EVA Series are controlled by Digital key panel console that displays KV, mA and mAs with APR meuu
programmed. System-self-diagnosis and self –diagnostic circuitry standard on each unit. Equpped with
closed loop for X-ray tube current as well as kVp, minimizing potential errors and need for re-adjustments.
Microprocessor controlled high frequency generator for all examinations in the area of the X- ray
diagnostics, prepared for special applications. The automatic exposure control (AEC) is integrated. This
generator features by a very compact design and can be integrated in the base of the examination table
as a space saving compact solution.
The foil coated control desk enables easy operation and hygienic possibilities for cleaning as well as easy
operation for the user. In the anatomical program operating mode can be stored programs individually.
An internal self-diagnosis centre monitors the X-ray unit and indicates errors on the screen. So the
generator is optimal protected against malfunctions.
T
e
c
h
n
i
c
a
l
D
a
t
a
T
e
c
h
n
i
c
a
T
e
c
Power output at 100 kV / 0,1 s 40 kW
Converter 40 kHz
Exposure techniques * kV-mAs
(alternatively with Automatic Exposure Control)* kV-mA-ms
The E 7252X is a double-focus rotating anode X-ray tube which is designed for general radiographic
procedures, according to the generators power data.
T
e
c
h
n
i
c
a
l
D
a
t
a
T
e
c
h
n
i
c
a
l
T
e
c
h
Maximal voltage 150 kV 125kV
Nominal power small focus 16 kW 21kW large focus 44 kW 45kW
Focal spot dimensions
small focus 0,6 x 0,6 mm 1.0x1.0mm
large focus 1,2 x 1,2 mm 2.0x2.0mm
The radiographic examination unit EVA featured by an optimised construction which permits to realise a
multitude of examinations on a minimised room. The table is prepared to integrate a special X-ray
generator into the base. The complete system can be installed quickly and simple without additional
special construction works.
The examination table is composed of a Bucky table with floating patient table top and an integrated
column which supports the tube assembly, collimator and the control board. The column doesn’t need any
additional fixing elements.
The arm which supports the tube and the control board permits the rotation of the radiogen system on the
axe of the arm itself and enable to realise special positions with different angles of incidence as well as the
operating of a Bucky wall stand. The vertical movement of the radiogen system is balanced by
counterweights. The transversal displacement of the tube as well as the rotation of the column is standard
All movements of the column and the table top are stopped by electromagnetic brakes.
The top panel of the Bucky table is in radio transparent laminated plastic with longitudinal profiles in
extruded aluminium and lateral guides of insertion and anchorage of the accessories of common use. The
movements of the table top are stopped by electromagnetic brakes which are active in the absence of
power. They keep it in a defined position as well the installation is off.
The Bucky diaphragms of the table as well as the wall stand can be equipped with AEC ionisation
measurement chambers.
T
e
c
h
n
i
c
a
l
D
a
t
a
T
e
c
h
n
i
c
a
l
T
e
c
h
Bucky table
Table top material grey-white colour laminated
plastic
Absorption approx. 0,9 mm Al-GW
Dimensions 200 x 74 cm
Longitudinal travel 60 cm (± 30 cm)
Transversal travel 16 cm (± 8 cm)
Table top upper edge to film distance 7,5 cm
Table top distance to the floor 70 cm
Bucky carriage longitudinal travel 38 cm
Grid 12:1, 36 L/cm, foc 110 cm
Acoustical signalling of the table’s central position.
The cassette tray is supported in a telescopic roller guide
Tube stand
Longitudinal displacement travel 185 cm
Column rotation around the vertical axis ± 90°
Min. focus distance to the floor (90°-Position) 73 cm
Vertical travel 113 cm
Max. FFD 120 cm
Tube rotation around the horizontal axis ± 90°
Transversal displacement travel of the tube 14 cm (± 7 cm)
Option: Single phase supply with battery
Fuse 3x 35 A delayed
Line impedance max. 0,3 Ω
Nominal power input 23 kVA
Short time pulse max. 68 kVA
Stand-by 120 W
Weights and dimensions
Bucky table with tube stand kg
Generator built-in into the
Examination table
kg
Control desk 230 x 410 x 80 mm
5 kg
Required space
You need an area of about 2,5 x 4 m.
Certification
CE Certificate L-0243314-02
EN ISO 9002:1994 / EN 46002:1996
10