P.3
▲
OPERATION
It is helpful to review operation before commencing operation. Steps (1) and (3) may be omitted once you have
verified that the unit is operating properly.
①
Connect a 50-Ohm dummy load (not supplied) to the “M” (“UHF”) jack.
②
Pressthe POWER switch to turn the unit on in the Normal mode.
Normal Mode Display Screen
Battery Level Meter (※2,3)
Frequency Battery Status
Band Scale
The amateur bands 1.8-430 MHzare shown with a blue
band (※1). Red underscore showscurrent band.
“U” appears to left of Battery icon when on UHF.
SWR Scale and Bar
The SWRis displayed graphically and numerically
here. The SWRis shown to two decimal places.
OHMSScale and Bar
The resistive component of the impedance is shown both
on the bar graph and numerically. Impedance (R, X) is
also displayed numerically. Above 190 MHz, R/X do not appear.
※
1 Note that not all the annotated amateur radio bandsare available in all countries.
※
2. If using Nickel-Metal Hydride batteries, the battery status meter may never show “Full.” You should recharge
the batteriesonce the Yellow indication appears. If using alkaline cells, replace when the Yellow indicator appears.
③
When connected to a 50-Ohm dummy load, the SWR indication should be 1.1 or less. If so, the meter is
performing correctly.
④
Perform measurements accordingto the discussion below.
▼
SWRMeasurement
●
Connect the coaxial cable to the meter; use the shortest possible cable between the antenna and meter, as any
cable between the antenna and meter can transform the measured impedance, and feedline loss can also cause the
displayed SWRto be better than it actually is. See FAQ later.
●
Set the BAND and FREQ knobsto the band and frequency to be tested on the antenna.
●
You may now read the SWR and resistive value of impedance (OHMS) on the display. If there is a subtle
difference between the mechanical and the LCD, the precision of the mechanical meter has priority.
About the Reactance Display
The display of impedance on the LCD includes both resistive (R) and reactance (X) indications. Note that the X
display is absolute value, so the polarity (+ or -) cannot be viewed. Because of diminished accuracy, the numerical
Rand Xdisplays do not appear above 190 MHz (but the impedance bar graph remainsoperational).
▼
Measurement of Antenna Resonant Frequency
■
Connect the antenna to the appropriate antenna connector on the meter, using the shortest possible cable. If you
use a long cable, the measurement will also include the effects of feedline loss, and impedance differences between
the antenna system and coaxial cable will not show the actual antennaimpedance.
■
While watching the frequency display, rotate the FREQ knob while observing SWR. If you rotate the FREQ knob to
the point where the frequency no longer changes, change the Band to set a new range.
■
The frequency at which the minimum SWRis observed is the resonant frequency for the system.
Always be sure that you connect your coaxial connector to the proper connector. If you are connected to the
opposite connector, measurements will be impossible and if the wrong connector is used, damage to the jack
may occur.
Full charge CcCharge
Nearly full
Battery low
Battery level
Critically low
You created this PDF from an application that is not licensed to print to novaPDF printer (http://www.novapdf.com)