-11, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Int’l Business Dept. Tel. : +82-31-7393-540~550 Fax. : +82-31-745-2133 Web site : www.commax.com
Printed In Korea / 2015
.07.104