Comet M1140, M1200, M1300, M1220, M1320 Instruction Manual

...
www.cometsystem.com
Instruction manual
M1140 M1200 M1220 M1300 M1320 M1321 M1322 M1323 M1440
2 IE-MLG-M1xxx-07
© Copyright: COMET System, s.r.o. It is prohibited to copy and make any changes in this manual, without explicit
agreement of company COMET System, s.r.o. All rights reserved. COMET System, s.r.o. makes constant development and improvement of
their products. Manufacturer reserves the right to make technical changes to the device without previous notice. Misprints reserved.
IE-MLG-M1xxx-07 3
Content
INTRODUCTION ........................................................................................ 4
GENERAL SAFETY PRECAUTIONS ......................................................... 6
GETTING STARTED .................................................................................. 7
Multilogger in portable application ................................................. 7
Multilogger in stationary application .............................................. 8
INSTALLATION INSTRUCTION ................................................................. 9
Recommendations for mounting ................................................. 12
INPUTS OF MULTILLOGER .................................................................... 15
Input connectors ......................................................................... 16
ALARM OUTPUT ..................................................................................... 19
POWER SUPPLY OF MULTILOGGER .................................................... 21
KEYBOARD OPERATION ........................................................................ 24
Control and indication parts ........................................................ 24
Basic display operation ............................................................... 25
Main menu .................................................................................. 27
CONNECTING WITH THE COMPUTER .................................................. 31
SOFTWARE COMET VISION .................................................................. 32
DEVICE SETUP USING SOFTWARE ...................................................... 34
Device configuration basics ........................................................ 34
Device configuration items .......................................................... 34
APPLICATIONS NOTES .......................................................................... 44
OPERATION AND MAINTENANCE RECOMMENDATIONS .................... 50
TECHNICAL PARAMETERS .................................................................... 53
Power supply .............................................................................. 53
ALARM OUT output .................................................................... 55
Communication interfaces ........................................................... 56
Measuring, data memory and real time clock .............................. 57
Electromagnetic compatibility ...................................................... 57
Operation and storage conditions ............................................... 58
Mechanical characteristics .......................................................... 58
Technical parameters of the inputs ............................................. 59
APPENDIXES .......................................................................................... 70
Appendix 1: Selected error messages ......................................... 70
Appendix 2: Output and power supply circuits ............................. 72
Appendix 3: Input circuits ............................................................ 73
Appendix 4: Connecting of Pt1000 RTD probe ............................ 74
4 IE-MLG-M1xxx-07
Introduction
The Multilogger has been designed for measuring and recording physical and electric quantities with an adjustable logging interval from 1 second to 24 hours. The device is equipped with 4 inputs (connectors) for external probes or signals. Moreover, some versions also include an internal atmospheric pressure and/or CO2 concentration sensor. The production line consists of several models with various combinations of the input ranges of the measured quantities. The Multilogger can be used either as stationary or portable device.
The device allows for:
- Measuring and processing input quantities; the maximum number of the measured and calculated quantities is 16. Some models can also acquire data from binary inputs (up to 2 inputs). Models for measuring barometric pressure allow sea level recalculation.
- Finding and storing minimal and maximal values for each measured quantity (since their last manual reset).
- Acquiring autonomous time record of measured values, independent of the mains supply. Recorded values can be stored as separate samples in the recording time or as average/ minimal/maximal values during the logging interval. Recording may be continuous, or only during an alarm optionally.
- Creating alarm states and perform actions (audio and optical signalization, alarm output signal, report messages via telephone dialer and/or various Ethernet interface protocols, etc.).
- On-line monitoring of the measured values and states. Accessing these values via the web.
- Collecting and processing data in the central database.
- Powering by batteries (NiMH accumulators or alkaline batteries).
- Powering by external power supply, which can be also used for charging NiMH accumulators. External power supply is required for full operation of the Ethernet interface. Some limited functions of this interface may be available during battery operation.
IE-MLG-M1xxx-07 5
Multilogger drawing:
* only certain models
6 IE-MLG-M1xxx-07
General safety precautions
The following list of precautions serves to reduce risk of injury or damage of described instrument. To prevent injuries use instrument accordingly with rules in this manual.
Installation and service need to be performed by qualified
persons only.
Use a suitable power supply. Use power supplies only with a voltage
recommended by the manufacturer (+5V DC) and approved in accordance with proper standards. Pay attention, the source has undamaged cables or cover.
The device and power supply must not be used in a humid and
hazardous environment (such as bathrooms), at locations exposed to direct sunlight and to other heat sources, thus preventing damages and deformation of the cover.
Do not use the instrument without its covers. Do not remove the
covers from the device.
Do not use the instrument if it does not work correctly. If you think,
that the device is not working correctly, let check it by qualified service person.
Never charge batteries that are not rechargeable.  Do not use the device in explosive environment.
IE-MLG-M1xxx-07 7
Getting started
Multilogger in portable application
Connect the appropriate probes and signals to the device - for the
DigiS/M, DigiL/M and Pt1000/M probes, follow the optical key for correct orientation of the connector. Probes of the DigiL/M-xx line with a cable can placed in the holder on the side of device (optional accessory). For more information, see Chapter “INPUTS OF MULTILOGGER.
Switch on the Multilogger - push the switch button on the device. The
display lights and the instrument starts working. If you will communicate with PC via the Ethernet interface, the power adapter +5V DC is necessary (supplied accessories). For more information, see Chapter “POWER SUPPLY OF MULTILOGGER”.
Install Comet Vision software to your computer - it is available at
www.cometsystem.com. If you purchased a license for full version,
register it.
Connect Multilogger with your PC - you can use the USB, Ethernet or
RS232 port (special cable necessary). The external power supply is needed for Ethernet interface communication. For more information, see Chapter “CONNECTING WITH THE COMPUTER”.
Device settings - use SW to set up the device in accordance with the
connected probes and your requirements. If you are not going to use the device soon, turn it off and place it in its transportation case. For more information, see Chapter DEVICE SETUP USING SOFTWARE.
Measurements - remove the Multilogger from the transportation case,
connect probes and external power adapter (if needed) and turn the device on. Some device settings can be changed from its keyboard. When measurement ends, move the device to your computer room and download the recorded data to PC. For more information, see Chapter KEYBOARD OPERATION.
Charging batteries - if the Multilogger includes rechargeable batteries,
you can recharge them when needed. For more information, see also Chapter “POWER SUPPLY OF MULTILOGGER”.
Maintenance and periodic inspections - In order to secure reliable
operation of the device, it is recommended to make regular inspections. For
more information, see part “OPERATION AND MAINTENANCE
RECOMMENDATIONS”.
8 IE-MLG-M1xxx-07
Multilogger in stationary application
Select a suitable location for placing of the Multilogger - pay attention
to parameters of ambient environment, minimize the number of cables, avoid sources of interferences. For more information, see Chapter INSTALLATION INSTRUCTIONS”.
Mounting of sensors and routing of cables - pay attention to rules of
theirs installation, use recommended working positions, keep enough distance from high power devices and cables.
Connect the appropriate probes and signals to the device - for the
DigiS/M, DigiL/M and Pt1000/M probes, follow the optical key for correct orientation of the connector. Probes of the DigiL/M-xx line with a cable can be placed in the holder on the side of device (optional accessory). For more information, see Chapter “INPUTS OF MULTILOGGER.
Check proper connection - before first switching on. If Multilogger
controls other action regulation elements, it is recommended to put them out of operation before configuration of Multilogger.
Switch on the Multilogger - push the switch button on the device. The
display lights and the instrument starts working. If you will communicate with PC via the Ethernet interface, the power adapter +5V DC is necessary (supplied accessories). For more information, see Chapter “POWER SUPPLY OF MULTILOGGER”.
Install the COMET Vision software to your computer - it is available at
www.cometsystem.com. If you purchased a license for the extended SW,
register it.
Connect Multilogger with your PC - you can use the USB, Ethernet or
RS232 port (special cable necessary). For Ethernet interface communication the external power supply is needed. For more information, see Chapter “CONNECTING WITH THE COMPUTER”.
Device settings - use COMET Vision software to set up the device in
accordance with the connected probes and your requirements. If needed, check the alarms functionality. For more information, see Chapter “DEVICE SETUP USING SOFTWARE.
Maintenance and periodic inspections - In order to secure reliable
operation of the device, it is recommended to make regular checks. For
more information, see part “OPERATION AND MAINTENANCE
RECOMMENDATIONS”.
IE-MLG-M1xxx-07 9
Installation instruction
Location of Multilogger must match the operating conditions and not allowed manipulations.
Working position of the device:
Multilogger placed on a horizontal, nonflammable surface (horizontal
working position is not suitable for devices with thermocouple inputs and for devices with cable-less probes).
Multilogger fixed on the wall from nonflammable materials or in a low-
current switchboard using the holder - the working position is with input connectors upwards.
Device dimensions:
10 IE-MLG-M1xxx-07
The way of mounting the holder to Multilogger and mounting holes dimensions:
MLP001
MLP002
IE-MLG-M1xxx-07 11
MLP004
12 IE-MLG-M1xxx-07
Recommendations for mounting
Do not mount Multilogger near sources of interference (Multilogger
must not be mounted directly to power switchboard nor to its nearness. Also do not mount data logger near power contactors, motors, frequency converters and other sources of strong interference).
In cable routing follow rules of standards for installation of low current
distribution (EN 50174-2), especially it is necessary to pay attention to avoid electromagnetic interference intrusion to the leads, transmitters, transducers and sensors. Do not locate cabling near sources of interference.
Do not use leads in parallel with power distribution network leads
IE-MLG-M1xxx-07 13
Do not use cables longer than 30 m generally; for RTD sensors is
recommended max. 15 m; cables must be installed only in interior areas.
Do not interconnect the inputs with other measuring systems.
Use shielded cables for the voltage and current inputs - connect the
shielding to the Multilogger but not to the end devices if these do not have a special terminal for it. In this case, shielding must not be connected to external metal parts of this device or to any other device. Never use shielding as a signal wire.
14 IE-MLG-M1xxx-07
Do not use common leads for several inputs.It is recommended to earth larger systems with Multilogger at one
point - the GND terminal of the alarm output can be used for this purpose. This grounding will work correctly if the system would not be grounded currently at another point too. One-point grounded systems are more resistant against electromagnetic disturbance.
IE-MLG-M1xxx-07 15
Inputs of Multilloger
Manufactured models differ from each other by their input ranges and by input connectors. Prior to connecting voltage and current input signals, you should check their parameters in the following part: “TECHNICAL
PARAMETERS OF THE MULTILOGGER INPUTS”.
Universal input with MiniDIN connector
temperature and humidity probe of Digi/M line temperature probe of Pt1000/M line temperature probe Pt1000 (custom RTD)
Universal input with miniature thermocouple connector
thermocouples of the K, J, S, B, T and N types DC Voltage - 18 mV to +18 mV DC Voltage - 60 mV to +140 mV
Universal input with terminals
DC current 0 to 20 mA DC voltage 0 to 10 V binary signals
M1140
M1200
M1220
M1300
M1320
M1321
M1322
M1323
M1440
Barometric pressure internal sensor CO2 concentration internal sensor CO2 external probe only
16 IE-MLG-M1xxx-07
Input connectors
The Multilogger does not have galvanically isolated inputs. For this reason, only devices that are not connected to any other circuit may be connected to the inputs of Multilogger. Simplified circuit diagram of inputs is shown in Appendix 3.
MiniDIN connector _______________________________
It is used for connecting the DigiS/M, DigiL/M, Pt1000/M probes and for external CO2 probe. Plug properly oriented probe connector to the input of Multilogger. You can follow the optical key for correct position. Make sure that the wires, including their shielding, are not connected to any other conductive parts! Universal MiniDin connectors are designed for DigiS/M…, DigiL/M… or Pt1000/M probes. Conventional (custom) probes with a Pt1000 sensor can also be connected, however with certain limitations (see Appendix 4). The input connector marked “CO2is not universal and it is designed for external CO2 probe only.
Universal thermocouple connector ___________________
Plug properly oriented connector to the input of Multilogger. The connector is mechanically coded using a wider pin. Use the white uncompensated connector for measuring voltage. Make sure that the wires are not connected to any other conductive parts! Electrical coupling among thermocouples can cause large measurement errors or instable values! To obtain correct measurement values, the Multilogger has to be temperature-stabilized!
Marking of subminiature thermocouple connectors and wires manufactured by OMEGA (in accordance with US standard):
Thermocouple type
Connector color
+ wire color
- wire color
K (NiCr-Ni)
Yellow
Yellow
Red
J (Fe-Co)
Black
White
Red
S (Pt10%Rh-Pt)
Green
Black
Red
B (Pt30%Rh-Pt)
White
Black
Red
T (Cu-CuNi)
Blue
Blue
Red
N (NiCrSi-NiSiMg)
Orange
Orange
Red
IE-MLG-M1xxx-07 17
Two voltage ranges are available on each universal thermocouple connector for measuring small voltages:
range -60 mV to +140 mV range -18 mV to +18 mV
Use the white uncompensated connector connected according to the picture.
Terminals ______________________________________
Inputs (0 to 20) mA, (0 to 10) V, binary inputs and the counter are equipped with a two-part, self-locking WAGO terminal block. Insert flat- bladed screwdriver to rectangle terminal hole and push screwdriver towards away from you - contact is released. Insert wire to released terminal (circular hole behind the rectangular one) and close the terminal by removing the screwdriver.
The terminal block can be removed from the device without the need to disconnect individual wires!
Connection of the binary and counter input:
18 IE-MLG-M1xxx-07
Connection of the transmitter with voltage output:
Connection of the transmitter with the current output (active):
Connection of the transmitter with the current output (passive):
IE-MLG-M1xxx-07 19
Alarm output
This output is accessible on terminals on the bottom side of the Multilogger. Here are available these signals: OUT - open collector transistor, +5V power supply from external power source and GND terminal (see also the diagram in Appendix 2).
As the default, this output transistor is switched during selected alarms. It is possible to set inverse behavior in the device's configuration (in this case loss of external power supply acts as an alarm state). Activity of this output can be canceled from the Multilogger keypad or remotely from a PC or web. Remember to set a suitable alarms delay to prevent false activity of this output.
It is possible to connect to this output: External audio alarm (optional accessory MP026) - connect the +5 V
terminal with the central lead of the CINCH connector of the audio alarm unit and the OUT terminal with the outer lead of CINCH connector (see the picture).
ATTENTION - the External audio alarm unit will not work if the
Multilogger does not have an external power supply.
Telephone dialer (optional accessory MP002) - in case of alarm
telephone dialer dials specified telephone number and announces voice message. Connect the dialer according to its operation manual.
External alarm with its own power supply - make the connection
according to the following picture. Be careful do not exceed the maximum allowed output current from OUT terminal (see the part Technical Parameters). Use an external relay for higher currents.
20 IE-MLG-M1xxx-07
IE-MLG-M1xxx-07 21
Power supply of Multilogger
The Multilogger is powered from three AA batteries and optionally also from an external +5 V DC power supply, which can be also used for charging the batteries. A simplified schematics of the internal supply circuits is showed in Appendix 2.
External power supply _____________________________
External power supply can be connected to the Multilogger in
several ways:
a) to the power supply connector on the bottom of Multilogger
(for example, AC adapter included in basic accessories)
b) to the terminal block of the RS232 interface (+5 V and GND
terminals)
c) via USB cable connected to PC (it will work only if the USB
port of the PC has current sufficiency to power the Multilogger)
No energy from the batteries is discharged if external source is
present.
The Ethernet interface works only if an external power source is
connected. Some limited functions of this interface may also be available during battery operation.
ATTENTION - never connect voltage greater than 5.5 V to
the external power supply input - it could damage the device.
Batteries _______________________________________
The new Multilogger is equipped with high-quality NiMH
rechargeable batteries; the device is preset to intelligent algorithm to their continuous charging if external power is present.
ATTENTION - if you insert the non-rechargeable batteries
to the Multilogger, change the battery model to Alkaline in the device configuration. Never charge batteries that are not rechargeabled!
Charging batteries:
a) batteries can be charged if an external power source is
connected
22 IE-MLG-M1xxx-07
b) the charging process will start if the NiMH charging
accumulators are selected in the device configuration
c) when the Multilogger is turned off, the batteries cannot be
charged
d) the Multilogger allows for two battery charging modes -
high current fast charging mode and low current slow charging mode
Slow low current charging mode is mainly used for stationary
installations. Batteries work as a backup source in case of external power failure. Charging is automatically activated after restoring of external power supply. During slow charging device operates normally. Recharging of completely discharged battery takes several days, however, in normal operation, do not expect a completely discharged battery. The charging process is indicated on the device display using the
corresponding symbol .
Fast high current charging mode is mainly used for portable
applications. Discharged batteries can be recharged within a few hours. The charging procedure has to be activated by the service personnel:
a) If the Multilogger is on, switch it off. Connect an external
power supply to the device.
b) Press down and hold (approximately 3 seconds) the red
On/Off key on the Multilogger until you see “NiMH charger” on the device display. The charging process is started.
c) During fast charging Multilogger does not work in the
normal operating mode, does not take measurements and does not communicate with the PC.
d) Once the charging process is completed, the Multilogger
displays the “Batteries charged” message and you can turn it off.
e) The charging can be proceed within a temperature range
of 0 °C to +40 °C only. If the temperature is out of this range, the charging process is automatically stopped.
IE-MLG-M1xxx-07 23
Battery replacement:
If you want to keep the correct time setting in your device, do not let the battery discharge completely.
When replacing the battery, do it quickly. The timer circuit runs without battery for max. 5 minutes if the device is
switched off. a) Switch the Multilogger off. b) Remove the battery cover, take out the old batteries and
insert new batteries (take care for correct polarity). Battery
polarity is shown on its holder. c) Switch the Multilogger on and make sure the set time is
correct (device menu View settings - General settings). d) If you change the battery chemistry (rechargeable <=> non-
rechargeable), you have to make this change in the device
configuration also.
Note: The Multilogger can operate with removed batteries powered externally. This is not a standard use of the device but it is possible. In this case, the full battery is displayed. Interruption of power supply causes loss of time setting.
Loading...
+ 51 hidden pages